K051180 · Vanguard Medical Concepts, Inc. · KTT · Jul 22, 2005 · Orthopedic
Device Facts
Record ID
K051180
Device Name
VANGUARD REPROCESSED EXTERNAL FIXATION DEVICES
Applicant
Vanguard Medical Concepts, Inc.
Product Code
KTT · Orthopedic
Decision Date
Jul 22, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Reprocessed External Fixation Devices are intended for the treatment of bone conditions that can be corrected or improved by external skeletal traction or fixation, including osteotomy, arthrodesis, fracture and reconstructive surgery.
Device Story
Reprocessed external fixation devices; previously used non-invasive orthopedic components cleaned, inspected, tested, and packaged for reuse. System comprises rods, couplings, clamps, rings, and ring segments; forms bridge constructs for bone stabilization. Surgeon attaches anchoring screws, pins, or wires to construct to immobilize bone fractures where soft tissue injury precludes internal fixation. Used in clinical settings for fracture management and reconstructive surgery. Benefits include rigid structural support for bone healing. Devices are non-sterile; require steam sterilization by healthcare facility prior to use.
Clinical Evidence
Bench testing only. Cleaning, packaging, and performance testing were conducted to demonstrate that reprocessed devices perform as intended and are safe and effective.
Technological Characteristics
Components include rods, rod-to-rod/pin couplings, rings, ring segments, and clamps. Materials: stainless steel, carbon fiber, aluminum, anodized aluminum alloys, and titanium alloys. Form factor: modular external fixation system. Connectivity: none. Sterilization: steam sterilization by healthcare facility. Software: none.
Indications for Use
Indicated for patients requiring external skeletal traction or fixation for bone conditions including osteotomy, arthrodesis, fracture, and reconstructive surgery. Contraindicated in patients with mental or neurologic impairment interfering with cooperative postoperative care; not for spinal screw attachment.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Smith & Nephew Ilizarov External Fixator
Smith & Nephew Taylor Spatial Frame Fixator
Vanguard Medical Concepts Reprocessed External Fixation Devices
Submission Summary (Full Text)
{0}------------------------------------------------
K051180 p 1/4
# JUL 2 2 2 2005 510(k) Summary of Safety & Effectiveness
| Submitter | Vanguard Medical Concepts, Inc.<br>5307 Great Oak Drive<br>Lakeland, FL 33815 | | | | | | | | |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|
| Contact | Mark Waddell<br>Project Manager, Research & Development<br>863-683-8680 ext. 263 [voice]<br>863-904-1604 [facsimile]<br>mwaddell@safe-reuse.com [email] | | | | | | | | |
| Date | April 6, 2005 | | | | | | | | |
| Device | Trade Name: Vanguard Reprocessed External Fixation Devices<br>Ilizarov System Taylor Spatial Frame Common Name: Reprocessed External Fixation Devices<br>Classification: 21 CFR, 888.3030<br>Classification Name: Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component<br>Device Class: Class II<br>Product Code: KTT | | | | | | | | |
| Predicate Devices | External Fixation Devices legally marketed by the following original equipment manufacturers (OEM) and third party reprocessor:<br>OEM / Reprocessor Trade Name Smith & Nephew®, Inc. Ilizarov External Fixator® Smith & Nephew®, Inc. Taylor Spatial Frame® Fixator Vanguard Medical Concepts, Inc. Reprocessed External Fixation Devices | | | | | | | | |
| Indications for Use | Reprocessed External Fixation Devices are intended for the treatment of bone conditions that can be corrected or improved by external skeletal traction or fixation, including osteotomy, arthrodesis, fracture and reconstructive surgery. | | | | | | | | |
| Contra-indications | External fixation devices are contraindicated in patients with mental or neurologic impairment that would interfere with cooperative postoperative care. These devices are not intended for attachment or fixation of screws to the spine. | | | | | | | | |
{1}------------------------------------------------
#### 510(k) Summary of Safety & Effectiveness, continued
Vanguard Reprocessed External Fixation Devices are previously used non-Device invasive orthopedic devices that have been cleaned, inspected, tested, and Description packaged by Vanguard Medical Concepts, Inc. External fixation device systems are comprised of various elements that, when used in conjunction with one another, form bridge constructs to which anchoring screws, wires and/or pins may be attached. Bridge elements are designed to provide a framework for stabilization of bone fractures where soft tissue injury may preclude the use of other fracture treatments such as IM rods, casting or other means of internal fixation.
> External fixation elements consist of components such as straight and curved rods, rod-to-rod, and rod-to-pin couplings and clamps, rings and ring segments, ring-to-rod, and ring-to-pin clamps. These components are provided in bulk, packaged in nylon pouches, non-sterile, with instructions for steam sterilization by the health care facility.
- Rods and Telescoping Rods ~ Straight rods are external fixation devices of varying lengths (adjustable lengths in the case telescoping rods) and diameters that are used with rod-to-rod and pin-to-rod clamps by the surgeon to connect anchoring pins, screws and wires together to form a rigid structure that immobilizes the affected bone or structures. Differences in length and diameter of the rods allow accommodation of a broad range of fracture scenarios and applied loads. All rods have a straight, solid, round design. Rods and telescoping rods are typically constructed of stainless steel, carbon fiber, or aluminum or a combination of materials.
Curved rods are multilateral devices bent in shapes useful for some constructs where straight rods are less suitable. These rods allow the surgeon to connect pins in various locations in the limb together to form a rigid structure. Curved rods are manufactured from either aluminum or stainless steel.
- Rod-to-Rod Couplings ~ These are multi-element components used to 체 connect one rod to another in a range of positions defined by the individual clamp configuration. They are designed to interconnect a specific size or range of sizes of rods. The devices are typically constructed from one or more of the following materials: anodized aluminum alloys, steel or stainless steel alloys and titanium alloys.
Continued on next page
Page 2 of 4
{2}------------------------------------------------
### 510(k) Summary of Safety & Effectiveness, continued
Device Description (continued)
- . Rod-to-Pin Couplings ~ These are multi-element components used to connect one rod to a pin or group of pins in a range of positions defined by the individual clamp configuration. They are designed to interconnect a specific size or range of sizes of rods to specific sizes or ranges of sizes of pins. The devices are typically constructed from one or more of the following materials: Anodized aluminum alloys, steel and/or stainless steel alloys and titanium alloys.
- I Rings and Ring Segments ~ These are circular or semicircular segments manufactured to surround the area of attachment to bone located centrally in the ring. Various attachments are made to the rings or ring segments to stabilize bone fractures or to reduce or extend the length of bones. Tensioned wires or pins are commonly attached to rings and ring segments using clamps designed for this purpose. Multiple ring or ring segment constructs are used to stabilize and structurally support the anatomical structures being treated using a range of configurations and attachments and connectors. The rings and ring segments are manufactured in a range of diameters to allow selection of a size most appropriate to the anatomy and application needed. Rings and ring segments are typically manufactured from one or more of the following materials: anodized aluminum alloys, steel and stainless steel alloys or carbon fiber composites.
- I Ring-to-Rod Clamps ~ Ring-to-rod clamps are utilized to connect a ring or ring segment to a rod. These attachments are made to form an external fixator frame construct as required for the particular biomechanical needs of the procedure. Ring to rod clamps are typically constructed from one or more of the following materials: anodized aluminum alloys, steel and/or stainless steel alloys and titanium alloys.
- . Ring-to-Pin or Ring-to-Wire Clamps ~ Ring-to-pin or ring-to-wire clamps are utilized to connect an external fixation ring or ring segment to a pin or wire that is normally affixed to the bone passing centrally through the ring. Wires are normally attached on one side of the ring, passed through the bone, and continue to an attachment point on the opposite side of the ring where they are affixed under tension to another ring to wire clamp. Ring-to-pin clamps are used to secure the ring to a pin or set of pins that are anchored in the bone passing centrally through the external fixation ring. Ring-to-pin and ring-to-wire clamps are typically constructed from one or more of the following materials: anodized aluminum alloys, steel and/or stainless steel alloys and titanium alloys.
V
{3}------------------------------------------------
V
# 510(k) Summary of Safety & Effectiveness, continued
| Technological<br>Characteristics | Vanguard Reprocessed External Fixation Devices are essentially identical to<br>the Original Equipment Manufacturer (OEM) devices. No changes are made<br>to the currently marketed OEM device specifications and the Reprocessed<br>External Fixation Devices possess identical technological characteristics. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test Data | Cleaning, packaging, and performance testing demonstrate that the<br>reprocessed devices perform as intended and are safe and effective. |
| Conclusion | Based upon the information provided herein and the 510(k) "Substantial<br>Equivalence" Decision Making Process Chart, we conclude that Vanguard<br>Reprocessed External Fixation Devices are substantially equivalent to their<br>predicate devices under the Federal Food, Drug and Cosmetic Act. |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the word "DEPARTMENT OF" in a bold, sans-serif font. The text is black and appears to be part of a larger document or heading. A horizontal line is visible below the text, possibly separating it from other content.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mark Waddell Project Manager, Research and Development Vanguard Medical Concepts, Inc. . . . . . 5307 Great Oak Drive Lakeland, Florida 33815
Re: K051180
K031160
Trade/Device Name: Vanguard Reprocessed External Fixation Devices (see enclosed list) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: KTT Product Code: II Dated: May 6, 2005 Received: May 9, 2005
Dear Mr. Waddell:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 9 rotty premainer is substantially equivalent (for the indications felerenced above and nave determined in a are and cate devices marketed in interstate for use stated in the encrosure/to regars actment date of the Medical Device Amendments, or to commerce production to May 20, 1976, the encordance with the provisions of the Federal Food, Drug, devices that have been receassined in accera approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, market the devices, bacjes, be courements for annual registration, listing of general controls provisions of the revy labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 above) also Existing major regulations affecting your device EDA may thay be subject to suell additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Oous overning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not meant Please be advised that I Dri 3 issualited or a baces and other requirements of the Act
that FDA has made a determination that your device complies with other requirements of that IDA has made a determination administered by other Federal agencies. You must of any I cuchares and regaranents, including, but not limited to: registration and listing (21 comply with an the 70c 31equirements, while and of CTN in the quality systems (DS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality Systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 -- Ms. Heather Crawford, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin maneting your and equivalence of your device to a legally premarket notification. THC PDF mailing of basis.ification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spocitie at no 10-10-10-10-12-0120. Also, please note the regulation entitled, Colliation of Compulance an (21 the entification" (21CFR Part 807.97). You may obtain Misoranding of Icrerchee to premainsvillities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Misiam C. Purvat
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Vanguard Reprocessed External Fixation Devices
. .
. .. .
Indications for Use:
Reprocessed External Fixation Devices are intended for the treatment of bone Neprocessed Extornal Pixation - or improved by external skeletal traction or conditions that out be comes arthrodesis, fracture and reconstructive surgery.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K051180
Page 1 of 1
0 1
iv
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.