The GenerOs SB is an implantable device for distraction osteogenesis techniques in the small bone of the extremities. GenerOs SB is used to treat conditions where small bones of the extremities are deficient. The types of deformities that fall into this category include, but are not limited to: - Congenital deficiencies of the bones of the forearms, wrists, ankles, hands and feet; - Post-traumatic deficiencies of the bones of the forearms, wrists, ankles, hands and feet; - Deficiencies of the bones of the forearms, wrists, ankles, hands and feet due to tumor resection. Each GenerOs SB is intended for single use only. The device is to be removed after distraction and bone stabilization are complete. It is to be used with other commercially-available accessory devices, such as bone screws for fixation to the bone surface. The device is not intended to be fixed to the bone with bone cement. However, it is possible that commercially-available bone cement may be used on the undersurface of the device to level or stabilize it on a curved surface.
Device Story
GenerOs SB is a surgical-grade stainless steel implantable distraction device for small bone extremities. Device consists of two telescoping bone plates distracted by a threaded drive shaft; internal gear articulates activation pin and drive shaft. Operated by physician in clinical setting; transcutaneous pin used for manual distraction activation; pin removed post-distraction. Device fixed to bone surface via accessory bone screws; removed after distraction and consolidation. Provides mechanical distraction to stimulate bone growth in deficient areas. Benefits patient by correcting bone deformities.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility, sterilization, packaging validation, and mechanical testing in conformance with standard guidelines for bone plate implant devices. No clinical data presented.
Indicated for distraction osteogenesis in small bones of extremities (forearms, wrists, ankles, hands, feet) to treat congenital, post-traumatic, or tumor-related bone deficiencies. Contraindicated for fixation with bone cement, though cement may be used for leveling/stabilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a sequence of alphanumeric characters. The characters are 'K', 'o', '5', '1', '1', '6', and '2'. The characters are written in a bold, handwritten style.
Image /page/0/Picture/1 description: The image shows the logo for OrthoNet Incorporated. The logo features a stylized eye graphic on the left, followed by the company name in a blocky, sans-serif font. Below the name, in smaller font, is the tagline "the osteoplastic surgery company".
510(k) Premarket Notification JUL 1 9 2005 GenerOs SB Small Bone Distraction Implant
## Section 1.3
#### Summary of Safety and Effectiveness
## REGULATORY AUTHORITY
Safe Medical Devices Act of 1990, 21 CFR 807.92
## COMPANY NAME/CONTACT
Terry Knapp, MD OrthoNetx, Inc. 1000 S. McCaslin Blvd, Suite 300 Superior, CO 80027 Voice: 303.494.1681 x 301 FAX: 303.494.1714
#### NAME OF DEVICE
| Trade Name: | GenerOs™ SB Small Bone Distraction Implant | | |
|---------------------|-------------------------------------------------------------------------------|--|--|
| Common Name: | Internal Bone Plate Distractor | | |
| Common Description: | Single/Multiple component metallic bone fixation<br>appliance and accessories | | |
#### Classification Names:
| Regulation<br>Number | Product<br>Code | Classification Name | Device<br>Class |
|----------------------|-----------------|--------------------------------------------------------------------------------|-----------------|
| 21 CFR<br>888.3030 | KTT | Appliance, fixation,<br>nail/blade/plate<br>combination, multiple<br>component | II |
| 21 CFR<br>888.3030 | HRS | Plate, Fixation, Bone | II |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for OrthoNet Incorporated. The logo features a stylized eye-like symbol on the left, followed by the word "OrthoNet" in a modern, sans-serif font. Below "OrthoNet," the word "Incorporated" is written in a smaller font, and below that is the tagline "the osteoplastic surgery company".
## DEVICE DESCRIPTION
GenerOs SB is an implantable device made of surgical grade stainless steel for deficiencies of small bones of the extremities. It features two telescoping component bone plates that are distracted apart by a threaded drive shaft. The activation pin and the drive shaft are articulated using an internal gear. The device has a fixation plates on the activation and sliding block, and if not used for fixation of the device, may be easily removed. Activation of the drive shaft occurs through a transcutaneous pin. which is removed once the distraction phase is complete. The GenerOs SB implant is removed after distraction and consolidation are complete. The GenerOs SB includes reusable instruments such as activation, insertion / removal tools and other surgical instruments.
GenerOs SB is a single-use device, sold non-sterile, and requires sterilization prior to use. It is. Sterilization instructions are included in the labeling.
## INDICATION FOR USE STATEMENT
The GenerOs SB is an implantable device for distraction osteogenesis techniques in the small bone of the extremities. GenerOs SB is used to treat conditions where small bones of the extremities are deficient. The types of deformities that fall into this category include, but are not limited to:
- Congenital deficiencies of the bones of the forearms, wrists, ankles, hands . and feet;
- . Post-traumatic deficiencies of the bones of the forearms, wrists, ankles, hands and feet:
- Deficiencies of the bones of the forearms, wrists, ankles, hands and feet due . to tumor resection.
Each GenerOs SB is intended for single use only. The device is to be removed after distraction and bone stabilization are complete. It is to be used with other commercially-available accessory devices, such as bone screws for fixation to the bone surface. The device is not intended to be fixed to the bone with bone cement. However, it is possible that commercially-available bone cement may be used on the undersurface of the device to level or stabilize it on a curved surface.
## PREDICATE DEVICES
- GenerOs Bone Generator; OrthoNetx, Inc. (formerly Inter-Os Technologies) . (#K993869)
- Limb Lengthener; OrthoNetx, Inc, (#K031875) ●
- Arthrex Small Fragment Plates and Screws; Arthrex, Inc., (#K040907) .
- Lorenz Small Fragment System (#K992961); Walter Lorenz Surgical, Inc.
- LCP Modular Foot Plates; Synthes (USA), (#K050110)
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for OrthoNet Incorporated. The logo features a stylized eye graphic on the left, followed by the word "OrthoNet" in a bold, sans-serif font. Below "OrthoNet" is the text "Incorporated X." and the tagline "the osteoplastic surgery company".
# SUBSTANTIAL EQUIVALENCE COMPARISON
The GenerOs SB is identical in features and technology to the GenerOs CF Craniofacial Bone Generator (#K993869); the only difference is the indication for use.
The features and indications for use are compared to OrthoNetx' Limb Lengthener (#K031875).
Indications for use are compared to Arthrex Small Fragment Plates and Screws by Arthrex, Inc., (#K040907); Lorenz Small Fragment System (#K992961) by Walter Lorenz Surgical, Inc. and LCP Modular Foot Plates by Synthes (USA), (#K050110).
The GenerOs SB is substantially equivalent to predicate devices based on the descriptive characteristics, similar intended use, and same principle operation of distraction osteogenesis.
#### PERFORMANCE
Performance of the GenerOs SB has been substantiated by biocompatibility, sterilization, packaging validation, and mechanical tests in conformance to standard testing quidelines for bone plate implant devices.
#### CONCLUSION
Based on the design, materials, function, intended use, the GenerOs SB is substantially equivalent to the devices currently cleared under the Federal Food, Drug and Cosmetic Act. The GenerOs SB Distraction Implant raises no new safety or effectiveness issues. Therefore, safety and effectiveness are reasonably assured, iustifying 510(k) clearance for the GenerOs SB Distraction Implant.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of two main elements: a symbol and text. The symbol is a stylized representation of a human figure with three heads, suggesting a focus on people and well-being. The text is arranged in a circular pattern around the symbol, spelling out "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 19 2005
Terry Knapp, M.D. CEO OrthoNetX, Inc. 1000 S. McCaslin Blvd. Suite 300 Superior, Colorado 80027
i
Re: K051162
Trade/Device Name: GenerOs™ SB Small Bone Distraction Implant Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT, HRS Dated: May 4, 2005 Received: May 5, 2005
Dear Dr. Knapp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Dr. Knapp
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin marketing of substantial equivalence of your device to a legally premarket nother in the r DAY midning of backand of ...
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your accept (240) 276-0120. Also, please note the regulation entitled, Comaci the Office of Compilance in (21 transmit (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-2041 or Manufacturers, Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Eun A. Kim
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
OrthoNet
Incorporated
*"the osteoplastic surgery company"*
| Section | 1.2 |
|---------|-----|
|---------|-----|
## Indications for Use Statement
510(k) Number:
Device Name:
GenerOs™ SB Small Bone Distraction Implant
Indications for Use:
The GenerOs SB is an implantable device for distraction osteogenesis techniques in t the GenerOs SB is an implantable downed to treat conditions where the small bone of the extremities are deficient. The types of deformities that fall into this ategory include, but are not limited to:
- Congenital deficiencies of the bones of the forearms, wrists, ankles, hands and feet:
- and leet,
Post-traumatic deficiencies of the bones of the forearms, wrists, ankles, hands and feet;
- and leet, to tumor resection.
Each GenerOs SB is intended for single use only. The device is to be removed after . Lach OcherOS OD is intonuba so ere complete. It is to be used with other distraction and bone stablization are semplotes, such as bone screws for fixation to the commercially-available accessery do heef on the bone with bone with bone with bone cement.
The bone surface. The device is not intended to be fixed to the word on the . However, it is possible that commercially-available bone cement may be used on the undersurface of the device to level or stabilize the GenerOs SB on a curved surface.
| Prescription Use | XX | or | Over-The-Counter Use |
|----------------------|----|----|----------------------|
| (Per 21 CFR 801.109) | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Confidential | K051162 1-2 |
|--------------|-------------|
|--------------|-------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.