MODEL TL-360 POWER SCOOTER

K051118 · Teh Lin Prosthetic & Orthopaedic, Inc. · INI · May 31, 2005 · Physical Medicine

Device Facts

Record IDK051118
Device NameMODEL TL-360 POWER SCOOTER
ApplicantTeh Lin Prosthetic & Orthopaedic, Inc.
Product CodeINI · Physical Medicine
Decision DateMay 31, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Device Story

TEH LIN Power Scooter, TL-360 is a battery-operated, four-wheeled indoor/outdoor mobility scooter. Rider operates device via hand controls located on steering column. Device features foldable frame, removable armrests, and onboard battery charger. Designed for transport via disassembly. Provides mobility for individuals restricted to seated position. Safety features include UL-certified batteries and chargers, and materials tested for resistance to ignition.

Clinical Evidence

No clinical data. Bench testing only. Compliance with EMC and safety standards: ANSI/RESNA WC/Vol.2-1998, CISPR 11:1990, EN61000-3-2:1995, and IEC61000-3-3:1995.

Technological Characteristics

Four-wheeled battery-operated scooter. Penny & Giles EGIS control system. UL-certified battery and charger. Foldable frame and upholstery materials passed resistance ignition testing (SGS).

Indications for Use

Indicated for persons restricted to a sitting position requiring mobility assistance.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 3 1 2005 ## 1051118 ## TEH LIN PROSTHETIC & ORTHOPAEDIC INC No. 7, Wu Chuan 7th Road, WuKu Industrial Park, Taipei County, Taiwan R.O.C. Telephone: 886-2-22991901 Fax: 886-2-22991030 E-mail: tlco(@ms2.hinet.net http://www.tehlin.com.tw #### 46 510(k) SUMMARY " Submitter's Name: TEH LIN Prosthetic & Orthopaedic Inc. No. 7, Wu Chuan 7th Road, WuKu Industrial Park, Taipei County, Taiwan R.O.C. April 26, 2005 Date summary prepared: Device Name: Proprietary Name: Common or Usual Name: Classification Name: TEH LIN Power Scooter, TL-360 Powered Scooter Powered Scooter, Class II, 21 CFR 890.3860 Indications for Use: The device is intended for medical purposes to provide mobility to persons rests to a seated position. Description of the device: The TEH LIN Powered Scooter, TL-360 is an indoor / outdoor Powered Scooter that It has a base with four-wheeled with a seat. is battery operated. The movement of the Scooter is controlled by the rider who uses hand controls located at the top of the The device can be disassembled for transport and is provided with steering column. an onboard battery charger. Performance Testing: EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically Powered Scooters, controller, and their chargers requirements and test methods) Legally marketed device for substantial equivalence comparison: TEH LIN POWER SCOOTER TL-821 (K042011) {1}------------------------------------------------ # 5051118 ## TEH LIN PROSTHETIC & ORTHOPAEDIC INC. のところです。 このため、 このため、 この 2/2 No. 7, Wu Chuan 7th Road, WuKu Industrial Park, Tainci County, Taiwan R.O.C. Telephone: 886-2-22991901 Fax: 886-2-22991030 E-mail: tlco@ms2.hinet.net http://www.tehlin.com.tw Summary for substantial equivalence comparison: The batteries used are the The intended use between the two devices is the same. same brand that is certified by UL. The control systems for the two devices are same brand i.e., Penny & Giles EGIS type for the two devices. The recharge for the two devices are used the same resource, and the recharge is certified by UL. Besides, the foldable frame, the maximum speed, the safe climbing angle, removable arm type, same warranty on component and frame, and back upholstery are the same material that also be passed the resistance ignition test by SGS. Thus the same safety level for the two devices is assured. The major differences existing of the two Power Scooters are the different overall dimension, tires, seat size, weight limit, weight capabilities, and cruising range between the two devices. Based on the above the information and the analysis, we know that the subject device and the predicate device have the same intended use, the same technological aspects and only minor differences exist. So the new device is substantially equivalent to the predicate devices in this aspect. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized eagle with three lines representing its wings and a wavy line representing its body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. MAY 3 1 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 TEH Lin Prosthetic and Orthopaedic Incorporated C/o Dr. Ke-Min Jen Chinese-European Industrial Research Society No. 58 Fu-Chiun Street Hsin Chu City, China (Taiwan) 300 Re: K051118 Trade/Device Name: TEH LIN Power Scooter, TL-360 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: II Product Code: INI Dated: April 26, 2005 Received: May 2, 2005 Dear Mr. Jen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ### Page 2 - Dr. Ke-Min Jen This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Stupt Rlwdn Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use K051118 510 (K) Number ( If Known ):__________________________________________________________________________________________________________________________________________________ Device Name: TEH LIN Power Scooter, TL-360 Indications for Use: The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. Prescription Use AND/OR Over-The-Counter Use _ V (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C ) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) rence of CDRH. Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________ Page_ 1__of __ 1 12 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...