MODIFICATION TO: MICROLIFE DIGITAL PACIFIER THERMOMETER, MODEL MT1751Q
K051100 · Microlife Intellectual Property GmbH · FLL · May 11, 2005 · General Hospital
Device Facts
Record ID
K051100
Device Name
MODIFICATION TO: MICROLIFE DIGITAL PACIFIER THERMOMETER, MODEL MT1751Q
Applicant
Microlife Intellectual Property GmbH
Product Code
FLL · General Hospital
Decision Date
May 11, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Microlife MT1751Q Digital Pacifier Thermometer is a non-sterile, reusable rno miorolime intended for the determination of oral body temperature in infants to children five years of age.
Device Story
Battery-powered digital thermometer; pacifier form factor; thermistor sensor located within silicone nipple; ABS plastic body. Principle of operation: thermistor resistance changes proportional to temperature; resistance changes converted to frequency via R-C oscillator circuit; fixed hardware offset applied to accelerate measurement time. Device used by caregivers to measure infant/child oral temperature. Output displayed on liquid crystal display (LCD). Provides rapid, accurate temperature readings to assist in clinical assessment of fever.
Clinical Evidence
Controlled human clinical studies conducted. Data evaluated clinical bias, clinical uncertainty, and clinical repeatability per Microlife Clinical Test Protocol. Results demonstrated performance consistent with clinical requirements for electronic thermometers.
Technological Characteristics
Materials: medical silicone rubber (nipple), ABS plastic (body). Sensing: thermistor-based. Energy: battery-powered. Connectivity: none. Standards: ASTM E1112, IEC60601-1, IEC60601-1-2. Processing: R-C oscillator circuit with fixed hardware offset for measurement speed.
Indications for Use
Indicated for determination of oral body temperature in infants and children up to five years of age.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Pro Check ™ Digital Pacifier Thermometer (K972259)
Submission Summary (Full Text)
{0}------------------------------------------------
# EXHIBIT #1
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in This outhinary of SE({{}, suirements of SMDA 1990 and 21 CFR §807.92.
The assigned 5l0(k) number is:_K O S ! ! OO
#### Submitter's Identification: 1.
Microlife Intellectual Property GmbH, Switzerland Max Schmidheiny-Strasse 201 9435 Heerbrugg / Switzerland
Contact: Mr. Gerhard Frick
Date Summary Prepared: April 27, 2005
#### Name of the Device: 2.
Microlife Digital pacifier Thermometer, Model MT1751Q
#### 3. Predicate Device Information:
Pro Check ™ Digital Pacifier Thermometer, K#972259, Micro Weiss Electronics
#### Device Description: 4.
The Microlife Digital Pacifier Thermometer, Model MT1751Q, is a batterypowered, liquid crystal display device using a thermistor inside the nipple. The patient contact portion is composed of medical silicone rubber. The body is of ABS plastic. This device is reusable and no components are disposable.
In addition, to speed up the measurement time, we have added the fixed offset into this thermometer, which is fulfilled by the hardware. With these characteristics, this thermometer can provide both a very high clinical accuracy and quick measurement time.
The basic principle of this thermometer is that change of thermistor resistance, caused by changes of temperature, are converted to changes of frequency of R-Coscillator circuit, Therefore, temperature can be given by measuring the frequency of the oscillator.
{1}------------------------------------------------
#### 5. Intended Use:
The Microlife MT1751Q Digital Pacifier Thermometer is a non-sterile, reusable rno miorolime intended for the determination of oral body temperature in infants to children five years of age.
#### Comparison to Predicate Devices: 6.
The Microlife Digital Pacifier Thermometer, Model MT1751Q, is identical to the Pro Check ™ Digital Pacifier Thermometer,K#972259, differing mostly in response time and, the position of the thermistor is fixed .
### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ASTM E1112, as well as IEC60601-1 and IEC60601-1-2 requirements.
Guidance documents included the "FDA Guidance on the Content of Premarket Notification (510(k)) Submissions for Clinical Electronic Thermometers".
#### Discussion of Clinical Tests Performed: 8.
Controlled human clinical studies were conducted using the Microlife Digital Pacifier Thermometer, Model MT1751Q. Clinical data was presented which evaluated clinical bias, clinical uncertainty and clinical repeatability per the Microlife Clinical Test Protocol outline.
#### റ്റ. Conclusions:
The Microlife Digital Pacifier Thermometer, Model MT1751Q, has the same intended use and similar technological characteristics as the predicate device, the Pro Check ™ Digital Pacifier Thermometer, K#972259. Moreover, bench testing contained in this submission demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the Microlife Digital Pacifier Thermometer, Model MT1751Q, is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/11 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the serpent entwined around a staff. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 1 2005
Microlife Intellectual Property GmbH C/O Ms. Susan D. Goldstein-Falk Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K051100
Trade/Device Name: Microlife MT1751Q Digital Pacifier Thermometer Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: April 27, 2005 Received: April 29, 2005
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your secures in ined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate commerce proc to tray as, seen reclassified in accordance with the provisions of Antichunches, or to do rough act (Act) that do not require approval of a premarket the rederal 1 00d; Drug, and Coou may, therefore, market the device, subject to the general approval application (1) = 1) = 1 = 1 = controls provisions of the restration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III If your device is classinod (soc ass 10) in a controls. Existing major regulations affecting (FMA), it may of Subject to Back and Federal Regulations, Title 21, Parts 800 to 898. In your device can be found in the Bear nouncements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Ms. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not Frease be auvisod that 1 Dr. o lostermination that your device complies with other requirements mean that I DA mas made a dotes and regulations administered by other Federal agencies. of the Act of ally I outer a base s requirements, including, but not limited to: registration 1 ou must comply with art 807); labeling (21 CFR Part 801); good manufacturing practice allu listing (21 OF Real vor), systems (QS) regulation (21 CFR Part 820); and if requirements as set form in als quadiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will anow yourse organ finding of substantial equivalence of your device to a premarket notification of a device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may other general Manufacturers, International and Consumer Assistance at its toll-free Drimber (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
slloo lof
### Exhibit B
of Page _ 1
510(k) Number (if known): _
Device Name: Microlife MT1751Q Digital Pacifier Thermometer
Indications For Use:
The Microlife MT1751Q Digital Pacifier Thermometer is a non-sterile, reusable The Microllie WTTPO & Digitaronthation of oral body temperature in infants to children five years of age.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | FOR DW |
|---------------------------------------------------------------------------------|--------|
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital, Infection Control, Dental Devices | |
Page 1 of 1510(k) Number KOSI | OO ______________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.