K051086 · Smith & Nephew, Inc. · KRR · May 31, 2005 · Orthopedic
Device Facts
Record ID
K051086
Device Name
SMITH & NEPHEW PATELLO-FEMORAL IMPLANT
Applicant
Smith & Nephew, Inc.
Product Code
KRR · Orthopedic
Decision Date
May 31, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.3540
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smith & Nephew Patello-Femoral Implant is indicated for replacement of the femoral side of the patello-femoral joint. It is intended to be used in patellofemoral arthroplasty in patients with degenerative arthritis in the distal femur and patients with a history of patellar dislocation or patellar fracture, and patients with failed previous surgery (arthroplasty, tibial tubercle elevation, and lateral release) where pain, deformity or dysfunction persists. The components are for single use only and are intended for implantation with bone cement.
Device Story
The Smith & Nephew Patello-Femoral Implant is a metallic prosthetic component designed to replace the articulating surface of the femoral trochlea. It is intended for use in orthopedic surgery to address degenerative arthritis or post-traumatic conditions of the patellofemoral joint. The device is manufactured from Oxinium and is available in multiple sizes (x-small, medium, large) for left and right configurations. It is implanted by a surgeon using bone cement. The trochlear groove geometry is derived from the previously cleared Genesis II Total Knee Femoral Component. By providing a smooth, artificial articulating surface, the implant aims to restore joint function, reduce pain, and correct deformity in patients who have failed conservative or prior surgical interventions.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material, and geometric comparisons to legally marketed predicate devices.
Technological Characteristics
Materials: Oxinium. Design: Semiconstrained, cemented prosthesis. Configuration: Available in x-small, medium, and large sizes, left and right. Trochlear groove geometry matches the Genesis II Total Knee Femoral Component (K951987).
Indications for Use
Indicated for patients requiring femoral side patello-femoral joint replacement due to degenerative arthritis of the distal femur and patella, history of patellar dislocation or fracture, or failed prior surgeries (arthroplasty, tibial tubercle elevation, lateral release) with persistent pain, deformity, or dysfunction.
Regulatory Classification
Identification
A knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis is a two-part device intended to be implanted to replace part of a knee joint in the treatment of primary patellofemoral arthritis or chondromalacia. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes a component made of alloys, such as cobalt-chromium-molybdenum or austenitic steel, for resurfacing the intercondylar groove (femoral sulcus) on the anterior aspect of the distal femur, and a patellar component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those devices intended for use with bone cement (§ 888.3027). The patellar component is designed to be implanted only with its femoral component.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Applications for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” and
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,”
(vii) ISO 7207-2:1998 “Implants for Surgery—Components for Partial and Total Knee Joint Prostheses—Part 2: Articulating Surfaces Made of Metal, Ceramic and Plastic Materials,” and
(viii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants,” and
(viii) F 1672-95 “Specification for Resurfacing Patellar Prosthesis.”
Predicate Devices
Howmedica Osteonics Corporation Avon Patello-Femoral Joint Prosthesis (K010100 and K041160)
Reference Devices
Smith & Nephew (formerly Richards Medical) Patella Type 1
Smith & Nephew Genesis II Total Knee Femoral Component (K951987)
Submission Summary (Full Text)
{0}------------------------------------------------
K051086 (pg 1 of 1)
### MAY 3 1 2005 510(K) Summary SMITH & NEPHEW PATELLO-FEMORAL KNEE IMPLANT
SUBMITTER'S NAME: SUBMITTER'S ADDRESS: SUBMITTER'S TELEPHONE NUMBER: CONTACT PERSON: DATE SUMMARY PREPARED: TRADE OR PROPRIETARY DEVICE NAME: COMMON OR USUAL NAME: CLASSIFICATION NAME:
DEVICE CLASS: PANEL CODE:
Smith & Nephew, Inc., Orthopaedic Division 1450 Brooks Road, Memphis, TN 38116 901-399-6707 Gino J. Rouss April 26, 2005 Smith & Nephew Patello-Femoral Implant Patello-Femoral Replacement Knee joint patellofemoral polymer/metal semiconstrained cemented prosthesis Class II Orthopedics/87 KRR
## DEVICE INFORMATION:
#### A. INTENDED USE:
The Smith & Nephew Patello-Femoral Implant is indicated for replacement of the femoral side of the patello-femoral joint. It is intended to be used in patellofemoral arthroplasty in patients with degenerative arthritis in the distal femur and patients with a history of patellar dislocation or patellar fracture, and patients with failed previous surgery (arthroplasty, tibial tubercle elevation, and lateral release) where pain, deformity or dysfunction persists. The components are for single use only and are intended for implantation with bone cement.
These indications are the same as found with the Howmedica Osteonics Avon" Patello-Femoral Joint Prosthesis cleared via K010100 and K041160.
#### B. DEVICE DESCRIPTION:
The Smith & Nephew Patello-Femoral Components are intended to replace the patellofemoral articulating surface of the femur. The components will be manufactured from Oxinium and will be initially manufactured in sizes x-small, medium, and large. Each size will be provided in both left and right configurations. The trochlear groove geometry is the same geometry as found with the Smith & Nephew Genesis II Total Knee Femoral Component subject of K951987.
#### SUBSTANTIAL EQUIVALENCE INFORMATION: C.
The Smith & Nephew Patello-Femoral Components are similar to the following commercially available devices regarding design features, overall indications, and materials:
- · Smith & Nephew (formerly Richards Medical) Patella Type 1 (preamendment)
- · Howmedica Osteonics Corporation Avon" Patello-Femoral Joint Prosthesis (K010100 and K041160)
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized representation of three human figures, depicted as abstract shapes with flowing lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 3 1 2005
Mr. Gino J. Rouss, M.S. Regulatory Affairs Specialist Smith & Nephew Incorporated Orthopaedic Division 1450 E. Brooks Road Memphis, Tennessee 38116
Re: -K051086
Trade/Device Name: Smith & Nephew Patello-Femoral Knee Implant Regulation Number: 21 CFR 888.3540 Regulation Name: Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis Regulatory Class: II Product Code: KRR Dated: April 26, 2005 Received: April 28, 2005
Dear Mr. Rouss:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bection of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the enclosale) (c region) the enactment date of the Medical Device Amendments, or to connineres prior to May 20, 1978, in encordance with the provisions of the Federal Food, Drug, devices that have been rochise in assessoval of a premarket approval application (PMA). and Costinette Ace (110.) market the device, subject to the general controls provisions of the Act. The r ou may, dicrerore, maines of the Act include requirements for annual registration, listing of general one of the resuring practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is clubsition (600 as controls. Existing major regulations affecting your device can may be subject to subject to submitted in the 21, Parts 800 to 898. In addition, FDA may be found in the Overnments concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Flease oe advised that I Dri 3 issuance or our device complies with other requirements of the Act that I DA has made a decerminations administered by other Federal agencies. You must or any it cacal statutes and regaraneers, including, but not limited to: registration and listing (21 Comply with an the Act 3 requirements, moreans, and manufacturing practice requirements as set CFN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Gino J. Rouss, M.S.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and wyou to obgen inding of substantial equivalence of your device to a legally premaince notification - nesults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrice for your 2009 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Miriam C. Provost, Ph.D.
Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# K051086 Indications for Use
510(k) Number (if known):
Device Name: Smith & Nephew Patello-Femoral Knee Implant
Indications for Use:
The Smith & Nephew Patello-Femoral Implant is indicated for replacement of the femoral side of the Shillin & Nephen Patello remoral to be used in patellofemoral arthroplasty in patients with:
- Degenerative arthritis in the distal femur and patella. 1.
- A history of patellar dislocation or patellar fracture. 2.
- Failed previous surgery (arthroplasty, tibial tubercle elevation, lateral release) where 3. pain, deformity or dysfunction persists.
The Smith & Nephew Patello-Femoral Implants are for single use only and are intended for implantation with bone cement.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X AND/OR |
|-------------------------------------------------|----------|
|-------------------------------------------------|----------|
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write Below This Line – Continue on another Page if Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hypt Church
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Division of General, Restorative, and Neurological Devices
510(k) Number Kos1086
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.