BIO-RAD VARIANT NBS SICKLE CELL PROGRAM

K051072 · Bio-Rad Laboratories, Inc. · GKA · May 12, 2005 · Hematology

Device Facts

Record IDK051072
Device NameBIO-RAD VARIANT NBS SICKLE CELL PROGRAM
ApplicantBio-Rad Laboratories, Inc.
Product CodeGKA · Hematology
Decision DateMay 12, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7415
Device ClassClass 2
AttributesReal-World Evidence, Pediatric, 3rd-Party Reviewed

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K051072 · May 12, 2005BIO-RAD VARIANT NBS SICKLE CELL PROGRAMBio-Rad Laboratories, Inc.Retrospective neonatal dried blood spot collection cardsRetrospective clinical samples were used to perform a method comparison study between the new device and the predicate device to establish substantial equivalence.Method comparison; Retrospective clinical samples; Neonatal dried blood spots

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective method comparisonNeonates; Sample Size: 1025Bio-Rad VARIANT™ Sickle Cell Short Program (K924813)Agreement in hemoglobin identification (F, A, S, D, C, E)

Indications for Use

The Bio-Rad VARIANT™ nbs Sickle Cell Program is intended as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high performance liquid chromatography (HPLC). The Bio-Rad VARIANT™ nbs Sickle Cell Program is intended for Professional Use Only. For In Vitro Diagnostic Use. The Bio-Rad VARIANT™ nbs Sickle Cell Program is intended for use only with the Bio-Rad VARIANT™ nbs Newborn Screening System.

Device Story

System uses HPLC to analyze neonatal dried blood spot samples; samples punched from filter paper into microwell plates; auto-sampler aspirates eluate; analytical cartridge with cation exchange resin separates hemoglobins via gradient elution (two buffers); dual-wavelength photometer (415 nm/690 nm) detects hemoglobin absorbance. GDM software processes data; generates reports with sample ID, retention times, peak heights/areas, and chromatograms. Optional pattern assignment feature applies literature-derived rules to identify hemoglobin profiles. Used in clinical laboratories by professional staff; output assists clinicians in screening for hemoglobinopathies. Benefits include automated, high-throughput qualitative screening for sickle cell and other hemoglobin variants in newborns.

Clinical Evidence

Bench testing only. Precision evaluated per NCCLS EP-05A/EP-05A2; within-run precision <1% for all hemoglobin peaks. Limit of detection for variants (S, D, C, E) confirmed at 1% peak area. Method comparison performed on 1025 retrospective neonatal dried blood spot samples; 99.8% agreement with predicate device. No clinical studies performed.

Technological Characteristics

Cation exchange HPLC; dual-wavelength photometer (415/690 nm); analytical cartridge with cation exchange resin; auto-sampler for 96-well plates. Standards met: EN375, EN591, EN980, EN1658, EN13485, EN13640, EN14971, EN61010-1, EN61010-2-101, EN61326, NCCLS EP-05A/EP-05A2. Connectivity: PC workstation. Power: 110/220V.

Indications for Use

Indicated for qualitative screening of hemoglobins F, A, S, D, C, and E in neonates using dried blood spot samples collected on filter paper.

Regulatory Classification

Identification

An abnormal hemoglobin assay is a device consisting of the reagents, apparatus, instrumentation, and controls necessary to isolate and identify abnormal genetically determined hemoglobin types.

Special Controls

*Classification.* Class II (special controls). A control intended for use with an abnormal hemoglobin assay is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K051072 B. Purpose for Submission: The Bio-Rad VARIANT™ nbs Sickle Cell Program system/device provides improvements over the a predicate chromatographic devices known as the: VARIANT™ Sickle Cell Short Program [K924813; cleared 01/14/93]. C. Analyte: Hemoglobins F, A, S, D, C and E D. Type of Test: IVD Qualitative, Hemoglobin HPLC. E. Applicant: Bio-Rad Laboratories, Inc., Clinical Systems Division 4000 Alfred Nobel Drive Hercules, CA, U.S.A, 94547-1803 F. Proprietary and Established Names: Bio-Rad VARIANT™ nbs Sickle Cell Program Hemoglobin variants determination by HPLC G. Regulatory Information: 1. Regulation section: 21 CFR 864.7415 [Abnormal Hemoglobin Assay] 2. Classification: Class II 3. Product Code: GKA 4. Panel: Hematology (81) H. Intended Use: 1. Intended use(s): The Bio-Rad VARIANT™ nbs Sickle Cell Program is intended as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high performance liquid chromatography (HPLC). The Bio-Rad VARIANT™ nbs Sickle Cell Program is intended for Professional Use Only. For In Vitro Diagnostic Use. The Bio-Rad VARIANT™ nbs Sickle Cell Program is intended for use only with the Bio-Rad VARIANT™ nbs Newborn Screening System. {1} Page 2 of 10 2. Indication(s) for use: This device, consisting of the reagents, apparatus, HPLC instrumentation, software and controls, is indicated for professional laboratory IVD use to isolate and identify genetically determined abnormal (S, D, C, E) and normal (F, A) hemoglobin types in neonatal blood samples. 3. Special condition for use statement(s): For professional IVD use only in clinical laboratory. 4. Special instrument Requirements: For Bio-Rad VARIANTnbs Newborn Screening System. I. Device Description: This complete device consists of the Bio-Rad VARIANT™ nbs Sickle Cell Program (VnbsSCP) reagent kit, the Bio-Rad VARIANT™ nbs Newborn Screening System (VNBSS) instrument, and the Bio-Rad Genetic Data Management Software (GDM). The VNBSS instrument consists of a VARIANTnbs Neonatal Auto Sampler (VNAS) module for microwell plates and a VARIANTnbs Neonatal Chromatography Station (VNCS) module containing the high performance liquid chromatography (HPLC) hardware. The VnbsSCP reagent kit includes a specific analytical HPLC cartridge containing cation exchange resin, as well as two (2) buffer reagents for establishing an HPLC gradient. The GDM software is designed to execute the VnbsSCP assay protocol on the VNBSS instrument using the VnbsSCP reagent kit components for the purposes of qualitatively screening for the presence of normal hemoglobins F and A, as well as the abnormal hemoglobins S, D, C and E from neonatal heel stick blood, as collected on filter paper that is punched and eluted with deionized water. The VNBSS processes each sample individually. An eluted sample is aspirated directly from a microwell plate in the VNAS module with the punched filter paper disc still present, and transferred into the sample loop in the VNCS module. The contents of the sample loop are subsequently injected into the flow path of the VNCS module. The hemoglobins of interest are retained on the analytical cartridge in the presence of Elution Buffer 1. The ionic strength is subsequently raised by adding increasing amounts of Elution Buffer 2. The pre-programmed gradient is designed to have the hemoglobins of interest elute from the cartridge with retention times that fall within pre-determined windows characteristic of known normal and abnormal hemoglobins. The eluted hemoglobins are sequentially detected with a dual-wavelength filter photometer, which monitors hemoglobin absorbance at 415 nm and corrects for any gradient induced absorbance changes at 690 nm. The software processed HPLC data is outputted in a printed report that contains: 1) sample identification, 2) date and time of analysis, 3) a table of peaks that includes: observed peak identification (hemoglobin type) name(s), retention time(s), peak height(s), peak area(s), and relative area percent(s), 4) total chromatogram area, 5) complete chromatographic display and 6) any error message(s) relating (if needed) to such data. Also reported is an optional "pattern assignment" for each hemoglobin based upon "pattern rules" derived from diagnostic hemoglobin literature. {2} Page 3 of 10 # J. Substantial Equivalence Information: 1. Predicate device name(s): Bio-Rad VARIANT™ Sickle Cell Short Program 2. Predicate K number(s): K924813 [cleared: 01/14/1993] 3. Comparison with predicate: | Summary of Technological Characteristic - Similarities to Predicate Device | | | | --- | --- | --- | | Features | New Device: Bio-Rad VARIANT™nbs Sickle Cell Program | Predicate Device: Bio-Rad VARIANT™ Sickle Cell Short Program (K#924813) | | Intended Use | The Bio-Rad VARIANTnbs Sickle Cell Program is intended as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high performance liquid chromatography (HPLC). | The VARIANT Sickle Cell Short Program is designed as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high performance liquid chromatography. | | | For In Vitro Diagnostic Use. | For In Vitro Diagnostic Use. | | | For Professional Use Only. | For Professional Use Only. | | Target Population | Neonates. | Neonates. | | Design – Assay principle | Cation exchange high performance liquid chromatography. | Cation exchange high performance liquid chromatography. | | Design – Assay Detection | Heme absorbance at 415 nm with background correction at 650 nm. | Heme absorbance at 415 nm with background correction at 650 nm. | | Design – Analytes Identified | Six retention time windows for hemoglobins F, A, E, D, S and C. | Six retention time windows for hemoglobins F, A, E, D, S and C. | | Design – Sample Type | Neonatal dried blood spots on filter paper collection cards. | Neonatal dried blood spots on filter paper collection cards. | | Design – Punched Disc | One 1/8” disc. | One 1/8” disc. | | Design – Manual Worklists | Accepts manual worklists. | Accepts manual worklists. | | Materials - Components | Elution Buffer 1. Elution Buffer 2. Wash Solution. Analytical Cartridge. Lyophilized Whole Blood Primer. Lyophilized Retention Time Marker 1 (FAES). Lyophilized Retention Time Marker 2 (FADC). | Elution Buffer 1. Elution Buffer 2. Wash Solution. Analytical Cartridge. Lyophilized Whole Blood Primer. Lyophilized Retention Time Marker 1 (FAES). Lyophilized Retention Time Marker 2 (FADC). | | Performance – Precision | Peak retention time precision is <1% for all hemoglobin peaks. | Peak retention time precision is <1% for all hemoglobin peaks. | | Compatibility with Environment | U.S. FCC EMI and E.U. EMC standard compliant. | U.S. FCC EMI and E.U. EMC standard compliant. | | Human Factors | For in vitro diagnostic use. For professional use only. | For in vitro diagnostic use. For professional use only. | | Energy Used | Auto-switching 110 V and 220 V. | 110 V and 220 V models. | {3} Page 4 of 10 | Summary of Technological Characteristic - Similarities to Predicate Device | | | | --- | --- | --- | | Features | New Device: Bio-Rad VARIANT™nbs Sickle Cell Program | Predicate Device: Bio-Rad VARIANT™ Sickle Cell Short Program (K#924813) | | Chemical Safety | Sodium azide concentration <0.05%. Gentamicin Sulfate concentration <0.1%. Tobramycin concentration <0.1%. Warnings provided in labeling as required, including State of California Proposition 65 Warning. | Sodium azide concentration <0.05%. Gentamicin Sulfate concentration <0.1%. Tobramycin concentration <0.1%. Warnings provided in labeling as required, including State of California Proposition 65 Warning. | | Electrical, Mechanical and Thermal Safety | System certification to US and Canadian product safety standards and EU low voltage safety standards. | System certification to US and Canadian product safety standards and EU low voltage safety standards. | | Standards Met | • EN375:2002 • EN591:2001 • EN980:2003 • EN1658:1996 • EN13485:2003 • EN13640:2002 • EN14971:2001 • EN61010-1:2001 • EN61010-2-101:2002 • EN61326:2001 | • EN375:2002 • EN591:2001 • EN980:2003 • EN1658:1996 • EN13485:2003 • EN13640:2002 • EN14971:2001 • EN61010-1:2001 • EN61010-2-101:2002 • EN61326:2001 | | Summary of Technological Characteristic - Differences to Predicate Device | | | | --- | --- | --- | | Features | New Device: Bio-Rad VARIANT™nbs Sickle Cell Program | Predicate Device: Bio-Rad VARIANT™ Sickle Cell Short Program (K#924813) | | Design – System Configuration | Separate chromatography and auto sampler modules and separate PC workstation with software. | Single integrated unit with chromatography, auto sampler and software functionalities. | | Design – Media | CD-ROM. | ROM Card. | | Design – Container, Elution Volume, Reconstitution Volume, Sample Loop, Column loading | Plastic 96 microwell plate. 250 μL sample elution volume. 500 μL primer and retention time marker reconstitution volume. 10 μL sample loop. Column loading is 1/25 of eluted sample volume and 1/50 of reconstituted material. | Plastic sample vial. 500 μL sample elution volume. 1000 μL primer and retention time marker reconstitution volume. 20 μL sample loop. Column loading is 1/25 of eluted sample volume and 1/50 of reconstituted material. | | Design – Aspiration Probe Tip, Punched Disc Disposition | Beveled aspiration probe tip. Dried blood spot punched disc left in microwell during sample aspiration. | Blunt aspiration probe tip. Dried blood spot punched disc removed before sample aspiration. | | Design – Additional retention time windows. | Seven (7) additional retention time windows: F1, “Other (1)”, “Other (2)”, “Other (3)”, “Other (4)”, “Other (5)” and “Other (6)”. | Feature not available. | | Design – Automated Worklists | Accepts automated worklists from spot punchers. | Feature not available. | | Design – Pattern Assignment | Optional pattern assignment feature uses pattern rules derived from literature. | Feature not available. | {4} Page 5 of 10 | Summary of Technological Characteristic - Differences to Predicate Device | | | | --- | --- | --- | | Features | New Device: Bio-Rad VARIANT™nbs Sickle Cell Program | Predicate Device: Bio-Rad VARIANT™ Sickle Cell Short Program (K#924813) | | Performance – Variants Limit of Detection | The limit of detection for S, D, C and E is 1% of the total area of the sample when the total area is 1.5 million microvolt·second. | Limit of detection for hemoglobins S, D, C and E is 1% of the total area when the total area of the sample is 1.0 million microvolt·second. | | Performance – Guideline for Interpretation of Results | Total area must be between 900,000 to 6.3 million microvolt·second. | Total area should range from 1,000,000–3,000,000 microvolt·second. | | Performance – Total Area Limit of Detection | 900,000 microvolt·second. | Not addressed in instruction manual. | | Performance – Bilirubin interference | Bilirubin up to 20 mg/dL does not interfere. | Not addressed in instruction manual. | | Performance – Triglyceride interference | Triglyceride up to 6000 mg/dL does not interfere. | Not addressed in instruction manual. | | Performance - Eluate stability | Eluates are stable for 48 hrs on the cooled auto sampler and at 2-8 °C and stable for 24 hrs at 15-30 °C. | Eluates are stable for 24 hrs at 2-8 °C. | | Guidances Met | FDA 2002 - General Principles of Software Validation; Final Guidance for Industry and FDA Staff. FDA 1999 - Guidance for Off-the-Shelf Software Use in Medical Devices; Final. FDA 1998 - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices; Final. FDA 2003 – Device Advice; Content of a 510(k) | Not addressed. | | Standards Met | NCCLS EP-05A 1999. NCCLS EP-05A2 2004. | Not addressed. | K. Standard/Guidance Document Referenced (if applicable): | Standards Met | EN375:2002 - Information supplied by the manufacturer with in vitro diagnostic reagents for professional use. EN591:2001 – Instructions for use in vitro diagnostic instruments for professional use. EN980:2003 - Graphical symbols for use in the labeling of medical devices. EN1658:1996 - Requirements for marking of in vitro diagnostic instruments. EN13485:2003 - Quality systems - Medical devices - Particular requirements for the application of EN ISO 9001:1994. EN13640:2002 - Stability testing of in vitro diagnostic medical devices. EN14971:2001 - Medical devices – Application of risk management to medical devices. EN61010-1:2001 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements EN61010-2-101:2002 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment. EN61326:2001 - Electrical equipment for measurement, control and laboratory use - | | --- | --- | {5} Page 6 of 10 | | EMC requirements • NCCLS EP05A 1999 – Evaluation of Precision Performance of Clinical Chemistry Devices. • NCCLS EP05A2 2004 – Evaluation of Precision Performance of Quantitative Measurement Methods. | | --- | --- | | Guidances Met | • FDA 2002 - General Principles of Software Validation; Final Guidance for Industry and FDA Staff. • FDA 1999 - Guidance for Off-the-Shelf Software Use in Medical Devices; Final. • FDA 1998 - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices; Final. • FDA 2003 – Device Advice; Content of a 510(k) | ## L. Test Principle: Cation exchange high performance liquid chromatography with visible light detection. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: The Bio-Rad VARIANT™ Sickle Cell Program device was based on NCCLS Protocol EP-05A (Vol. 19, No. 2 [1999]) and EP-05A2 (Vol. 24, No. 25 [2004]). Two analytical runs were performed per day on 20 days for a total of 40 runs on each of 3 separate systems. Each run included 4 replicates of two retention time positional QC controls. Within-run precision and within-device precision (formerly total precision) were determined. The reported predicate device retention time precision protocol included the same two retention time positional QC controls. However, the protocol did not strictly conform to NCCLS Protocol EP5-A2 guidelines, nor was within laboratory (or within-device) precision reported, as this was strictly a qualitative assay. The results of within run precision for both predicate and new devices was less than 1% for Hemoglobins F, A, E, D, S and C, and are presented in the following tables: | Average Retention Time Summary | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | RT Window | 0.59-0.71 | 0.79-0.89 | 0.95-1.03 | 1.03-1.13 | 1.15-1.25 | 1.63-1.77 | | | | | Retention Time Within-Run Precision (CV %) | | | | | | | Device | Replicates | Sample | Peak F | Peak A | Peak E | Peak D | Peak S | Peak C | | New System 1 | 160 | QC Control 1 | 0.63 | 0.83 | 0.98 | | 1.21 | | | New System 2 | 160 | QC Control 1 | 0.63 | 0.83 | 0.98 | | 1.21 | | | New System 3 | 160 | QC Control 1 | 0.63 | 0.83 | 0.98 | | 1.20 | | | New System 1 | 160 | QC Control 2 | 0.63 | 0.83 | | 1.08 | | 1.70 | | New System 2 | 160 | QC Control 2 | 0.63 | 0.83 | | 1.08 | | 1.70 | | New System 3 | 160 | QC Control 2 | 0.63 | 0.83 | | 1.08 | | 1.70 | {6} Page 7 of 10 | Retention Time Within Run Precision Summary | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Device | Replicates | Sample | Retention Time Within-Run Precision (CV%) | | | | | | | | | | Peak F | Peak A | Peak E | Peak D | Peak S | Peak C | | New System 1 | 160 | QC Control 1 | 0.3 | 0.3 | 0.4 | | 0.3 | | | New System 2 | 160 | QC Control 1 | 0.5 | 0.4 | 0.3 | | 0.3 | | | New System 3 | 160 | QC Control 1 | 0.4 | 0.4 | 0.3 | | 0.2 | | | Predicate | 10 | QC Control 1 | 0.7 | 0.0 | 0.0 | | 0.0 | | | New System 1 | 160 | QC Control 2 | 0.3 | 0.3 | | 0.2 | | 0.3 | | New System 2 | 160 | QC Control 2 | 0.5 | 0.4 | | 0.3 | | 0.2 | | New System 3 | 160 | QC Control 2 | 0.5 | 0.3 | | 0.3 | | 0.1 | | Predicate | 10 | QC Control 2 | 0.6 | 0.0 | | 0.4 | | 0.3 | | Retention Time Within Device Precision (Formerly Total Precision) | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Device | Replicates | Sample | Retention Time Within-Run Precision (CV %) | | | | | | | | | | Peak F | Peak A | Peak E | Peak D | Peak S | Peak C | | New System 1 | 160 | QC Control 1 | 0.3 | 0.4 | 0.5 | | 0.5 | | | New System 2 | 160 | QC Control 1 | 0.6 | 0.6 | 0.4 | | 0.6 | | | New System 3 | 160 | QC Control 1 | 0.6 | 0.6 | 0.6 | | 0.5 | | | Predicate | Not Reported | | | | | | | | | New System 1 | 160 | QC Control 2 | 0.3 | 0.4 | | 0.4 | | 0.2 | | New System 2 | 160 | QC Control 2 | 0.7 | 0.6 | | 0.4 | | 0.3 | | New System 3 | 160 | QC Control 2 | 0.6 | 0.6 | | 0.5 | | 0.3 | | Predicate | Not Reported | | | | | | | | {7} Page 8 of 10 b. Linearity/assay reportable range: Refer to method correlation results in Section 2a, c. Traceability (controls, calibrators, or method): Not applicable d. Detection limit: The Bio-Rad VARIANT™ Sickle Cell Program new device peak area limit of detection for hemoglobin variants (E, D, S and C) was determined using a total of 102 sample measurements bracketing the 1% peak area limit of detection for each variant, when the total chromatogram area was 1.5 million microvolt x second (= μvolts·sec). The reported predicate device peak area limit of detection was 1%, when the total chromatogram area was 1.0 million μvolts·sec. The results re: peak area limit of detection for hemoglobin variants (E, D, S and C) was 1% for both predicate and new device, and is documented in the following table: | Variant Peak Area – Limit of Detection | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Device | Chromatogram Total Area (microvolt-second) | Peak Area % Limit of Detection | | | | | | | | Peak E | Peak D | Peak S | Peak C | | | New System 1 | 1.5 | 1% | 1% | 1% | 1% | | | Predicate | 1.0 | 1% | 1% | 1% | 1% | | e. Analytical specificity: Refer to method correlation results in Section 2a. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Method correlation between predicate and -Rad VARIANT™nbs Sickle Cell Program new device was evaluated using 1025 unknown samples prepared from retrospective neonatal dried blood spot collection cards. Using a total area range of 0.90–6.30 million μvolts·sec, there was 99.8% (1023/1025) agreement between test and predicate devices for all sample hemoglobin identifications. There was 100.0% (250/250) agreement between test and predicate devices for identifying hemoglobin S (HbS) in those samples that predicate device identified as containing HbS. In the two disagreements, the new device identified all hemoglobins identified by the predicate as well as one additional peak in each case. [See explanation below.] These results indicate satisfactory correlation between the predicate and test devices. The results are presented in the following tables {8} Page 9 of 10 | Hemoglobin Identification Correlation Summary | | | | | | --- | --- | --- | --- | --- | | | Number of Samples | Predicate Device | New device | | | | | F, A, E, D, S and (or) C Identified | Agree | Disagree | | | 591 | FA | 590 | 1 (FAS) | | | 52 | FAE | 52 | 0 | | | 38 | FAD | 37 | 1 (FADC) | | | 247 | FAS | 247 | 0 | | | 90 | FAC | 90 | 0 | | | 3 | FSC | 3 | 0 | | | 2 | FC | 2 | 0 | | | 1 | FE | 1 | 0 | | | 1 | F | 1 | 0 | | Total 1025 | | | 1023 | 2 | In one case of disagreement the predicate device reported FA with an "unknown" peak in the S window while the test device reported FAS. Independent iso-electric focusing results identified FAS in support of the test device results. | FA and FAS | Test Device | | | | --- | --- | --- | --- | | | | FA | FAS | | Predicate Device | FA | 591 | 1 | | | FAS | 0 | 247 | In the other case of disagreement the predicate device reported FAD while the test device reported FADC with a $2.6\%$ peak area of HbC. Inspection of the test device chromatogram showed a low broad background in the C window without a resolved peak suggesting baseline noise. | FAD and FADC | Test Device | | | | --- | --- | --- | --- | | | | FAD | FADC | | Predicate Device | FAD | 37 | 1 | | | FADC | 0 | 0 | 2. Comparison studies (continued): b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence [SE] decision. {9} Page 10 of 10
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...