FLUOROSCAN INSIGHT MINI C-ARM FLUOROSCOPIC IMAGING SYSTEM
K051025 · Hologic, Inc. · JAA · May 13, 2005 · Radiology
Device Facts
Record ID
K051025
Device Name
FLUOROSCAN INSIGHT MINI C-ARM FLUOROSCOPIC IMAGING SYSTEM
Applicant
Hologic, Inc.
Product Code
JAA · Radiology
Decision Date
May 13, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations.
Device Story
Mini C-Arm fluoroscopic imaging system; provides real-time visualization of patient anatomy. Inputs: X-ray radiation passed through patient anatomy; captured by image intensifier. Transformation: X-ray signals converted to digital images via digital camera; processed by system computer (2.4 GHz processor, 512 MB RAM). Output: Digital fluoroscopic images displayed on medical-grade 2 Mega Pixel 19.8" monochrome flat LCD. Used in OR, clinic, or emergency settings; operated by physicians/technicians. Healthcare providers use images for intraoperative guidance and diagnostic visualization; facilitates surgical procedures and clinical decision-making. Benefits: Real-time imaging for orthopedic/podiatric interventions.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and technological comparison.
Indicated for general fluoroscopic visualization of patients, including surgical orthopedic, podiatry, critical care, emergency care, and light anatomy imaging procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
### F. 510(k) Summary
## F.1 Manufacturing Establishment and Contact Information
F.1.1 Manufacturer Name and Address:
Hologic, Inc. 35 Crosby Drive Bedford, MA 01730
F.1.2 Establishment Registration Number:
1221300
F.1.3 Name, Title, and Telephone Number of Contact:
Daniel F. Phelan Senior Regulatory Affairs Specialist (781) 999-7300
#### F.2 Device Identification
- F.2.1 Device Trade Name:
Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System
- F.2.2 Common / Usual Name:
Fluoroscopic Imaging System
- F.2.3 Proposed Intended Use:
The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations.
### F.3 Device Classification
- F.3.1 Classification:
Class II
- F.3.2 Classification Name and Rule:
Image-Intensified Fluoroscopic X-Ray System, 21 CFR 892.1650
- F.3.3 Classification Panel:
Radiology
- F.3.4 Product Code:
90 JAA
- F.3.5 Predicate Device
510(k) No.: K974058
Trade Name: FS IV Mini C- Arm System
January 12, 1998. SE Date:
Manufacturer: Hologic, Inc.
{1}------------------------------------------------
-
## F.4 Substantial Equivalence
| | Fluoroscan Premier(FS-IV) | Fluoroscan InSight |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| | Image: Fluoroscan Premier(FS-IV) | Image: Fluoroscan InSight |
| 510(k) Number | K974058 | Same |
| Indications for Use | Mini C-Arm Fluoroscopic Imaging System<br>designed to provide physicians with general<br>fluoroscopic visualization of a patient,<br>including, but not limited to surgical<br>orthopedic and podiatry use, critical and<br>emergency care procedures, and light<br>anatomy imaging situations | Same |
| Dimensions (HxWxD) | 60" (152cm) x 36" (91cm) x 32" (81cm) | 65" (165.1cm) x 35" (88.9cm) x 35" (88.9cm) |
| Weight (maximum) | 500 lbs (227 kg) | 650 lbs (295.46 kg) |
| Voltage (nominal) | 100/120/220/240 Single Phase | Same |
| Frequency | 50/60 Hz | Same |
| Source-Image receptor<br>distance | 17.5" (44cm) | Same |
| X-Ray Source | Grounded anode x-ray tube 0.005" (0.127<br>mm) Beryllium | Same |
| Beam Filtration | Stainless Steel with Al equivalence ><br>2.5mm | Same |
| Focal Spot | 0.0018" (0.045mm) @ 7.5 watts | Same |
| Field of View | Operator selectable, 4" or 6" collimation<br>(10.16 cm or 15.24 cm) at plane of Image<br>Intensifier | Same |
| Rated peak tube<br>potential | 75kVp | Same |
| Tube kVp range | 40 to 75kVp | Same |
| Tolerances | kVp ± 5%<br>mA ± 8% ≤ 0.035mA<br>mA ± 5% ≥ 0.035mA | Same |
| Maximum duty cycle<br>at 75 kVp/0.100mA | 50% | Same |
| Video Camera | Analog | Digital |
| System computer | 850 MHz processor | 2.4 GHz processor |
| minimum requirements | 128 MB RAM | 512 MB RAM |
| | CD-ROM | Serial 80 GB HDD |
| | 20 GB HDD | DVD RAM drive |
| | 3.5" Floppy | 3.5" Floppy drive |
| Operating System | Windows NT Embedded | Windows XP |
| Display | Dual 15" Hi-Res CRT | Medical grade 2 Mega Pixel 19.8" Monochrome<br>Flat LCD |
| Input Devices | Keyboard, Footswitch, Touch Screen | Same |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized bird in the center. The text around the perimeter appears to be the name of a department or organization. The bird is depicted with three curved lines representing its body and wings, giving it a sense of motion. The logo has a simple, clean design, likely intended to represent the organization's values or mission.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 13 2005
Mr. Daniel F. Phelan Senior Regulatory Affairs Specialist HOLOGIC, Inc. 35 Crosby Drive BEDFORD MA 01730
Re: K051025
Trade/Device Name: Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System Regulation Number: 21 CFR §892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA r K Dated: April 21, 2005 Received: April 25, 2005
Dear Mr. Phelan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the r carrar states and regarding, but not limited to registration and listing (21 CFR Part 807); labeling / 21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 801); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will anow you to ogin hances of substantial equivalence of your device to a legally premarket notification. The I Dri mailig of oation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de restled in the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, prease note the regulation entersely formation on your responsibilities under the Act from the 007.97). Tou may obtain outer general missance at its toll-free number (800) DIVISION of Stilan Manufactures, Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## A.2 Indications for Use Statement
510(k) Number (if known):K0510 25 ____________________________________________________________________________________________________________________________________________
Device Name: _Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System
Indications for Use:
The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide physicians TTle Fluoroscan frisualization of a patient, including, but not limited to surgical orthopedic and with general hoorosopic visualization of a patient, wis and light anatomy imaging situations.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter-Use
(Per 21 CFR 801.109)
(Optional Format 1)
Nancy Beaudoin
(Division Sign Division of Reproductive, Ab and Radiological Device 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.