K051002 · Ethicon Endo-Surgery, Inc. · GDW · May 17, 2005 · General, Plastic Surgery
Device Facts
Record ID
K051002
Device Name
ENDOPATH ENDOCUTTER 60 ENDOSCOPIC LINEAR CUTTER
Applicant
Ethicon Endo-Surgery, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
May 17, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ENDOPATH Endocutter 60 Endoscopic Linear Cutter is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials.
Device Story
ENDOPATH Endocutter 60 is a sterile, single-patient-use mechanical surgical stapler; used in open or endoscopic procedures. Device delivers staples while simultaneously dividing tissue. Features include 60mm cartridge design, enhanced compression system, rigid anvil, E-beam, and wide jaw aperture for tissue manipulation. Available in three shaft lengths (standard, compact, long). Cartridges provided in four sizes (white, blue, gold, green) corresponding to tissue thickness. Operated manually by surgeon; one-handed design allows clamping, firing, and automatic knife return. Provides consistent staple line formation; facilitates tissue transection and anastomosis. Benefits include improved tissue positioning and ergonomic handling during surgery.
Clinical Evidence
Bench testing and preclinical laboratory evaluations were performed to demonstrate that the device will perform as intended. No clinical data provided.
Technological Characteristics
Mechanical surgical stapler; 60mm cartridge design; enhanced compression system; rigid anvil and E-beam; wide jaw aperture; ergonomic one-handed operation with automatic knife return. Available in three shaft lengths. Four cartridge sizes (white, blue, gold, green) for varying tissue thicknesses. Gold cartridge features longer staple leg length for thicker tissue. Sterile, single-patient-use.
Indications for Use
Indicated for transection, resection, and creation of anastomoses in general, gynecologic, thoracic, and pediatric surgical procedures; applicable for both open and minimally invasive surgery; compatible with staple line or tissue buttressing materials.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
ENDOPATH Linear Cutters and Staplers
Submission Summary (Full Text)
{0}------------------------------------------------
Ko5/002
Ethicon Endo-Surgery, Inc. 510(k) Premarket Notification ENDOPATH Endocutter 60 Endoscopic Linear Cutter
## 510(k) Summary
| Company | Ethicon Endo-Surgery, Inc. |
|----------------------|----------------------------|
| 4545 Creek Road | |
| Cincinnati, OH 45242 | |
Georgia C. Abernathy, MBA, RAC Contact Senior Regulatory Affairs Associate Telephone: (513) 337-3179 (513) 337-1444 Fax: --Email: gabernat@eesus.jnj.com
Date Prepared April 19, 2005
Trade Name: ENDOPATH® Endocutter 60 Endoscopic Linear Cutter Device Name Common or Usual Name: Linear Cutter Classification Name: Endoscope and Accessories and Implantable Staples [21 CFR 876.1500 (KOG) and 21 CFR 878.4750 (GDW)]
ENDOPATH® Linear Cutters and Staplers Predicate Device
These instruments are all mechanical surgical stapling devices. The Device Description ENDOPATH Endocutter 60 Endoscopic Linear Cutter models are sterile, single patient use instruments that deliver staples while simultaneously dividing tissue between rows. These instruments may be used in either open or endoscopic procedures. The devices will be available in three shaft lengths: standard, compact and long, and will be furnished without a cartridge. Cartridges for the Endocutter 60 devices will be available in sizes that correspond to the thickness of tissue that is to be transected or resected: white (vascular/thin), blue (standard), gold (thick, less dense) and green (thick).
Indications for Use The ENDOPATH Endocutter 60 Endoscopic Linear Cutter is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials.
Technological Characteristics The device incorporates a new 60mm cartridge design; new enhanced compression system; rigid anvil and E-beam with robust, one-piece design; wide jaw aperture for easy tissue positioning and manipulating; and ergonomic features for onehanded use (ability to clamp, fire and release with automatic knife return). The ENDOPATH Endocutter 60 device will be available in three shaft lengths: standard, compact and long. Compatible cartridges come in four sizes to accommodate various tissue thicknesses. The gold cartridge is an additional size added to the product line. The gold cartridge staple wire diameter is identical to the existing blue cartridge staple wire diameter, with a longer staple leg length for thicker indicated tissue.
Performance Data Bench testing and preclinical laboratory evaluations were performed to demonstrate that the device will perform as intended.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA", which is arranged in a circular fashion around the top of the logo. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 7 2005
Ms. Georgia C. Abernathy, MBA, RAC Senior Regulatory Affairs Associate Ethicon Endo-Surgery Incorporated 4545 Creek Road Cincinnati, Ohio 45242-2839
Re: K051002
Trade/Device Name: ENDOPATH® Endocutter60 Endoscopic Linear Cutter Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: April 19, 2005 Received: April 20, 2005
Dear Ms. Abernathy:
We have reviewed your Section 510(k) premarket notification of intent to market the device w t nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent of the enactment date of the Medical Device Amendments, or to conniner of they 2011-12-11 in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dierere, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may or subject to back additions, Title 21, Parts 800 to 898. In addition, FDA can be found in ther announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I DIT has made a and regulations administered by other Federal agencies. You must or any I ederal statutes and registents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IT rat 0077, laooiing (21 CFR Part 820); and if applicable, the electronic forth in the qualis) 35001 provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
## Page 2 - Ms. Georgia C. Abernathy, MBA, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will anow you to oegin maneting your with a equivalence of your device to a legally prematics notification: "The PDF intentig stification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your act (240) 276-0115 . Also, please note the regulation entitled, Colliact the Office of Comphalled as (21 transmit (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Kos/oc2
Ethicon Endo-Surgery, Inc. Ethicon Endo-Surgery, The.
510(k) Premarket Notification ENDOPATH Endoeutter 60 Endoscopic Linear Cutter
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
ENDOPATH® Endocutter 60 Endoscopic Linear Cutter Device Name:
Indications for Use:
The ENDOPATH Endocutter 60 Endoscopic Linear Cutter is intended for transection, The END OF FITT Entron of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, thoracic, and pediatric open of millinary in were general with staple line or tissue buttressing materials.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
.
Ka51062
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.