AWQ-104

K050989 · Lhasa Oms, Inc. · GZJ · Apr 6, 2006 · Neurology

Device Facts

Record IDK050989
Device NameAWQ-104
ApplicantLhasa Oms, Inc.
Product CodeGZJ · Neurology
Decision DateApr 6, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AWQ-104 is intended for use in the symptomatic relief of chronic intractable pain, post traumatic acute pain, and post surgical pain. It can be used by a physician on his patients. This is a prescription device and should be used under continued medical supervision.

Device Story

AWQ-104 is a battery-powered, four-channel transcutaneous electrical nerve stimulator (TENS). Device delivers square waveform pulses to patient skin via self-adhesive electrodes; includes hand-held probe for site-specific treatment and grounding pole. Operated by physician under continued medical supervision for symptomatic pain relief. Controls allow adjustment of output polarity, frequency, intensity, and voltage. Device does not provide curative treatment; suppresses pain signals. Clinical utility involves symptomatic management of chronic intractable, post-traumatic, or post-surgical pain.

Clinical Evidence

Bench testing only. Device conforms to ISO 14971:2000, EN ISO 10993 series (biocompatibility), and IEC 60601 series (electrical safety and electromagnetic compatibility).

Technological Characteristics

Battery-powered (9V) TENS device; four-channel output; square waveform pulses. Includes hand-held probe and grounding pole. Electrodes (K946230) are self-adhesive. Conforms to IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for symptomatic relief of chronic intractable pain, post-traumatic acute pain, and post-surgical pain. Contraindicated for use transcerebrally, in the carotid sinus area, during pregnancy, or in patients with demand cardiac pacemakers. Not recommended for patients who find electrical stimulation sensation unpleasant.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # F. 510(k) Summary K050989 Submitters information: Lhasa OMS, Inc. 230 Libbey Parkway Weymouth, MA 02189 Contact Person: Mark W. Sheehan Telephone: 781-340-1072 ext. 20 Fax: 781-340-1637 Date Summary Prepared: April 3, 2006 Device name: Proprietary name: Common or usual name: Classification name: AWQ-104 transcutaneous electrical nerve stimulator (TENS) Transcutaneous electrical nerve stimulator, Class II, 21 CFR 882.5890. Legally marketed device for substantial equivalence comparison: TX-3 TENS, K913532 ### Description of the device: The device consists of a battery powered instrument with four channel outputs. A hand held probe allows treatment at individual sites. Output polarity, frequency, intensity, and voltage are controlled by four independent channels. Each channel drives a pair of self adhesive electrodes. A grounding pole electrically grounds the device when using the hand held probe. #### Intended use of the device: The AWQ-104 is intended for use in the symptomatic relief of chronic intractable pain, post traumatic acute pain, and post surgical pain. It can be used by a physician on his patients. This is a prescription device and should be used under continued medical supervision. The intended use of the TX-3 TENS is identical. It does not have curative value, but relives pain symptoms. The AWQ-104 cannot be used transcerebrally, in the carotid sinus area or during pregnancy. Patients suspected of having heart disease should consider adequate precautionary measures prior to administration. Severe spasm of the laryngeal and pharyngeal muscles may occur when an electrode is placed across the neck or mouth. This may be enough to close off the airway. Stimulation will inhibit the output of some demand cardiac pacemakers and therefore, is not recommended for patients with this type of pacemaker. Electrical nerve stimulation is a symptomatic treatment, and as such may suppress the progress of pain which would otherwise serve as a protective influence {1}------------------------------------------------ on the outcome of a disease process. The potential for physical and/or psychological dependence upon nerve stimulation as a means of relieving pain has not been determined. It has been noted that some patients find the sensation of electrical stimulation extremely unpleasant and should probably be excluded from further use of the stimulator. ### Technological Characteristics: Both devices use one 9 volt battery. During stimulation, the electrical output for both devices is square waveform pulses. Details of electrical output are different but new safety concerns are not raised by the differences. Controls on each device modulate pulse outputs and regulate stimulation intensity. Material composition of the electrodes is identical. Electrode 510(k) number: K946230. #### Performance: The AWQ- 104 has been designed and tested to conform to the following standards: ISO 14971:2000 EN ISO 10993.1:1997/EN ISO 10993.5:1999/ EN ISO 10993.10.2002 EN 60601-1:1990+A1:1993+A2:1995+A13:1996/ IEC 60601-1:1988+A1:1991+A2:1995 EN 60601-1-2:2001/IEC 60601-1-2:2001 EN 980:2003 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a human figure with outstretched arms. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 6 2006 APR Lhasa OMS. Inc. c/o Mr. Mark W. Sheehan 230 Libbey Parkway Weymouth, Massachusetts 02189 Re: K050989 Trade/Device Name: AWQ-104 Digital TENS Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: II Product Code: GZJ Dated: January 27, 2006 Received: January 30, 2006 Dear Mr. Sheehan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally {3}------------------------------------------------ Page 2- Mr. Mark W. Sheehan marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, 4 Mark N. Melkerson, M.S Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # G. Indications for Use 510(k) Number (if known): K050989 Device Name: AWQ-104 Indications For Use: The AWQ-104 is intended for use in the symptomatic relief of chronic intractable pain, post traumatic acute pain, and post surgical pain. Prescription Use x AND/OR 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) Over-The-Counter Use (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division S ign-Oitl Division of General, Restorative, and Neurological Devices Page 1 of 1 **510(k) Number** K050989
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...