PUREVISION MULTI-FOCAL (BALAFILCON A) VISIBILITY TINTED CONTACT LENS

K050948 · Bausch & Lomb, Inc. · LPL · May 18, 2005 · Ophthalmic

Device Facts

Record IDK050948
Device NamePUREVISION MULTI-FOCAL (BALAFILCON A) VISIBILITY TINTED CONTACT LENS
ApplicantBausch & Lomb, Inc.
Product CodeLPL · Ophthalmic
Decision DateMay 18, 2005
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or notaphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D. Replacement schedules may vary from patient to patient, and should be decided by eye care practitioners in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eve and discarded after the recommended wearing period prescribed by the eye care professional. The lens may be disinfected using a chemical disinfection system. FREQUENT/PLANNED REPLACEMENT WEAR When prescribed for Frequent/Planned Replacement Wear, the PureVision™ Multi-Focal Contact Lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care professional. The lens may be disinfected using a chemical disinfection system. DISPOSABLE WEAR When prescribed for Disposable Wear, the PureVision™ Multi-Focal Contact Lens is to be discarded after each removal.

Device Story

Hemispherical flexible silicone-hydrogel contact lens; covers cornea and portion of sclera. Corrects refractive errors and presbyopia via aspheric, concentric center-near multifocal design. Manufactured via cast molding using balafilcon A material. Used by patients under guidance of eye care practitioners. Lens requires cleaning, rinsing, and chemical disinfection for frequent/planned replacement wear, or disposal after each removal for disposable wear. Benefits include vision correction for presbyopic patients with astigmatism.

Clinical Evidence

Bench testing only. Safety and effectiveness data for balafilcon A material previously established in K944895 and K972454. No new clinical data presented.

Technological Characteristics

Material: balafilcon A (silicone-hydrogel); FDA Group III (low water/ionic); 36% water content. Design: Aspheric multifocal, concentric center-near. Diameter: 13.5-15.5 mm. Base curve: 7.5-9.5 mm. Manufacturing: Cast mold. Color additive: Reactive Blue Dye 246 (21 CFR 73.3106).

Indications for Use

Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) and presbyopia in aphakic/notaphakic persons with non-diseased eyes; astigmatism ≤ 2.00 diopters; power range +20.00D to -20.00D; add power +0.75D to +5.00D.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ka50948 ### 510(k) PREMARKET NOTIFICATION BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens April 2005 MAY 1 8 2005 ## 510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens #### 1. Submitter Information: Bausch & Lomb Incorporated 1400 North Goodman Street Rochester, NY 14609 | Contact Person: | Anne Zavertnik | | |-----------------|-----------------------------|--| | | Manager, Regulatory Affairs | | | Telephone No.: | (585) 338-5816 | | # 2. Device Identification: | Common Name: | Soft (hydrophilic) contact lens | |------------------------|-----------------------------------------------------------------------------------------| | Proprietary Name: | BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A)<br>Visibility Tinted Contact Lens | | Classification Name: | Daily Wear, Soft (hydrophilic) Contact Lens | | Device Classification: | Class II, 21 CFR 886.5925 | #### 3. Description of Device: The PureVision Multi-Focal Contact Lens is a hemispherical flexible shell which covers the cornea and may cover a portion of the adjacent sclera. The lens material is the same silicone-hydrogel material as the visibility tinted balafilcon A contact lens described in K972454 and P980006. The color additive conforms with 21 CFR Part 73.3106. | Parameter Ranges | Specification | |-------------------|-----------------------------------| | Diameter: | 13.5 to 15.5 mm | | Base Curve Range: | 7.5 to 9.5 mm | | Center Thickness: | varies with power, 0.05 to 0.50mm | | Powers: | +20.00D to -20.00D | | Add Powers: | +0.75 to +5.00D | {1}------------------------------------------------ #### 4. Indications for Use: BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or notaphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D. Replacement schedules may vary from patient to patient, and should be decided by eye care practitioners in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eve and discarded after the recommended wearing period prescribed by the eye care professional. The lens may be disinfected using a chemical disinfection system. # FREQUENT/PLANNED REPLACEMENT WEAR When prescribed for Frequent/Planned Replacement Wear, the PureVision™ Multi-Focal Contact Lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care professional. The lens may be disinfected using a chemical disinfection system. ## DISPOSABLE WEAR When prescribed for Disposable Wear, the PureVision™ Multi-Focal Contact Lens is to be discarded after each removal. #### 5. Predicate Devices: The predicate devices were selected to address material type, lens design and indications for use. ### Lens material: Bausch & Lomb® PureVision™ (balafilcon A) Visibility Tinted Contact Lens, FDA Group III, low water, ionic contact lens, is marketed under K972454 and P980006. Lens Design and Indications for use: Bausch & Lomb® SofLens® Multi-Focal (polymacon) Visibility Tinted Contact Lens is an aspheric lens design and is marketed for daily wear of correction of presbyopia under K020927. {2}------------------------------------------------ # 6. Substantial Equivalence: | Material Similarities | | | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------| | | PureVisionTM<br>Multi-Focal Contact Lens<br>(subject device) | PureVisionTM<br>Contact Lens | | Material | Balafilcon A | Balafilcon A | | Water Content | 36% | 36% | | FDA Group | Group III<br>(low water/ionic) | Group III<br>(low water/ionic) | | Oxygen Permeability<br>(Dk)† | 91*<br>101** | 91*<br>101** | | Color Additive | Reactive Blue Dye 246 | Reactive Blue Dye 246 | | Manufacturing Method | Cast Mold | Cast Mold | | Lens Design and Indications Similarities | | | | | PureVisionTM<br>Multi-Focal Contact Lens<br>(subject device) | SofLens® Multi-Focal<br>Contact Lens | | Lens Design | Aspheric Multifocal<br>Concentric - Center Near | Aspheric Multifocal<br>Concentric - Center Near | | Intended Use | Multifocal - daily wear for<br>the correction of<br>refractive ametropia<br>(myopia, hyperopia, and<br>astigmatism) and<br>presbyopia | Multifocal - daily wear for<br>the correction of<br>refractive ametropia<br>(myopia, hyperopia, and<br>astigmatism) and<br>presbyopia | 1 Dk Units = x10 1 1 cm 0 (STP)xcm]/(secxcm2xmm Hg)@35°C (Polarographic Method) *Boundary and Edge Corrected ** Boundary and Non-Edge Corrected By reference, data from the series of non-clinical laboratory tests and clinical studies previously performed to assess the safety and effectiveness of the balafilcon A contact lens can be found in K944895, cleared December 19, 1994 and K972454, cleared August 8, 1997. # Conclusion The PureVision Multi-Focal Contact Lens is substantially equivalent to predicate devices currently legally marketed for the lens material, design and intended uses. The risks and benefits of the PureVision Multi-Focal Contact Lens are the same as those normally attributed to the wearing of soft (hydrophilic) contact lenses on a daily wear basis. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAY 1 8 2005 Bausch & Lomb, Inc. c/o Ms. Ann Zavertnik 1400 North Goodman St. Rochester, NY 14609 Re: K050948 KUJU946 Trade/Device Name: Baush & Lomb® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens = Regulatory Class: Class II Product Code: LPL Dated: April 13, 2005 Received: April 18, 2005 Dear Ms. Ann Zavertnik: We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your becaon 910(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regally manative personal provice Amendments, or to commerce provided in accordance with the provisions of the Federal Food, Drug, devices that have occh reculised in assee approval of a premarket approval application (PMA). and Cosment Act (Act) that Go not requence subject to the general controls provisions of the Act. The You may, mercrore, market the act include requirements for annual registration, listing of general controls provisions of the 120 labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is olassinod (600 wor of ols. Existing major regulations affecting your device can Inay be subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Obacements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a and regulations administered by other Federal agencies. You must of any I coural statutes und regulatents including, but not limited to: registration and listing Comply with an the rece requirements, monthly good manufacturing practice requirements as set (21 CFR Part 607), laborang (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) I his iciter will anow you to ought finding of substantial equivalence of your device to a legally prematicated predicated." in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spocific advisor Jan (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionnation on Joan Copsumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, David M. Whipple David M. Whipple Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### 510(k) PREMARKET NOTIFICATION ว ใน(K) PREMARKET NOTH 1071-10711-11-11 BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens April 2005 Bausch & Lomb 1400 North Goodman Street Rochester, NY 14609 ### Indications for Use Statement 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Device Name: Visibility Tinted Contact Lens ### Indications for Use: BAUSCH & LOMB® PureVision™ Multi-Focal (balafilcon A) Visibility Tinted Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-(11) play 11) persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D. Replacement schedules may vary from patient to patient, and should be decided by eye care practitioners in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care professional. The lens may be disinfected using a chemical disinfection system. # FREQUENT/PLANNED REPLACEMENT WEAR When prescribed for Frequent/Planned Replacement Wear, the PureVision™ Multi-Focal Contact Lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care professional. The lens may be disinfected using a chemical disinfection system. #### DISPOSABLE WEAR When prescribed for Disposable Wear, the PureVision™ Multi-Focal Contact Lens is to be discarded after each removal. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF #### NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use V Over-The-Counter-Use _ OR | ::vision Sign-Off) | |----------------------------| | Division of Ophthalmic Ear | | Nose and Throat Devises | 510(k) Number K050948 Page 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...