← Product Code [LDJ](/productcode/LDJ) · K050936

# ON CALL HOME DRUG TEST FOR MARIJUANA AND ON CALL HOME DRUG TEST FOR MARIJUANA AND COCAINE (K050936)

_ACON Laboratories, Inc. · LDJ · Jun 24, 2005 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050936

## Device Facts

- **Applicant:** ACON Laboratories, Inc.
- **Product Code:** [LDJ](/productcode/LDJ.md)
- **Decision Date:** Jun 24, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3870
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The On Call™ Home Drug Test for Marijuana (THC) is only the first step in a two-step process for determining the presence of the THC in human urine. The On Call™ Home Drug Test for Marijuana (THC) and Cocaine (COC) is only the first step in a two-step process for determining the presence of the marijuana and cocaine in human urine. The On Call Home Drug Test for Marijuana is a screening test for the rapid detection of Marijuana (THC) and its metabolites in urine at a designated cut-off concentration of 50 ng/mL for Marijuana (THC). The test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers with information concerning the presence of THC or its metabolites in a urine sample. Information, along with the materials for shipping a portion of the urine specimen to the laboratory for confirmation testing of a preliminary positive result, the second step in the process, is provided. This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The On Call Home Drug Test for Marijuana & Cocaine is a screening test for the rapid detection of Marijuana (THC), Cocaine (COC) and their metabolites in urine at a designated cut-off concentration of 50 ng/mL for Marijuana (THC) and 300 ng/mL for Marijuana & Cocaine (THC, COC). The test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers with information concerning the presence of THC, COC or their metabolites in a urine sample. Information, along with the materials for shipping a portion of the urine specimen to the laboratory for confirmation testing of a preliminary positive result, the second step in the process, is provided. This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

## Device Story

Lateral flow immunochromatographic assay; detects THC and/or COC metabolites in urine. Principle: competitive binding antigen-antibody immunochemistry; mouse monoclonal antibodies. Input: urine sample; Output: visual colored-line in test region (negative) or absence of line (positive); control line confirms proper wicking. Used by lay consumers in home setting; no instrumentation required. Provides preliminary screening; positive results require laboratory GC/MS confirmation. Benefits: rapid, accessible drug screening for consumers.

## Clinical Evidence

No clinical data provided. Substantial equivalence based on technological identity to predicate devices and established performance of lateral flow immunoassay methodology for drug screening.

## Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle. Mouse monoclonal antibodies. Cut-offs: 50 ng/mL (THC), 300 ng/mL (COC). Visual readout (colored lines). No instrumentation required. Standalone test device.

## Regulatory Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

## Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Accu-Stat™ Home Drug Test for Marijuana (THC), or Marijuana & Cocaine (THC, COC) ([K040327](/device/K040327.md))
- ACON THC One Step Marijuana Test Device ([K003557](/device/K003557.md))
- ACON One Step Multi-Drug Multi-Line Screen Test Card/Device ([K020313](/device/K020313.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k050936

B. Purpose for Submission:
New Device

C. Measurand:
Marijuana, or Marijuana and Cocaine

D. Type of Test:
Qualitative

E. Applicant:
ACON Laboratories, Inc.

F. Proprietary and Established Names:
On Call® Home Drug Test for Marijuana
On Call® Home Drug Test for Marijuana and Cocaine

G. Regulatory Information:
1. Regulation section:
21 CFR 862.3870 – Cannabinoid test system
21 CFR 862.3250 – Cocaine and cocaine metabolite test system
2. Classification:
Class II
3. Product code:
LDJ, DIO
4. Panel:

{1}

2

Toxicology (91)

H. Intended Use:

1. Intended use(s):

The On Call™ Home Drug Test for Marijuana (THC) is only the first step in a two-step process for determining the presence of the THC in human urine.

The On Call™ Home Drug Test for Marijuana (THC) and Cocaine (COC) is only the first step in a two-step process for determining the presence of the marijuana and cocaine in human urine.

2. Indication(s) for use:

The On Call Home Drug Test for Marijuana is a screening test for the rapid detection of Marijuana (THC) and its metabolites in urine at a designated cut-off concentration of 50 ng/mL for Marijuana (THC). The test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers with information concerning the presence of THC or its metabolites in a urine sample. Information, along with the materials for shipping a portion of the urine specimen to the laboratory for confirmation testing of a preliminary positive result, the second step in the process, is provided.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

The On Call Home Drug Test for Marijuana &amp; Cocaine is a screening test for the rapid detection of Marijuana (THC), Cocaine (COC) and their metabolites in urine at a designated cut-off concentration of 50 ng/mL for Marijuana (THC) and 300 ng/mL for Marijuana &amp; Cocaine (THC, COC). The test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers with information concerning the presence of THC, COC or their metabolites in a urine sample. Information, along with the materials for shipping a portion of the urine specimen to the laboratory for confirmation testing of a preliminary positive result, the second step in the process, is provided.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive

{2}

results are indicated.

3. Special conditions for use statement(s):

The On Call Home Drug Test for Marijuana and the On Call Home Drug Test for Marijuana &amp; Cocaine are for OTC use.

These assays provide only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. The materials necessary for confirmation testing and the confirmation testing itself are provided to the consumer at no additional cost. The consumer, however, pays for shipment of the sample to the laboratory.

4. Special instrument requirements:

None

I. Device Description:

The On Call® Home Drug Test kits include the following materials: 1 drug test with dropper, 1 instructional booklet, 1 urine collection cup with confirmation label, 1 shipping box, and 1 shipping bag with absorbent. The test device is composed of a test strip, containing mouse monoclonal antibody, enclosed in plastic housing.

With the exception of the brand name, the On Call® Home Drug Tests for Marijuana or Marijuana and Cocaine are identical to the Accu-Stat™ Home Drug Test for Marijuana (THC) and the Accu-Stat™ Home Drug Test for Marijuana &amp; Cocaine (THC, COC) (k040327), the ACON THC One Step Marijuana Test Device (k003557), and the ACON One Step Multi-Drug Multi-Line Screen Test Card/Device (k020313).

J. Substantial Equivalence Information:

1. Predicate device name(s):

Accu-Stat™ Home Drug Test for Marijuana (THC), or Marijuana &amp; Cocaine (THC, COC), ACON THC One Step Marijuana Test Device, First Check® Home Drug Test for THC, and Phamatech Quick Screen THC Drug Screening Tests

ACON Multi-Drug Multi-Line Test Device, First Check® Home Drug Test for Marijuana &amp; Cocaine, and Phamatech At Home™ Drug Test for Marijuana &amp; Cocaine

2. Predicate 510(k) number(s):

k040327, k003557, k993663, and k973122, respectively

{3}

k020313, k994139, and k991641, respectively

## 3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicates  |
|  Intended Use | Qualitative detection of marijuana or both marijuana and cocaine | Same  |
|  Specimen | Urine | Same  |
|  Test Principle | Immunochromatographic lateral flow assay | Same  |
|  Cut-off | 50 ng/mL (THC)
300 ng/mL (COC) | Same  |
|  Target User Population | OTC / laypersons | Same (k040327, k993663, k973122, k994139, and k991641)  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Target User Population | OTC / laypersons | Professionals (k003557 & k020313)  |

## K. Standard/Guidance Document Referenced (if applicable):

None referenced

## L. Test Principle:

The tests employ lateral flow immunochromatographic technology, which involves the recognition and formation of a specific antibody/target drug complex.

Drug in the sample and drug-labeled conjugate compete for antibody binding sites. Absence of a line in the test area is a presumptive positive result, and the presence of a line in the test area is a negative result.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. Precision/Reproducibility:

{4}

Refer to the consumer study data provided in k040327.

b. Linearity/assay reportable range:
Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable

d. Detection limit:
This was addressed in the previous Acon submissions listed in Section I.

e. Analytical specificity:
This was addressed in the previous Acon submissions listed in Section I.

f. Assay cut-off:
See “Detection limit” above.

2. Comparison studies:
a. Method comparison with predicate device:
Refer to the consumer study data provided in k040327.

b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable

b. Clinical specificity:
Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):
See Method comparison with predicate device section for information on consumer studies.

{5}

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
The expected values were based on literature.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda-staging.innolitics.com/device/K050936](https://fda-staging.innolitics.com/device/K050936)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com