European Market Post-Market Review; Retrospective review of post-market data; Follow-up/Duration: Two years; Study Period: Two years prior to submission
European users of the NMS-28 device
Not applicable for this study
Device performance and safety
Indications for Use
TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is used to provide symptomatic pain relief for chronic, acute or post-operative pain. For the EMS function, the NMS-28 is used for: Relaxation of muscle spasm; Increasing local blood circulation; Muscle re-education; Prevention or retardation of disuse atrophy; Prevention of venous thrombosis of the calf muscles immediately after surgery; Maintaining or increasing range of motion
Device Story
EasySTIM NMS-28 is a portable, dual-channel, battery-operated device combining TENS and EMS functions. It features a large LCD, adjustable frequency/pulse width/timing, 9 modes, and a doctor lock facility. Input is via user-selected settings; device transforms these into electrical stimulation pulses delivered through electrodes. Used in clinical or home settings; operated by patients or clinicians. Output is electrical stimulation for pain relief or muscle stimulation. Mechanical integrity prevents simultaneous TENS/EMS operation. Benefits include pain management, muscle re-education, and circulation improvement. Device is non-volatile and includes open circuit detectors.
Clinical Evidence
Bench testing only. Includes risk analysis, FMEA, and performance testing to verify output energy within safety ranges. Post-market feedback from two years of European market experience indicates the device performs as intended.
Technological Characteristics
Portable, dual-channel, battery-operated (2 AA alkaline) electrical stimulator. Features LCD display, open circuit detection, and non-volatile memory. Software-controlled pulse parameters (frequency, width, timing). Mechanical interlock prevents simultaneous TENS/EMS output. Class II device.
Indications for Use
Indicated for patients requiring symptomatic relief of chronic, acute, or post-operative pain (TENS) or muscle re-education, relaxation of muscle spasms, increased local blood circulation, prevention/retardation of disuse atrophy, prevention of post-surgical venous thrombosis of calf muscles, and maintenance/increase of range of motion (EMS).
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "EasyMed" in bold, black font. The word is underlined with a thin, black line. The background is white.
依士文电子仪器有限公司
vour suffering .
EasyMed Instrument Co.,Ltd
6
7
8
9
K050921
DEC I 2 2005
# 510(k) Summary
| Date of Summary prepared: | January 21, 2005 |
|---------------------------|--------------------------------------------------------------|
| Submitter: | EasyMed Instrument Co., Ltd |
| Address : | No.2 Bei Hai Da Rd Lunjiao, ShunDe, Foshan, Guangdong, China |
| Tel : | +86-757-27727282 27727283 |
| Fax : | +86-757-27727868 |
| E-Mail : | info@EasyMed.com.cn |
| Contact person: | Frances Lon (Long Shuming) |
Jeffery Wu (Wu Tingjie)
Address of the manufacturing facility: The same as above
Address of American Representative:
Norman F. Estrin, PhD, RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac MD 20854 U.S.A Tel: 001-301-279-2899 Fax: 001-301-294-0126
| Submitted Device: | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Generic name: | Powered Muscle Stimulator (PMS) (or Electrical Muscle Stimulator (EMS)) and Transcutaneous Electrical Nerve Stimulator (TENS) |
| Trade name: | EasySTIM NMS-28 |
| Common name: | For the TENS function—TENS device<br>For the EMS function—Powered Muscle Stimulators |
| Classification name: | For TENS functions<br>Stimulator, Nerve, Transcutaneous, for Pain Relief - GZJ;<br>21 CFR 882.5890. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "EasyMed" in a bold, sans-serif font. The word is black and appears to be underlined. The letters are closely spaced, giving the word a compact appearance.
**Easy for the relief of your suffering**
### EasyMed Instrument Co.,Ltd
For EMS functions Powered Muscle Stimulator for re-education of muscles - IPF;
21 CFR 890.5850
Device Classification: For both TENS & EMS functions - Class II
Predicate Device: For the TENS function EasyMed TENS Unit TN-28C (K040253) For the EMS function TensCare Ultima NMS/EMS Model XL-A3 - (K023997)
The class of the predicate Device:
Class II
Device Description:
For the TENS function A portable TENS device for pain relief.
#### For the EMS function
A portable EMS device for the re-education of muscles.
Features:
· Innovative design
· Large LCD display
• Dual output channels
• 2 AA Alkaline Batteries
· Adjustable frequency, pulse width, and timing parameters
• 9 different modes
• Timer option
• Doctor lock/unlock facility
· Open circuit detectors
• Non-volatile
The intended use of the device:
1. TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is used to provide symptomatic pain relief for chronic, acute or post-operative pain.
2. For the EMS function, the NMS-28 is used for:
· Relaxation of muscle spasm
•Increasing local blood circulation
· Muscle re-education
· Prevention or retardation of disuse atrophy
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "EasyMed" in bold, black font. The word is split into two parts: "Easy" and "Med". A horizontal line underlines the entire word, emphasizing the brand name. The font is sans-serif and appears to be a logo or brand identifier.
依士文电子仪器有限公司
EasyMed Instrument Co.,Ltd suffering
•Prevention of venous thrombosis of the calf muscles immediately after surgery
· Maintaining or increasing range of motion
The intended use and indications for use of the new device are the same with that of the predicate devices.
Technological Comparison:
- 1. The marketed device EasyMed TN-28C (K040253) is a digital TENS unit which has been well exploited the digital technology, supplying the user with full ranges of parameters, traditional TENS functions, and non-traditional new functions to be selected.
- 2. In assembly, except software, the new device NMS-28 is exactly the same with the marketed device TN-28C, they have the same enclosure, the same circuit diagram, the same PCB layout, the same components and the same working principle;
- 3. The marketed device TensCare Ultima NMS/EMS Model XL-A3 -(K023997) combines the functions of a TENS device and an EMS (Electrical Muscle Stimulator) or Powered Muscle Stimulator, into one package. It is also a digital unit with similar electronic working principle with TN-28C.
- The new device NMS-28 also combines the functions of a TENS ধ device and an EMS into one package. It is also substantial equivalent to the marketed device TensCare Ultima NMS/EMS Model XL-A3 - (K023997), their features and performances in muscle controlling as well as pain relief therapy are very similar substantially.
- 5. The software of the new device NMS-28 can be divided into two parts: one for TENS function, the another for EMS function. The part of software for TENS function of the new device NMS-28 is only a subset of that of the marketed device TN-28C, what is more,
- the considerations on safety and reliability in TN-28C are also introduced into the EMS function of the new device; While the part of software for EMS function of the new device NMS-28 is very similar to that of the marketed device TensCare Ultima NMS/EMS Model XL-A3, by applying the similar control methods for muscle contraction-relaxation.
- The accessories of the new device NMS-28 are almost the same ్.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "EasyMed" in a bold, sans-serif font. The letters are black, and there is a black line underneath the entire word. The word is likely a logo or brand name.
Easy for the relief of your suffering
## EasyMed Instrument Co.,Ltd
with those of the marketed device TN-28C, and substantially equivalent to marketed device TensCare Ultima NMS/EMS Model XL-A3.
The Labeling is substantially equivalent to that of the predicate Labeling Comparison: devices.
Safety information:
Design to comply with relevant safety applicable recognized consensus standards; the output energy is well controlled in the safety and effectiveness ranges specified by relevant FDA guidance's. Testing has been carried out in very detailed and strict. · Test results, Risk Analysis, and FMEA analysis show that the new unit NMS-28 is safe with no any hazard.
NMS-28 combines the functions of TENS and EMS into one package, it is not possible to use the TENS and the EMS functions simultaneously. Mechanical integrity ensures that only one function can be selected at any one time.
NMS-28 has been on the European Market for the past two years. During this time a review of Customer Complaints, Returned Product and the results of Post Market Feedback, has demonstrated that the product has performed as Intended, to it's Specified Requirements.
Submitted times: It is the first submission to FDA for this new device
Conclusions:
The NMS-28 EMS/TENS unit is substantially equivalent to the predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure embracing a bird, which is meant to convey the department's mission of protecting and promoting the health and well-being of Americans.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### DEC 1 2 2005
Jeffrey Wu EasyMed Instrument Co., Ltd No. 2 Bei Hai Da Road LunJiao, ShunDe Guangdong, China
Re: K050921
K050921
Trade/Device Name: EasyMed EMS/TENS Model NMS-28 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: II Product Codes: IPF, GZJ Dated: June 17, 2005 Received: November 23, 2005
Dear Mr. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 3 D(t), plenial is substant for the indications for the indications for referenced above and have delemined the devices marketed in interstate commerce
use stated in the enclosure) to legally marketed predicate or to devices that use stated in the enclosure) to tegally hiar Recol prece Amendments, or to devices that prior to May 28, 1976, the enactment date of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Federal PMA) You may. have been reclassified in accordance with the provisions of the Act. The Act. The peneral Act (Act) that do not require approval or a prematic controls provisions of the Act. The general
therefore, market the device, subject to the general controls provisions of therefore, market the device, subject to the general on annual registration, listing of devices, good
controls provisions of the Act include requirements for anning mind and controls provisions of the Act mende requirements for and adulteration.
manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into cliffer the stime aftecting your device can be
may be subject to such additional controls. Existing major regulations FDA may may be subject to such additions, Title 21, Parts 800 to 898 . In addition, FDA may publish
found in the Code of Federal Regulations, Title 21, Parts 800 to 898 . found in the Code of Federal resganaties in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that Please be advised that FDA's issualice of a substance requirements of the Act or any
FDA has made a determination that your device complex . You must comply with all FDA has made a determination inal your tee lice concess. You must comply with all
Federal statutes and regulations administerial and listing (21 CFR Part 807); Federal statutes and regulations administered or our registration and listing (21 CFR Part 807);
the Act's requirements, including, but not limited to: registration as set the Act s requirements, melading, but not maring practice requirements as set
labeling (21 CFR Part 801); good manufactures (200); and if annlice labeling (2) CFR Part 801); good inaliation (2) CFR Part 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000-1050. forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-
This letter will allow you to begin marketing your device as described in your Section 510(k)
. It is a legin and the first of a local started assissalence of your device to This letter will allow you to begin marketing your device as assisted by and one of your device to a legally
{5}------------------------------------------------
Page 2 -- Mr. Jeffrey Wu
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs of the many of the many of 10, 2019 - Alancy lass mote the regulation e If you desire specific advice for your de vice on on our nooning note the regulation entitled,
contact the Office of Compliance at (240) 276-0120. Also, please note the regul contact the Office of Compilance at (240) 210 of 2019 Part 807.97). You may obtain other "Misbranding by reference to premaired nother the Act from the Division of Small Manufactures,
general information on your responsibilities under the Act 1 - 1000 - 3301 May general information on your responsionities under (800) 638-2041 or (301) 443-6597 or
International and Consumer Assistance at is this and works of the l international and Consumer Treaver Tra.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
2
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use
·· K050921 510(k) Number
EasyMed EMS/TENS Model NMS-28 Device Name :
Indications for Use :
TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is used to provide symptomatic pain relief for chronic, acute or post operative pain.
PMS/EMS stands for Powered Muscles Stimulator/Electrical Muscles Stimulator. This PMS/EMS system is indicated for:
- · Relaxation of muscle spasm
- · Increasing local blood circulation
- Muscle re-education
- · Prevention or retardation of disuse atrophy
- · Prevention of venous thrombosis of the calf muscles immediately after surgery
- Maintaining or increasing range of motion
Prescription Use_ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**(Division Sign-Off)**
Division of General, Restorative, and Neurological Devices
page 1/1
Page 9 of 205
**510(k) Number** K050421
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.