SAFE-CROSS RADIO FREQUENCY TOTAL OCCLUSION CROSSING SYSTEM [CORONARY]

K050915 · Intra Luminal Therapeutics, Inc. · DQX · Jun 7, 2005 · Cardiovascular

Device Facts

Record IDK050915
Device NameSAFE-CROSS RADIO FREQUENCY TOTAL OCCLUSION CROSSING SYSTEM [CORONARY]
ApplicantIntra Luminal Therapeutics, Inc.
Product CodeDQX · Cardiovascular
Decision DateJun 7, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1330
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Safe-Cross® Radio Frequency Total Occlusion Crossing System is indicated for use in facilitating the placement of devices used in percutaneous interventions in native coronary arteries with total occlusions.

Device Story

System facilitates placement of interventional devices in native coronary arteries with total occlusions; consists of RF crossing wire (175cm or 275cm), console with display, and footswitch. Wire connects to console via Y-site hub for optical visualization and RF energy delivery. Practitioner uses optical feedback to navigate vessel structures; applies discrete RF energy to distal wire tip to cross occlusions. Used in clinical settings by physicians. Benefits include improved ability to cross total occlusions, enabling subsequent percutaneous interventions.

Clinical Evidence

Bench testing and in vivo animal studies conducted. Bench testing evaluated performance characteristics (tip flexibility, shaping, retention). Animal studies confirmed components function properly together and satisfy intravascular performance requirements. No human clinical trial data reported.

Technological Characteristics

System includes RF crossing wires, console, display, and footswitch. Wires feature optical fiber and RF connectors. Sterilized via ethylene oxide. Single-use. Connectivity involves wire-to-console interface for optical visualization and RF energy delivery. Design modifications focus on tip flexibility and shape retention.

Indications for Use

Indicated for patients undergoing percutaneous interventions in native coronary arteries with total occlusions.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 7 - 2005 Ko509/5 ## 510(K) SUMMARY #### SUBMITTER INFORMATION | A. | Company Name: | IntraLuminal Therapeutics, Inc. | |----|--------------------|------------------------------------------------------------------------------| | B. | Company Address: | 6354 Corte Del Abeto - Suite A<br>Carlsbad, CA 92009 | | C. | Company Phone: | (760) 918-1820 | | D. | Company Facsimile: | (760) 603-9615 | | E. | Contact Person: | Pamela Misajon<br>Vice President of Regulatory Affairs and Quality Assurance | #### DEVICE IDENTIFICATION | A | Device Trade Name: | Safe-Cross® Radio Frequency Total Occlusion<br>Crossing System | |----|----------------------|----------------------------------------------------------------| | B. | Device Common Name: | Catheter Guide Wire | | C. | Classification Name: | Catheter Guide Wire | | D. | Device Class: | Class II (per 21 CFR 870.1330) | ### IDENTIFICATION OF PREDICATE DEVICE The predicate device is the Safe-Cross Radio Frequency Total Occlusion Crossing System, manufactured by IntraLuminal Therapeutics and cleared under Premarket Notification 510(k) K040037. ### DEVICE DESCRIPTION The Safe-Cross Radio Frequency Total Occlusion Crossing System consists of the following: - 175cm Working Length 0.014" Safe-Cross RF Crossing Wire Straight and . Angled Tip (with Torquer and Tip Shaping Tool) - 275cm Working Length 0.014" Safe-Cross RF Crossing Wire Straight and . Angled Tip (with Torquer and Tip Shaping Tool) - Safe-Cross RF System Console with Display and Footswitch ● {1}------------------------------------------------ The modified Safe-Cross RF System is similar to the predicate Safe-Cross System. The proximal end of the RF Crossing Wire is connected to a Y-Site hub that houses the optic fiber connector and the RF connector. The optical connector is connected to the OCR input on the console to allow the medical practitioner to visualize structures within the vessel for navigation purposes. The RF connector is connected to the RF output on the console. This allows the medical practitioner to provide discrete RF energy to the distal tip to assist in moving the wire tip through the occlusion in the vessel. The RF Crossing Wire is packaged in a Tyvek® sealed plastic tray with the Tip Shaping Tool. A Torquer is provided in a separate peel pouch. The packaged RF Crossing Wire is provided "STERILE" (ethylene oxide) and non-pyrogenic, and is intended for single use only. ## INTENDED USE The Safe-Cross® Radio Frequency Total Occlusion Crossing System is indicated for use in facilitating the placement of devices used in percutaneous interventions in native coronary arteries with total occlusions. # TECHNOLOGICAL CHARACTERISTICS The components of the Safe-Cross System are similar in basic materials, design, construction and performance to the predicate device. The RF Crossing Wires have been modified to improve performance characteristics (i.e., tip flexibility, tip shaping and shape retention). The safety and performance of the modified Safe-Cross System has been verified through biocompatibility testing, bench testing and in vivo animal studies. ## BIOCOMPATIBILITY AND PERFORMANCE DATA Biocompatibility testing has been conducted to verify that the materials in the modified RF Crossing Wires are biologically safe. In vitro bench testing was conducted to evaluate the performance characteristics of the modified RF Crossing Wires. Benchtop performance test results indicate that the modified RF Crossing Wires satisfy safety and performance requirements of the device specifications and do not raise additional safety issues. In vivo animal studies have shown that the components of the system function properly together and satisfy intravascular performance requirements in an animal model. ### CONCLUSIONS DRAWN FROM STUDIES On the basis of the testing conducted on the modified Safe-Cross System it may be concluded that the device satisfies safety and performance requirements when used in accordance with the Instructions for Use for the indicated patient population. The modified Safe-Cross System is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three horizontal lines curving upwards, resembling a symbol or abstract representation of a flag. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 7 - 2005 Ms. Pamela Misajon Vice President of Regulatory Affairs and Quality Assurance Intraluminal Therapeutics, Inc. 6354 Corte del Abeto, Suite A Carlsbad, CA 92009 K050915 Re: Trade/Device Name: Safe-Cross® Radio Frequency Total Occlusion Crossing System Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: II Product Code: DQX Dated: May 5, 2005 Received: May 9, 2005 Dear Ms. Misajon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 – Ms. Pamela Misajon forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic s (21 CFR) (2006) (2006) (2006) (21 CFR (21 (2) 642 - 6th Anti) 21 CFR 1000forth in the quality systems (QS) regulation (21 -51 -542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections 531-542 of the Action Section 5 product radiation control provisions (Sections 95 described in your Section 510(k) This letter will allow you to begin marketing your device to a legal This letter will allow you to begin marketing your antial equivalence of your device to a legally premarket notification. The FDA imulig of substantial equivalies by . marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please r the light and the many of the 10, 276, Alles were the regulation entitled If you desire specific advice for your ac not one of a lise, please note the regulation entitled, contact the Office of Compliance at (240) 276-0120. Also, please note the re contact the Office of Comphalec at (240) 276 or Frant 807.97). You may obtain "Misbranding by reference to premarket notification (21CFR Part 807.97). You may obtain "Misbranding by reference to premarkst notificancer the Act from the Division of Small other general information on your responsibilities under that the gumber (800) 638-200 other general information on your responsion.com in total-free number (800) 638-2041 or Manufacturers, Imemational and Consumer Assistance a no vol an an an an an an an and and and and and and a Sincerely yours, Duna R. Vicines Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K 6 50915 Device Name: Safe-Cross® Radio Frequency Total Occlusion Crossing System The Safe-Cross® Radio Frequency Total Occlusion Indications For Use: Crossing System is indicated for use in facilitating the placement of devices used in percutaneous interventions in native coronary arteries with total occlusions. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Donna P. Vachal (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_K050915
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...