# BARRIER SURGICAL GOWNS (K050786)

_Molnlycke Health Care · FYA · May 9, 2005 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050786

## Device Facts

- **Applicant:** Molnlycke Health Care
- **Product Code:** [FYA](/productcode/FYA.md)
- **Decision Date:** May 9, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

BARRIER® surgical gowns are intended for single use to be worn by the Surgical Team Members in the operating room during surgical procedures to protect both the patient and the operating room personnel from the transfer of microorganisms, body fluids and particulate material.

## Device Story

BARRIER® Surgical Gowns are single-use protective apparel designed for surgical team members in operating rooms; function as physical barrier to prevent transfer of microorganisms, body fluids, and particulate material between patient and personnel. Gowns consist of various fabric assortments tailored for general surgical procedures or high-risk procedures requiring increased protection against bloodborne pathogens. Healthcare providers wear gowns during surgery to maintain sterile fields and protect staff. Benefits include reduced risk of cross-contamination and infection transmission during invasive procedures.

## Clinical Evidence

Bench testing only. Performance evaluated according to ANSI/AAMI PB-70:2003 standard for liquid barrier performance and classification of protective apparel. Gowns were classified into AAMI Levels 2, 3, or 4 based on barrier performance testing.

## Technological Characteristics

Single-use surgical gowns constructed from various fabrics. Barrier performance classified per ANSI/AAMI PB-70:2003. Levels of protection range from Level 2 to Level 4 depending on the specific gown family.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- BARRIER® Standard and Standard Fabric Reinforced Surgical Gowns ([K760903](/device/K760903.md))
- BARRIER® Extra and Ultra Protection Gowns ([K760903](/device/K760903.md))
- BARRIER® Extra and Ultra Protection Plus Surgical Gowns ([K932731](/device/K932731.md))
- BARRIER® Surgical Gown II and BARRIER® Surgical Gown II, Fabric Reinforced ([K920996](/device/K920996.md))
- BARRIER® Fluid Protection Plus Surgical Gowns ([K990395](/device/K990395.md))
- BARRIER® Urology Gown ([K760903](/device/K760903.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

**510(k) SUMMARY**

K030786

| Applicant:               | Mölnlycke Health Care<br>826 Newtown Yardley Rd.<br>Newtown, PA 18940                                                                                                                                                                                                                                                                                                                                                                           |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:          | John Clay<br>Regulatory Affairs Officer                                                                                                                                                                                                                                                                                                                                                                                                         |
| Tel.:                    | 267-685-2078                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Fax:                     | 267-685-2010                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Name:             |                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Proprietary Name:        | Barrier® Surgical Gowns                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Common/Usual Name:       | Surgical Gowns                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Classification:   | Class II - 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Substantial Equivalence: | For the purpose of Section 510(k) of the Federal Food, Drug and<br>Cosmetic Act, Mölnlycke Health Care considers the BARRIER®<br>Surgical Gowns are substantially equivalent in composition,<br>function and intended use to the previously marketed BARRIER®<br>Surgical Gowns.                                                                                                                                                                |
| Intended Use:            | BARRIER® surgical gowns are intended for single use to be worn<br>by the Surgical Team Members in the operating room during<br>surgical procedures to protect both the patient and the operating<br>room personnel from the transfer of microorganisms, body fluids<br>and particulate material.                                                                                                                                                |
| Description:             | The BARRIER® Surgical Gowns have been designed with an<br>assortment of fabrics and have been on the market for a number of<br>years cleared under predicate 510(k) notifications. BARRIER<br>surgical gowns have been developed for general surgical<br>procedures and for procedures which represent possible exposure<br>to bloodborne pathogens and other potentially infectious materials<br>which require increased levels of protection. |

{1}------------------------------------------------

#### BARRIER 510(k) Summary, Page 2

The following chart highlights the current BARRIER surgical gown families and the predicate 510(k) notifications under which they are currently marketed.

| Description                                                                       | Predicate 510(k) |
|-----------------------------------------------------------------------------------|------------------|
| BARRIER® Standard and Standard<br>Fabric Reinforced Surgical Gowns                | K760903          |
| BARRIER® Extra and Ultra Protection<br>Gowns                                      | K760903          |
| BARRIER® Extra and Ultra Protection<br>Plus Surgical Gowns                        | K932731          |
| BARRIER® Surgical Gown II and<br>BARRIER® Surgical Gown II , Fabric<br>Reinforced | K920996          |
| BARRIER® Fluid Protection Plus<br>Surgical Gowns                                  | K990395          |
| BARRIER® Urology Gown                                                             | K760903          |

# Summary of

Testing:

BARRIER® Surgical Gowns have been evaluated using the ANSI/AAMI PB-70:2003 Standard. (Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities) The chart represents the AAMI levels claimed for each gown family.

| Description                                                                       | Level of Protection according to AAMI |
|-----------------------------------------------------------------------------------|---------------------------------------|
| BARRIER® Standard and Standard<br>Fabric Reinforced Surgical Gowns                | Level 2                               |
| BARRIER® Extra and Ultra Protection<br>Gowns                                      | Level 3                               |
| BARRIER® Extra and Ultra Protection<br>Plus Surgical Gowns                        | Level 4                               |
| BARRIER® Surgical Gown II and<br>BARRIER® Surgical Gown II , Fabric<br>Reinforced | Level 3                               |
| BARRIER® Fluid Protection Plus<br>Surgical Gowns                                  | Level 3                               |
| BARRIER® Urology Gown                                                             | Level 4                               |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.

MAY - 9 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. John Clay Regulatory Affairs & Quality Officer-US Mölnlycke Health Care 826 Newtown-Yardley Road, Suite 300 Newtown, Pennsylvania 18940

Re: K050786

Trade/Device Name: BARRIER® Surgical Gowns Regulation Number: 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: April 19, 2005 Received: April 22, 2005

Dear Mr. Clay:

We have reviewed your Section 510(k) premarket notification of intent to market the device we have ro rowed your and have determined the device is substantially equivalent (for the relerenced above and nave attermclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Ameliuments, or to act roos that nortic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls providents or annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III If your device to such additional controls. Existing major regulations affecting (I MA), it may be subject to back of Federal Regulations, Title 21, Parts 800 to 898. In your device can be found in the reannouncements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not I least to advisod that 1 Da determination that your device complies with other requirements Incall that I D'I may Federal statutes and regulations administered by other Federal agencies. of the Act of any I oderal statuse and submitted to: but not limited to: registration 1 ou must comply with and 1); labeling (21 CFR Part 801); good manufacturing practice and listing (21 CF RP art 867), laceling (QS) regulation (21 CFR Part 820); and if requirements as bet ferm nroduct radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow yourse bogan made of substantial equivalence of your device to a premarket notified.com - The PDF results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at not 100 your at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other generational and Consumer Assistance at its toll-free Divisor (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Clair Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K050786

Device Name: BARRIER® Surgical Gowns

Indications for Use:

BARRIER® surgical gowns are intended for single use to be worn by the Surgical Team Members in the operating room during surgical procedures to protect both the patient and the operating room personnel from the transfer of microorganisms, body fluids and particulate material.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

### Concurrence of CDRH, Office of Device Evaluation (ODE)

Shula H. Murphy, MD

of Anesthesiology, General Hospital, n Control, Dental Devices

Number: K050 786

Page __ of

---

**Source:** [https://fda-staging.innolitics.com/device/K050786](https://fda-staging.innolitics.com/device/K050786)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).