The Lanx Anterior Cervical Plate System is intended for anterior interbody fixation of the cervical spine. The Lanx Anterior Cervical Plate System is suitable for use to provide temporary stabilization of the anterior spine while awaiting bony fusion (healing) in patients with degenerative disc disease (neck or radicular pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), or pseudoarthrosis and/or failed previous fusion between C1 and C6.
Device Story
Anterior cervical plate system; provides temporary stabilization of anterior spine during bony fusion. Anatomically curved metallic plates; titanium alloy construction. Used in surgical setting by surgeons. Device implanted to provide structural support to vertebrae; facilitates healing in patients with degenerative disc disease, trauma, tumors, or deformity. Output is mechanical stabilization of spinal segments. Benefits patient by maintaining alignment and stability while bone fusion occurs.
Clinical Evidence
Bench testing only. Fatigue and static testing performed according to engineering and scientific principles. Results demonstrate performance equivalent to previously cleared anterior cervical plates.
Technological Characteristics
Anatomically curved metallic plates; Titanium Alloy (ASTM F136-92, ISO 5832-3) and Titanium (ASTM F67 GR2-95, ISO 5832-2). 1-level to 5-level configurations; 18mm-118mm lengths. Manually applied locking mechanism. Non-sterile; requires steam autoclave sterilization (270°F/134°C for 30 minutes).
Indications for Use
Indicated for anterior interbody fixation of the cervical spine (C1-C6) for temporary stabilization during bony fusion in patients with degenerative disc disease, trauma (fractures), tumors, deformity (kyphosis, lordosis, scoliosis), pseudoarthrosis, or failed previous fusion. Contraindicated for screw attachment to posterior elements (pedicles) of cervical, thoracic, or lumbar spine.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
KOSO774
# APR 1 1 2005
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>. www.virtu.run |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Summary Information | LANX, LLC. |
| | |
| Premarket Notification, Section 510(k) | MARCH 21, 2005 |
| | |
Safe Medical Devices Act of 1990, 21 CFR 807.92 Regulatory Authority:
#### 1. Device Name:
Lanx Anterior Cervical Plate System Trade Name:
Common Anterior cervical spine system Name(s):
Classification Spinal intervertebral body fixation orthosis. Name(s):
### Establishment Name & Registration Number: 2.
| Name: | LANX, LLC. |
|---------|------------|
| Number: | 3004485144 |
### 3. Classification(s):
\$ 888.3060 Spinal intervertebral body fixation orthosis. (a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct "sway back", scoliosis (lateral curvature of the spine), or other conditions. (b) Classification. Class 11.
| Device Class: | Class II for all requested indications |
|-----------------------|----------------------------------------------|
| Classification Panel: | Orthopaedic and Rehabilitation Devices Panel |
| Product Code(s): | KWQ |
### Equivalent Predicate Device: 4.
LANX, LLC. believes that the Lanx Anterior Cervical Plate System is substantially equivalent to the following:
Equivalence can be seen in the design, material composition, surgical technique and intended use.
### 5. Device Description:
## General description of the new plate.
### Materials:
| Titanium Alloy | ASTM F136-92 | ISO 5832-3 |
|----------------|-----------------|------------|
| Titanium | ASTM F67 GR2-95 | ISO 5832-2 |
Cervical plates are offered for the following levels/configurations. These are offerings as detailed in the original submission.
One Level Two Level Three Level Four Level Five Level
### Cervical Special.doc
31
{1}------------------------------------------------
Fatigue and static testing is complete. Samples were tested according to accepted Testing Summary. engineering and scientific principals. Test results demonstrate that the new plates perform in a manner equivalent to the previously cleared anterior cervical plates.
### Applicant Name & Address: 6.
Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501 fulton@lanx.us
#### 7. Company Contact:
Michael Fulton, M.D. Lanx. LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501 fulton@lanx.us
### Submission Correspondent: 8.
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax
#### 9. Performance Standards:
United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized. Voluntary standards utilized include ASTM, Standard Operating Procedures, vendor & process certification procedures, Quality Systems Regulations, ISO materials standards and ISO 9000 series quality regulations.
LANX, LLC. also meets appropriate general controls authorized under Sections 501, 502, 518, 518, 519, and 520 of the Food, Drug, and Cosmetic Act.
### Storage, Packaging & Sterilization Information: 10.
The Lanx Anterior Cervical Plate System is supplied "NON-STERILE" and must be sterilized prior The recommended sterilization process is high temperature steam autoclave sterilization. The to use. referenced sterilization cycle produces a Sterility Assurance Level (SAL) of at least 10°
The validated cycle is: Method: Steam Gravity Cycle: Temperature: 270°F (134°C) Exposure Time: 30 minutes
All packages containing implants or instruments should be intact upon receipt. Damaged packaging may indicate the presence of unsafe product. If the package or product is damaged, the product should not be used and should be returned.
Cervical Special.doc
32
{2}------------------------------------------------
Product must be handled, stored and opened in such a way that it is protected from inadvertent damage or I rough mused, the product must be placed into use following cleaning, sterilization and accepted surgical sterile technique.
| FEATURE | Lanx Anterior Cervical Plate System | Modified Anterior Cervical Plate. | SE? |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----|
| Indications for Use: | Temporary stabilization of the anterior spine while awaiting<br>bony fusion (healing) in patients with degenerative disc disease<br>(neck or radicular pain of discogenic origin with degeneration of<br>the disc confirmed by patient history and radiographic studies),<br>trauma (including fractures), tumors, deformity (kyphosis,<br>lordosis or scoliosis), or pseudoarthrosis and/or failed previous<br>fusion. | SAME | YES |
| Design: | Anatomically curved metallic plate | SAME | YES |
| Sterile: | No - Must be Autoclaved | SAME | YES |
| Plate Sizes: | 1-level to 5-level<br>18mm - 118mm length in 2-5mm increments | EQUIVALENT | YES |
| Material: | Titanium Alloy | SAME | YES |
| Locking<br>Mechanism: | Manually applied locking mechanism | Intergral locking mechanism - no intra-<br>operative assembly required | NO |
| Manufacturer: | Lanx, LLC | Lanx, LLC | YES |
| Product Code: | KWQ | KWQ | YES |
| K Number: | K040401 | Pending | YES |
### Summary Comparison Table: 12.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - U.S.A." is arranged in a circular pattern around the eagle.
APR 1 1 2005
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Lanx, LLC C/o Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, California 94523
Re: K050774
Trade/Device Name: Lanx Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: KWQ Dated: March 20, 2005 Received: March 25, 2005
Dear Mr. Schlerf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. David W. Schlerf
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally presticated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general militiremational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Stup Rhodes
Miriam Provost, Ph.D.
Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
KOSO774
Page l
**510(k) Number**: K050774
Lanx Anterior Cervical Plate System Device Name(s):
## Indications For Use:
The Lanx Anterior Cervical Plate System is intended for anterior interbody fixation of the cervical spine. The Lanx Anterior Cervical Plate System is suitable for use to provide temporary stabilization of the anterior spine while awaiting bony fusion (healing) in patients with degenerative disc disease (neck or radicular pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), or pseudoarthrosis and/or failed previous fusion between C1 and C6.
Warning: This device is not cleared for screw attachment to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
Prescription Use OR Over-The-Counter Use PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE)
Stypk Pluder
(Division Sign-Off Division of General, Restorative, and Neurological Devices
**510(k) Number** K050774
(Optional format 1-2-96)
(Per 21 CFR 801.109)
Cervical Special.doc
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.