BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM, LEVOFLOXACIN (STREP) 0.25-16 UG/ML

K050747 · Becton, Dickinson & CO · LON · Apr 29, 2005 · Microbiology

Device Facts

Record IDK050747
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM, LEVOFLOXACIN (STREP) 0.25-16 UG/ML
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateApr 29, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

Device Story

System performs automated antimicrobial susceptibility testing (AST) and identification (ID) of Streptococcus species. Input: pure culture bacterial isolates suspended in AST-S broth (cation-adjusted, non-blood) at 0.5 McFarland density. Process: inoculated polystyrene trays with 136 micro-wells containing dried reagents/antibiotics; continuous incubation at 35°C; colorimetric oxidation-reduction redox indicator detects bacterial growth (blue to pink/colorless). Instrument reads wells every 20 minutes; software interprets biochemical reactions and MIC values. Output: ID result and MIC value/category interpretation (SIR). Used in clinical microbiology laboratories; operated by technicians. Healthcare providers use MIC data to guide antibiotic therapy decisions. Benefits: rapid, automated determination of susceptibility to levofloxacin (0.25-16 ug/mL) to inform treatment.

Clinical Evidence

Clinical performance evaluated by comparing Phoenix System results for clinical and challenge isolates against the CLSI reference broth microdilution method. Study included 1055 isolates. Metrics: Essential Agreement (EA) and Category Agreement (CA). Reproducibility assessed at three sites with streptococcal isolates, showing >90% intra-site and >95% inter-site reproducibility.

Technological Characteristics

Materials: molded polystyrene trays. Principle: broth microdilution with colorimetric oxidation-reduction (redox) indicator. Energy: automated incubation at 35°C. Connectivity: networked instrument. Software: rule-based 'EXPERT' system using CLSI guidelines. Inoculum: 0.5 McFarland standard. Detection: automated optical reading of color change every 20 minutes.

Indications for Use

Indicated for in vitro rapid identification and antimicrobial susceptibility testing (MIC) of Streptococcus species from pure culture. Prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} Page 1 of 6 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k050747 B. Purpose for Submission: Addition of the antibiotic levofloxacin to the Phoenix™ SMIC/ID and SMIC Panels C. Measurand: Levofloxacin 0.25-16 ug/mL D. Type of Test: Antimicrobial Susceptibility Test (Quantitative and Qualitative) colorimetric oxidation-reduction, growth-based E. Applicant: Becton, Dickinson &amp; Company F. Proprietary and Established Names: BD Phoenix™ Automated Microbiology System – Levofloxacin (strep) 0.25-16 ug/mL G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial 2. Classification: Class II 3. Product Code: LON 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): BD Phoenix™ Automated Microbiology System: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae and gram-positive bacteria belonging to the genera Staphylococcus, Enterococcus and Streptococcus. {1} Page 2 of 6 The BD Phoenix™ SMIC/ID and SMIC Panel is intended for the *in vitro* rapid identification (ID) of bacteria from pure culture belonging to the genera *Streptococcus*. The BD Phoenix™ SMIC/ID and SMIC panels are also intended for the quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most bacteria isolates from pure culture belonging to the genera *Streptococcus* when used with the BD Phoenix™ Automated Microbiology System. 2. Indication(s) for use: This submission is for the addition of the antibiotic levofloxacin at concentrations of 0.25-16 ug/mL to the Phoenix™ SMIC/ID and SMIC Panels. 3. Special condition for use statement(s): Prescription Use Only 4. Special instrument Requirements: Not Applicable I. Device Description: The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST-S Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST-S broth, which contains an AST-S indicator, prior to inoculating the panel. The AST-S broth is a non-blood, cation-adjusted broth containing purified water, Tween 80, pancreatic digest of casein, peptones and other additional supplements for optimization of streptococcal growth. After adding the indicator solution to the AST-S inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The resulting AST has a final inoculum of 5 × 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the Clinical and Laboratory Standards Institute (CLSI). {2} Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible. ## J. Substantial Equivalence Information: 1. Predicate device name(s): VITEK® System 2. Predicate K number(s): N50510 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | Intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most bacteria. | same | | Isolates | Isolated colonies from culture used | Isolated colonies from culture used | | Results | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | | Incubation conditions | <16 hours | <16 hours | | Differences | | | | Item | Device | Predicate | | Inoculum preparation | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 1.0 McFarland standard | | Reading algorithm | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions | | Technology | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner. | ## K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S15) "Methods {3} for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard." # L. Test Principle: The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in "growth control wells" which contain no antibiotic. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Twenty eight isolates were evaluated for site to site and inter site reproducibility demonstrating $&gt;95\%$ reproducibility. The ten isolate study described in the guidance document was used (10 organisms tested 3 times on 3 days at 3 sites). # b. Linearity/assay reportable range: Not applicable # c. Traceability, Stability, Expected values (controls, calibrators, or method): CLSI recommended Quality Control strain was tested at the concentrations listed (see table below). The Phoenix™ results demonstrated that the system can produce QC results in the recommended range. The modes for the reference method and the Phoenix™ were the same. | Organism | Concentration ug/mL | Reference results | PhoenixTM results | | --- | --- | --- | --- | | | | | | | S. pneumoniae ATCC 49619 Expected range 0.5-2 ug/mL | 0.5 | 6 | 41 | | | 1 | 118 | 79 | | | 2 | 1 | 1 | | | 4 | | | | | 8 | | | | | 16 | | | Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of {4} Page 5 of 6 testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used. Additional testing was performed on 5 streptococcal strains to demonstrate acceptable performance for streptococcal species. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: The CLSI recommended broth dilution reference panel with 2-5% lysed horse blood was prepared according to the CLSI recommendation and used to compare with the Phoenix™ results. Clinical testing was performed at four sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. The study included 1955 isolates tested with the following performance. There is a slight trend for the Phoenix™ result to be more susceptible than the reference but still have very good EA. This was also apparent in the QC test results. | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 1823 | 1777 | 97.5 | 1740 | 1707 | 98.1 | 1810 | 99.3 | 20 | 8 | 4 | 1 | | Challenge | 132 | 131 | 99.2 | 121 | 120 | 99.2 | 132 | 100 | 8 | 0 | 0 | 0 | | Combined | 1955 | 1908 | 97.6 | 1861 | 1827 | 98.2 | 1942 | 99.3 | 28 | 8 | 4 | 1 | EA-Essential Agreement CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation. Evaluable (Eval) is when the results are on scale. The test device had a growth rate of &gt;95%. b. Matrix comparison: Not applicable {5} Page 6 of 6 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: S. pneumoniae and other Streptococcus spp. Interpretive criteria = ≤ 2 (S), 4 (I), ≥ 8 (R) N. Proposed Labeling: The expected value range, interpretive criteria and QC are the same as recommended by FDA and CLSI. All values will be included in the package insert O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...