← Product Code [MOB](/productcode/MOB) · K050715

# QUANTA PLEX ANCA PROFILE (K050715)

_Inova Diagnostics, Inc. · MOB · May 12, 2005 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050715

## Device Facts

- **Applicant:** Inova Diagnostics, Inc.
- **Product Code:** [MOB](/productcode/MOB.md)
- **Decision Date:** May 12, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5660
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K050715 · May 12, 2005 | QUANTA PLEX ANCA PROFILE | Inova Diagnostics, Inc. | Clinically defined patient samples; Normal blood donor samples; Technical service evaluation samples | Retrospective clinical samples were used to perform method comparison studies against predicate ELISA devices and to establish reference ranges/expected values for the new assay. | Method comparison; Clinical samples; Normal blood donors; Retrospective analysis |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Method comparison with predicate device; Retrospective method comparison | 327 clinically defined patient samples (including rheumatoid arthritis, infectious diseases, and atypical ANCA) and 256 normal blood donors; Sample Size: 583 total samples | Predicate ELISA devices (QUANTA Lite MPO/PR3) | Positive and negative percent agreement |

## Indications for Use

The QUANTA Plex™ ANCA Profile is a fluorescent immunoassay for the semi-quantitative detection of anti-MPO and anti-PR3 autoantibodies in human serum. The presence of these antibodies can be used in conjunction with clinical findings and other laboratory aids to aid in the diagnosis of the autoimmune vasculitides microscopic polyarteritis, crescentic glomerulonephritis, and Wegener’s granulomatosis.

## Device Story

QUANTA Plex™ ANCA Profile is a fluorescent immunoassay; detects anti-MPO and anti-PR3 antibodies in patient samples. Used by laboratory professionals to aid in diagnosing autoimmune vasculitides (microscopic polyarteritis, crescentic glomerulonephritis, Wegener's granulomatosis). Results are interpreted by clinicians alongside clinical findings and other laboratory data to support diagnostic decision-making.

## Clinical Evidence

No clinical prospective/retrospective studies performed. Evidence consists of method comparison testing using 583 serum samples (256 normal, 327 clinical). Primary endpoints were agreement with predicate ELISA results. MPO assay showed 95.9% positive agreement and 98.6% negative agreement. PR3 assay showed 98.2% positive agreement and 98.1% negative agreement. Analytical performance included precision (CVs 3-15%), linearity (0-25,000 MFI), and interference studies showing little to no interference from hemolysis, icterus, lipemia, or high IgG.

## Technological Characteristics

Multiplexed fluorescent immunoassay using color-coded microspheres coated with purified MPO, PR3, and IgG control. Detection via PE-labeled goat anti-human IgG conjugate. Instrument: Luminex 100 flow cytometer (vs. 2.2). Energy source: optical (fluorescence). Connectivity: standalone instrument. Software: algorithm for bead discrimination and fluorescence intensity quantification.

## Regulatory Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k050715

B. Purpose for Submission:
This is a new submission.

C. Measurand:
Antineutrophil Cytoplasmic Antibodies (ANCA)

D. Type of Test:
ELISA (Semi-quantitative)

E. Applicant:
INOVA Diagnostics, Inc.

F. Proprietary and Established Names:
QUANTA Plex™ ANCA Profile

G. Regulatory Information:
1. Regulation section:
21 CFR§ 866.5660 Multiple Antibodies Immunological Test System
2. Classification:
Class II
3. Product code:
MOB, Anti-neutrophil cytoplasmic antibodies (ANCA)
4. Panel:
(82) Immunology

H. Intended Use:
1. Intended use(s):
The QUANTA Plex™ ANCA Profile is a fluorescent immunoassay for the semi-quantitative detection of anti-MPO and anti-PR3 autoantibodies in human serum. The presence of these antibodies can be used in conjunction with clinical findings and other laboratory aids to aid in the diagnosis of the autoimmune vasculitides microscopic polyarteritis, crescentic glomerulonephritis, and Wegener’s granulomatosis.
2. Indication(s) for use:
Same as above
3. Special conditions for use statement(s):
The device is for prescription use only.
4. Special instrument requirements:
Luminex 100 tm Flow Cytometer vs.2.2

I. Device Description:
The device consists of the following: a foil package containing 12 (1 X 8) microwell strips with holder (each microwell contains 3 different color beads and each color bead is coated with a different purified antigen (MPO, PR3 and an IgG control), a negative control, a low positive control, a high positive control, HRP sample diluent, PBS concentrate, PE-labeled goat anti-human IgG conjugate (fc-specific) and

{1}

conjugate diluent.

## J. Substantial Equivalence Information:

1. Predicate device name(s):
QUANTA Lite™ MPO IgG ELISA
QUANTA Lite™ PR3 IgG ELISA

2. Predicate 510(k) number(s):
k981330
k981328

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | QUANTA Plex™ ANCA | QUANTA Lite™ MPO and QUANTA Lite™ PR3  |
|  Intended Use | Detection of anti-MPO and anti-PR3 autoantibodies | Same  |
|  Sample type | Serum | Same  |
|  Type of test | Semi-quantitative | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Assay type | Flow Cytometer based | ELISA  |
|  Assay Format | Multiplexed | Individual analytes  |
|  Instrument | Luminometer (Luminex vs. 2.2) | Spectrophotometer  |
|  Detection method | Fluorescence | Colorimetric  |
|  Conjugate | Phycoerythrin | HR  |
|  Solid Phase Capture | Color-coded microspheres | Microwells  |

## K. Standard/Guidance Document Referenced (if applicable):

None referenced.

## L. Test Principle:

Native MPO and PR3 are each bound to different, fluorescently "colored" beads. The two different antigen coated beads, as well as an IgG-coated bead, are mixed together, and put into wells of a microwell plate under conditions that preserve the antigens in their native state. Pre-diluted controls and diluted patient sera are added to separate microwells, allowing any MPO and PR3 autoantibodies present to bind to the immobilized antigen. Then an anti-human IgG conjugated to a fluorescent probe is added to each microwell. A second incubation allows the anti-human IgG fluorescent conjugate to bind to any patient autoantibodies that have become attached to the antigen on the beads. The samples are then measured in the Luminex™ 100 flow Cytometer. This flow Cytometer can discriminate the color of each bead from the others as well as measure the fluorescent intensity of the conjugate on each bead. The

2

{2}

conjugate fluorescent intensity is proportional to the amount of labeled anti-human IgG bound to the patient autoantibodies on the bead. Each antigen can be semi-quantitated by comparing the fluorescent intensity of the patient sample with the fluorescence of the corresponding Low Positive. An IgG-coated control bead is included in each microwell to ensure that false negative results due to operational errors are detected.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

The automated pipetting method precision testing was not performed on this application. On previous applications for the ENA Profile 4, ENA Profile 5, ENA Profile 6, and SLE Profile 8, it was shown that automated method showed same results as the manual method. Only the manual method was performed on this application.

Intra-assay reproducibility was determined by assaying 10 samples 9 times on one plate. A representative data of the data are shown on the table below.

|  Antigen | Average LU | Mean %CV  |
| --- | --- | --- |
|  MPO | 26 | 10%  |
|  MPO | 62 | 5%  |
|  MPO | 161 | 5%  |
|  PR3 | 41 | 15%  |
|  PR3 | 76 | 9%  |
|  PR3 | 281 | 5%  |

Another form of intra-assay variation was tested by running one sample on an entire plate. The processed data for that sample showed a 7% CV for the MPO bead, 5% CV for the PR3 bead and 3% CV for the IgG control bead.

Inter-assay reproducibility was determined by assaying 12 samples 5 times for 5 days. A representative data of the data are shown on the table below.

|  Antigen | Average LU | Mean %CV  |
| --- | --- | --- |
|  MPO | 24 | 14%  |
|  MPO | 63 | 15%  |
|  MPO | 183 | 7%  |
|  PR3 | 41 | 10%  |
|  PR3 | 85 | 10%  |
|  PR3 | 423 | 7%  |

#### b. Linearity/assay reportable range:

Dynamic range of the assay is 0 MFI (median fluorescence intensity) to 25,000 MFI.

#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Negative, low positive and high positive controls are included.

Accelerated stability studies were performed for the ANCA profile beads and the Low and High controls specific for the ANCA profile. Results were

{3}

acceptable. Stability testing was not repeated for components used in other products such as the Quanta Plex negative control, HRP sample diluent, PE conjugate and Conjugate diluent because data for all of these products have been submitted in previous applications.

d. Detection limit:

The Luminex Users manual states that the limit of detection is 1000 fluorochromes of phycoerythrin per microsphere. Serial dilutions of positive sera to end point were performed using the QUANTA Plex™ ANCA and detection limit were set at:

- MPO 5.3 LU
- PR3 3.3 LU

e. Analytical specificity:

Cross-reactivity

To assess cross-reactivity of other autoantibodies, 90 sera samples were tested. The disease control group consisted of 7 patients with rheumatoid arthritis, 3 with Lyme disease, 3 with hepatitis C, 4 with toxoplasmosis, 3 with cytomegalovirus, and 3 with herpes simplex virus. 67 samples that yielded an atypical ANCA pattern or an X-ANCA pattern by immunofluorescence were also included. This class of samples yields positive ANCA immunofluorescence. However, they are expected to be negative for anti-MPO and anti-PR3 because they do not have the typical pANCA or cANCA pattern associated with the latter autoantibodies.

Comparison of samples from the Other Controls on the MPO portion of the ANCA Profile and the MPO ELISA

|  Other Controls n=90 |   | Luminex ANCA Profile (MPO)  |   |
| --- | --- | --- | --- |
|  MPO ELISA
98.8% the same |  | Negative | Positive  |
|   |  Negative | 89 (98.9%) | 1* (1.1%)  |
|   |  Positive | 0 (0%) | 0 (0%)  |

* This sample was negative for ANCA by IFA.

Comparison of samples from the Other Controls on the PR3 portion of the ANCA Profile and the MPO ELISA

|  Other Controls n=90 |   | Luminex ANCA Profile (PR3)  |   |
| --- | --- | --- | --- |
|  PR3 ELISA
96.7% the same |  | Negative | Positive  |
|   |  Negative | 81 (90.0%) | 3* (3.3%)  |
|   |  Positive | 0 (0%) | 6 (6.7%)  |

*One sample was 19 U on ELISA and cANCA by IFA, so was likely a true positive. 2 samples were 11 U by ELISA and atypical ANCA by IFA.

Interfering substances

Interference studies were performed by mixing each of 6 serum samples: 1 normal, 1 hemolyzed (hemoglobin 1000 mg/dL), 1 icteric (bilirubin 29.7) -2 lipemic (one with cholesterol at 369 and one with triglycerides at 1016) and 1 high IgG, in equal volumes with a sera positive for MPO and PR3. Little or no

{4}

interference was observed.

f. Assay cut-off:

The cut-off value was determined using 256 normal samples and 20 arbitrary units were chosen for continuity with the predicate devices. The fluorescence unit that was assigned 20 units was based on non-parametric statistical analysis of the data.

2. Comparison studies:

a. Method comparison with predicate device:

Clinical samples:

Three hundred twenty seven clinically defined patient samples were tested with the new device and the predicate device. These samples included 237 samples from INOVA Technical Service for evaluation of ANCA, 7 patients with rheumatoid arthritis, 16 with infectious diseases, and 67 with atypical ANCA. Age and sex information were provided for 216 samples. There were 125 males and 91 females. The average age of the males was 39.4 years, while the average age for females was 35.9 years. Discrepant results were re-tested by both devices and the retest result was used as the final result.

|  Total sample N = 327 | Both Pos | Both Neg | E Pos Q Neg | E Neg Q Pos | % Positive Agreement | % Negative Agreement | %Total Agreement  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  MPO | 247 | 71 | 3 | 6 | 95.9% | 97.6% | 97.2%  |
|  PR3 | 212 | 106 | 1 | 8 | 99.1% | 96.4% | 97.2%  |

All Samples:

Five hundred eighty three serum samples were tested with the new device and the predicate device. The samples included 256 normal, and the 327 clinical samples.

|  All Samples N=583 | Both Neg | Both Pos | ELISA Neg Q.Plex Pos | ELISA Pos Q.Plex Neg | %Positive Agreement (95% CII) | %Negative Agreement (95% CI) | Percent Agreement  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  MPO | 502 | 71 | 7 | 3 | 95.9% (88.6-99.2%) | 98.6% (97.2-99.4%) | 98.3%  |
|  PR3 | 465 | 107 | 9 | 2 | 98.2% (93.5-99.8%) | 98.1% (96.2-99.0%) | 98.1%  |

b. Matrix comparison:

Serum is the only recommended matrix.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

See assay cut-off.

5. Expected values/Reference range:

Two hundred fifty six from normal blood donors fro INOVA Technical Service

{5}

Library was run on the QUANTA Plex™ ANCA Profile test. The normal range results were shown below.

|  Analyte | % Negative | Range LU* | Mean ±SD (LU)  |
| --- | --- | --- | --- |
|  MPO | 99.6 | 0-41 | 2.1±2.8  |
|  PR3 | 99.2 | 0-11.5 | 1.4±1.6  |

# N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

# O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K050715](https://fda-staging.innolitics.com/device/K050715)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).