ATLANTIS SR PRO2 CORONARY IMAGING CATHETER, MODEL 39014
K050577 · Boston Scientific Corp · DQO · Mar 30, 2005 · Cardiovascular
Device Facts
Record ID
K050577
Device Name
ATLANTIS SR PRO2 CORONARY IMAGING CATHETER, MODEL 39014
Applicant
Boston Scientific Corp
Product Code
DQO · Cardiovascular
Decision Date
Mar 30, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Atlantis SR Pro² catheter is intended for ultrasound examination of coronary intravascular pathology ONLY. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Device Story
Intravascular ultrasound (IVUS) imaging catheter; used with compatible imaging systems (e.g., Clearview Ultra, Galaxy). Input: ultrasonic signals from 40 MHz transducer at distal tip. Operation: rotating drive shaft transmits ultrasound; electromechanical interface connects to Motor Drive Unit (MDU) and imaging system. Output: real-time intravascular ultrasound images. Used in clinical settings by physicians during coronary interventions. Catheter features distal/proximal lumens, telescoping section for 15 cm linear movement, and hydrophilic coating (Bioslide) for lubricity. Flush port with one-way valve requires heparinized saline for acoustic coupling. Output allows visualization of coronary pathology to guide interventional decision-making and improve procedural outcomes.
Clinical Evidence
Bench testing only. Included dimensional testing, sheath bond tensile testing, and functional performance testing of sheath and telescope assemblies. Biocompatibility testing performed per ISO 10993. No clinical data required.
Technological Characteristics
40 MHz rotating ultrasonic transducer; hi-torque flexible drive shaft; hydrophilic (Bioslide) coating on distal 230mm; Dymax 191 M-T adhesive; radiopaque marker at 0.5 cm from distal tip; heparinized saline acoustic coupling; electromechanical MDU interface; sterile, single-use.
Indications for Use
Indicated for patients who are candidates for transluminal coronary interventional procedures requiring ultrasound examination of coronary intravascular pathology.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
# MAR 3 0 2005
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## Boston Scientific/Scimed (BSS) Atlantis™ SR Pro2 Coronary Imaging Catheter
| Submitted by: | Boston Scientific/Scimed<br>IVUS Technology Center<br>47900 Bayside Parkway<br>Fremont, CA 94538 | | | | | | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| Contact Person: | Veronica M. Kocken<br>Regulatory Affairs Specialist II<br>Tel: (510) 624-1431<br>Fax: (510) 624-1449<br>kockenv@bsci.com | | | | | | |
| Date prepared: | March 4, 2005 | | | | | | |
| Proprietary Name: | BSS Atlantis™ SR Pro² Coronary Imaging<br>Catheter | | | | | | |
| Common Name(s): | Ultrasound Diagnostic Imaging Catheter<br>Diagnostic Intravascular Catheter (74DQO)<br>Diagnostic Ultrasonic Transducers (90ITX) | | | | | | |
| Classification Name(s): | Diagnostic Intravascular Catheter, 21 CFR Part<br>870.1200 (74DQO)<br>Diagnostic Ultrasonic Transducer, 21 CFR Part<br>892.1570 (90ITX) | | | | | | |
| Predicate Device(s): | The BSS Atlantis™ SR Pro² Coronary Imaging<br>Catheter is substantially equivalent to the<br>following device:<br>Product 510(k) Clearance<br>Date Atlantis™ SR Plus<br>Coronary Imaging K010707 March 29,<br>2001 | | | | | | |
.
Coronary Imaging Catheter
{1}------------------------------------------------
#### Description of the Device:
The catheters included in the Atlantis™ family of intravascular ultrasound coronary imaging catheters (i.e., SR, SR Plus, SR Pro, and SR Pro') are designed to be used with the Clearview® Ultra with High Frequency (HF) option, Galaxy™ or Galaxy™ 2 intravascular ultrasound imaging instruments. These catheters are comprised of two main assemblies:
- . Imaging core
- . Catheter body
The catheter body is comprised of three sections:
- Distal lumen .
- Proximal lumen .
- . Telescoping section
The imaging core is composed of a hi-torque, flexible, rotating drive shaft with a radially-looking 40 MHz ultrasonic transducer at the distal tip. An electromechanical connector interface at the proximal end makes the connection to the Motor Drive Unit (MDU) which is in turn connected to the imaging system (e.g., Galaxy™ system). The MDU-Catheter interface consists of an integrated mechanical drive hub and an electrical connection.
The catheter body has a distal guidewire lumen with proximal exit port at approximately 1.5 cm from the distal end. The catheter body is attached to the telescope section via a male/female luer connection. A radiopaque (RO) marker is embedded in the catheter body at 0.5 cm from the distal tip. In addition, an insertion depth indicator is located on the catheter body at 105 cm, corresponding to femoral insertions.
Within the catheter body, the distal lumen and proximal lumen sections comprise the "working length" of the catheter, and the telescoping section remains outside
{2}------------------------------------------------
of the quiding catheter. The telescoping section allows the imaging core to be advanced and retracted for 15 cm of linear movement. The corresponding movement of the transducer occurs from the proximal end of the wire exit port, to the proximal end of the distal lumen.
A flush port with a one-way valve is used to displace air near the transducer. The catheter must be flushed with heparinized saline prior to use, as this provides the acoustic coupling media required for ultrasonic imaging. The oneway valve helps retain saline in the catheter during use.
#### Atlantis SR Pro2 catheter
The Atlantis SR Pro2 catheter modifications are being implemented to improve performance and functionality of the catheter. The Atlantis™ SR Pro2 catheter is the same as the Atlantis™ SR Plus catheter, except for the following changes:
- Inclusion of a hydrophilic coating (Bioslide™) to the distal 230mm portion of . the catheter sheath. The purpose of the hydrophilic-coating is to improve catheter lubricity, thus reducing catheter frictional resistance during introduction into the coronary vessel.
- A change is made to switch from Dymax adhesive 1-20271-G to Dymax . adhesive 191 M-T in the Female Telescope assembly. The Dymax adhesive 191 M-T is currently being used within the same product for the Male Telescope assembly. The reason for the change is to reduce material variability in manufacturing of the catheter, and to limit the types of adhesives used in the device. The change is being implemented based on acceptable results from verification testing, in which the bond strength met the acceptance criteria. The female telescope assembly is a section of catheter that remains external to the patient during the application of the device.
{3}------------------------------------------------
#### Intended Use/Indications:
The Atlantis SR Pro2 catheter is intended for ultrasound examination of coronary intravascular pathology ONLY. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
#### Device Technology Characteristics an Comparison to Predicate Device:
The BSS Atlantis™ SR Pro2 Coronary Imaging Catheter utilizes the same basic catheter design as the predicate device, the BSS Atlantis™ SR Plus Coronary Imaging Catheter, cleared in March 29, 2001, and the current Atlantis™ SR Pro, cleared via internal documentation (letter-to-file dated, February 14, 2003 and January 27, 2004). These devices have the same intended use, use the same operating principal, incorporate that same basic catheter design, have the same shelf life, and are packaged using the same materials and processes. The two modifications from the Atlantis™ SR Pro to the Atlantis™ SR Pro2, is the addition of lubricous hydrophilic coating (Bioslide™) to distal segment (230mm minimum) of the catheter sheath, and the use of Dymax 191 M-T adhesive on the telescope assembly.
#### Non-clinical Test Results:
Bench and biological safety testing demonstrate that the Atlantis™ SR Pro2 imaging catheter meets or exceeds the performance requirements and is safe and effective for its intended use.
#### Bench Testing:
Bench testing was performed to evaluate the physical integrity and functionality of the catheter. This testing included dimensional testing, sheath bond tensile testing, and a variety of functional and performance testing of the sheath assembly, the telescope assembly, and the final sterile device. The results demonstrated that the device satisfies all performance, physical, and functional requirements.
{4}------------------------------------------------
#### Biological Safety Testing:
The Atlantis™ Pro2 catheter was subjected to a series of biocompatibility tests per ISO 10993, bioburden, endotoxin, sterility assurance, and latex testing. The results of the biocompatibility testing demonstrate that the catheter is acceptable for its intended use.
### Acoustic Output Testing:
Acoustic Output testing was not required for the Atlantis™ SR Pro2, as the addition of hydrophilic coating to the distal sheath of the catheter will not change the Acoustic Output measurement of the transducer, thus undetectable to the performance of the catheter.
Per the FDA Guidance, Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (September 30 1997), Acoustic Output measurements for the Atlantis™ SR Pro2 catheter remains unchanged from its predicate device, the Atlantis™ SR Plus catheter and the Atlantis™ SR Pro catheter for which a letter-to-file was documented.
Table 2 of this submission provides the Acoustic Output equivalency comparison, followed by a rationale for not conducting Acoustic Output testing on the Atlantis™ SR Pro2 coronary catheter.
#### Conclusion:
The BSS Atlantis™ SR Pro2 Coronary Imaging Catheter utilizes the same design features and has the same intended use as the predicate device, the Atlantis™ SR Plus Coronary Imaging Catheter. The tests support a determination of substantial equivalence of the modified device, the Atlantis™ SR Pro2 Coronary Imaging Catheter to its predicate device.
< 050577
」 : /
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping wave-like shapes, often interpreted as representing the human form. The logo is black and white.
MAR 3 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Boston Scientific Corp. IVUS Technology Center c/o Ms. Veronica Kocken Regulatory Affairs Specialist II 47900 Bayside Parkway Freemont, CA 94538-6515
Re: K050577
Trade Name: Atlantis SR Pro2 Coronary Imaging Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: II (two) Product Code: DQO Dated: March 04, 2005 Received: March 07, 2005
Dear Ms. Kocken:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Ms. Veronica Kocken
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
B/Simimas for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use Statement
| 510(k) Number: | K050577 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Boston Scientific/Scimed, Atlantis™ SR Pro² Coronary<br>Imaging Catheter |
| Indications for Use: | The Atlantis SR Pro² catheter is intended for ultrasound<br>examination of coronary intravascular pathology ONLY.<br>Intravascular ultrasound imaging is indicated in patients<br>who are candidates for transluminal coronary interventional<br>procedures. |
| Prescription Use<br>X | AND/OR Over the counter Use |
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Part 21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummore
(Part 21 CFR 801 Subpart D)
ardiovascular Devices
Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.