The Medicon IMF Orion Screws are intended for use as a bone screw in temporary fixation of the maxilla and mandible, providing indirect stabilization of fractures of the maxilla, mandible or both.
Device Story
Medicon IMF-Orion Screws are titanium alloy (Ti6Al4V) bone screws used for temporary intermaxillary fixation (IMF). Screws feature 2.0 mm shaft diameter, 4.0 mm head diameter, and lengths of 10.0–16.0 mm. The head is grooved for wire/rubber band attachment and includes a tapered inner hexagonal fitting for blade engagement. Used in healthcare facilities (hospitals, surgical centers) by surgeons/dentists to provide rapid, indirect stabilization of maxillary or mandibular fractures. Screws are applied to bone; wires or bands are then attached to the screw heads to stabilize the jaw. Benefits include reduced surgical time for fracture stabilization. Device is provided sterile.
Clinical Evidence
Clinical evidence provided via literature review, including German language publications (OP-Journal 2003, FAMI-Schraube study) and Addenbrooke's NHS study regarding the use of bicortical bone screws for temporary intermaxillary fixation. No specific clinical trial metrics (e.g., sensitivity/specificity) reported; evidence supports safety and effectiveness of the material and application method.
Technological Characteristics
Material: Ti6Al4V titanium alloy per ASTM F136-02a (DIN ISO 5832-3). Dimensions: 2.0 mm shaft diameter, 4.0 mm head diameter, 10-16 mm lengths. Features: Self-tapping/self-drilling, grooved head for wire/band attachment, tapered inner hexagonal drive. Sterilization: Steam (pre-vacuum, 132°C, 4 min hold, 10 min dry) validated per ISO 14937:2000.
Indications for Use
Indicated for temporary fixation of the maxilla and mandible to provide indirect stabilization of fractures of the maxilla, mandible, or both in patients requiring intermaxillary fixation.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Traditional 510(k) - DZL
Medicon eG
.
Image /page/0/Picture/2 description: The image shows a handwritten string of characters. The string appears to be "K050478". The characters are written in a casual, slightly slanted style, with varying stroke thicknesses. The image is in black and white.
# JUN 1 5 2005
## 2.3 Device Names: [807.92(a)(2)]
| Device Names: [807.92(a)(2)] | |
|------------------------------|-------------------------------|
| Proprietary | MEDICON IMF-Orion Screws |
| Common | Small Bone Screw |
| Classification | Screw, Fixation, Intraosseous |
- 2.4. Reason for Submission: [807.81(2)] New Device
## 2.5 Predicate Device [807.92(a)(3)]
| Manufacturer | Synthes (USA); Paoli.<br>K010527 |
|------------------|--------------------------------------------------------------------------------|
| Proprietary Name | Synthes (USA) IMF Screws |
| Catalogue #'s | "IMF Screw Set" GP1852-B<br>see exhibit 4a |
| Manufacturer | Walter Lorenz Surgical, Inc.<br>K983728 |
| Proprietary Name | Lorenz IMF Screws |
| Catalogue #'s | « 2.0/2.4 MANDIBULAR FRACTURE SYSTEM »<br>00-2117, p. 4<br>see exhibit 4b |
| Manufacturer | KLS-Martin L.P.<br>K980760 |
| Proprietary Name | KLS-Martin MMF Screw |
| Catalogue #'s | « FAST-FIX IMF-SYSTEM » 90-848-02<br>see exhibit 4c |
| Manufacturer | Howmedica Leibinger, Inc.<br>K963030 |
| Proprietary Name | Leibinger IMF Screws |
| Catalogue #'s | « IMF Screw System for Intermaxillary Fixation »<br>90-01590<br>see exhibit 4d |
{1}------------------------------------------------
### 2.6 Device Description [807.92(a)(4)]
The MEDICON IMF-Orion Screws are available in different lengths. Threadlengths within the range of 10.0, 12.0, 14.0 mm and 16.0 mm. The diameter of the shafts is 2.0 mm.
The head shape is in a diameter of 4.0 mm. It is grooved for fastening of wires and rubber bands. For a self-retaining fit to the blade the top is tapered inner hexagonal. The screws are manufactured of Ti6Al-4V according to ASTM F 136-02a.
The MEDICON IMF-Orion Screws provide temporary occlusal and fracture stabilization which can be achieved within a matter of minutes. These screws may be applied prior to or after exposure of the fracture.
The shapes, sizes and materials used are substantially equivalent to those of SE devices
### 2.7 Intended Use: [807.92 (a)(5)]
The Medicon IMF Orion Screws are intended for use as a bone screw in temporary fixation of the maxilla and mandible, providing indirect stabilization of fractures of the maxilla, mandible or both.
#### 2.8 Environment of Use
The MEDICON IMF-Orion Screws are intended for use in healthcare facilities, including hospitals, medical clinics and surgical centers.
Within these facilities the MEDICON IMF-Orion Screws may be located in areas where sterile surgical/dental instruments are used such as operating rooms for surgery.
#### 2.9 Difference in Design and Technological Characteristics when Compared to SE Devices [807.92(a)(6)]
- Material: Like the SE devices, the MEDICON IMF-Orion Screws are manufactured from nonmagnetic, high quality Titanium Alloy (Ti6Al4V) acc. to ASTM F136-02a (DIN ISO 5832-3).
- The design is very similar. Both MEDICON and SE devices Design: consist of various sizes regarding length, measurement and fitting of the screw head.
{2}------------------------------------------------
#### Discussion of Safety and Effectiveness [807.92(b)] 2.10
Clinical results of the MEDICON IMF-Orion Screws made out of Titanium Alloy (Ti6Al4V) and manufactured after ASTM F136-02a have shown their safety and effectiveness:
- 1. see exhibit 5a: German Literature OP-Journal 2003 "Verwendung von IMF-Schrauben zur mandibulo-maxillären Fixation"
- 2. see exhibit 5b: German Literature with English abstract p. 361 "Die FAMI-Schraube für die temporäre intermaxillare Fixation"
- 5c: Literature Addenbrooke's NHS "THE exhibit 3. see INTERMAXILLARY SCREW -- A DEDICATED BICORTICAL BONE SCREW FOR TEMPORARY INTERMAXILLARY FIXATION"
#### Industry Standards: [807.92 (d)] 2.11
MEDICON certifies compliance with required ISO/EN/ASTM and other device-related standards that apply to the manufacure, packaging, labeling, sterilization, and reprocessing of subject devices including the validation of these processes.
### Substantially Equivalent
#### Information Bearing on the Safety and Effectiveness: 2.12
[807.92 (b)(3)]
The MEDICON IMF-Orion Screws have the same intended use as predicate devices. They are made of identical material according to ASTM F136-02a.
The slight differences in design and size do not adversely affect the safety and effectiveness of this device.
The results of design validation and clinical testing raise no new issues of safety and effectiveness.
2-4
{3}------------------------------------------------
| | 2.13 Comparison with predicate devices (table) | | | | | |
|--|------------------------------------------------|--|--|--|--|--|
|--|------------------------------------------------|--|--|--|--|--|
| | MEDICON | Walter<br>Lorenz | KLS-Martin | Synthes | Howmedica<br>Leibinger |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------------------------------------|--------------------------|--------------------------------|--------------------------|
| INTENDED USE:<br>The Medicon IMF Orion Screws are<br>intended for use as a bone screw in<br>temporary fixation of the maxilla and<br>mandible, providing indirect stabilization of<br>fractures of the maxilla, mandible or both. | Yes | Yes | Yes | Yes | Yes |
| METHODS OF STERILIZATION:<br>Steam Sterilization to SAL of 10-6<br>(For Medicon IMF Orion Screws:<br>3 times pre-vacuum, 132° C, holding time 4<br>minutes [full cycle], drying time 10 minutes.<br>Validated according to ISO 14937:2000) | Yes | Yes | Yes | Yes | Yes |
| MATERIAL: | Ti6Al4V ASTM<br>F136-02a | Ti6Al4V ASTM<br>F136-02a | Ti6Al4V ASTM<br>F136-02a | 316-L | Ti6Al4V ASTM<br>F136-02a |
| DESIGN:<br>Color, Coding: | Natural silver | Gold,<br>Blue (5.0mm) | Gold | Natural silver | Gold |
| Thread length (mm) | 10, 12, 14, 16 | 5, 7, 9, 11 | 9, 11, 13, 15 | 8, 12 | 8, 14 |
| Thread function | self-tapping &<br>selfdrilling | selfdrilling (blue<br>colored 5.0mm)<br>self-tapping (7,<br>9, 11) | self-tapping | self-tapping &<br>selfdrilling | Self-tapping |
| Headshape secures fast and safe gripping | Yes | Yes | Yes | Yes | Yes |
| Labeling (Packaging shows indications for<br>screw identification) | Assumed | Assumed | Assumed | Assumed | Assumed |
| manufactured after ASTM F 136-02a<br>see Exhibit 3 | Yes | Yes | Yes | Assumed | Yes |
2-5
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure with outstretched arms.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUN 1 5 2005
Mr. Joachim Schmid General Manager Medicon, E.G. Gaensaecker 15 Tuttlingen, GERMANY D-78532
Re: K050478
Trade/Device Name: Medicon IMF Orion Screws Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Implant Regulatory Class: II Product Code: DZE Dated: June 1, 2005 Received: June 2, 2005
Dear Mr. Schmid:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becases and his device is substantially equivalent (for the Indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinone, or to active act (Act) that do not require approval of a premarket the Federal I vou, Drag, Drag, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls providens of a registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Nr.), It hay of be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device can or roundsh further announcements concerning your device in the Eederal Register.
{5}------------------------------------------------
### Page 2 - Mr. Schmid
Please be advised that FDA's issuance of a substantial equivalence determination does not I least that FDA has made a determination that your device complies with other requirements Incall that I DA mas made a aves and regulations administered by other Federal agencies. of the Act of ally I oderal business requirements, including, but not limited to: registration 1 out must comply with and are labeling (21 CFR Part 801); good manufacturing practice allu listing (21 CFR Part 877), labelity systems (QS) regulation (21 CFR Part 820); and if requirements as set form in als quadiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter wifi are in you The FDA finding of substantial equivalence of your device to a premarket notification - results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 11 you desire specific at Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Querie Y. Michaid DMD
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE
## 510(k) Number (if known): K050478
.
## Device Name: Medicon IMF Orion Screws
Indications for Use: The Medicon IMF Orion Screws are intended for use as a Indications for USC. The Mediconf inn. Onxilla and mandible, providing indirect bono of fractures of the maxilla, mandible or both.
Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runoe
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K050478
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.