K050457 · Merete Medical GmbH · KTT · Mar 14, 2005 · Orthopedic
Device Facts
Record ID
K050457
Device Name
MERETE METAFIX SMALL FRAGMENT BONE PLATE SYSTEM
Applicant
Merete Medical GmbH
Product Code
KTT · Orthopedic
Decision Date
Mar 14, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Merete MetaFix Small Fragment Locking Plate System is used for adult and pediatric patients as indicated for small bone fracture fixation. Indications for use include fixation of fractures, osteotomies, non unions of the clavicle, scapula, olecranon, radius, ulner, fibula, metacarpals, metatarsals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone.
Device Story
System consists of profile T-Oblique plates and 3.0 mm locking screws; titanium alloy construction. Locking mechanism creates fixed-angle construction to stabilize fracture or osteotomy reduction. Used by surgeons in clinical settings for internal bone fixation. Benefits include stable fixation in osteopenic bone.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Titanium alloy (Ti-6Al-4V) per ASTM F-136 and ISO 5832/3. Locking plate and screw system providing fixed-angle construction. Small fragment system.
Indications for Use
Indicated for adult and pediatric patients requiring small bone fracture fixation, including fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, radius, ulna, fibula, metacarpals, metatarsals, and middle hand/foot bones, particularly in osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes Small Fragment Locking Compression Plate (LCP) (K000684)
Smith & Nephew Locking Bone Plate System (K033669)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Submission - Merete MetaFix™ Small Fragment Locking Bone Plate System
-
-
. .
| 510(k) Submission - Merete MetaFix™ Small Fragment Locking Bone Plate System | K050457 |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Merete Medical GmbH<br>Alt Lankwitz 102, 12247 Berlin<br>Germany |
| FDA Registration Number: | 3002949614 |
| Contact Person: | Jenik Radon,<br>269 West Seventy-First Street<br>New York, N.Y. 10023<br>Tel. 212- 496-2700 Fax 212- 724-3393 |
| Device Name: | Merete MetaFix Small Fragment Locking Bone Plate System |
| Device Classification: | 21 CFR 888.3030 Single/multiple component Metallic bone<br>fixation appliances and accessories and 888.3040 Smooth<br>/threaded metallic bone fixation fastener. |
| Product Code: | KTT |
| Proposed Regulatory Class: | Class II |
| Predicate Device: | • Synthes Small Fragment Locking Compression Plate (LCP)<br>K000684<br>• Smith & Nephew Locking Bone Plate System<br>K033669 |
| Description of Device: | The Merete MetaFix Small Fragment Locking Plate System<br>consists of profile T-Oblique plates, right or of various left, of<br>various sizes and 3.0 mm locking screws. The system is<br>available in titanium (ASTM F-136). Locking plates/screws<br>incorporate a screw-to-plate locking feature which creates a<br>locked, fixed angle construction to hold fracture or osteotomy<br>reduction. |
| Intended use: | The Merete MetaFix Small Fragment Locking Plate System is<br>used for adult and pediatric patients as indicated for small<br>bone fracture fixation. Indications for use include fixation of<br>fractures, osteotomies, non unions of the clavicle, scapula,<br>olecranon, radius, ulner, fibula, metacarpals, metatarsals,<br>Hallux Valgus osteotomy corrections, middle hand and<br>middle foot bones, particular in osteopenic bone. |
| Technological Characteristics: | The components of the Merete MetaFix Small<br>Fragment Locking Plate System are similar to legally<br>marketed predicate devices listed above in that they share<br>similar indications for use, are manufactured from similar<br>materials and incorporate similar technological<br>characteristics. |
| Potential Risks: | The risks associated with this device are the same as with any<br>metallic internal fixation device. These include but not<br>limited to the following: Delayed or nonunion which may<br>lead to breakage the implant. Bending or fracture of the<br>implant. Metal sensitivity, or allergic reaction to a foreign<br>body. Pain, discomfort, or abnormal sensation due to the<br>presence of the device. |
{1}------------------------------------------------
## 3. Standards
The MetaFix-system is produced from titanium alloy Ti-6Al-4V according to ASTM F-136 and ISO 5832/3.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 4 2005
Merete Medical GMBH c/o Mr. Jenik Radon 269 West Seventy- First Street New York, New York 10023
Re: K050457
Trade/Device Name: Merete MetaFix Small Fragment Locking Bone Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: February 21, 2005 Received: February 23, 2005
Dear Mr. Radon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we nave levelewed your occirco. I substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predices marketed in interstate commerce prior to May 28, 1976, cherosure) to regarry manoledical Device Amendments, or to devices that have been reclassified in inc clance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to approval of a promance approvat affect . The general controls provisions of the Act include the general connols provisions of the read of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be If your device is elassified (600 ac regulations affecting your device affecting your device can be found in the Sublot to sam a Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR requirements, merating, out neguirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 -- Mr. Jenik Radon
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket I list letter will anow you to ogen manential equivalence of your device to a legally marketed predicate nothication. The I DA imaing of substantaries arrathus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact If you desire specific advice tor your do not 100 . Also, please note the regulation entitled, "Misbranding by the Office of Comphance ut (210) 216 - 115 - 115 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 relerence to premainer nontroutent (on the Division of Small Manufacturers, International and your responsibilities under the rion the Driver (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H. Milliken
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K056457/A'
## Indications for Use Merete MetaFix Small Fragment Locking Bone Plate System
510(k) Number (if known):_
Device Name: Merete MetaFix Small Fragment Locking Bone Plate
Indications For Use:
The Merete MetaFix Small Fragment Locking Bone Plate System is used for The Mercic wotan's onlairs as indicated for small bone fracture fixation. Indications for use include fixation of fractures, osteotomies, non unions of the influications for aco intrado intradius, ulnar, fibula, metacarpals, Halllux Valgus Clavicle, Scapara, Shoranon, Tauland and middle foot bones, particular in osteopenic bone.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Si)
Division of Ceneral, Restorative,
and Neurological Devices
510(k) Number K050457
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.