K050453 · Ivoclar Vivadent, Inc. · GEX · May 27, 2005 · General, Plastic Surgery
Device Facts
Record ID
K050453
Device Name
ODYSSEY 2.4G, DENLASER ELITE, MODEL 002-00050
Applicant
Ivoclar Vivadent, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 27, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to be used for a variety of surgical procedures on soft tissue within the oral cavity. Dental Soft Tissue Indications: Sulcular debridement of diseased or fibrous tissue Excision and biopsy Gingivectomy and gingivoplasty Lesion (tumor) removal Fibroma removal Tissue retraction (troughing) Aphthous ulcers Gingival hyperplasia (excision and recontour) Crown Lengthening Operculectomy Frenectomy Photocoagulation Laser Periodontal procedures, including: Laser soft tissue curettage Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.
Device Story
Odyssey 2.4G Diode Laser delivers 810nm laser energy for oral soft tissue surgery; max 5W output. System uses solid-state diode; energy transmitted via proprietary optical fiber to metal handpiece with disposable single-use tips. Operator controls power and pulse settings; activation via foot-switch. Used in dental clinics by professionals. Provides precise tissue cutting/ablation; benefits include reduced bleeding and controlled tissue removal during periodontal or surgical procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with 21 CFR 1040.10, 1040.11, IEC 60601-1:1998+A1, IEC 60601-2-22:1995, and FCC Part 15 verified.
Technological Characteristics
810±20nm solid-state diode laser; 5W max output; fiber optic delivery; fan-cooled; 630-650nm aiming beam; digital emission control; 100-240 VAC power. Complies with IEC 60601-1, IEC 60601-2-22, and 21 CFR 1040.10/11.
Indications for Use
Indicated for dental, oral, and soft tissue surgery in patients requiring soft tissue procedures, including sulcular debridement, excision, biopsy, gingivectomy, gingivoplasty, lesion/fibroma removal, tissue retraction, aphthous ulcer treatment, gingival hyperplasia recontouring, crown lengthening, operculectomy, frenectomy, photocoagulation, and periodontal soft tissue curettage.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name, "U.S. Food & Drug Administration," written below in a smaller font.
November 3, 2021
Ivoclar Vivadent, Inc. Donna Harnett Director 175 Pineview Dr. Amherst, New York 14228
Re: K050453
Trade/Device Name: Odyssey 2.4g, Denlaser Elite, Model 002-00050 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX
Dear Donna Harnett:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 27, 2005. Specifically, FDA is updating this SE Letter as an administrative correction. A second product code was inadvertently included.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Michael Adjodha, OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, 301-796-6276. Michael. Adjodha@fda.hhs.gov.
Sincerelv.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director for Restorative and Surgical Dental Devices DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white circular seal. The seal contains the symbol of the United States Department of Health and Human Services, which is an abstract image of an eagle. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, INC." is written around the edge of the circle. The text is written in a small, sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 7 2005
Ms. Donna Marie Hartnett Director of QA/ Regulatory Affairs and Assistant Corporate Counsel Ivoclar Vivadent Incorporated 175 Pineview Drive Amherst, New York 14228
Re: K050453
Trade/Device Name: ODYSSEY 2.4G DIODE LASER Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX, LYB Dated: February 15, 2005 Received: February 22, 2005
Dear Ms. Hartnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assessment the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The rou may, arerers, maxis of the Act include requirements for annual registration, listing of general vontrols profice, labeling, and prohibitions against misbranding and adulteration.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device it may be sucjoct to bash of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r rease oe advisou that I Dr. Brosan that your device complies with other requirements of the Act that : 19 ederal statutes and regulations administered by other Federal agencies. You must or any I cactar statutes and regaranents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI R rate 077, accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality by over sions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
#### Page 2 – Ms. Donna Marie Hartnett
This letter will allow you to begin marketing your device as described in your Section 510(k) This icher wifi anow you to organ manies of substantial equivalence of your device to a legally premation notification - results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your acon 15 . Also, please note the regulation entitled, Colliation of Complians at (210) = i 6 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = Misolanding by reference to premaritor nibilities under the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Colisal http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Rela
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known): K050453
Device Name: ODYSSEY 2.4G DIODE LASER Indications For Use:
Dental Soft Tissue Indications:
Dental, oral and soft tissue surgery including:
Sulcular debridement of diseased or fibrous tissue Excision and biopsy Gingivectomy and gingivoplasty Lesion (tumor) removal Fibroma removal Tissue retraction (troughing) Aphthous ulcers Gingival hyperplasia (excision and recontour) Crown Lengthening Operculectomy Frenectomy Photocoagulation
Laser Periodontal procedures, including:
Laser soft tissue curettage Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket penoval of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.
Prescription Use __ _ _ _ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
Page 1 of 1
Image /page/3/Picture/14 description: The image shows a handwritten text that appears to be a combination of letters and numbers. The text reads 'keso453' in a cursive style. The characters are written in black ink on a white background, and the handwriting is somewhat slanted.
Testorative
{4}------------------------------------------------
#### MAY 2 7 2005
#### 510(k)
## Summary of Safety and Effectiveness
Ivoclar Vivadent, Inc. 175 Pineview Dr., Amherst, NY 14228 Tel: 716-691-0010 Fax: 716-691-2294
Donna Hartnett, Director of QA/Regulatory Affairs Preparation Date: May 16, 2005
#### Device Name:
| Trade Name: | Odyssey™ 2.4G Diode Laser |
|-------------------------|-------------------------------------|
| Common Name: | 810nm Diode Laser |
| Product Classification: | Laser Instrument, Surgical, Powered |
## Legally Marketed Predicate Devices for Substantial Equivalence:
LaserSmile, Manufactured by BioLase Technology, Inc. (K030539)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
DenLaser 800, Manufactured by CAO Group, Inc. (K003541)
Opus 10 Dental Diode Laser, Manufactured by OpusDent, Inc. (K000990)
Aurora SL Diode Laser, Manufactured by Premier Laser Systems, Inc. (K993285).
#### Rationale for Substantial Equivalence:
The aforementioned laser devices and their accompanying delivery systems share similar indications for use in the oral environment, similar design features including wavelength, mulcations for use in allo off. The devices share similar methods of control operating controls, and neer den 1007
systems, safety features, and performance monitoring. The devices share similar performance specifications including power output and energy type.
{5}------------------------------------------------
## Description of Submitted Device:
The Odyssey 2.4G Diode Laser is a device for delivering laser energy to surfaces within the Odyssey 2.40 Diode Lase is a dones for a do not to reasons of crowth orovide a consistent and the oral cavity. This ellergy is generation of bonus chooming of 5 watts of energy output.
reliable generation of laser energy at 810 ± 20nm for a maximum of 5 water reliable generation of last chergy at 010 = 2016. 10. and 10. a proprietary optical fiber system,
The laser energy is delivered to the surgical site by means of a proprietary I he laser energy is defrected to the safe the site without creating undue risk to which allows for the sale transmission or collateral laser emissions. The device features some the patient of operatory start of certail of conceral hast since and adjustable power output for user demiable settings, mending a swittenant of thinks of there or pulse delivery options.
The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use of the denvery noor so commission with the device. Activation of the disposative single ass tap ompleted by use of a foot-activated switch.
## Intended Uses of the Odyssey 2.4G Diode Laser:
The device is intended to be used for a variety of surgical procedures on soft tissue within the oral cavity.
Dental Soft Tissue Indications:
- Sulcular debridement of diseased or fibrous tissue Excision and biopsy Gingivectomy and gingivoplasty Lesion (tumor) removal Fibroma removal Tissue retraction (troughing) Aphthous ulcers Gingival hyperplasia (excision and recontour) Crown Lengthening Operculectomy Frenectomy Photocoagulation
Laser Periodontal procedures, including:
Laser soft tissue curettage Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket perfodomal pocket
Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium,
{6}------------------------------------------------
# Technological Characteristics and Substantial Equivalence:
The DenLaser 800 uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system of I his system uses a noer denvery system to transmit and that allow for adjusting the output of also leatures a 050mm animing between a continuous or pulsed-mode laser emission. The maximum output of the unit 5 watts.
The Opus 10 Diode Laser uses solid-state diodes to generate laser energy in the infrared region. The system features timing controls that allows for variation in duration or intervals of legion. This system fourses a 630-650nm aiming beam.
The Aurora Laser System uses solid-state diodes to generate laser energy in the infrared region. This system features similar power output, spot size, and equivalent pulse duration to that of the submitted device.
The LaserSmile System uses solid state diodes to generate laser energy in the infrared region. The system features similar power output to that of the submitted device.
#### Performance Standards:
The Odyssey 2.4G Diode Laser complies with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated July 26, 10-0.10 and 10-0.11, with IEC 60601-1:1998+A1, IEC 60601-2-22:1995, and IEC 2001. The device also complies with FCC regulations Part 15 for wireless devices.
{7}------------------------------------------------
# PERFORMANCE DATA - Comparison Table
| | Ivoclar Vivadent, Inc.<br>Odyssey 2.4G Diode<br>Laser | CAO Group, Inc.<br>DenLaser 800 | OpusDent, Inc.<br>Opus 10 Dental<br>Diode Laser | Premier Laser<br>Systems, Inc.<br>Aurora SL Diode<br>Laser | LaserSmile<br>Diode Laser<br>Biolase<br>Technology,<br>Inc. |
|-----------------------|-------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|
| Wavelength | 810±20 nm | 810±20 nm | 810 nm | 810 nm | 810nm |
| Power | 5 watts | 5 watts | 10 watts | Unavailable | 10 watts |
| Aiming Beam | 630-650 nm, 3mW | 630-650 nm, 3mW | 630-680 nm,<br>power unavailable | 630-680 nm,<br>power unavailable | 630-670 nm |
| Cooling System | Fan air cooled | Fan air cooled | Fan air cooled | Fan air cooled | Fan air cooled |
| Pulse Control | Digital emission<br>control | Digital emission<br>control | Digital emission<br>control | Digital emission<br>control | 20 ms-9.9 sec |
| Laser Source | Solid-state diode | Solid-state diode | Solid-state diode | Solid-state diode | Solid-state<br>diode |
| Power<br>Requirements | 100-240 VAC @ 50-<br>60 Hz, 1.5A<br>(switchable) | 100-240 VAC @<br>50-60 Hz, 1.5 A<br>(switchable) | 110-120VAC @<br>50-60 Hz, 2.0A or<br>220-240VAC @<br>50-60 Hz, 1.2A | 115VAC @ 50-60<br>Hz, 0.8A<br>Info on 220VAC<br>unavailable | Not available |
| Dimensions | 10" x 8" x 4" | 10" x 8" x 4" | unavailable | 12" x 12" x 6" | 8.5" x 9" x<br>12.5" |
| 510(k) Number | Pending this<br>application | K003541 | K000990 | K993285 | K030539 |
#### Conclusion
The Odyssey 2.4G Diode Laser is substantially equivalent to the listed laser surgical devices without raising any issues of safety or effectiveness. This device shares similar intended devices without raising any issues of baracteristics. The device is designed to comply uses, and similar ranscroninternational safety and performance standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.