When intended to promote fusion of the cervical spine and occipito-cervicothoracic junction (occiput-T3), the PROTEX™ CT Spinal System is indicated for: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, atlanto/axial fracture with instability, occipitocervical dislocation, revision of previous cervical spine surgery, and tumors. The use of polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine. Occipital bone screws are limited to occipital fixation; they are not intended for fixation of the posterior cervical spine. The PROTEX™ CT System can also be linked to titanium rod systems ranging in diameter from 3.7mm to 6.5mm, including the PROTEX™ System, using parallel connectors.
Device Story
PROTEX™ CT Cervicothoracic Spinal System; spinal fixation hardware including rods, polyaxial screws, hooks, locking caps, t-connectors, lateral connectors, and occipital clamps. Titanium alloy construction. Used by surgeons in clinical settings to stabilize and promote fusion of cervical spine and occipito-cervicothoracic junction. System allows linkage to other titanium rod systems (3.7mm-6.5mm) via parallel connectors. Provides mechanical stabilization to address degenerative conditions, fractures, and tumors. Output is rigid structural support for spinal segments.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Titanium alloy implants; includes rods, polyaxial screws, hooks, and various connectors. Designed for spinal fixation. No software or electronic components.
Indications for Use
Indicated for patients requiring fusion of the cervical spine and occipito-cervicothoracic junction (occiput-T3) due to degenerative disc disease, spondylolisthesis, spinal stenosis, fracture, dislocation, atlanto/axial fracture with instability, occipitocervical dislocation, revision surgery, or tumors. Polyaxial screws contraindicated for cervical spine; limited to T1-T3 thoracic conditions. Occipital screws limited to occipital fixation.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
K056391 9/2
JUL 2 1 2005
#### = 510(K) Summary
#### SUBMITTED BY:
Globus Medical Inc. 303 Schell Lane Phoenixville, PA 19460 (610) 415-9000 x218 Contact: Kelly J. Baker
### DATE PREPARED:
February 15, 2005
### DEVICE NAME:
PROTEX™ CT Cervicothoracic Spinal System
# CLASSIFICATION:
Per CFR as follows: 21 CFR §888.3050 Spinal Interlaminal Fixation Orthosis 21 CFR §888.3070 Pedicle Screw Spinal System Product codes KWP and MNI. Panel code is 87. Device classification Class II.
## PREDICATE DEVICES:
Danek Vertex (K003780, K042524) Depuy Summit (K002733, K041203) Howmedica Osteonics (Stryker) Oasys (K032394) Synthes Cervifix/Axon (K023675)
#### DEVICE DESCRIPTION:
The PROTEX™ CT Spinal System consists of rods, polyaxial screws, hooks, locking caps, t-connectors, lateral connectors, parallel connectors, and occipital clamps. The implants are composed of titanium alloy.
#### INTENDED USE:
When intended to promote fusion of the cervical spine and occipito-cervicothoracic junction (occiput-T3), the PROTEX™ CT Spinal System is indicated for: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, atlanto/axial
{1}------------------------------------------------
Koscill 1½
fracture with instability, occipitocervical dislocation, revision of previous cervical spine surgery, and tumors.
The use of polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine. Occipital bone screws are limited to occipital fixation; they are not intended for fixation of the posterior cervical spine.
The PROTEX™ CT System can also be linked to titanium rod systems ranging in diameter from 3.7mm to 6.5mm, including the PROTEX™ System, using parallel connectors.
#### PERFORMANCE DATA:
Mechanical testing was conducted to evaluate performance as a basis for substantial equivalence.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines representing movement or progress.
Public Health Service
JUL 2 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kelly Baker, Ph.D. Project Manager, Regulatory Affairs Globus Medical, Inc. 303 Schell Lane Phoenixville, Pennsylvania 19460
-
Re: K050391
Trade/Device Name: PROTEX™ CT Cervicothoracic Spinal System Regulation Number: 21 CFR 888.3070 and 21 CFR 888.3050 Regulation Name: Pedicle screw spinal system and Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP and MNI Dated: July 8, 2005 Received: July 11, 2005
Dear Dr. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regally the enactment date of the Medical Device Amendments, or to commerce provided in accordance with the provisions of the Federal Food, Drug, de vices that have been require approval of a premarket approval application (PMA). allu Cosmetic rice (71ct) that to not requent to the general controls provisions of the Act. The Tou may, therefore, thanks of the Act include requirements for annual registration, listing of general controls provisions of wactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may be subject to sach additions, Title 21, Parts 800 to 898. In addition, FDA can be found in the Could cements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Drive issuates of your device complies with other requirements of the Act that I Dri has made a acterinalations administered by other Federal agencies. You must or any I catales and the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K I at 607), laboring (21 CFR Part 820); and if applicable, the electronic forth in the quality by tickles (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Kelly Baker, Ph.D.
This letter will allow you to begin marketing your device as described in your section 510(k) I his letter will anow you to obgin marketing your antial equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivalence of your premarked notification. "The PDT Intamg of casification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your act (240) 276-0120 . Also, please note the regulation entitled, p contact the Office of Comphance at (210) = + = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = " Misbranding by reference to premarket notification (in i se Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Consulter Abouter Provinces on on the Propport/index.html.
Sincerely yours,
Em
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement II.
| 510(k) Number: | K050391 |
|----------------|------------------------------------------|
| Device Name: | PROTEX™ CT Cervicothoracic Spinal System |
#### Indications:
When intended to promote fusion of the cervical spine and occipito-cervicothoracic junction (occiput-T3), the PROTEX™ CT Spinal System is indicated thorable junotion (oose ase (as defined by neck pain of discogenic origin for: degenerative also also confirmed by patient history and radiographic win't degeneration of the is, spinal stenosis, fracture, dislocation, atlanto/axial Studies), Spondylollothools, Spinal real dislocation, revision of previous cervical spine surgery, and tumors.
The use of polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed (11-19) in trouting theracted bone screws are limited to occipital fixation; they in the our noal of fixation of the posterior cervical spine.
The PROTEX™ CT System can also be linked to titanium rod systems ranging The PROTEX - GT System Smm, including the PROTEX™ System, using parallel connectors.
Prescription Use × (Per 21 CFR §801.109) Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
E. A.
.
Siston Sign-Off) Tivision of General, Restorative ad Neurological Devices
113 Minininer_________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.