INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODEL IS2000
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Apr 29, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intuitive Surgical da Vinci Surgical System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing and delivery and placement of microwave ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.
Device Story
The da Vinci Surgical System is a computer-controlled endoscopic instrument system used in operating rooms by trained physicians. It consists of a Surgeon Console, a Patient Side Cart (PSC), and the Insite Vision System. The surgeon operates from the console using Master Tool Manipulators (MTM) to control endoscopic instruments and an endoscope held by Patient Side Manipulators (PSM) and an Endoscope Camera Manipulator (ECM) on the PSC. The vision system provides 3-D or 2-D images of the surgical field to the surgeon. Commands are relayed from the console to the PSC via cable. The system facilitates precise tissue manipulation, including grasping, cutting, dissection, suturing, and ablation. By providing enhanced visualization and ergonomic control, the system assists surgeons in performing complex minimally invasive procedures, potentially improving surgical accuracy and patient outcomes.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is supported by design analysis, comparison to predicate devices, and risk analysis activities.
Technological Characteristics
Computer-controlled endoscopic instrument system. Components: Surgeon Console (Model IS2000-SSC), Patient Side Cart (Model IS2000-PSC), and Insite Vision System (Model VS1000). Features stereo endoscope, camera, light source, and manipulators (PSM, ECM). Connectivity via cable between console and cart. System utilizes electromechanical components for instrument control. Software-driven interface for surgeon control of robotic arms.
Indications for Use
Indicated for trained physicians performing urologic, general laparoscopic, gynecologic laparoscopic, general noncardiovascular thoracoscopic, and thoracoscopically assisted cardiotomy procedures, including coronary anastomosis via adjunctive mediastinotomy, using endoscopic instruments and accessories for tissue manipulation.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K990144)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K002489)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K011002)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K013416)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K021036)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K022574)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K040237)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K043153)
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments (K050005)
Submission Summary (Full Text)
{0}------------------------------------------------
### Section II
# APR 2 9 2005
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
### 510(k) Number: K050369
| Date | April 26, 2005 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Intuitive Surgical, Inc.<br>950 Kifer Road<br>Sunnyvale, CA 94086 |
| ER Number | 2955842 |
| Contact | Mike Yramategui<br>Director, Regulatory Affairs<br>Telephone: (408) 523 - 2145<br>Fax: (408) 523 - 1390<br>e-mail: mike.yramategui@intusurg.com |
| Subject<br>Device | Name: Intuitive Surgical® da Vinci® Surgical System and Endowrist™<br>Instruments<br>Classification Name: System, Surgical, Computer Controlled Instrument<br>(21 CFR 876.1500)<br>Common Name: Endoscopic Instrument Control System, Endoscopic<br>Instruments, and accessories |
| Predicate<br>Devices | Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments<br>(legally marketed under K990144 / K002489 / K011002 / K013416 /<br>K021036 / K022574 / K040237 / K043153 / K050005) |
| Device<br>Description | The da Vinci Surgical System, consists of three integrated sub-systems as<br>follows: |
| | Endoscopic Instrument Control System: The Endoscopic Instrument<br>Control System is comprised of two sub-systems - the Surgeon Console,<br>Model IS2000-SSC, and Patient Side Cart. Model IS2000-PSC. While<br>seated at the Surgeon Console, the surgeon controls critical aspects of the<br>procedure, including movement of the endoscopic instruments and<br>endoscope, within the operative field. Endoscopic instrument and camera<br>movements are controlled by the surgeon through |
{1}------------------------------------------------
#### Device Description (continued)
use of the Master Tool Manipulators (MTM), two hand operated mechanisms residing within the Surgeon Console. The endoscopic instruments are held in a lixed position (with respect to the patient) by cither two (or optionally three) unique manipulators known as Patient Side Manipulators (PSM), which are located on the Patient Side Cart (PSC). The endoscope is also held in a fixed position (with respect to the patient) by another manipulator, similar to the PSM, known as the Endoscope Camera Manipulator (ECM) and also located on the PSC). The PSM and ECM are attached to surgical arms on the PSC known as Set-up Joint (SUJ) arms. Commands from the Surgeon Console arc relayed to the PSC, which is located immediately adjaccnt to the patient, via a cable. Instrument and endoscope changes are performed by another provider positioned adjacent to the PSC.
Stereo View Endoscopic Vision System: The endoscopic vision system used with the da Vinci Surgical System, also known as Intuitive Surgical Insite Vision System. Model VS1000, consists of a stereo endoscope, endoscopic camera, and various accessories, including a light source and light guides. The Insite Vision System provides two independent images that are relayed to the viewer located in the Surgeon Console, where they are fused to form a 3-D (or alternatively a 2-D image) image of the surgical field.
#### Intended Use
The Intuitive Surgical da Vinci Surgical System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, clectrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing and delivery and placement of microwave ablation probos and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.
{2}------------------------------------------------
| Comparison to<br>Predicate<br>Device | This 510(k) notification is submitted for design modifications to the<br>Endoscopic Instrument Control System. These modifications affect<br>primarily the PSM, ECM, and the Patient Side Cart in order to enhance<br>case of use. There are also associated minor changes to the Surgeon<br>Console and endoscopic instruments. The user interface design is<br>essentially identical to the predicate device except for minor<br>modifications to accommodate the aforementioned changes. |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device operation, functionality, and methods of use for the subject device<br>are identical to the predicate device. The technology, materials,<br>manufacturing methods, and performance are essentially the same for the<br>predicate device as the subject device. The primary differences are 1)<br>changes to the PSM and ECM form factor and associated<br>electromechanical components; 2) changes to the Patient Side Cart<br>mechanics and circuitry to accommodate the modified PSM and ECM;<br>and 3) changes to the Surgeon Console and software to accommodate the<br>modified Patient Side Cart. Minor mechanical modifications to the<br>endoscopic instruments to interface with the modified PSM are also<br>described. |
| Technological<br>Characteristics | The technological characteristics of the subject devices are essentially the<br>same as for the predicate devices. |
| Performance<br>Data | Design analysis and comparison, as well as bench testing and risk<br>analysis activities, have been conducted to confirm that the<br>characteristics of the modified device are substantially equivalent to the<br>predicate devices cited. |
| Conclusion | Based upon the device's general specifications, intended use, and results<br>of risk analysis and performance testing provided in this pre-market<br>notification, the da Vinci Surgical System described herein has been<br>shown to be substantially equivalent to current legally marketed predicate<br>device. |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
1
、
:
:
·
---------
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 2 9 2005
Mr. Michael H. Yramategui Director, Regulatory Affairs Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, California 94086
Re: K050369
K050507
Trade/Device Name: Intuitive Surgical da Vinci Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NA Y Dated: February 10, 2005 Received: February 14, 2005
Dear Mr. Yramategui:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 310(x) premainter is substantially equivalent (for the indications referenced above and nave determined the arrest are devices marketed in interstate for use stated in the enclosure) to tegally manced profithe Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 20, 1970, the enactition with the provisions of the Federal Food, Drug, devices that have been recrassified in accordance while approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act . The and Cosmetic Act (Act) that do not require apperal controls provisions of the Act. The Act. The
You may, therefore, market the device, subject to the general consucted pos You may, inerefore, market the device, belyer to meet general controls provisions of the Fet merade requirements of the management misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mis chairs are regulations affecting your device can
may be subject to such additional controls. Existing major regulation FDA may may be subject to such additional controlis. Entroling and on 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to Register be found in the Code of Peacharies.
publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualice of a substitive only other requirements of the Act
that FDA has made a determination that your device acteral agencies. You must that FDA has made a determination that your arrest by other Federal agencies. You must and or any Federal statutes and regulations and united to: registration and listing (21)
comply with all the Act's requirements, including, but not one provine moutrements as set comply with all the Act s requirements, mendanturing practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR Part 807); labeling (21 CFR 1 at 601); god manat 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000-1050 forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000
{4}------------------------------------------------
Page 2 – Mr. Michael H. Yramategui
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your article equivalence of your device to a legally
premarket notification. The FDA finding of sybstantial end this permits yo premarket notification. The FDA multing of substantial equive and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs of the may of the more of 0115. All any desse not the requiries entitled If you desire specific advice for your ac not on on success note the regulation entitled, the regulation entitled, contact the Office of Complaned at (210) 276 - 1 - 1 - (21)FR Part 807.97). You may obtain "Misbranding by reference to premarket nothleadon (en oet from the Division of Small
other general information on your responsibilities under the Act from number (800) 638-20 other general information on your responsion.com its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-findex html Manufacturers, international and Consulter Abolokation and Consultances of the new onlines.html.
Sincerely yours,
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K050369
Intuitive Surgical® da Vinci® Surgical System and Endowrist 60 Device Name: Instrumicats
Indications For Use:
The Intuitive Surgical" da Vinci" Surgical System is intended to assist in the accurate control of Intuitive Surgical® Endoscopic Instruments including rigid endoscopes, blunt and sharp cadoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, clectrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and shapp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave ablation probos and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to norform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating toom environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDCO)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
tive
Ko 50369
Page 1 of 1
Intuitive Surgical" da Vinci Surgical System
SIO(k) Section III, Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.