Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. Revision of previously failed total hip arthroplasty. Cemented and Uncemented Applications
Device Story
E-Poly™ Acetabular Liners are orthopedic hip implants fabricated from ultra-high molecular weight polyethylene (UHMWPE) infused with medical-grade Vitamin E. The device is intended for use in total hip arthroplasty. The manufacturing process involves high cross-linking of the polyethylene to improve wear characteristics. The liners are designed for both cemented and uncemented applications. Surgeons implant these components during hip replacement procedures to restore joint function and alleviate pain associated with degenerative joint disease or fractures. The device serves as the bearing surface within the acetabular component of the hip prosthesis, facilitating articulation with the femoral head. By utilizing Vitamin E-infused, highly cross-linked polyethylene, the device aims to provide enhanced wear resistance compared to standard UHMWPE, potentially increasing the longevity of the hip implant and reducing the need for revision surgery.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical laboratory testing confirming device functionality within its intended use.
Technological Characteristics
Material: Ultra-high molecular weight polyethylene (UHMWPE) per ASTM F-648. Modification: Higher cross-linking process and infusion with medical-grade Vitamin E. Application: Acetabular liners for hip arthroplasty. Compatibility: Cemented and uncemented applications.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, non-union, femoral neck/trochanteric fractures of the proximal femur, or those requiring revision of failed total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
ArCom® Polyethylene Liners and Components (K023357)
{0}------------------------------------------------
K050327 page '98
SEP 2 7 2005
BIOMET
## 510(k) Summary
| Sponsor: | Biomet Manufacturing Corp.<br>P.O. Box 587<br>Warsaw, IN 46581-0587 |
|-----------------|---------------------------------------------------------------------------------------------------|
| Contact Person: | Tracy Bickel Johnson, RAC<br>Regulatory Associate<br>Biomet Manufacturing Corp.<br>(574) 267-6639 |
Vitamin E Acetabular Liners (E-Poly™) Proprietary Name:
Common Name: UHMWPE Liners
prosthesis, hip, semi-constrained, metal/polymer, cemented (888.3350);
comented on Classification Name(s): prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or
prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemi-constrained prosthesis, hip, semi-constrained, mictainsis, hip, semi-constrained,
non-porous, uncemented (888.3353); prosthesis, hip, semi-constrained, non-porous, uncemented (606.355); prosultsion inp.
uncemented metal/ polymer, non-porous, calcium phosphate (888.3358); uncemented metal/ polymer, non-porous, ourner, porous (888.3358);
prosthesis, hip, semi-constrained, metal/polymer, porous (888.3358); prosthesis, hip, semi-constrained, metal, polymer, por, while has and (888.3358)
Substantially Equivalent Devices: -ArCom® Polyethylene Liners and Components (K023357)
- Arcon Polyers - - Binch - Binch - - Bingl -RingLoc® 36mm Liners (K032396)
Device Description: Biomet Manufacturing Corp. is modifying the manufacturing process of ultra-high Device Description: Biomet Manufacturing Corp. Is modifylene acetabular components.
molecular weight polyethylene (UHMVPE) used in the fabrication of polyethylene. The high molecular weight polyethylene (UHWVP) dised in the fabrication of polyethylene. The highly cross-linked
The modified manufacturing process results in a higher cross-linked p UHMWPE is infused with medical grade Vitamin E.
## Indications for Use:
- tions for Use:
Non-inflammatory degenerative joint disease including osteoarthritis and avascular 1) necrosis.
- Rheumatoid arthritis. 2)
- Correction of functional deformity. 3)
- Correction of functional delomity.
Treatment of non-union, femoral neck fracture, and trochanteric fractures of the 4) Treatment of non-unfor, remoral neck itserent, unmanageable using other techniques.
proximal femur with head involvement, unmanageable using other techniques.
The management,
- Revision of previously failed total hip arthroplasty. 5)
Cemented and Uncemented Applications
MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587 SHIPPING ADDRESS 56 E. Bell Drive Warsaw, IN 46582
100
OFFICE 574.267.6639
FAX 574.267.8137
E-MAIL biomet@biomet.com
{1}------------------------------------------------
page 2 of 8
.
## 510(k) Summary- Page 2 of 3 910(K) Manufacturing, Corp. Blomet Manufacturing, 2017
Vitamin E Acetabular Liners (E-Poly™)
| Summary of Technologies: | The intended use, indications, contraindications, and design specifications of the subject components remain identical to their predicate component counterparts. The raw material being utilized in the manufacture of both the subject and the predicate devices remains a ultra-high molecular weight polyethylene (UHMWPE) per ASTM F-648. The modifications to the manufacturing process of this polyethylene will be introduced in order to create a higher cross-linked polyethylene. The safety and effectiveness of this cross-linked polyethylene in acetabular applications, as well as the proposed wear claims, are adequately supported by the substantial equivalence information, materials data, and testing results provided within this Premarket Notification. |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-Clinical Testing: | Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use. |
| Clinical Testing: | None provided as a basis for substantial equivalence. |
All trademarks are property of Biomet, Inc.
,
・
:
·
and the comments of the comments of
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes extending from its head, representing health, services, and human needs. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.
Public Health Service
SEP 2 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Tracy Bickel Johnson, RAC Regulatory Associate Biomet Manufacturing Corporation 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46582
Re: K050327
Trade/Device Name: E-Poly™ Acetabular Liners Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH, JDI, LWJ, MAY Dated: July 27, 2005 Received: July 28, 2005
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to conimered pror to riay 2011-11-11 in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic For (110) in the device, subject to the general controls provisions of the Act. The r ou may, ateres, and on of the Act include requirements for annual registration, listing of general controls profice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may or sations to dates the regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i that FDA has made a determination that your device complies with other requirements of the Act that I Dr Has Intatutes and regulations administered by other Federal agencies. You must of any I edetar statutes and states and the mot limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF N Fart 677), accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Ms. Tracy Bickel Johnson, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your actres of your device of your device to a legally
premarket notification. The FDA finding of substantial equivaleond thus, premarket notification. The FDA Inding of substantial equivaliation of your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rst led If you desire specific advice for your de not one of a cases note the regulation entitled,
contact the Office of Compliance at (240) 276-0120. contact the Office of Compliance at (240) 270 of the PR Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (214) at first the Division of Sma "Misbranding by relerence to premarkst noutheation" (1993 from the Division of Small
other general information on your responsibilities under the Act from pumber (800) 638-20 other general information on your responsionites at its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-idemanan httpl Manufacturers, International and Consultion Pressource agov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark Milliken
Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation . Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(**k**) Number (if known):
Device Name: E-Poly™ Acetabular Liners
Indications For Use:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular 1) necrosis.
- 2) Rheumatoid arthritis.
- 2) Rheumatold aframational deformity.
- 3) Correction of functional deformity.
4) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the
- Treatment of non-union, femoral neck rracture, and dochanteriories with the more of the manager.
proximal femur with head involvement, unmanageable using other techniques.
Th proximal remur with nead involvanians, 5) Revision of previously.
5) Revision of previously failed total hip arthroplasty.
Cemented and Uncemented Applications
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
er-The-Counter Use
(CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark M. Millman
(Division Sign-Off) (Division Sigand Neurological Devices
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.