The Aequalis Reversed Shoulder Prosthesis is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation. When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis. When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Device Story
Aequalis Reversed Shoulder Prosthesis is a semi-constrained total shoulder system; designed to medialize the center of rotation and lengthen the deltoid muscle lever arm to restore shoulder function in patients with non-repairable rotator cuff tears. System comprises humeral and glenoid components; humeral components are cemented; glenoid components are non-cemented, fixed with 4 screws. Modification adds multidirectional screws and compatible glenoid baseplate; adds 'fish-scales' to humeral insert peg; updates polyethylene packaging. Used by orthopedic surgeons in clinical settings for primary or revision arthroplasty. Provides mechanical joint replacement to reduce pain and restore mobility when rotator cuff muscles are non-functional.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design and bench testing.
Technological Characteristics
Semi-constrained shoulder prosthesis. Materials: Metaphyseal inserts (UHMWPE, ISO 5834-2); glenoid base and anchoring screws (Titanium alloy, ISO 5832-3). Features: Multidirectional screw fixation, humeral insert with 'fish-scales' for fixation. Non-cemented glenoid fixation; cemented humeral fixation.
Indications for Use
Indicated for patients with functional deltoid muscle requiring total shoulder replacement for pain relief and disability due to arthropathy associated with massive, non-repairable rotator cuff tears; also indicated for prosthetic revisions involving such tears.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
MAR 16 2005
TORNIER
Implants Chirurgicaux
**Summary of Safety and Effectiveness information**
Special 510(k) – AEQUALIS Reversed Shoulder Prosthesis
Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92
| 1) Device name | |
|----------------------|-------------------------------------------------------------------|
| Trade name: | AEQUALIS Reversed Shoulder Prosthesis |
| Common name: | Total-Shoulder System and Hemi-Shoulder System |
| Classification name: | Shoulder joint metal/polymer semi-constrained cemented prosthesis |
## 2) Submitter
Tomier S.A. B.P. 11 - Ruc Doyen Gosse 38330 Saint Ismier - France
#### 3) Company contact
Tomier S.A. Mrs Mireille Lémery Regulatory affairs & Quality Engineer ZIRST - 161, rue Lavoisier 38330 Montbonnot - France Tel: 00 33 4 76 61 38 98 Fax: 00 33 4 76 61 35 33 e-mail : mireille.lemery@tomier.fr
4) Classification Device class: Class II Classification panel: Orthopedic KWS Product code: 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.
5) Equivalent / Predicate device Aequalis Reversed Shoulder Prosthesis, TORNIER SA, K030941, K041873 Acqualis Shoulder System, TORNIER SA, K952928 Bipolar Shoulder Prosthesis, BIOMET Inc, K991585 Delta Shoulder, DePuy Inc. K021478
### 6) Device description
The Aequalis Reversed Shoulder Prosthesis is intended to relieve pain and significant disability following massive and non repairable cuff-tear associated to arthropathy and following massive cuff-tear arthropathy. In this case, the rotator muscles of the shoulder (supraspinatus, teres minor and long head of the biceps) are no more useful for mobility, and only the dettoid (for abduction and external rotation) and the subscapularis (for internal rotation) are functional.
Page 1/ page 2
Image /page/0/Picture/14 description: The image shows a logo inside of a hexagon. The logo appears to be the letter 'T' with some additional design elements. The 'T' is stylized with a textured pattern, and there are additional lines and shapes surrounding it, all within the hexagon's border. The image is in black and white.
TORNIER S.A 161, rue Lavoisier 38330 MONTBONNOT FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A. au capital de 1 800 000 F SIRET : 070 501 275 000 13 R.C.S. : B 070 501 275 CODE APE : 331 B
SIEGE SOCIAL : B.P. 11 - rue du Doyen Gosse - 38330 SAINT-ISMIER - FRANCE
{1}------------------------------------------------
# Implants Chirurgicaux
Therefore, the usual goal of such surgery is to restore the shoulder joint to facilitate its working condition And to reduce or eliminate pain. The Aegualis Reversed Shoulder Prosthesis is intended to accomplish these goals. Is reversed design allows to medialize the center of rotation of the shoulder, lengthening the deltoid muscle lever arm.
The Aequalis Reversed Shoulder Prosthesis is a semi-constrained system composed of a humeral and a glenoid parts.
The present device modification submission consists in the following changes:
1ª change: addition of components in order to have the possibility to use multidirectional screws with the glenoid baseplate
2 types of components are added to the components of the Aequalis Reversed Shoulder prosthesis:
- Multidirectional screws in various length, l
- Specific glenoid baseplate for multidirectional screws.
- 244 change: addition of "fish-scales" on the peg of the humeral insert
- 314 change: modification of packaging of polyethylene components
The indications for use already covered by the previous 510(k) clearance are not modified.
# 7) Materials
Metaphyseal inserts are made of ultra high molecular weight polyethylene (UIMWPE) according to ISO standard 5834-2. The base of the glenoid implant is manufactured from Titanium alloy according to ISO standard 5832-3 and the anchoring screws are manufactured from Titanium alloy according to ISO standard 5832-3.
# 8) Indications
The Aequalis Reversed Shoulder Prosthesis is indicated for patients with a functional deltoid musche as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when gleaoid bone fracture occurs during the surgical procedures, the hemiprosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemiprosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Page 2/ page 2
Image /page/1/Picture/18 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is bold and appears to be constructed from thick, slightly irregular lines, giving it a textured or hand-drawn feel. The hexagon provides a clear, geometric border around the letter, making the logo distinct and easily recognizable.
TORNIER S.A 16), rue Lavoisier 38330 MONTBONNOT FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A. au capital de 1 800 000 F SIRET : 070 501 275 000 13 R.C.S. : B 070 501 275 CODE APE : 331 B
STEGE SOCIAL : B.P. 11 - rue du Doycn Gosse - 3830 SAINT-ISMIER - FRANCE
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, which is a traditional symbol of medicine, but in this case, it is a more abstract representation.
Public Health Service
MAR 1 6 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mrs. Mireille Lémery Regulatory Affairs and Quality Engineer Tornier S.A. 161 rue Lavoisier 38330 Montbonnot France
Re: K050316
Trade/Device Name: Aequalis Reversed Shoulder Prosthesis Regulation Number: 888.3660 Regulation Name: shoulder joint metal / polymer scmi-constrained cemented prosthesis Regulatory Class: II Product Code: KWS Dated: February 4, 2005 Received: February 14, 2005
Dear Mrs. Lemery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mrs. Mireille Lémery
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter witi anow you to begin maneting of substantial equivalence of your device to a legally premits that notification. The PDA mainly of your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you ucate specific advice io. Jour 240) 276-0210. Also, please note the regulation entitled, comaci the Office of Comphalled as (21 millication" (21CFR Part 807.97). You may obtain of Small other general information on your respense at its toll-free number (800) 638-2041 or Manufacturers, Internet and Consantip://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
L. Mark N. Mikkelsen
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
:11 koso316
# Indications for Use
510(k) Number (if known):
Device Name: Aequalis Reversed Shoulder Prosthesis
#### Indications For Use:
The Aequalis Reversed Shoulder Prosthesis is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRA, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K050316
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.