MILLENNIUM SERIES, INCLUDING MILLENNIUM B, B+, B2 AND BU
K050263 · M.O. Com S.R.L. · FLE · Nov 21, 2006 · General Hospital
Device Facts
Record ID
K050263
Device Name
MILLENNIUM SERIES, INCLUDING MILLENNIUM B, B+, B2 AND BU
Applicant
M.O. Com S.R.L.
Product Code
FLE · General Hospital
Decision Date
Nov 21, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The device is intended to be used in medical and dental practices, hospitals, clinics, nursing homes, laboratories and other facilities to sterilize re-usable surgical instruments (including dental handpieces) and medical materials, heat and moisture resistant and compatible with steam sterilization process.
Device Story
Millennium series (B, B+, B2, Bμ) are steam sterilizers (autoclaves) for medical/dental instruments; chamber volumes 5.5, 17, or 21 liters. Operation: user loads instruments; selects program; device performs pre-heating, fractionated or single vacuum air removal, steam pressure/temperature rise, sterilization hold, steam discharge, vacuum drying, and sterile air venting. Features advanced fractionated vacuum system for hollow/porous loads and final vacuum drying. Operated by trained professionals (dentists, nurses) in clinical settings. Output: sterilized instruments. Benefits: effective, fast sterilization of heat/moisture-resistant tools; ensures thermodynamic stability and safety.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Steam sterilizer; chamber volumes 5.5-21L; fractionated (3 vacuum/pressure pulses) or single vacuum air removal; vacuum drying phase; steam generation system; digital control system; stainless steel chamber; electrical energy source.
Indications for Use
Indicated for sterilization of re-usable surgical instruments (including dental handpieces) and medical materials that are heat and moisture resistant and compatible with steam sterilization. Intended for use by trained professionals in medical/dental practices, hospitals, clinics, nursing homes, and laboratories. Contraindicated for sterilization of fluids, liquids, or pharmaceutical products.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K050263" in a handwritten style on the top line. Below that is the logo for "mocom" in a bold, sans-serif font. Underneath the logo is the text "Sistemi innovativi di sterilizzazione" in a smaller font size.
The information required by 21 CFR § 807.92 is listed below.
| Submitter: | M.O.COM. s.r.l.<br>1, Via delle Azalee<br>20090 Buccinasco (MI) - Italy | NOV 21 200 | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------|--|
| Contact Person: | Daniele Tosi<br>Technical Manager<br>Phone: (+39) 02 45701505<br>Fax: (+39) 02 48840514 | | |
| Date of Submission: | 2005, February 1st | | |
| Device Name: | Proprietary: MILLENNIUM series (B, B+, B2, Bμ)<br>Common: Steam Sterilizer (or Autoclave)<br>Classification: 880.6880 FLE: Sterilizer, Steam | | |
| Devices to which the<br>Millennium series<br>claims equivalence: | W&H Lisa (510(k) number K022816)<br>Midmark M9 UltraClave (510(k) number K023348) | | |
### Device Description
MILLENNIUM series are sophisticated equipments but of very friendly use, adaptable to the different demands thanks to the wide possibilities of configuration and choice of the cycles.
The equipments are able to treat in fast way every type of load through various patented systems and hi-tech solutions.
MILLENNIUM series sterilizers are equipped with a 5.5, 17 or 21 lites sterilization chamber and are characterized by an advanced fractionated vacuum system for the complete air removal also from hollow and porous materials as well as by an effective final vacuum drying phase able to eliminate any trace of condensation from the load.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for "mocom," with the "m" in a bold, stylized font and the rest of the letters in a simpler font. Below the logo, there is text that reads "Sistemi innovativi di sterilizzazione." The logo is in black and white.
An exclusive steam generation system, combined with the advanced control system, guarantees an high process speed and high stability of the thermodynamic parameters during the whole sterilization process and consequently full safety and perfect result.
The equipment offers sterilization programmes optimized for an effective, fast sterilization of different tools and materials used in medical environment, particularly the dental one.
The sterilization program of a MILLENNIUM series sterilizer can properly be described by a sequence of phases, each with a well defined activity.
Considering a standard program (i.c. program at 273 °F – 4' for porous load), after arranged the material in the chamber, closed the door, selected the program, started the cycle by the START key (with activation of the door locking device), the following sequence will run:
- 1. Pre-heating of the steam gencrator and sterilization chamber;
- 2. Chamber air removal and steam penctration inside the load through a series of vacuum stages and pressure phases;
- 3. Pressure rise-up, with consequent steam temperature increasing up to the preset sterilization conditions (for the example, 273 %F);
- 4. Stabilization of the pressure and temperature conditions inside the sterilization chamber;
- 5. Run of the sterilization process for the preset time (for the example, 4 minutes);
- 6. Chamber pressure decreasing through steam discharge;
- 7. Vacuum drying phase;
- 8. Venting phase through sterile air:
- 9. Chamber pressure leveling up to atmospheric value.
As reached this last phase the door locking mechanism will be released and the door can be opened to recovery the load from the sterilization chamber paying attention to use proper precautions.
### Intended Use
The device is intended to be used in medical and dental practices, hospitals, clinics, nursing homes, laboratories and other facilities to sterilize re-usable surgical instruments (including dental hand-pieces) and medical materials, heat and moisture resistant and compatible with steam sterilization process.
- NOTE: make reference to the Table of the Available Programs on the following pages for detailed information on each model.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for MOCOM, a company that specializes in innovative sterilization systems. The logo consists of the letter "m" in a bold, stylized font, followed by the letters "ocom" in a similar font. Below the logo is the text "Sistemi innovativi di sterilizzazione", which translates to "Innovative sterilization systems" in English.
### Table of the Available Programs (Millennium B)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum (1)<br>(F = fract.; S = single) | Standard drying (2)<br>(L = long; S = short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|------------------------------|---------------------|----------------------------|-----------------------|--------------------------------------------|----------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | L | 41' | 525 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.00<br>0.75<br>0.60<br>3.00<br>1.50 | 0.30<br>0.25<br>0.20<br>1.00<br>0.50 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | L | 56' | 550 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.00<br>0.75<br>0.60<br>3.00<br>1.50 | 0.30<br>0.25<br>0.20<br>1.00<br>0.50 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 31' | 525 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 45' | 550 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | L | 34' | 300 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 48' | 325 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 26' | 300 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 40' | 325 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 16' | 300 | 0.45 | Unpackaged solid material | 0.50 | 0.50 | 0.50 |
| 134°C or 121°C<br>CUSTOM (4) | 134<br>or<br>121 | 2.10<br>or<br>1.10 | >4<br>or<br>>20 | F/S | L/S | 56'<br>(max) | 550<br>(max) | 0.9<br>(max) | Unpackaged solid material | 6.00<br>(max) | 1.20<br>(max) | 0.50<br>(max) |
| BOWIE & DICK TEST | 273 | 2.10 | 3,5 | F | S | 22' | - | - | Test pack only (without any other load) | | | |
| VACUUM TEST | - | -0.80 | - | - | - | 22' | - | - | Empty chamber | | | |
Image /page/2/Picture/4 description: The image shows a front view of a laboratory autoclave. The autoclave is white and has a digital display and several buttons on the right side. The door of the autoclave is round and has a handle on the left side.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Mocom, a company that specializes in innovative sterilization systems. The logo features the word "mocom" in a stylized font, with the "m" in a solid black box. Below the logo is the text "Sistemi innovativi di sterilizzazione", which translates to "Innovative sterilization systems" in English.
# Table of the Available Programs (Millennium B+)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum (1)<br>(F= fract.; S= single) | Standard drying (2)<br>(L= long; S= short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|------------------------------|---------------------|----------------------------|-----------------------|------------------------------------------|--------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|-----------------------------------------------|------------------------|---------------------------|------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | L | 33' | 525 | 0.8 | Unpackaged porous material | 1.00 | 0.30 | 0.30 |
| | | | | | | | | | Porous material in single package | 0.75 | 0.25 | 0.25 |
| | | | | | | | | | Porous material in double package | 0.60 | 0.20 | 0.20 |
| | | | | | | | | | Solid material / handpieces in single package | 3.00 | 1.00 | 0.25 |
| | | | | | | | | | Solid material / handpieces in double package | 1.50 | 0.50 | 0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | L | 48' | 550 | 0.8 | Unpackaged porous material | 1.00 | 0.30 | 0.30 |
| | | | | | | | | | Porous material in single package | 0.75 | 0.25 | 0.25 |
| | | | | | | | | | Porous material in double package | 0.60 | 0.20 | 0.20 |
| | | | | | | | | | Solid material / handpieces in single package | 3.00 | 1.00 | 0.25 |
| | | | | | | | | | Solid material / handpieces in double package | 1.50 | 0.50 | 0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 25' | 525 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 39' | 550 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | L | 25' | 300 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 39' | 325 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 18' | 300 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 32' | 325 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 12' | 300 | 0.45 | Unpackaged solid material | 0.50 | 0.50 | 0.50 |
| 134°C or 121°C<br>CUSTOM (4) | 134<br>or<br>121 | 2.10<br>or<br>1.10 | >4<br>or<br>>20 | F/S | L/S | 48'<br>(max) | 550<br>(max) | 0.9<br>(max) | Unpackaged solid material | 6.00<br>(max) | 1.20<br>(max) | 0.50<br>(max) |
| BOWIE & DICK TEST | 273 | 2.10 | 3,5 | F | S | 20' | - | - | Test pack only (without any other load) | | | |
| VACUUM TEST | - | -0.80 | - | - | - | 18' | - | - | Empty chamber | | | |
Image /page/3/Picture/4 description: The image shows a front view of a medical autoclave. The autoclave has a round door on the left side and a control panel on the right side. The control panel has a digital display and four buttons.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "mocom Sistemi innovativi di sterilizzazione". The logo features a stylized lowercase "m" in a black box, followed by the text "ocom" in a simple sans-serif font. Below the logo is the text "Sistemi innovativi di sterilizzazione" in a smaller font size.
### Table of the Available Programs (Millennium B2)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum (1)<br>(F= fract.; S= single) | Standard drying (2)<br>(L=long; S= short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|----------------------------------|---------------------|----------------------------|-----------------------|------------------------------------------|-------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | I. | 43' | 525 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.25<br>1.00<br>0.75<br>4.00<br>2.00 | 0.40<br>0.30<br>0.25<br>1.25<br>0.60 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | I. | 58' | 550 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.25<br>1.00<br>0.75<br>4.00<br>2.00 | 0.40<br>0.30<br>0.25<br>1.25<br>0.60 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 35' | 525 | 0.7 | Unpackaged handpieces | 7.50 | 1.50 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 50' | 550 | 0.7 | Unpackaged handpieces | 7.50 | 1.50 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | I. | 32' | 300 | 0.6 | Solid material in single package | 4.00 | 1.25 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 46' | 325 | 0.6 | Solid material in single package | 4.00 | 1.25 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 25' | 300 | 0.5 | Unpackaged solid material | 7.50 | 1.50 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 39' | 325 | 0.5 | Unpackaged solid material | 7.50 | 1.50 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 14' | 300 | 0.45 | Unpackaged solid material | 0.50 | 0.50 | 0.50 |
| 134°C or 121°C<br>CUSTOM (4) (5) | 134<br>or<br>121 | 2.10<br>or<br>1.10 | > 4<br>or<br>> 20 | F/S | L/S | 58'<br>(max) | 550<br>(max) | 0.9<br>(max) | Unpackaged solid material | 7.50<br>(max) | 1.50<br>(max) | 0.50<br>(max) |
| BOWIE & DICK TEST | 273 | 2.10 | 3,5 | F | S | 24' | - | - | Test pack only (without any other load) | - | - | - |
| VACUUM TEST | - | -0.80 | - | - | - | 18' | - | - | Empty chamber | - | - | - |
Image /page/4/Picture/4 description: The image shows a front-facing view of a white dryer. The dryer has a round door on the left side and a control panel on the right side. The control panel has a digital display and four buttons.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Mocom, a company that specializes in innovative sterilization systems. The logo features the word "mocom" in a stylized font, with the "m" in a bold, blocky style and the rest of the letters in a thinner, more rounded font. Below the logo is the tagline "Sistemi innovativi di sterilizzazione," which translates to "Innovative sterilization systems" in English. The logo is simple and modern, and the tagline clearly communicates the company's focus.
# Table of the Available Programs (Millennium Bu)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum (1)<br>(F= fract.; S= single) | Standard drying (2)<br>(L=long; S= short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|------------------------------|---------------------|----------------------------|-----------------------|------------------------------------------|-------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|-----------------------------------------------|------------------------|---------------------------|------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | L | 19' | 200 | 0.6 | Unpackaged porous material | 0.30 | 0.30 | 0.30 |
| | | | | | | | | | Porous material in single package | 0.30 | 0.30 | 0.25 |
| | | | | | | | | | Porous material in double package | 0.30 | 0.30 | 0.20 |
| | | | | | | | | | Solid material / handpieces in single package | 1.00 | 0.50 | 0.25 |
| | | | | | | | | | Solid material / handpieces in double package | 0.50 | 0.25 | 0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | L | 34' | 210 | 0.6 | Unpackaged porous material | 0.30 | 0.30 | 0.30 |
| | | | | | | | | | Porous material in single package | 0.30 | 0.30 | 0.25 |
| | | | | | | | | | Porous material in double package | 0.30 | 0.30 | 0.20 |
| | | | | | | | | | Solid material / handpieces in single package | 1.00 | 0.50 | 0.25 |
| | | | | | | | | | Solid material / handpieces in double package | 0.50 | 0.25 | 0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 17' | 200 | 0.6 | Unpackaged handpieces | 1.80 | 0.90 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 32' | 210 | 0.6 | Unpackaged handpieces | 1.80 | 0.90 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | L | 16' | 125 | 0.5 | Solid material in single package | 1.00 | 0.50 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 31' | 130 | 0.5 | Solid material in single package | 1.00 | 0.50 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 13' | 125 | 0.4 | Unpackaged solid material | 1.80 | 0.90 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 28' | 130 | 0.4 | Unpackaged solid material | 1.80 | 0.90 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 9' | 125 | 0.4 | Unpackaged solid material | 0.20 | 0.20 | 0.20 |
| 134°C or 121°C<br>CUSTOM (4) | 134<br>or<br>121 | 2.10<br>or<br>1.10 | >4<br>or<br>>20 | F/S | L/S | 34'<br>(max) | 210<br>(max) | 0.7<br>(max) | Unpackaged solid material | 1.80<br>(max) | 0.90<br>(max) | 0.25<br>(max) |
| BOWIE & DICK TEST | 273 | 2.10 | 3,5 | F | S | 16' | - | - | Test pack only (without any other load) | | | |
| VACUUM TEST | - | -0.80 | - | - | - | 16' | - | - | Empty chamber | | | |
Image /page/5/Picture/4 description: The image shows a white device with a digital display and several buttons on the top right. The device has a boxy shape with a slightly curved front. The display shows some text, but it is not clear enough to read. There are four buttons below the display, arranged in a row. The device appears to be some kind of electronic instrument or appliance.
Page 6 of 8 pages
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for "Mocom". The logo has a stylized "m" in a black box, followed by the letters "ocom" in a simple font. Below the logo, the text "Sistemi innovativi di sterilizzazione" is written in a smaller font.
#### GENERAL NOTES (for the Tables of the Available Programs above)
- FRACTIONATED = Pre-vacuum stage made by a sequence of 3 vacuum pulses + 3 pressure pulses. (1) "Fractionated vacuum cycles" are dedicated to the sterilization of porous materials or hand-pieces.
SINGLE = Pre-vacuum stage made by 1 vacuum + 1 pressure pulse. "Single vacuum cycles" are dedicated to the sterilization of solid materials
- (2) LONG = Drying stage for porous material and/or hand-pieces and/or soltd material in single/double package. The validated LONG drying times (STANDARD option) are 16.5 min for Millemium B, 12.5 min for B+, 16.5 min for B' and 6 min for Bu
The EXTRA and INTELLIGENT options have not been validated.
SHORT = Drying stage for unpackaged solid instruments and/or unpackaged hand-pieces. The validated "SHORT drying" times (STANDARD option) are 7 min for Millennium B, 6 min for B+, 8 min for B+ and 4 min for Bu.
The FAST option, having drying times of 2.5 min for Millennium B (up to a load of 1.0 kg max), 2 min for B+ (up to a load of 1.0 kg max), 2.5 min for B2 (up to a load of 1.0 kg max) and 1.5 min for Bp (up to a load of 0.4 kg max) has not been validated.
- (3) The Total Cycle Time is only indicative. It represents the time required for the entire program, from the cycle START to the cycle END and is calculated with a half-load within the chamber
- The 134°C / 12°C CUSTOM program has holding times of 4 minutes (or greater) and 20 minutes (or greater) respectively at (4) 134℃ and 121℃
Pre-vacuum type and Drying type settable according to the indications given in the notes (1) and (2) above.
The 134°C / 121°C CUSTOM program has not been validated
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for MOCOM, a company that specializes in innovative sterilization systems. The logo features the word "mocom" in a bold, sans-serif font, with the "m" in a black box. Below the logo is the tagline "Sistemi innovativi di sterilizzazione", which translates to "Innovative sterilization systems" in English.
### Technological Characteristics Comparison
The comparison shows that the new devices (M.O.COM. s.r.l. Millennium series Steam Sterilizers), which are intended to market, and the predicate devices:
- 1. Have the same intended use, principles of opcration and energy sources.
- 2. Have the same target population (user group).
- 3. Have the same (or better) technological characteristics, safeguards and performances.
The physical design, control system, materials of construction and method of manufacture differ slightly from the predicate devices.
There are not any new questions about safety and effectiveness and the new device is safe and effective as the predicate devices.
The M.O.COM. s.r.l. Millennium series Steam Sterilizers are substantially equivalent to the W&H Lisa Steam Sterilizer and Midmark M9 UltraClave Steam Sterilizer.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Daniele Tosi Technical Manager M.O.C.O.M S.r.l. VIA Delle Azalee 1 Buccinasco (MI), ITALY 20090
NOV 2 1 2006
Re: K050263
Trade/Device Name: MILLENNIUM Series Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: November 6, 2006 Received: November 8, 2006
Dear Mr. Tosi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Mr. Tosi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Cacey Wilson for
Chiu Lin. Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
510(k) Number: K050263 (if known)
Applicant: M.O.COM. S.r.l.
Device Name: MILLENNIUM series (including Millennium B, Millennium B+, Millennium B2 and Millennium Bu)
#### Indication For Use:
The device is intended to be used in medical and dental practices, hospitals, clinics, nursing homes, laboratories and other facilities to sterilize re-usable surgical instruments (including dental handpieces) and medical materials, heat and moisture resistant and compatible with steam sterilization process.
Typical users of this system are trained professionals including, but not limited to, dentists, nurses and other specific medical professionals.
- NOTE: The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
Also see the "Table of the Available Programs", one for each model, on the next pages.
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use X (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shila H. Murphy, MD
താവിടേത്തിയും, General Hospital, Control, Dental Devices
057263
{11}------------------------------------------------
### Table of the Available Programs MILLENNIUM B (temperature in °C and pressure in bar)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum (1)<br>(F= fract.; S= single) | Standard drying (2)<br>(L= long; S= short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|------------------------------|---------------------|----------------------------|-----------------------|------------------------------------------|--------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | L | 41' | 525 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.00<br>0.75<br>0.60<br>3.00<br>1.50 | 0.30<br>0.25<br>0.20<br>1.00<br>0.50 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | L | 56' | 550 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.00<br>0.75<br>0.60<br>3.00<br>1.50 | 0.30<br>0.25<br>0.20<br>1.00<br>0.50 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 31' | 525 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 45' | 550 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | L | 34' | 300 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 48' | 325 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 26' | 300 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 40' | 325 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 16' | 300 | 0.45 | Unpackaged solid material | 0.50 | 0.50 | 0.50 |
| 134°C or 121°C<br>CUSTOM (4) | 134<br>or<br>121 | 2.10<br>or<br>1.10 | >4<br>or<br>>20 | F/S | L/S | 56'<br>(max) | 550<br>(max) | 0.9<br>(max) | Unpackaged solid material | 6.00<br>(max) | 1.20<br>(max) | 0.50<br>(max) |
| BOWIE & DICK TEST | 273 | 2.10 | 3.5 | F | S | 22' | - | - | Test pack only (without any other load) | - | - | - |
| VACUUM TEST | | -0.80 | - | | | 22' | - | - | Empty chamber | - | - | - |
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).
: 上一
Concurrence of CDRH, Office of Device Evaluation (ODE)
.
. 1 ・・・
{12}------------------------------------------------
### Table of the Available Programs MILLENNIUM B+ (temperature in ℃ and pressure in bar)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum (1)<br>(F= fract.; S= single) | Standard drying (2)<br>(L= long; S= short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|------------------------------|---------------------|----------------------------|-----------------------|------------------------------------------|--------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | L | 33' | 525 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.00<br>0.75<br>0.60<br>3.00<br>1.50 | 0.30<br>0.25<br>0.20<br>1.00<br>0.50 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | L | 48' | 550 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.00<br>0.75<br>0.60<br>3.00<br>1.50 | 0.30<br>0.25<br>0.20<br>1.00<br>0.50 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 25' | 525 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 39' | 550 | 0.7 | Unpackaged handpieces | 6.00 | 1.20 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | L | 25' | 300 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 39' | 325 | 0.6 | Solid material in single package | 3.00 | 1.00 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 18' | 300 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 32' | 325 | 0.5 | Unpackaged solid material | 6.00 | 1.20 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 12' | 300 | 0.45 | Unpackaged solid material | 0.50 | 0.50 | 0.50 |
| 134°C or 121°C<br>CUSTOM (4) | 134<br>or<br>121 | 2.10<br>or<br>1.10 | > 4<br>or<br>> 20 | F/S | L/S | 48'<br>(max) | 550<br>(max) | 0.9<br>(max) | Unpackaged solid material | 6.00<br>(max) | 1.20<br>(max) | 0.50<br>(max) |
| BOWIE & DICK TEST | 273 | 2.10 | 3,5 | F | S | 20' | | | Test pack only (without any other load) | | | |
| VACUUM TEST | | -0.80 | | | | 18' | | | Empty chamber | | | |
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).
Concurrence of CDRH, Office of Device Evaluation (ODE)
{13}------------------------------------------------
## Table of the Available Programs MILLENNIUM B2 (temperature in ℃ and pressure in bar)
| PROGRAM<br>DESCRIPTION | Temperature<br>(°C) | Pressure<br>(bar relative) | Holding time<br>(min) | Pre-vacuum(1)<br>(F= fract.; S= single) | Standard drying (2)<br>(L= long; S= short) | Total cycle time (3)<br>(approx.) | Average H2O<br>consumption<br>(ml/cycle) | Average energy<br>consumption<br>(kWh/cycle) | TYPE OF LOAD | MAX TOTAL<br>MASS (kg) | MAX MASS<br>PER TRAY (kg) | MAX MASS<br>PER ARTICLE (kg) |
|---------------------------------|---------------------|----------------------------|-----------------------|-----------------------------------------|--------------------------------------------|-----------------------------------|------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|
| 134°C POROUS | 134 | 2.10 | 4 | F | L | 43' | 525 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.25<br>1.00<br>0.75<br>4.00<br>2.00 | 0.40<br>0.30<br>0.25<br>1.25<br>0.60 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 121°C POROUS | 121 | 1.10 | 20 | F | L | 58' | 550 | 0.8 | Unpackaged porous material<br>Porous material in single package<br>Porous material in double package<br>Solid material / handpieces in single package<br>Solid material / handpieces in double package | 1.25<br>1.00<br>0.75<br>4.00<br>2.00 | 0.40<br>0.30<br>0.25<br>1.25<br>0.60 | 0.30<br>0.25<br>0.20<br>0.25<br>0.25 |
| 134°C HOLLOW | 134 | 2.10 | 4 | F | S | 35' | 525 | 0.7 | Unpackaged handpieces | 7.50 | 1.50 | 0.50 |
| 121°C HOLLOW | 121 | 1.10 | 20 | F | S | 50' | 550 | 0.7 | Unpackaged handpieces | 7.50 | 1.50 | 0.50 |
| 134°C WRAPPED | 134 | 2.10 | 4 | S | L | 32' | 300 | 0.6 | Solid material in single package | 4.00 | 1.25 | 0.25 |
| 121°C WRAPPED | 121 | 1.10 | 20 | S | L | 46' | 325 | 0.6 | Solid material in single package | 4.00 | 1.25 | 0.25 |
| 134°C SOLID | 134 | 2.10 | 4 | S | S | 25' | 300 | 0.5 | Unpackaged solid material | 7.50 | 1.50 | 0.50 |
| 121°C SOLID | 121 | 1.10 | 20 | S | S | 39' | 325 | 0.5 | Unpackaged solid material | 7.50 | 1.50 | 0.50 |
| 134°C EMERGENCY | 134 | 2.10 | 3 | S | S | 14' | 300 | 0.45 | Unpackaged solid material…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.