← Product Code [LNJ](/productcode/LNJ) · K050236

# FISHVIEW (K050236)

_Applied Spectral Imaging , Ltd. · LNJ · Jun 27, 2005 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050236

## Device Facts

- **Applicant:** Applied Spectral Imaging , Ltd.
- **Product Code:** [LNJ](/productcode/LNJ.md)
- **Decision Date:** Jun 27, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5260
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Pediatric

## Indications for Use

The FISHView system is intended to be used for karyotyping with real-time microscope images from cultured and stained cell specimens in their metaphase. The system works with bright field and fluorescent samples. Specimens suitable for banding analysis are: amniotic fluid, peripheral blood, chorionic villus, bone marrow and solid tumor. Karyotyping is normally applied for the pre and postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (such as Down's syndrome), cancer, and for the follow up of cancer treatment. The FISHView system does not locate metaphase spreads; it does not rank the given cells according to quality; nor does it automatically classify chromosomes. In addition the FISHView is intended as an aiding tool to the pathologist or cytogeneticist for digital visualizing, processing, counting and classification of stained cells and for storage of FISH multi - dye images of the following specimens: amniotic fluid, peripheral blood, chorionic villus, bone marrow and solid tumor. The FISHView system does require and relies on the operator to analyze the digitized microscope images.

## Device Story

FISHView is an integrated digital platform comprising a microscope, camera, frame grabber, and workstation. It enables cytological experts to acquire, view, enhance, archive, and print images of cultured/stained cells under brightfield or fluorescent illumination. The system supports FISH multi-dye imaging. Operation is manual; the user interactively controls the microscope to scan slides and identify cells of interest. The software facilitates image acquisition, editing, relocation, manual counting, and classification. The device does not automatically locate metaphase spreads, rank cell quality, or classify chromosomes; it serves as an aiding tool for the pathologist or cytogeneticist, who must perform the final analysis of the digitized images. By digitizing the workflow, it assists in the diagnosis of genetic conditions and cancer.

## Clinical Evidence

Bench testing only. A comparison study evaluated 115 slides across various specimen types (amniotic fluid, blood lymphocytes, chorionic villus, bone marrow, solid tumors) using various DNA probes. Two expert cytogeneticists compared FISHView images against direct microscopic examination. Parameters included image quality, number of FISH signals, clinical adequacy, data loss, image alteration, and safety. No parameters were disapproved for the FISHView system.

## Technological Characteristics

Integrated digital platform consisting of a microscope, camera, frame grabber, and workstation. Features 1280 x 1024 spatial resolution. Supports brightfield and fluorescent illumination. Software-controlled image acquisition, enhancement, and storage. Manual operation.

## Regulatory Identification

An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)

## Predicate Devices

- BANDView system ([K012103](/device/K012103.md))
- Duet system ([K040591](/device/K040591.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE

A. 510(k) Number:
K050236

B. Purpose for Submission:
Addition of fluorescent in-situ hybridization (FISH) imaging to the BANDView System

C. Manufacturer and Instrument Name:
Applied Spectral Imaging, Ltd., FISHView

D. Type of Test or Tests Performed:
N/A

E. System Descriptions:

1. Device Description:
The FISHView System is an integrated digital platform constructed of a microscope, camera, frame grabber and a workstation. It is designed to acquire images of cells and enables identification and examination of cells of interest. Cytological analysis experts can view, manually scan cells and record the image, using both bright field and fluorescent illumination. The acquired images can be enhanced, archived, retrieved and printed. The digital visualization, processing, and storage of FISH multi dye images of the FISHView System are additions to the legally marketed BANDView System.

2. Principles of Operation:
Microscopic slides are prepared using standardized cell preparation methods. The user interactively controls the microscope under brightfield or fluorescent illumination. Each field of view is visualized using several fluorescent filters. The device generates and displays a single layer or a combined image for each field of view. The work station is software controlled and includes features such as: acquisition of images, views, editing relocation, image enhancement, manual counting and classification, printing, export of images and backups. This methodology is well established and has been used in other cleared devices.

3. Modes of Operation:

{1}

The FISHView system is manually operated. The user has interactive control over the scan of the slide under either bright field or fluorescent illumination.

4. Specimen Identification:
Patient name and individual specimen slide case details are entered in the Case Data Manager.

5. Specimen Sampling and Handling:
Standardized cell preparations of amniotic fluid, chorionic villus, peripheral blood, bone marrow and solid tumor are applied to microscope slides.

6. Calibration:
N/A

7. Quality Control:
No special quality control is required; however annually preventive maintenance by ASI service personnel is recommended.

8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ or No ☑

F. Regulatory Information:

1. Regulation section:
21 CFR 864.5260, Automated cell-locating device

2. Classification:
Class II

3. Product code:
LNJ, Analyzer, Chromosome, Automated

4. Panel:
88 (Pathology)

{2}

G. Intended Use:

1. Indication(s) for Use:

The FISHView System is intended to be used for karyotyping with real-time microscope images from cultured and stained cell specimens in their metaphase. The system works with bright field and fluorescent samples. Specimens suitable for banding analysis are: amniotic fluid, peripheral blood, chorionic villus, bone marrow and solid tumor. Karyotyping is normally applied for the pre and postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (such as Down’s Syndrome), cancer, and for the follow up of cancer treatment. The FISHView system does not locate metaphase spreads; it does not rank the given cells according to quality; nor does it automatically classify chromosomes.

In addition the FISHView is intended as an aiding tool to the pathologist or cytogeneticist for digital visualizing, processing, counting and classification of stained cells and for storage of FISH multi-dye images of the following specimens: amniotic fluid, peripheral blood, chorionic villus, bone marrow and solid tumor.

The FISHView system does require and relies on the operator to analyze the digitized microscope images.

2. Special Conditions for Use Statement(s):

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:

BANDView System (k012103)

BioView, Ltd., Duet™ System (k040591)

2. Comparison with Predicate Device:

|  Similarities  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | BANDView | Duet  |
|  Cell Source | Amniotic fluid, chorionic villus, bone marrow, peripheral blood and solid tumor | Amniotic fluid, chorionic villus, bone marrow, peripheral blood and solid tumor | Peripheral blood, bone marrow, and amniotic fluid  |
|  Preparation techniques | Same | Same | Same  |

{3}

|  Similarities  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | BANDView | Duet  |
|  Hardware | Same | Same | Same  |
|  Differences  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | BANDView | Duet  |
|  Automated scanning | No | No | Yes  |
|  Spatial Resolution | 1280 x 1024 | 1280 x 1024 | 640 x 480  |

I. Special Control/Guidance Document Referenced (if applicable):

N/A

J. Performance Characteristics:

1. Analytical Performance:

a. Accuracy:

N/A

b. Precision/Reproducibility:

N/A

c. Linearity:

N/A

d. Carryover:

N/A

e. Interfering Substances:

N/A

2. Other Supportive Instrument Performance Data Not Covered Above:

A comparison study was performed between the FISHView System and routine examination of the same cells directly through the microscope. The evaluation was performed with a total of 115 slides from different specimens (amniotic fluid,

{4}

blood lymphocytes, chorionic villus, bone marrow, and solid tumors) using various DNA probes by 2 expert cytogeneticists. The following parameters were compared between the two methods: image quality, number of FISH signals, clinical adequacy, data loss, image alteration, and safety problems. The cytogeneticists either approved or disapproved each of these parameters for each specimen type and accompanying DNA probe. No parameters were disapproved for the FISHView System.

## K. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

## L. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

5

---

**Source:** [https://fda-staging.innolitics.com/device/K050236](https://fda-staging.innolitics.com/device/K050236)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
