The HearTwave II Cardiac Diagnostic System is intended for the recording of electrocardiograms, vector cardiograms and measurement of Microvolt T-Wave Alternans* at rest and during ECG stress testing. The presence of Microvolt T-wave Alternans as measured by the HearTwave II Cardiac Diagnostic System in patients with known, suspected or at risk of ventricular tachyarrhythmia predicts increased risk of a cardiac event (ventricular tachyarrhythmia or sudden death). The HearTwave II Cardiac Diagnostic System should be used only as an adjunct to clinical history and the results of other non-invasive and/or invasive tests. The interpretive results of the Alternans Processing System should be reviewed by a qualified physician. The predictive value of T-wave Alternans for cardiac events has not been established in patients with active, untreated ischemia. *Microvolt T-wave Alternans is defined as T-wave alternans which (a) is measured from high-resolution multi-segment sensors, (b) is present in leads X, Y, Z, VM or two adjacent precordial leads, (c) is at the level of 1.9 microvolts after signal optimization and subtraction of the background noise level, (d) is at least three standard deviations greater than the background noise level, (e) has an onset heart rate at or below 110 beats per minute, and (f) is sustained for all heart rates above the onset heart rate.
Device Story
HearTwave II Cardiac Diagnostic System records ECGs, vector cardiograms, and measures Microvolt T-Wave Alternans (MTWA). Input: digitized ECG signals from seven standard stress test electrodes and seven proprietary multi-segment Micro-V Alternans sensors via a belt-worn patient module. Operation: Analytic Spectral Method processes signals to detect beat-to-beat T-wave amplitude variations. Used in stress labs (treadmill, ergometer, pharmacologic) by qualified personnel. Output: MTWA measurements and diagnostic reports. Clinical use: adjunct to clinical history and other tests to assess risk of sudden cardiac death or ventricular tachyarrhythmia. Physician reviews output to inform clinical decision-making regarding patient risk stratification.
Clinical Evidence
No clinical data provided in this summary. Device relies on established Analytic Spectral Method performance previously cleared in predicate submissions.
Technological Characteristics
System includes belt-worn patient module, standard stress electrodes, and proprietary multi-segment Micro-V Alternans sensors. Complies with ANSI/AAMI EC11-1991, EN60601-1, EN60601-1-2, EN60601-1-4, IEC 60601-2-25, and IEC 60601-2-51. Analytic Spectral Method used for signal processing.
Indications for Use
Indicated for patients with known, suspected, or at risk of ventricular tachyarrhythmia to predict risk of cardiac events (ventricular tachyarrhythmia or sudden death). Contraindicated in patients with active, untreated ischemia.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
05-0225
APR 7 2005
# 510(k) Summary
January 31, 2005
# Submitter: Cambridge Heart, Inc 1 Oak Park Bedford, Ma 01730 (781) 271-1200 (781) 275-8431
Contact: Dave Chazanovitz
### 510(k) Numbers and Product Codes of equivalent devices:
Cambridge Heart, Inc .; HearTwave II Cardiac Diagnostic System
510(k) Number: # K022149
Product Code: 74 DPS
CFR Section: 870.2340
### Indications for Use and Intended Population
The HearTwave II Cardiac Diagnostic System is intended for the recording of electrocardiograms, vector cardiograms and measurement of Microvolt T-Wave Alternans* at rest and during ECG stress testing.
The presence of Microvolt T-wave Alternans as measured by the analytic spectral method of the HearTwave II Cardiac Diagnostic System in patients with known, suspected or at risk of ventricular tachyarrhythmia predicts increased risk of a cardiac event (ventricular tachyarrhythmia or sudden death).
The Cambridge Heart HearTwave II Cardiac Diagnostic System should be used only as an adjunct to clinical history and the results of other non-invasive and/or invasive tests.
The predictive value of T-wave Alternans for cardiac events has not been established in patients with active, untreated ischemia.
*Microvolt T-wave Alternans is defined as T-wave alternans which (a) is measured from high-resolution multi-segment sensors, (b) is present in leads X, Y, Z, VM or two adjacent precordial leads, (c) is at the level of 1.9 microvolts after signal optimization and subtraction of the background noise level, (d) is at least three standard deviations greater than the background noise level, (e) has an onset heart rate at or below 110 beats per minute, and (f) is sustained for all heart rates above the onset heart rate.
#### Device Description
The Cambridge Heart HearTwave II Cardiac Diagnostic System is intended for the measurement and recording of T-Wave alternans. The alternans levels reported in K983012, K001034, K003492 and K022149 were measured using the Analytic Spectral Method. This method consists of several computational steps that combine to form a
> HearTwave II Cardiac Diagnostic System Special 510(k) Pre-market Submission Section 4.0 - 510(k) Summary page 1
Ong
{1}------------------------------------------------
# K050225
unique analytical process. The Microvolt T-wave Alternans measurement, the output of this specific process, has been shown to be useful in predicting ventricular tachyarrhythmias and sudden cardiac death.
The Cambridge Heart HearTwave II Cardiac Diagnostic System provides T-wave alternans diagnostic capabilities to standard stress labs. The Analytic Spectral Method of Alternans Processing used in the Cambridge Heart HearTwave II is intended for the measurement of microvolt T-Wave alternans at rest and during treadmill, ergometer and pharmacologic stress testing.
The Alternans test using the Cambridge Heart HearTwave II is performed with seven standard stress test electrodes and seven proprietary multi-segment Micro-V Alternans™ Sensors. The electrodes and sensors are attached through a leadwire set to the beltworn Patient module, which provides digitized data to the HearTwave II.
### Patient Electrodes:
Patient electrodes designed and approved specifically for use during exercise stress testing should be used at all times with the Cambridge Heart Model HearTwave II Cardiac Diagnostic System.
Measurement of alternating beat to beat T-wave amplitude (alternans) requires the use of the Cambridge Heart Hi-Res Electrode (Ref: # K962115) or The Cambridge Heart Micro-V Alternans Sensor (Ref: #K002230) in conjunction with other Patient electrodes designed and approved specifically for use during exercise stress testing.
### Performance Standards
The Cambridge Heart HeartWave II Cardiac Diagnostic System is designed to meet the following Performance Standards:
- ANSI/AAMI EC11-1991 .
- EN60601-1: 2004, "Medical Electrical Equipment, Part 1: General Requirements . for Safety" including Amendments A1 and A2
- EN60601-1-1: 2000, "Medical Electrical Equipment, Part 1: General . Requirements for Safety - Section 1.1 Collateral standard: Safety requirements for medical electrical systems"
- EN60601-1-2: 2004, "Medical Electrical Equipment, Part 2: Collateral Standard: . Electromagnetic Compatibility - Requirements and Tests"
- EN60601-1-4: 2000, "Medical Electrical Equipment, Programmable Medical . Systems"
- IEC 60601-2-25: 1993, "Particular requirements for the safety of ● electrocardiographs" including Amendment 1
- . IEC 60601-2-51: 2003 "Particular requirements for safety, including essential performance, of recording and analyzing single and multichannel electrocardiographs
HearTwave II Cardiac Diagnostic System Special 510(k) Pre-market Submission Section 4.0 - 510(k) Summary page 2
{2}------------------------------------------------
- FDA Diagnostic ECG Guidance: 1998, Version 1.0 .
- FDA Electrocardiograph Lead Switching Adaptor Guidance: 1997, Version 1.0 .
- FDA Guidance For Industry: 2000 " General Principles of Software Validation" .
- FDA Guidance For Industry: 1999 " Guidance for Off-The -Shelf Software for use . in Medical Devices"
- ISO 14971: 2000 " Application of risk management to medical devices" .
K050225
# Similarities and Differences to Predicates
The Model HearTwave II Cardiac Diagnostic System (new) is essentially the same device as in K02149 with the exception of the modification described in this premarket submission. The Model HearTwave II Cardiac Diagnostic System uses the same Analytic Spectral Method as the CH2000 (K02149) for measuring T-Wave Alternans.
### Conclusion
There are more similarities than differences between the predicate device and the Cambridge Heart HearTwave II Cardiac Diagnostic System. Both the predicate devices use the Analytic Spectral Method of Alternans Processing. When used in accordance with the directions for use, by qualified personnel, the Cambridge Heart HeartWave II Cardiac Diagnostic System is safe and effective, as indicated, for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
2005 APR 7
Cambridge Heart, Inc. c/o Mr. John Greenbaum 20310 SW 48th Street Ft. Lauderdale, Florida 33332
K050225 Re:
K050223
Trade Name: HearTwave II Cardiac Diagnostic System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: 74 DQK Dated: March 11, 2005 Received: March 15, 2005
Dear Mr. Greenbaum:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 510(x) premaint is substantially equivalent (for the indications
referenced above and have determined the devices marketed in interstate referenced above and have determined the acvee is sedicate devices marketed in interstate for use stated in the enclosure to tegally interest date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enatures with the provision of the Federa commerce prior to May 28, 1976, the eliaculance with the provisions of the Federal Food, Drug, devices that have been reclassined in accordance warane approval application (PMA).
and Cosmetic Act (Act) that do not require approval approval applisions of the Act. The and Cosmetic Act (Act) that do not require approval or the general controls of the Act. The Act. The You may, therefore, market the device, subject to the gents for annual registration, listing of
general controls provisions of the Act include requires misbranding and general controls provisions of the Act mender requirements of a seainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can
may be subject to such additional controls. Exist. 21 - 80 - In addition F may be subject to such additions. Laisting may regalato 898 In addion, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addion, FDA m be found in the Code of Federal Regalations, The Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. John Greenbaum
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 Toast be advised that 1 Dris assumes over device complies with other requirements of the Act that I Dri has made a actuations administered by other Federal agencies. You must or any I cochar statutes and regalazents and united to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR Part 807); adoning (21 CFR Part 820); and if applicable, the electronic forth in the quality bybeline (Sections 531-542 of the Act); 21 CFR 1000-1050. product lauation control provisions (overing your device as described in your Section 510(k) This letter wif anow you to ogen finding of substantial equivalence of your device to a legally prematication: The PDF intentigation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrice ior Jour as 1) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Ivilsoranding of reference to premeesponsibilities under the Act may be obtained from the Oinci general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
B. Summerfor
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number(if known): K050225
Device Name: HearTwave II Cardiac Diagnostic System
Indications For Use:
The HearTwave II Cardiac Diagnostic System is intended for the recording of electrocardiograms, vector cardiograms and measurement of Microvolt T-Wave Alternans* at rest and during ECG stress testing.
The presence of Microvolt T-wave Alternans as measured by the HearTwave II Cardiac Diagnostic System in patients with known, suspected or at risk of ventricular tachyarrhythmia predicts increased risk of a cardiac event (ventricular tachyarrhythmia or sudden death).
The HearTwave II Cardiac Diagnostic System should be used only as an adjunct to clinical history and the results of other non-invasive and/or invasive tests. The interpretive results of the Alternans Processing System should be reviewed by a qualified physician.
The predictive value of T-wave Alternans for cardiac events has not been established in patients with active, untreated ischemia.
*Microvolt T-wave Alternans is defined as T-wave alternans which (a) is measured from high-resolution multi-segment sensors, (b) is present in leads X, Y, Z, VM or two adjacent precordial leads, (c) is at the level of 1.9 microvolts after signal optimization and subtraction of the background noise level, (d) is at least three standard deviations greater than the background noise level, (e) has an onset heart rate at or below 110 beats per minute, and (f) is sustained for all heart rates above the onset heart rate.
Prescription Use_X_ (Per 21 CRF 801.109) Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kllummer
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K050225
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.