← Product Code [LPL](/productcode/LPL) · K050157

# MAXSIGHT (POLYMACON) SPORT-TINTED CONTACT LENS (K050157)

_Bausch & Lomb, Inc. · LPL · Apr 26, 2005 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050157

## Device Facts

- **Applicant:** Bausch & Lomb, Inc.
- **Product Code:** [LPL](/productcode/LPL.md)
- **Decision Date:** Apr 26, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in non-aphakic persons with non-diseased eyes exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00 to -20.00 diopters. The MAXSIGHT (polymacon) Sport-Tinted contact lens helps protect against the transmission of harmful UV radiation to the cornea and into the eye. Replacement schedules may vary from patient to patient, and should be decided by eye care professionals in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care professionals. The lens may be disinfected using a chemical disinfection system.

## Device Story

MAXSIGHT is a hemispherical, flexible soft contact lens composed of polymacon (HEMA and EGDMA polymer) with 38.6% water content. Lens features a sport-specific tint using FDA-conforming dyes to enhance contrast and visual clarity in outdoor athletic settings. Device is worn daily by patients; requires cleaning, rinsing, and chemical disinfection after removal. Prescribed by eye care professionals to correct refractive ametropia and provide UV radiation protection. Lens dimensions range 13.5-15.5mm diameter, 8.4-9.0mm base curve, and 0.06-0.175mm center thickness.

## Clinical Evidence

Bench testing only. In vitro and in vivo toxicology and biocompatibility testing performed per FDA 1994 guidance. Physicochemical properties, extractables, and cytotoxicity evaluated. Results showed no detectable monomer/cross-linker/initiator/diluent, dye extractables < 10 ppm, and no toxicity/irritation. Device confirmed compatible with chemical disinfection systems.

## Technological Characteristics

Material: Polymacon (HEMA/EGDMA polymer). Water content: 38.6%. Refractive index: 1.43. Specific gravity: 1.13. Oxygen permeability (Dk): 8.4x10^-11. UV absorption: >95% (UVA/UVB). Manufacturing: Spin-cast/lathe cut. Tinting: In-monomer visibility tinting with dyes conforming to 21 CFR 73.3106/73.3121. Dimensions: 13.5-15.5mm diameter, 8.4-9.0mm base curve, 0.06-0.175mm center thickness.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- NaturalTint® (polymacon) Contact Lens (N16-895)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SUMMARY OF SAFETY AND EFFECTIVENESS

# FOR MAXSIGHT™ (polymacon) Sport-Tinted Contact Lens

### 1. Submitter Information:

Bausch & Lomb 1400 North Goodman Street Rochester, NY 14609

Contact Person:Lisa Graney Manager, Global Regulatory Affairs (585) 338-6612 Telephone No.: (585) 338-0702 Fax No.:

### 2. Device Name:

Soft (hydrophilic) contact lens Classification Name: MAXSIGHT (polymacon) Sport-Tinted Contact Lens Proprietary Name:

#### Predicate Device: 3.

NaturalTint® (polymacon) Contact Lens

N16-895

### DESCRIPTION OF DEVICE 4.

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is a hemispherical The MPONSTY (polymaton), Spersea and may cover a portion of the adjacent fiexible Sholl Which of a polymer of Hydroxyethyl Methacrylate (HEMA) and sciera. It consists of a polymor Dimethacrylate (EGDMA), and is 38% water by weight when immersed in a sterile saline solution. This lens is tinted in weight which immorou nlue Dye 246 (1, 4-Bis[4-(2-methacryloxyethyl) monomer with Productione), which conforms to 21 CFR Part 73.3106. Finished lenses are subsequently tinted to achieve different colors using r linished lenood are olor additives that conform to 21 CFR Part 73.3121.

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is designed to aid The MAXSIGHT (polymaten), Spitings. MAXSIGHT (polymacon) Sport-Tinted Visual penonmance in athlette set in a wide range of outdoor light conditions. Contact lenses chhanos contract in a created objects with greater clarity than with the naked eye.

{1}------------------------------------------------

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is manufactured with the following dimensions:

| Diameter:                                          | 13.5 to 15.5 mm                                                                   |
|----------------------------------------------------|-----------------------------------------------------------------------------------|
| Center Thickness:                                  | 0.06 to 0.175mm                                                                   |
| Base Curve:                                        | 8.4-9.0mm                                                                         |
| Powers (Spherical):                                | +20.00D to -20.00D                                                                |
| The physical / optical properties of the lens are: |                                                                                   |
| Specific Gravity:                                  | 1.13                                                                              |
| Refractive Index:                                  | 1.43                                                                              |
| Light Transmittance:                               | ≥30%                                                                              |
| UV Absorption:                                     | >95%                                                                              |
| UVA absorption:                                    | >95%                                                                              |
| UVB absorption:                                    | >95%                                                                              |
| Water Content:                                     | 38.6%                                                                             |
| Oxygen Permeability (Dk):                          | 8.4x10-11[cm³O2 (STP) x cm]/(sec x cm x cm² x mmHg)@35°C<br>(Polargraphic Method) |

Each MAXSIGHT (polymacon) Sport-Tinted Contact Lens is supplied in a plastic Each M. Roler . (por saline solution. The lid of the container is marked with blieter oonlainer wower, diameter, manufacturing lot number, and expiration date.

### 5. INDICATIONS FOR USE

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in notwear for the sonovith non-diseased eyes exhibiting astigmatism of 2.00 diopters aphartis por done not interfere with visual acuity. The lens may be prescribed in or loos, than wers ranging from +20.00 to -20.00 diopters. The MAXSIGHT (polymacon) Sport-Tinted contact lens helps protect against the transmission of harmful UV radiation to the cornea and into the eye.

Replacement schedules may vary from patient to patient, and should be decided replacement ochouals in consultation with their patients. The lens is to be by eye care professionals in ochsalt time it is removed from the patient's eye cleaned, finded and theinfoccour ended wearing period prescribed by the eye and ulscarded after the roommonded working a chemical disinfection system.

{2}------------------------------------------------

# 6. DESCRIPTION OF SAFETY AND SUBSTANTIAL EQUIVALENCE

A series of non-clinical laboratory tests was performed to demonstrate the safety and effectiveness of the MAXSIGHT (polymacon) Sport-Tinted Contact Lens. A summary of results of these tests are provided below.

## Safety: Non-Clinical Laboratory Testing

A series of in vitro and in vivo toxicology and biocompatibility testing was performed to assess the safety and effectiveness of the contact lens device. Testing was performed in accordance with FDA guidance Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses, May 1994. All non-clinical laboratory studies were conducted in compliance with the GLP regulation.

The results of the testing of the MAXSIGHT (polymacon) Sport-Tinted Contact Lenses demonstrate that:

- The physicochemical properties of the MAXSIGHT (polymacon) Sport-. Tinted Contact Lens are equivalent to the predicate device, Bausch & Lomb NaturalTint (polymacon) Contact Lens.
- The physicochemical properties of the MAXSIGHT (polymacon) Sport-. Tinted Contact Lens are equivalent to the base lens prior to tinting.
- The extracts of the MAXSIGHT (polymacon) Sport-Tinted Contact Lens do . not show any detectable quantities of monomer components, cross-linker, initiator, or diluent. The levels of extracts of the dye are less than 10 ppm, and cytotoxicity testing results of the MAXSIGHT (polymacon) Sport-Tinted Contact Lens demonstrated no toxicity or irritation.
- The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is compatible with . several different chemical disinfection systems.

### Substantial Equivalence

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is similar to the Bausch & Lomb NaturalTint (polymacon) Contact Lens, in that both:

- Fall into the same FDA material classification group (Group I), .
- Have the same USAN name (polymacon) .
- Are manufactured by the same visibility tinted lens manufacturing . process (Spin-cast/Lathe cut with in monomer visibility tinting)
- Have equivalent lens color tinting processes; both processes involve . tinting a fully cured soft contact lens.

{3}------------------------------------------------

Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans.

Public Health Service

Food and Drug Administration Rockville MD 20857

APR 2 6 2005

Ms. Lisa C. Graney Manager, Global Regulatory Affairs Bausch & Lomb, Inc. 1400 North Goodman Street Rochester, NY 14609

Re: K050157

Trade/Device Name: MAXSIGHT™ (polymacon) Sport-Tinted Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: March 31, 2005 Received: April 1, 2005

Dear Ms. Graney

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Lisa C. Graney

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ

:

... ..

David M. Whipple

Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo and address of Bausch & Lomb. The address is 1400 North Goodman Street, Rochester, NY 14609. The logo is a stylized version of the company name.

### Indications for Use Statement " "

510(k) Number (if known):_K ο δύ | 57

Device Name: MAXSIGHT™ (polymacon) Sport-Tinted Contact Lens

Indications for Use:

The MAXSIGHT (polymacon) Sport-Tinted Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in notaphakic persons with non-diseased eyes exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00 to -20.00 diopters. The MAXSIGHT (polymacon) Sport-Tinted contact lens helps protect against the transmission of harmful UV radiation to the cornea and into the eve.

Replacement schedules may vary from patient to patient, and should be decided by eye care professionals in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care professionals. The lens may be disinfected using a chemical disinfection system.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

*Ni-Chun Shu*

(Division Sign-Off)

Division of Ophthalmic Ear,

Nose and Throat Devises

| 510(k) Number | K050157 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda-staging.innolitics.com/device/K050157](https://fda-staging.innolitics.com/device/K050157)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
