POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)

K050154 · Zibo Zhuangyuan Plastic & Rubber Co., Ltd. · LYZ · Feb 7, 2005 · General Hospital

Device Facts

Record IDK050154
Device NamePOWDER-FREE VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)
ApplicantZibo Zhuangyuan Plastic & Rubber Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateFeb 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable vinyl patient examination glove; worn on examiner's hand or finger. Acts as physical barrier to prevent contamination between patient and examiner. Used in clinical settings by healthcare providers. Device is non-sterile, powder-free, and white (non-colored). Benefits patient and examiner by reducing risk of cross-contamination during examinations.

Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing including physical property testing, pinhole AQL testing per 21 CFR 800.20, and biocompatibility testing (primary skin irritation and dermal sensitization).

Technological Characteristics

Vinyl material; powder-free. Meets ASTM D5250-00e4 for dimensions, physical properties, and powder residual. Meets 21 CFR 800.20 for freedom from pinholes. Biocompatibility confirmed via ISO 10993-10 (skin irritation and sensitization).

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## C Summary FEB - 7 2005 "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is: K05 0 34 ." (applicant leave blank) ## Premarket Notification {510(k)} Summary [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared : | Submitter's name : | ZIBO ZHUANGYUAN PLASTIC & RUBBER CO.,LTD | |--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's address : | THE DEVELOPING TECHNICAL DISTRICT IS IN THE<br>CROSS OF JIQING AND ZHANGTIAN EXPRESSWAY,<br>ZIBO CITY,SHANDONG PROVINCE, 255087, P.R.CHINA | | Phone number : | (86) 533-3811696 | | Fax number : | (86) 533-3811696 | | Name of contact person: | Ms. GENG Xiaocong | | Date the summary was prepared: | 9 January 2005 | ## [(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known | Device Name: | Powder Free Vinyl Patient Examination Gloves, white(non-colored) | | |-------------------------|--------------------------------------------------------------------------------|--| | Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves<br>Other clients private labeling | | | Common Name: | Patient examination glove | | | Classification Name: | Patient examination glove | | | Device Classification: | I | | | Regulation Number: | 21 CFR 880.6250 | | | Panel: | General Hospital (80) | | | Product Code: | LYZ | | ## [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence . Class I* powder free vinyl patient examination gloves , white(non-colored) that meets all of the requirements of ASTM standard D 5250-0064. Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 . #### [(a)(4)] A description of the device {1}------------------------------------------------ Device Description : powder free vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-0064 ## [(a)(5)] The summary describes the intended use of the device Device Intended Use: powder free vinyl patient examination glove, white(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's band or 'inger to prevent contamination between patient and examiner. ## [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device. The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | Device performance | |-----------------------|-------------------------------------------|--------------------| | Dimension | ASTM standard D 5250-00e4 | Meets | | Physical Properties | ASTM standard D 5250-00e4 | Meets | | Freedom from pinholes | 21 CFR 800.20 | Meets | | Powder Residual | ASTM standard D 5250-00e4 | Meets | | Biocompatability | and D6124-01 | <2mg/glove | | | Primary Skin Irritation in<br>rabbits | Passes | | | Dermal sensitization in the<br>guinea pig | Passes | ## [(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Powder free vinyl patient examination gloves , white(non-colored) meet requirements per ASTM D5250-006, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10. ## [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. ## [(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3). It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for watericak test on pinhole AQL, meet labeling claims . {2}------------------------------------------------ # Indications for Use Applicant: ZIBO ZHUANGYUAN PLASTIC & RUBBER CO., LTD. 510(k) Number (if known): K050154 Device Name: _ Powder Free Vinyl Patient Examination Gloves, White (Non-colored) Indications For Use: Powder free vinyl patient examination glove is a disposable device intended for medical purposes I owder rece miner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of ____________________________________________________________________________________________________________________________________________________________________ {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a circular seal or logo. The seal contains the emblem of the U.S. Department of Health & Human Services (HHS) in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the emblem in a circular fashion. FEB - 7 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Zibo Zhuangyuan Plastic & Rubber Company Limited C/O Ms. Chen Yuhong TUV Rheinland Beijing Office Unit 707, AVIC Building, No. 10B Central Road, East 3rd Ring Road, Chaoyang District, Beijing, P.R. CHINA 100022 Re: K050154 Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, White, (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: January 9, 2005 Received: January 24, 2005 Dear Ms. Yuhong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Yuhong Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Cuts Chiu-Ling, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Applicant: ZIBO ZHUANGYUAN PLASTIC & RUBBER CO., LTD. 510(k) Number (if known): K050154 Device Name: _ Powder Free Vinyl Patient Examination Gloves, White (Non-colored) Indications For Use: Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ZIBO ZHUANGYUAN PLASTIC & RUBBER CO.,LTD YANG Yanhong CHAIRMAN OF BOAR 7180 李 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cittic Proku Joneice Mashwood Tivisio · Sian-Off) 1 Anesthesiology, General Hospital, 1950 Control, Dental Dental Devices Number: K050154
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...