K050099 · Kontron Medical Systems Sas · IYN · Jan 28, 2005 · Radiology
Device Facts
Record ID
K050099
Device Name
SIGMA 5000 SERIES IMAGIC ULTRASOUND SYSTEM
Applicant
Kontron Medical Systems Sas
Product Code
IYN · Radiology
Decision Date
Jan 28, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Diagnostic ultrasound investigations: Imaging (B-mode), Time motion (M-mode), Pulsed wave Doppler (PW Doppler), Continous wave Doppler (CW Doppler), Color Flow Mapping (CFM) adn Color Time motion (CM mode).
Device Story
The Sigma 5000 series, Imagic is a mobile ultrasound console (approx. 60x95x130 cm) featuring a TFT screen and keyboard control panel. It utilizes various transducers to perform B-mode, M-mode, PW/CW Doppler, Color Flow Mapping, and Color Time motion imaging. Operated by clinicians in a diagnostic setting, the device processes ultrasonic echoes to generate real-time images and flow data. These outputs assist healthcare providers in visualizing internal structures and blood flow, supporting clinical diagnosis and patient management. The system includes image storage and hard-copy capabilities.
Clinical Evidence
No clinical data. The device relies on substantial equivalence to existing, legally marketed ultrasound systems using the same technology and principles. Non-clinical bench testing confirmed compliance with safety standards for acoustic output, biocompatibility, and electrical/mechanical safety.
Technological Characteristics
Mobile ultrasound console with TFT display. Supports B-mode, M-mode, PW/CW Doppler, CFM, and CM modes. Includes various transducers (e.g., 2-4 PA, 2-5 CA, 5-12 LA, 3-8 PA, 3-8 TEM, 2 MHz Pen, 8 MHz Pen). Design conforms to 21 CFR 820 and ISO 13485. Connectivity includes image storage and hard-copy output.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, pediatric, small organs (thyroid, breast, testicle), cardiac, transesophageal, peripheral vascular, and musculo-skeletal applications, depending on the specific transducer used.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K050099
Kontron Medical 52, rue Pierre Curie F-78373 Plaisir, Yvelines
Image /page/0/Picture/2 description: The image shows the logo for Kontron Medical. The logo has a stylized letter K on the left, followed by the word "KONTRON" in all caps. Below "KONTRON" is the word "MEDICAL", also in all caps. The font is sans-serif.
# Summary of Safety and Effectiveness
Prepared in accordance with 21 CFR Part 807.92(c)
- 1. Submitter Information JAN 2 8 2005 Kontron Medical SAS a. Submitter: 52, rue Pierre Curie 78373 Plaisir, Yveline France b. Contact Person: Ms. Georgina Fabian Kontron Medical LLC. 9j Brookside Heights Wanaque, NJ 07465 Phone: 973-839-8669 c. Date Prepared: 15 September 2004 2. Name of device a. Trade name: Sigma 5000 series, Imagic b. Common name: Medical Diagnostic Ultrasound Imaging System and transducers Ultrasonic Pulsed Doppler Imaging System 21 CFR 892.1550 c. Classification name: 90-IYN Ultrasonic Pulsed Echo Imaging System 21 CFR 892.1560 90-IYO Diagnostic Ultrasonic Transducer 21 CFR 892.1570 90-ITX
3. Equivalent Legally-Marketed Devices:
Kontron Medical Sigma 110/330, K002239
The technological characteristics of the predicate device are the same as those of the new device.
- 4. Description
The Sigma 5000 series, Imagic is an ultrasound instrument intended to perform the following diagnostic ultrasound investigations: Imaging (B-mode), Time motion (M-mode), Pulsed wave Doppler (PW Doppler), Continuous wave Doppler), Color Flow Mapping (CFM) and Color Time motion (CM mode).
The submission also includes the transducers necessary for these procedures.
The system is a mobile console approximately 60 cm wide, 95 cm deep and 130 cm high equipped with a keyboard control panal, a large TFT screen, assorted transducers and image storage or hard-copy devices
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized letter "K" on the left, followed by the word "KONTRON" in bold, uppercase letters. Below "KONTRON" is the word "MEDICAL" in a smaller font size. The logo is simple and professional, and the use of bold letters makes it easily recognizable.
#### 5. Intended use
Diagnostic ultrasound investigations: Imaging (B-mode), Time motion (M-mode), Pulsed wave Doppler (PW Doppler), Continous wave Doppler (CW Doppler), Color Flow Mapping (CFM) adn Color Time motion (CM mode).
#### 6. Performance Data
- a. Non-clinical tests: The device has been evaluated for acoustic output, biocompatibility and thermal, electrical and mechanical safety, and has been found conform with applicable medical device safety standards.
- b. Clinical tests: Since the Sigma 5000 series Imagic uses the same technology and principles as existing devices, clinical tests are not required.
- c. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820 Quality System Regulation and ISO13485 quality system standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore it is the opinion of Kontron Medical that the Sigma 5000 series Imagic is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
۰۰:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a bird in flight, composed of three curved lines.
JAN 2 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kontron Medical S.A.S. % Mr. Robert Mosenkis President CITECH Medical Device Testing and Consulting 5200 Butler Pike PLYMOUTH MEETING PA 19462-1298
Re: K050099
Trade Name: Sigma 5000 Series, Imagic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: January 17, 2005 Received: January 18, 2005
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Sigma 5000 Series, Imagic Ultrasound System, as described in your premarket notification:
Transducer Model Number
2-4 PA 2-5 CA
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a list of items with numbers and abbreviations. The first item is "5-12 LA", followed by "3-8 PA" and "3-8 TEM". The last two items are "2 MHz Pen" and "8 MHz Pen", with MHz likely referring to megahertz.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your de vice is elassified (360 acre) arols. Existing major regulations affecting your device and including the subject to such it inay be subject to sach additions, Title 21, Parts 800 to 898. In addition, FDA can be found in the Oode of Feactis concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I feast be advised that I Drisones or our device complies with other requirements of the Act that I DA has made a acteriminations administered by other Federal agencies. You must of ally it cochar statutes and regulations, but not limited to: registration and listing (21 Compry with an the Hec STOQuenters)01); good manufacturing practice requirements as set CFN i at 607), adoling (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping I ms device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers In Appendix U, (cheroood) of the Sixinostic Ultrasound Systems and Transducers." If the special Secking Manteling Orcalians unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket I mis lotter will and in your ding of substantial equivalence of your device to a legally marketed nonicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please conact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small
{4}------------------------------------------------
Page 3 - Mr. Mosenkis
Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized letter K on the left, followed by the word "KONTRON" in bold, sans-serif letters. Below "KONTRON" is the word "MEDICAL" in a thinner, sans-serif font. The overall design is simple and modern.
#### System: Sigma 5000 series, Imagic
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|------|------|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PW D | CW D | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | N | |
| Abdominal | | N | N | N | | N | N | | N | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | N | |
| Small organs (specify) | | N | N | N | | N | N | | N | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | N | |
| Transesophageal | | N | N | N | N | N | N | | N | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | N | N | N | | N | |
| Laparascopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | N | N | N | | N | N | | N | |
| Musculo-skeletal Superficial | | N | N | N | | N | N | | N | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
- · Small organs: Thyroid, Breast, Testicle
- Combined modes: B + M, B + PWD, Color Doppler + PWD, Amplitude Doppler + PWD
…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
што такшини и и примении и и причини и и причини и и присини и и присини и пристични и принисимии и принистични и принистични при п ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
(Division Sign Division of Reproductive, and Radiological Devic
KITNEMber Sigma 5000 series, IMAGIC : Summary of Safety and Effectiven
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Kontron Medical. The logo consists of a stylized letter "K" on the left, followed by the word "KONTRON" in bold, uppercase letters. Below "KONTRON" is the word "MEDICAL" in a thinner, uppercase font. The overall design is simple and professional.
### System: Sigma 5000 series, Imagic
Transducer: 2-4 PA
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-----------------------------------|-------------------|---|---|---------|---------|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small organs (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | N | N | N | N | N | N | N | | N | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparascopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal Super-<br>ficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
• Combined modes: B + M, B + PWD, Color Doppler + PWD, Amplitude Doppler + PWD
…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
15 Sept 04
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number ..
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized letter K on the left, followed by the word "KONTRON" in bold, sans-serif font. Below "KONTRON" is the word "MEDICAL" in a thinner, sans-serif font. The logo is black and white.
#### System: Sigma 5000 series, Imagic
#### Transducer: 2-5 CA
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------------|-------------------|---|---|---------|---------|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | N | |
| Abdominal | | N | N | N | | N | N | | N | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | N | |
| Small organs (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparascopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Super-<br>ficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
• Combined modes: B + M, B + PWD, Color Doppler + PWD, Amplitude Doppler + PWD
……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… …………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy C Hodgdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K050090
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized "K" symbol on the left, followed by the word "KONTRON" in bold, sans-serif font. Below "KONTRON" is the word "MEDICAL" in a thinner, sans-serif font. The overall design is simple and professional.
# System: Sigma 5000 series, Imagic
# Transducer: 5-12 LA
Dysteint Digital Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------------|-------------------|---|---|------|------|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PW D | CW D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | | N | N | | N | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | N | |
| Small organs (specify) | | N | N | N | | N | N | | N | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | N | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | N | N | N | | N | |
| Laparascopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | N | N | | N | N | | N | |
| Musculo-skeletal Super-<br>ficial | | N | N | N | | N | N | | N | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
- Small organs: Thyroid, Breast, Testicle
- Combined modes: B + M, B + PWD, Color Doppler + PWD, Amplitude Doppler + PWD
| Concurrence of CDRH, Office of Device Evaluation (ODE) | <img alt="Nancy Brogdon" src="signature.png"/> |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Prescription Use (Per 21 CFR 801.109) | (Division Sign-Off)<br>Division of Reproductive, Abdominal,<br>and Radiological Devices |
| 510(k) Number | 2050099 |
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness 15 Sept 04 Page 6Sigma 5000 series, IMAGIC : Summary of Safety and Effectivencss
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized "K" symbol on the left, followed by the word "KONTRON" in bold, uppercase letters. Below "KONTRON" is the word "MEDICAL" in a smaller, uppercase font. The overall design is simple and professional.
# System: Sigma 5000 series, Imagic
Transducer: 3-8 PA
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-----------------------------------|-------------------|---|---|---------|---------|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | | N | | |
| Small organs (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | N | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparascopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal Super-<br>ficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
• Combined modes: B + M, B + PWD, Color Doppler + PWD, Amplitude Doppler + PWD
…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy C. Bogdan
15 Sept 04
(Division Sian-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness
Page 7
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized "K" symbol on the left, followed by the word "KONTRON" in bold, uppercase letters. Below "KONTRON" is the word "MEDICAL" in a smaller font size, also in uppercase letters. The overall design is simple and professional, suggesting a company in the medical technology field.
## System: Sigma 5000 series, Imagic
## Transducer: 3-8 TEM
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------------|-------------------|---|---|---------|---------|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small organs (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | N | |
| Transesophageal | | N | N | N | N | N | N | | N | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparascopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Super-<br>ficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
• Combined modes: B + M, B + PWD, Color Doppler + PWD, Amplitude Doppler + PWD
この他にある
Y
.............................................................................................................................................................................. ..............................................................................................................................................................................
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
ance (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 5 ! Olk) Number ______________________________________________________________________________________________________________________________________________________________
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Kontron Medical. The logo features a stylized letter "K" on the left, followed by the word "KONTRON" in bold, uppercase letters. Below "KONTRON" is the word "MEDICAL" in a smaller, non-bold font.
## System: Sigma 5000 series, Imagic
### Transducer: 2 MHz Pen
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------------|-------------------|---|---|---------|---------|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small organs (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | N | N | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparascopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Super-<br>ficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: .........................................................................................................................................................
……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
.............................................................................................................................................................................. ..............................................................................................................................................................................
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Choydon
15 Sept 04
Prescription Use (Per 21 CFR 801.109)
(Division Sigh-Off) Division of Reproductive, Abdor ana Radiological Devices 510(k) Number
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness
Page 9
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for Kontron Medical. The logo consists of a stylized letter K on the left, followed by the word "KONTRON" in bold, uppercase letters. Below "KONTRON" is the word "MEDICAL" in a smaller font size, also in uppercase letters. The overall design is simple and professional.
## System: Sigma 5000 series, Imagic
### Transducer: 8 MHz Pen
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------------|-------------------|---|---|---------|---------|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neuro-<br>logical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small organs (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | N | N | | | | | |
| Laparascopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Super-<br>ficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: .........................................................................................................................................................
……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
.............................................................................................................................................................................. ..............................................................................................................................................................................
..............................................................................................................................................................................
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Swogdon
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Reproductive, Abdominal, and Padiological Devices 2050094 51 Jan Number _
Sigma 5000 series, IMAGIC : Summary of Safety and Effectiveness
15 Sept 04
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.