← Product Code [LON](/productcode/LON) · K050089

# BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - AMPICILLIN-SULBACTAM (GP) 2-32 UG/ML (K050089)

_Becton, Dickinson & CO · LON · Feb 28, 2005 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050089

## Device Facts

- **Applicant:** Becton, Dickinson & CO
- **Product Code:** [LON](/productcode/LON.md)
- **Decision Date:** Feb 28, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

## Device Story

The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes sealed, self-inoculating polystyrene panels containing dried biochemicals and antimicrobial agents. Input consists of a pure culture bacterial isolate prepared in Phoenix ID or AST broth. The system uses a redox indicator to detect bacterial growth in the presence of antimicrobial agents via turbidity measurements. Panels are continuously incubated at 35°C, with the instrument taking readings every 20 minutes. The system software interprets these readings to determine Minimum Inhibitory Concentration (MIC) values and provides category interpretations (S, I, R, or N). Used in clinical microbiology laboratories by trained technicians, the output assists clinicians in selecting appropriate antibiotic therapy for patients. The system provides rapid results compared to traditional manual methods, potentially improving patient outcomes through timely targeted treatment.

## Clinical Evidence

Bench testing only. Performance evaluated using 1,240 clinical and stock isolates (Staphylococcus and Enterococcus). Compared against CLSI broth dilution reference method. Results: 97.2% Essential Agreement (EA) and 97.3% Category Agreement (CA). Reproducibility >95%. Quality control performed using S. aureus ATCC 29213, S. aureus ATCC 25923, E. coli ATCC 25922, and E. coli ATCC 35218.

## Technological Characteristics

Automated microbiology system using sealed, self-inoculating molded polystyrene trays (136 micro-wells). Employs colorimetric oxidation-reduction (redox) indicator for growth detection. Cation-adjusted Mueller-Hinton broth with 0.01% Tween 80. Incubation at 35°C. Software-driven 'EXPERT' system applies CLSI-derived rules for interpretation. Connectivity: barcode scanning of panels.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK® System (PMA No. N50510)
- BD Phoenix™ Automated Microbiology System with Gatifloxacin ([K020323](/device/K020323.md))
- BD Phoenix™ Automated Microbiology System with Ofloxacin ([K020323](/device/K020323.md))
- BD Phoenix™ Automated Microbiology System with Levofloxacin ([K020322](/device/K020322.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k050089

B. Purpose for Submission:
Addition of ampicillin-sulbactam to the Phoenix™ Automated Microbiology System

C. Measurand:
Ampicillin-Sulbactam 2/1 – 32/16 μg/mL

D. Type of Test:
Antimicrobial Susceptibility Test (Quantitative and Qualitative) colorimetric
oxidation-reduction, growth-based

E. Applicant:
Becton, Dickinson &amp; Company

F. Proprietary and Established Names:
BD Phoenix™ Automated Microbiology System – Ampicillin-Sulbactam 2/1 – 32/16
μg/mL Gram-Positive Panel

G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle
Antimicrobial Susceptibility System
2. Classification:
Class II
3. Product Code:
LON
4. Panel:
83

H. Intended Use:
1. Intended use(s):
BD Phoenix™ Automated Microbiology System:
The BD Phoenix™ Automated Microbiology System is intended for
in vitro quantitative determination of antimicrobial susceptibility by minimal
inhibitory concentration (MIC) of gram-negative aerobic and facultative
anaerobic bacteria belonging to the family Enterobacteriaceae and
non-Enterobacteriaceae and gram-positive bacteria belonging to the genera
Staphylococcus and Enterococcus.

{1}

Page 2 of 6

The BD Phoenix™ GP Panel:

The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-positive bacteria from pure culture belonging to the genera *Staphylococcus* and *Enterococcus*.

2. Indication(s) for use:

This submission is for the addition of the antibiotic Ampicillin-Sulbactam at concentrations of $2/1 - 32/16\ \mu\mathrm{g/mL}$ to the gram-positive susceptibility panel.

3. Special condition for use statement(s):

Prescription use only

4. Special instrument Requirements:

Not applicable

I. Device Description:

The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram-positive or gram-negative. Colonies are then suspended in broth, and equated to a $0.5\ \mathrm{McFarland}$ with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing $0.01\%$ Tween 80. After adding the indicator solution to the AST inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at $35^{\circ}\mathrm{C}$. The AST has a final inoculum of $5\times 10^{5}$ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the CLSI standards.

Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible.

{2}

Page 3 of 6

# J. Substantial Equivalence Information:

1. Predicate device name(s): VITEK® System
2. Predicate K number(s): N50510
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  1. | Isolated colonies from culture used | Isolated colonies from culture used  |
|  2. | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR)  |
|  3. | <16 hours | <16 hours  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  1. | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions  |
|  2. | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 1.0 McFarland standard  |
|  3. | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner.  |

# K. Standard/Guidance Document Referenced (if applicable):

"Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S15) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard."

# L. Test Principle:

The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in "growth control wells" which contains no antibiotic.

# M. Performance Characteristics (if/when applicable):

1. Analytical performance:

{3}

Page 4 of 6

a. Precision/Reproducibility:

Fifteen gram positive isolates were evaluated for site to site and inter site reproducibility demonstrating &gt;95% reproducibility. The ten isolate study described in the guidance document was used (10 organisms tested 3 times on 3 days at 3 sites).

b. Linearity/assay reportable range:

Not applicable

c. Traceability (controls, calibrators, or method):

The CLSI recommended QC isolates for this drug were both gram-negative organisms which would not be appropriate for the panel tested therefore, S. aureus ATCC 29213 was used for this purpose. It was tested on every test occasion with the reference method and the BD Phoenix™. Since this QC strain is not recommended by CLSI, results obtained for this QC organism with the reference system was not used to determine the acceptability of the data obtained from isolates tested on a given day. The Phoenix™ was tested a sufficient number of times to demonstrate that the system can produce QC results in the expected range most of the time. The BD Phoenix™ and the reference device had the same mode for the QC organism. Additional QC organisms were tested with the reference method as shown in the table below to determine the acceptability of the data obtained from isolates tested on that given day.

|  ORGANISM | conc. | Reference |   | Phoenix™  |   |
| --- | --- | --- | --- | --- | --- |
|  |   |   |   |   |   |
|  S. aureus
ATCC 29213
Expected Result:
≤2 μg/mL | ≤2 |  | 165 |  | 161  |
|   | 4 |  |  |  |   |
|   | 8 |  |  |  | 1  |
|  |   |   |   |   |   |
|  |   |   |   |   |   |
|  S. aureus | ≤2 |  |  |  | 163  |
|  ATCC 25923 |  |  |  |  |   |
|  Expected Result: |  |  |  |  |   |
|  ≤2 μg/mL |  |  |  |  |   |
|  |   |   |   |   |   |
|  E. coli | ≤2 |  | 27 |  |   |
|  ATCC 25922 | 4 |  | 136 |  |   |
|  Expected Result: |  |  |  |  |   |
|  2 - 8 μg/mL |  |  |  |  |   |
|  |   |   |   |   |   |
|  E. coli | 16 |  | 88 |  |   |
|  ATCC 35218 | 32 |  | 74 |  |   |
|  Expected Result: | >32 |  | 2 |  |   |
|  8 - 32 μg/mL |  |  |  |  |   |

{4}

Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the  $\mathrm{BBL}^{\mathrm{TM}}$  CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the  $\mathrm{BBL}^{\mathrm{TM}}$  CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used.

d. Detection limit: Not applicable
e. Analytical specificity: Not applicable
f. Assay cut-off: Not applicable

# 2. Comparison studies:

a. Method comparison with predicate device: The CLSI recommended broth dilution reference panel was prepared according to the CLSI standards. Clinical testing was performed at four sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. Since Oxacillin Resistant isolates will be reported as Resistant, only Oxacillin Sensitive isolates were evaluated. The test device had a growth rate of greater than  $90\%$ . A comparison was provided to the reference method with the following agreement.

Staphylococcus spp. is presented in the table below because it has different breakpoints than the Enterococcus spp. The overall performance of this drug with these two organism groups is acceptable.

Summary Table for Staphylococcus spp.

|   | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Staphylococcus | 693 | 672 | 97 | 143 | 126 | 88 | 665 | 96 | 8 | 27 | 0 | 1  |
|  Enterococcus | 547 | 533 | 97.4 | 42 | 36 | 85.7 | 541 | 98.9 | 216 | N/A | 6 | 0  |
|  Combined | 1240 | 1205 | 97.2 | 185 | 162 | 87.6 | 1206 | 97.3 | 224 | 27 | 6 | 1  |

EA-Essential Agreement

maj-major discrepancies

CA-Category Agreement

vmj-very major discrepancies

R-resistant isolates

min- minor discrepancies

N/A - No intermediate range therefore no minor errors possible

Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference

{5}

Page 6 of 6

method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation.

b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Enterococcus spp. ≤8(S), ≥16(R)
Staphylococcus spp. ≤8/4(S), 16/8(I), ≥32/16(R)

The expected value range, interpretative criteria and QC are the same as recommended by CLSI. All values will be included in the package insert.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K050089](https://fda-staging.innolitics.com/device/K050089)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com