REHABILICARE IF 3 WAVE INTERFERENTIAL STIMULATOR SYSTEM, MODEL 7110S
K050046 · Compex Technologies, Inc. · IPF · Apr 13, 2005 · Physical Medicine
Device Facts
Record ID
K050046
Device Name
REHABILICARE IF 3 WAVE INTERFERENTIAL STIMULATOR SYSTEM, MODEL 7110S
Applicant
Compex Technologies, Inc.
Product Code
IPF · Physical Medicine
Decision Date
Apr 13, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Interferential stimulation can be used for the following applications: - Relieve acute pain . - . Relieve and manage chronic pain - Relax muscle spasms . - Maintain and increase the range of motion . - Increase local blood circulation . Neuromuscular stimulation can be used for the following applications: - Relax muscle spasms . - Prevent or retard disuse atrophy . - Increase local blood circulation . - Re-educate muscles . - Maintain or increase the range of motion . - Prevent venous thrombosis using immediate postsurgical stimulation on calf . muscles Pulsed direct current stimulation can be used for the following applications: - Reduction of edema (under negative electrode) t - Reduction of muscle spasm . - Influencing local blood circulation (under negative electrode) . - Retardation or prevention of disuse atrophy . - Facilitation of voluntary motor function . - Maintenance or increase of range of motion . Interferential and Neuromuscular combination stimulation can be used for the following applications: - Reduction of edema (under negative electrode) . - Reduction of muscle spasm .
Device Story
IF 3Wave (Model 7110S) is a microprocessor-controlled, dual-channel, non-invasive electro-stimulator. It delivers four modalities: Interferential (IF), Neuromuscular Electrical Stimulation (NMES), Pulsed Direct Current (PDC), and combined IF/NMES. Device stores settings and compliance data for 90 days; data is downloaded via a battery charger/modem base for physician review. Physicians use reports to adjust treatment regimens. Device features include automatic overload/no-load trips, automatic shut-off, and patient override controls. Powered by a lithium-ion polymer rechargeable battery. Used in clinical or home settings under physician prescription to manage pain, muscle function, and circulation.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, comparison of technical specifications, and functional equivalence to predicate devices.
Technological Characteristics
Microprocessor-controlled dual-channel stimulator. Case: ABS/PC blend. Power: Lithium-ion polymer rechargeable battery. Waveforms: Symmetrical biphasic square wave (0 net DC) and PDC. Output: 0-50 mA (IF) or 0-100 mA (NMES/PDC) into 500-1000 ohm loads. Features: Automatic overload/no-load trip, automatic shut-off, patient override. Connectivity: Modem-based data download for compliance reporting. Sterilization: Not applicable (non-invasive).
Indications for Use
Indicated for patients requiring pain relief (acute/chronic), muscle spasm relaxation, range of motion maintenance/increase, local blood circulation improvement, disuse atrophy prevention/retardation, muscle re-education, venous thrombosis prevention (postsurgical calf stimulation), edema reduction, and voluntary motor function facilitation.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
Predicate Devices
Staodyne, Inc. EMS+2 Neuromuscular Stimulation System (K926510)
Medical Devices, Inc. Model IF-II Interferential Stimulation System (K923914)
Staodyne, Inc. Model SporTX Pulsed Direct Current and TENS Stimulator (K921668)
Axelgaard Manufacturing Company TENS Pals Plus Electrodes (K872976)
Axelgaard Manufacturing Company Valutrode Electrodes (K970426)
Submission Summary (Full Text)
{0}------------------------------------------------
KC50046
#### APR 1 3 2005
**Summary of Safety and Effectiveness:**
## Summary of Safety and Effectiveness
#### Applicant Information
Compex Technologies, Inc. 1811 Old Highway 8 New Brighton, MN 55112 Phone: (612) 638-0590 (612) 638-0479 FAX:
#### Contact Person
Brian J. Edwards, M.S. Director of Regulatory Affairs & Quality Assurance
## Date Summary Prepared
September 1, 2004
## Device identification
| Trade / Proprietary Name: | IF 3Wave Interferential Muscle Stimulator System,<br>Model 7110S |
|---------------------------|------------------------------------------------------------------|
| Common / Usual Name: | Muscle and Interferential Current |
| Classification Name: | 21 CFR 890.5850 "Powered Muscle Stimulator" |
| Product Classification: | Class II |
| Product Code: | IPF |
{1}------------------------------------------------
### Legally Market Predicate Devices
| Predicate Device | 510(k) Number / Clearance Date |
|--------------------------------------------------------------------------|--------------------------------|
| Staodyne, Inc. EMS+2 Neuromuscular<br>Stimulation System | K926510 / August 11, 1993 |
| Medical Devices, Inc. Model IF-II<br>Interferential Stimulation System | K923914 / November 6, 1992 |
| Staodyne, Inc. Model SporTX Pulsed<br>Direct Current and TENS Stimulator | K921668 / August 3, 1992 |
| Axelgaard Manufacturing Company<br>TENS Pals Plus Electrodes | K872976 / September 22, 1987 |
| Axelgaard Manufacturing Company<br>Valutrode Electrodes | K970426 / May 9, 1997 |
The IF 3Wave Interferential Stimulator System, Model 7110S is comparable in type and function and are substantially equivalent to the following predicate devices.
#### Device Description
The IF 3 Wave is a non-invasive medium/low frequency interferential muscle rie it -> wave is a intended for treating patients that have acute and chronic pain, edema, tight musculature, muscle spasms and muscular weakness due to disuse atrophy. The IF 3 Wave stimulator is a microprocessor controlled dual channel Interferential electro-stimulator with neuro-muscular and pulse direct current micronital expabilities. It will offer 4 separate electro-therapy stimulation modalities. Stimulation capacificies: 17 Will on P), neuromuscular electrical stimulation (NMES), r neso direct current (PDC) electrical stimulation, and a combination of interferential and neuromuscular electrical stimulation (IF/NMES).
The IF 3 Wave will be capable of storing device settings, use, and compliance data for up to 90 days. It will be capable of down loading the above compliance date through a up to 90 days. which is contained within the battery charger base. Once per phone mic modelin, will be instructed to download the stored data from their device, month, the patient will over compliance report which is forwarded to their physicians for which will ereate a paysician will review the report and provide adjustment to the patient treatment regimen and device setting accordingly.
Accessories provided with the IF 3 Wave include lead wires, electrode pads, Accessories proTracu wik, battery charger/modem, phone cable, power supply, and a users manual.
{2}------------------------------------------------
# Technical Characteristics
The IF functional mode of the IF 3Wave is substantially equivalent to the functional mode of the Rehabilicare IF-II device.
| Interferential Modality. | | |
|--------------------------------------------------|----------------------------------------------------------|----------------------------------------------|
| | IF-II<br>Model 7100S | IF 3Wave<br>Model 7110S |
| 510(k) Number | K923914 | |
| Manufacturer | Medical Devices, Inc. | Compex Technologies, Inc. |
| Power Source | 4 AA rechargeable batteries<br>Line Powered Repack | Lithium ion polymer rechargeable<br>battery. |
| Weight | 12 oz, with battery | 12 oz, with battery |
| Dimensions | 3.2 x 5.2 x 1.15 in. | 3.89 x 6.33 x 1.42 in |
| Case Materials | ABS/PC Blend | Same |
| Patient Leakage Current* | | |
| Normal | Unknown | < 0.1 microamperes |
| Single Fault | Unknown | 2.4 microamperes |
| Output Modes | 1 | 3 |
| Number of Output<br>Channels | 2 | Same |
| Method of Channel<br>Isolation | Transformer isolation | Same |
| Regulated Current or<br>Voltage | Voltage | Same |
| Software, Firmware,<br>Microprocessor Controlled | Yes | Same |
| Automatic Overload Trip | No | Yes, 5 kohms |
| Automatic No-Load Trip | Yes | Same |
| Automatic Shut-off | Yes | Same |
| Patient Override Controls | Yes | Same |
| Indicator Display<br>On/Off Status | Yes | Same |
| Low Battery | Yes | Same |
| Voltage, Current Level | Yes | Same |
| Waveform | Symmetrical biphasic square<br>wave with 0 net DC | Same |
| Pulse Width | 125 microseconds | Same |
| Carrier Pulse Frequency<br>(Channel 1) | 4000 Hz, fixed | Same |
| Adjustable Pulse<br>Frequency<br>(Channel 2) | 4001 — 4150 Hz | Same |
| Interferential Beat<br>Frequency | 1 - 150 Hz | Same |
| Pulse Amplitude | 0 - 50 mA, adjustable in 1 mA<br>steps into 500 ohm load | Same |
{3}------------------------------------------------
| Indications for Use | Relieve acute pain<br>Relieve and manage chronic<br>pain<br>Relax muscle spasms<br>Maintain and increase the range<br>of motion<br>Increase local blood circulation | Same |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Number of Continuous<br>Treatments | 1 | Same |
| Continuous Treatment<br>Time Selection | 20 minutes, fixed | 10 - 60 minutes, selectable in 5<br>minute steps |
| Number of Variable Sweep<br>Programs | 3 | 4 |
| Sweep Treatment Time<br>Selection | 20 minutes, fixed | 10 - 60 minutes, selectable in 5<br>minute steps |
| Sweep Program Cycle<br>Times | 1 sec Up / 1 sec Down<br>6 sec Up / 6 sec Down | 6 sec Up / 6 sec Down<br>12 sec Up / 12 sec Down<br>24 sec Up / 24 sec Down |
| Number of Fixed<br>Parameter Sweep<br>Programs | 0 | 7 |
| Sweep Program Frequency<br>Range | 1 - 150 Hz, selectable<br>continuously using a rheostat | 1 - 150 Hz, selectable in 1 Hz steps |
| Maximum Charge per<br>Pulse | 120 microcoulombs | Same |
| Maximum Output Voltage | 50 volts, peak-to-peak, 500 ohm<br>load | Same |
| Peak Output Current | 100 mA, peak-to-peak, 500<br>ohm load | Same |
| Net Charge per Pulse | 0 microcoulombs, biphasic<br>waveform | Same |
| Max Phase Charge | 6.25 microcoulombs | Same |
| Max Current Density# | 1.937 mA / cm2 (average) | Same |
| Max Power Density# | 0.097 W/cm2 (average) | Same |
| Burst Modes | N/A | N/A |
| Compliance with Voluntary<br>Standards | Unknown | Yes, refer to page 43 of the<br>submission for list |
| NMES Modality. | | |
| EMS+2<br>Model 6840S | IF 3Wave<br>Model 7110S | |
| 510(k) Number | K926510 | |
| Manufacturer | Staodyne, Inc. | Compex Technologies, Inc. |
| Power Source | 9 volt battery - alkaline or nickel<br>metal hydride rechargeable | Lithium ion polymer<br>rechargeable battery. |
| Weight | 7.97 oz, with battery | 12 oz, with battery |
| Dimensions | 5.4 x 3.2 x 1.2 in. | 3.89 x 6.33 x 1.42 in |
| Case Materials | ABS/PC Blend | Same |
| Waveform | Symmetrical biphasic with zero<br>net DC component | Same |
| Patient Leakage Current* | | |
| Normal | Unknown | < 0.1 microamperes |
| Single Fault | Unknown | 2.4 microamperes |
| Output Modes | 1 | 3 |
| Number of Output Channels | 2 | Same |
| Method of Channel Isolation | N/A | Same |
| Regulated Current or Voltage | Current | Same |
| Software, Firmware,<br>Microprocessor Controlled | Yes | Same |
| Automatic Overload Trip | No | Same |
| Automatic No-Load Trip | Yes | Same |
| Automatic Shut-off | Yes | Same |
| Patient Override Controls | Yes | Same |
| Indicator Display | | |
| On/Off Status | Yes | Same |
| Low Battery | Yes | Same |
| Voltage, Current Level | Yes | Same |
| Number of Preset Programs | 0 | 6 |
| Pulse Width | 50 - 300 microseconds, adjustable | 70 - 300 microseconds in 10<br>microsecond steps |
| Pulse Frequency | 4 - 80 Hz, adjustable | 10 – 80 Hz in 5 Hz steps |
| Pulse Amplitude | 0 – 100 mA across a 1000 Ohm<br>load, adjustable | 0 - 100 mA in 1 mA steps<br>across a 1000 Ohm load |
| Indications for Use | Relax muscle spasms<br>Prevent or retard disuse atrophy<br>Increase local blood circulation<br>Re-educate muscles<br>Maintain or increase the range of<br>motion | Same |
| | | |
| | Prevent venous thrombosis using<br>immediate postsurgical stimulation<br>on calf muscles | |
| Treatment Time Selection | 15, 30, or 60 minutes, or<br>Continuous, selectable | 10 - 60 minutes, in 5 minute<br>increments |
| Cycle ON Time | 0.5 - 30 seconds, adjustable | 1 - 30 seconds, in 1 second<br>increments |
| Cycle OFF Time | 0.1 - 60 seconds, adjustable | 1 - 60 seconds, in 1 second<br>increments |
| Ramp-Up Time | 0.1 - 6, adjustable | 0 -10 seconds, in 1 second<br>increments |
| Ramp-Down Time | 0.1 - 6 seconds, adjustable | 0 - 10 seconds, in 0.5 second<br>increments |
| Maximum Output Voltage | 100 volts, 0-to-peak, 1000 ohm<br>load | Same |
| Peak Output Current | 100 mA, 0 -to-peak, 1000 ohm<br>load | Same |
| Maximum Charge per Pulse | 30 microcoulombs | Same |
| Net Charge per Pulse | 0 microcoulombs, biphasic<br>waveform | Same |
| Max Phase Charge | 30 microcoulombs | Same |
| Max Current Density# | 0.19 mA/cm2 (average) | Same |
| Max Power Density# | 0.019 W/cm2 (average) | Same |
| Burst Modes | N/A | N/A |
| Compliance with Voluntary<br>Standards | Unknown | Yes, refer to page 43 of the<br>submission for list |
| PDC Modality. | | |
| | SporTX<br>Model 7700S | IF 3Wave<br>Model 7110S |
| 510(k) Number | K921668 | |
| Manufacturer | Staodyne, Inc. | Compex Technologies, Inc. |
| Power Source | 9 volt alkaline or rechargeable<br>battery | Lithium ion polymer<br>rechargeable battery. |
| Weight | 4.4 oz, with battery | 12 oz, with battery |
| Dimensions | 2.1 x 4.2 x 0.9 in | 3.89 x 6.33 x 1.42 in |
| Case Materials | ABS/PC Blend | Same |
| Patient Leakage Current* | | |
| Normal | Unknown | < 0.1 microamperes |
| Single Fault | Unknown | 2.4 microamperes |
| Output Modes | 1 | 3 |
| Number of Output Channels | 2 | Same |
| Method of Channel Isolation | N/A | Same |
| Regulated Current or<br>Voltage | Current | Same |
| Software, Firmware,<br>Microprocessor Controlled | Yes | Same |
| Automatic Overload Trip | No | Same |
| Automatic No-Load Trip | Yes | Same |
| Automatic Shut-off | Yes | Same |
| Patient Override Controls | Yes | Same |
| Indicator Display | | |
| On/Off Status | Yes | Same |
| Low Battery | Yes | Same |
| Voltage, Current Level | Yes | Same |
| Waveform 1 | Symmetrical biphasic squarewave<br>with zero net DC | Same |
| Waveform 2 | Unbalanced Triphasic with<br>positive net DC | Same |
| Number of Preset Programs | 0 | 3 |
| Pulse Width | Biphasic: 60 microseconds<br>Triphasic: 60 microseconds | Same |
| Pulse Frequency | Biphasic: 278 Hz<br>Triphasic: 222 Hz | Same |
| Pulse Amplitude | 0 - 100 mA across a 1000 Ohm<br>Load, adjustable | 0 - 100 mA in 1 mA<br>increments across a 1000<br>Ohm load |
| Indications for Use | Reduction of edema (under<br>negative electrode) | Same |
| | | |
| | Reduction of muscle spasm | |
| | Influencing local blood circulation | |
| | Retardation or prevention of<br>disuse atrophy | |
| | Facilitation of voluntary motor<br>function | |
| | Maintenance or increase of range<br>of motion | |
| Treatment Time Selection | 38 minutes or Continuous,<br>selectable | 30 - 60 minutes, in 5 minute<br>increments |
| Maximum Output Voltage | 100 Volts, 100 ohm load | Same |
| Maximum Charge per Pulse | 6 microcoulombs | Same |
| Average DC Current | 266 microamperes | 300 microamperes |
| Net Charge per Pulse | 0 microcoulombs, biphasic<br>waveform | Same |
| Max Phase Charge | 60 microcoulombs per pulse | Same |
| Max Current Density# | 0.13 mA/cm² (average) | Same |
| Max Power Density# | 0.013 W/cm² (average) | Same |
| Burst Modes | N/A | N/A |
| Compliance with Voluntary<br>Standards | Unknown | Yes, refer to page 43 of the<br>submission for list |
*Patient Leakage Current was determined following test procedure within IEC 601-1
#These values were calculated using the average current over time, not pulse current, with a 500 ohm load.
{4}------------------------------------------------
The NMES functional mode of the IF 3Wave is substantially equivalent to the functional mode of the Rehabilicare EMS+2 device.
{5}------------------------------------------------
*Patient Leakage Current was determined following test procedure within IEC 601-1
#These values were calculated using the average current over time, not pulse current, with a 500 ohm load.
{6}------------------------------------------------
The PDC functional mode of the IF 3Wave is substantially equivalent to the functional mode of the Rehabilicare SporTX device.
{7}------------------------------------------------
## *Patient Leakage Current was determined following test procedure within IEC 601-1
#These values were calculated using the average current over time, not pulse current, with a 500 ohm load.
{8}------------------------------------------------
### Indications for use
The IF 3Wave Interferential Stimulator System is intended for:
Interferential stimulation can be used for the following applications:
- Relieve acute pain .
- . Relieve and manage chronic pain
- Relax muscle spasms .
- Maintain and increase the range of motion .
- Increase local blood circulation .
Neuromuscular stimulation can be used for the following applications:
- Relax muscle spasms .
- Prevent or retard disuse atrophy .
- Increase local blood circulation .
- Re-educate muscles .
- Maintain or increase the range of motion .
- Prevent venous thrombosis using immediate postsurgical stimulation on calf . muscles
Pulsed direct current stimulation can be used for the following applications:
- Reduction of edema (under negative electrode) t
- Reduction of muscle spasm .
- Influencing local blood circulation (under negative electrode) .
- Retardation or prevention of disuse atrophy .
- Facilitation of voluntary motor function .
- Maintenance or increase of range of motion .
Interferential and Neuromuscular combination stimulation can be used for the following applications:
- Reduction of edema (under negative electrode) .
- Reduction of muscle spasm .
{9}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/9/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract image of an eagle.
APR 1 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Compex Technologies, Inc. c/o Mr. Daniel Lehtonen Intertek Testing Services 70 Codman Hill Road Boxborough, Massachusetts 01719
Re: K050046
Trade/Device Name: IF 3 Wave Interferential Muscle Stimulator System, Model 7110S Regulation Numbers: 21 CFR 890.5850, 21 CFR 882.5890 Regulation Name: Powered muscle stimulator, Transcutaneous electrical nerve stimulator for pain relief and Interferential current therapy Regulatory Class: II Product Codes: GZJ, IPF and LIH Dated: March 28, 2005 Received: March 29, 2005
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reaso or admised a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2 - Mr. Daniel Lehtonen
This letter will allow you to begin marketing your device as described in your Section 510(k) This icher will anow you to begin mailing of substantial equivalence of your device to a legally premailed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at no 101-11-1120. Also, please note the regulation entitled, Colliact the Office of Compilance in (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Renele K
Miriam Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### 050046 510(k) Number (if knowm):
#### Device Name
Compex Technologies IF 3Wave Interferential Stimulator System, Model 7110S.
#### Indications for Use
Interferential stimulation can be used for the following applications:
- Relieve acute pain .
- Relieve and manage chronic pain .
- Relax muscle spasms ●
- Maintain and increase the range of motion .
- Increase local blood circulation .
Neuromuscular stimulation can be used for the following applications:
- . Relax muscle spasms
- Prevent or retard disuse atrophy .
- Increase local blood circulation .
- Re-educate muscles .
- Maintain or increase the range of motion .
- Prevent venous thrombosis using immediate postsurgical stimulation on calf . muscles
Pulsed direct current stimulation can be used for the following applications:
- Reduction of edema (under negative electrode) .
- Reduction of muscle spasm .
- Influencing local blood circulation (under negative electrode) .
- Retardation or prevention of disuse atrophy .
- Facilitation of voluntary motor function .
- Maintenance or increase of range of motion .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Restorative
Ko50046
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.