← Product Code [GCJ](/productcode/GCJ) · K050027

# LAPAROCISION SCOPE CONTROLLER SYSTEM (K050027)

_Gmp/Companies, Inc. · GCJ · Jan 27, 2005 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K050027

## Device Facts

- **Applicant:** Gmp/Companies, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 27, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Laparocision™ Scope Controller System is an electromechanical arm whose intended use is to allow a surgeon to hold and control a laparoscope's movement and position during laparoscopic surgical procedures by means of either a foot controller or a hand controller.

## Device Story

Electromechanical arm system for holding/positioning laparoscopes during minimally invasive surgery. Input: operator commands via hand or foot controller. Transformation: electromechanical actuation of pinch rollers in Disposable Zoom Assembly (DZA) to move scope in three axes (in/out, head/foot, medial/lateral). Output: physical movement of laparoscope. Used in OR by surgeons. System includes reusable arm, DZA, hand/foot controls, sterile polyethylene sleeves, and power supply. Benefits: provides stable, surgeon-controlled visualization of surgical cavity without manual holding.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence consists of in-vitro electrical safety testing (UL 60601-1, CSA C22.2 NO 601.1-M90, IEC 60601-2-18) and electromagnetic compatibility (EMC) testing (IEC 60601-1-2). Mechanical and functional testing confirmed design and performance requirements were met.

## Technological Characteristics

Electromechanical arm assembly; pinch roller mechanism for scope manipulation. Materials include sterile polyethylene sleeves. Power: 110 VAC to 9 VDC conversion. Connectivity: None (non-computerized). Standards: UL 60601-1, CSA C22.2 NO 601.1-M90, IEC 60601-2-18, IEC 60601-1-2. Sterilization: Sterile barrier sleeves used for reusable components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Computer Motion Inc. AESOP System and Accessories ([K931783](/device/K931783.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K050027 1/3

# 510(k) Summary

# GMP|Surgical Solutions Inc. Laparocision™ Scope Controller System

Classification Name: Endoscope and Accessories 21 CFR 876.1500 Device Class: II Product Code: GCJ

> GMP Companies, Inc. One East Broward Boulevard Suite 1701 Fort Lauderdale, FL 33301

> > Phone: (954) 745-3510 Fax: (954) 745-3511

Contact: Ralph Jugo, Prepared: October 25, 2004

#### LEGALLY MARKETED PREDICATE DEVICE A.

The GMP|Surgical Solutions Inc. Laparocision™ Scope Controller System is substantially equivalent to the Computer Motion Inc. AESOP System and Accessories that was cleared by FDA under K931783 on November 22, 1993.

#### B. DEVICE DESCRIPTION

The GMP|Surgical Solutions Laparocision™ Scope Controller System allows for surgeons to control the movement and position of conventional laparoscopes or endoscopes during thoracic, gynecologic, orthopedic, and abdominal laparoscopic procedures by means of an electromechanical arm which is controlled by either a hand controller or a foot controller. The purpose of an endoscope is to allow visualization of a surgical cavity during minimally invasive laparoscopic procedures.

The device is comprised of the following five main components:

1) The Laparocision™ Reusable Arm Assembly is the device's main component. It is attached via a clamp to an operating table and it serves to hold and control the position and movement of a laparoscope.

2) The Disposable Zoom Assembly (DZA) is attached to the distal end of the Reusable Arm Assembly, and its purpose is to hold 5mm and 10 mm diameter laparoscopes between its pinch rollers. The pinch rollers turn and act to move a laparoscope in three axes (i.e. in and out, towards the head or foot of the patient, and medially/laterally). The pinch rollers in the DZA are under the direct control of an operator via the Foot Control and the Hand Control.

{1}------------------------------------------------

3) The Foot Control and the Hand Control plug into the Reusable Arm, and by pressing control buttons on the Foot and Hand Controls, the operator can move the laparoscope in three axes.

4) The Steri-Sleeves are sterile polvethylene material bags which act to provide for protection and a sterile barrier between the reusable components (i.e. the Reusable Arm Assembly, and the Foot and Hand Controls) and the sterile field.

5) The Power Supply converts 110 alternating current voltage (VAC) to 9 volts of direct current (9VDC) required to operate the Reusable Arm Assembly.

#### C. INDICATIONS FOR USE

The Laparocision™ Scope Controller System is indicated for thoracic, gynecologic, orthopedic, and abdominal procedures for the purpose of holding and controlling the movement of standard laparoscopes or rigid endoscopes within inflated surgical cavities during minimally invasive laparoscopic surgery.

#### D. INTENDED USE

The Laparocision™ Scope Controller System is an electromechanical arm whose intended use is to allow a surgeon to hold and control a laparoscope's movement and position during laparoscopic surgical procedures by means of either a foot controller or a hand controller.

The trade name or proprietary name of the device which is the subject of the 510(k) premarket notification is the "GMP|Surgical Solutions Inc. Laparocision™ Scope Controller System". The common or usual name for the device is an "Endoscope Accessory".

#### E. SUBSTANTIAL EQUIVALENCE SUMMARY

The GMP|Surgical Solutions Inc. Laparocision™ Scope Controller System is an electromechanical medical device that has essentially the same indications for use as the predicate device. Both devices are indicated for laparoscopic surgical procedures. The only main differences between the two devices is that the AESOP device uses a computer and software to control the devices operations, while the GMP Laparocision device does not use a computer or software to control the device's operations. The differences do not alter the intended diagnostic effect. Both devices have essentially the same intended use. A comparison of the descriptive characteristics of the GMP Laparocision device and the AESOP predicate device is sufficiently precise to assure equivalence. In-vitro performance data are provided in the form of electrical safety testing and electromagnetic compatibility (EMC) testing that was performed on both devices. The results of the performance testing demonstrate compliance with the applicable electrical safety and the data also support a determination of substantial equivalence.

{2}------------------------------------------------

As recommended by the 3-17-95 (Draft) Guidance for the Content of Premarket Notifications for Endoscopes Used in Gastroenterology and Urology, electrical safety and Electromagnetic Compatibility (EMC) testing as per the applicable standards should be conducted and provided in a 510(k) submission.

#### F. TECHNOLOGICAL CHARACTERISTICS

Both devices use electronic components to control the position and movement of laparoscopes during minimally invasive laparoscopic surgical procedures. While the details differ regarding how the control of the movement of laparoscopes is achieved, given that the AESOP device uses a computer and software while the GMP device does not, the basic technology is very similar. The differences in technology do not raise any significant questions or issues regarding the safety and effectiveness of the devices,

#### G. TESTING

In-vitro electrical safety testing was conducted and met all the requirements in accordance with the standards, "UL 60601-1-UL Standard for Safety of Medical Device Equipment, Part 1: General Requirements for Safety First Edition", CSA C22.2 NO 601.1-M90 -Medical Electrical Equipment - Part 1: General Requirements for Safety General Instructions No 1; Supplement 1; 1994 R(1997)", IEC 60601-2-18 (1996-08) - Medical Electrical Equipment -- Part 2: Particular Requirements for the Safety of Endoscopic Equipment Second Edition; Amendment 1, 07-2000".

In-vitro Electromagnetic Compatibility (EMC) testing was performed and met all the requirements as per "IEC 60601-1-2 (2001-09) - Medical Electrical Equipment Part 1-2; General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests Second Edition". In-vitro mechanical and functional testing was also performed and demonstrated that the device's design and performance requirements were met. This functional testing also demonstrates and supports a determination of substantial equivalence to the predicate device.

#### H. CONCLUSIONS

This pre-market notification has demonstrated Substantial Equivalence as defined and understood in Sections 513(f)(1) and 513(i)(1) of the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### JAN 2 7 2005

GMP/Companies, Inc. c/o Mr. Daniel W. Lehtonen Intertek Testing Services 70 Codman Hill Road Boxborough, Massachusetts 01719

Re: K050027

Trade/Device Name: Laparocision Scope Controller System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: January 21, 2005 Received: January 24, 2005

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for asso mated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FIJA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You nust comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -- Mr. Daniel W. Lehtonen

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and with your finding of substantial equivalence of your device to a legally premated predicated device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you desire specific arrestiance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Muriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name:

GMP|Surgical Solutions Inc. Laparocision™ Scope Controller System

Indications for Use:

The GMP|Surgical Solutions Inc. Laparocision™ Scope Controller System is indicated for thoracic, gynecologic, orthopedic, and abdominal procedures for the purpose of holding and controlling the movement of standard laparoscopes or rigid endoscopes within inflated surgical cavities during minimally invasive laparoscopic surgery.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter-Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE JS NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative. and Neurological Devices

510(k) Number_405 0037

Posted 13 Nov 2003

Page 1 of T

---

**Source:** [https://fda-staging.innolitics.com/device/K050027](https://fda-staging.innolitics.com/device/K050027)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
