MICROSCAN DRIED GRAM-POSITIVE MIC/COMBO PANELS - ERTAPENEM

K043589 · Dade Microscan, Inc. · JWY · Feb 14, 2005 · Microbiology

Device Facts

Record IDK043589
Device NameMICROSCAN DRIED GRAM-POSITIVE MIC/COMBO PANELS - ERTAPENEM
ApplicantDade Microscan, Inc.
Product CodeJWY · Microbiology
Decision DateFeb 14, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive cocci. After inoculation, panels are incubated for 16-20 hours at 35℃ +/- 1℃ in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert. This particular submission is for the addition of the antimicrobial Ertapenem at concentrations of 0.002 to 32 mcg/ml to the test panel. The gram-positive organisms which may be used for Ertapenem susceptibility testing in this panel are: Staphylococcus aureus (methicillin-susceptible strains) Streptococcus agalactiae

Device Story

MicroScan® Dried Gram-Positive MIC/Combo Panels contain dehydrated Mueller Hinton Broth and nutrients with microdilutions of antimicrobial agents. Panels are rehydrated and inoculated with bacterial suspension (turbidity or Prompt® method); incubated 16-20 hours at 35°C in non-CO2 incubator. Growth is measured visually or automatically via WalkAway® or autoScan® 4 instruments. Instruments use optics systems and growth algorithms to determine minimum inhibitory concentration (MIC) by identifying lowest concentration inhibiting growth. Results assist clinicians in selecting appropriate antimicrobial therapy for gram-positive cocci infections.

Clinical Evidence

Bench testing only. External evaluation compared performance of Dried Gram-Positive Panel with Ertapenem against NCCLS frozen Reference Panel using fresh/stock Efficacy isolates and stock Challenge strains. Overall Essential Agreement >96%. Reproducibility and precision testing performed using Turbidity and Prompt inoculum methods on autoSCAN-48 and WalkAway® instruments. Quality Control testing met acceptance criteria.

Technological Characteristics

Microdilution broth susceptibility test miniaturized and dehydrated in panels. Contains Ertapenem (0.002 - 32 mcg/ml). Compatible with visual reading or automated instrumentation (autoSCAN-48, WalkAway®).

Indications for Use

Indicated for determining quantitative and/or qualitative antimicrobial susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive cocci. Includes addition of ertapenem (0.002-32 ug/mL). Contraindicated for methicillin-resistant staphylococci and Listeria monocytogenes.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K043589 B. Purpose for Submission: To add Ertapenem to the Dried Gram-Positive MIC/Combo Panels C. Measurand: Ertapenem at concentrations from 0.002-32 ug/mL D. Type of Test: Quantitative and Qualitative growth based detection E. Applicant: Dade Behring Dade MicroScan Inc. F. Proprietary and Established Names: MicroScan® Dried Gram-Positive MIC/Combo Panels G. Regulatory Information: 1. Regulation section: 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: LRG-Instrument for Auto Reader & Interpretation of Overnight Antimicrobial Susceptibility Systems JWY - Manual Antimicrobial Susceptibility Test Systems LTT – Panels, Test, Susceptibility, Antimicrobial LTW – Susceptibility Test Cards, Antimicrobial 4. Panel: {1} 83 Microbiology H. Intended Use: 1. Intended use(s): For use with MicroScan® Dried Gram Positive MIC/Combo, Dried Gram Positive Breakpoint Combo and Dried Gram Positive ID Type 2 panels. MicroScan® Positive panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultatively gram-positive cocci, some fastidious aerobic gram positive cocci and *Listeria monocytogenes*. Refer to Limitation of Procedure Section for use with fastidious streptococci. 2. Indication(s) for use: The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram positive cocci. This indication is for the addition of the antimicrobial ertapenem at concentrations of 0.002 to 32 ug/mL to the test panel. 3. Special conditions for use statement(s): The Prompt® method of inoculation is an alternate method of inoculation preparation that is supported in the methodology along with the turbidity method. The stationary and log inoculum methods should not be used with this antibiotic. Results should be reported for Methicillin susceptible *Staphylococcus spp.* and Beta hemolytic streptococcus. 4. Special instrument requirements: These panels can be read at ≥ 16 hours of incubation either manually, automatically on the autoScan® 4, or with the WalkAway® instrument systems. I. Device Description: The MicroScan® Dried Gram-Positive MIC/Combo Panel contains microdilutions of each antimicrobial agent in various concentrations with Mueller Hinton Broth and various nutrients which are dehydrated and dried in panels. Each panel contains two control wells: a no-growth control well (contains water only/no nutrients or broth), and a growth control well (contains test medium without antibiotic). The panel is rehydrated and inoculated at the same time with 0.1 ml of suspension prepared by the turbidity method (inoculum prepared in water, then 0.1 ml transferred to 25 ml of inoculum water containing pluronic-D/F-a wetting solution) for a final inoculum of 3-7 X 10⁵. The Prompt® method of inoculation is also recommended as an alternate {2} means of preparing the inoculum. The panels are incubated at 35° C in a non-CO₂ incubator for 16-20 hours and read by visual observation of growth. Panels may also be read automatically with the WalkAway® or the AutoScan®4. ## J. Substantial Equivalence Information: 1. Predicate device name(s): MicroScan Dried Gram-Positive and Gram-Negative MIC/Combo Panels 2. Predicate 510(k) number(s): k862140 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | See above | Same | | Inoculum preparation | Inoculum prepared from isolated colonies using either the Turbidity method or Prompt® system | Same | | Technology | Growth based after 16 hours incubation | Same | | Results | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | | | Instrument | autoScan® -4 or WalkAway® | Same | | Differences | | | | Item | Device | Predicate | | Antibiotic | Ertapenem at 0.002-32 ug/mL | Different concentrations depending on the antibiotic | | Test organism | Methicillin susceptible Staphylococcus aureus and Streptococcus agalactiae | Varies according to the antibiotic | | Limitations | Do not report methicillin resistant staphylococci and Listeria monocytogenes. | Varies according to the antibiotic | ## K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; CLSI M7 (M100-S15) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”. {3} L. Test Principle: After incubation in a non-CO₂ incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organisms are read by determining the lowest antimicrobial concentration showing inhibition of growth. The panels are read either manually using a touchScan® SR, or with the autoScan 4® or the WalkAway® instrument, which uses an optics systems with growth algorithms to directly measure organism growth. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Reproducibility was demonstrated using 10 isolates tested at 3 sites on 3 separate days in triplicate. All ten isolates had a mode that was on scale. The mode was determined by the method used and therefore is not always the same for each method. The study included the testing of the following inoculum and reading variables; turbidity inoculum method and Prompt® method of inoculation with reading performed manually using a touchScan SR, autoScan 4® or the WalkAway® instrument. The following table provides the overall results for all combinations of these variables | Difference in the number of dilutions between the mode of the MicroScan result and the actual result with each different variable | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Inoculation method | Read method | ≥ Minus 2 dilutions | Minus 1 dilution | Exact | Plus 1 dilution | ≥ Plus 2 dilutions | | Turbidity | Manual | | 16 | 230 | 30 | 1 | | Turbidity | WalkAway® | | 21 | 230 | 20 | 5 | | Turbidity | autoScan® 4 | | 19 | 221 | 36 | 1 | | Prompt® | Manual | 1 | 49 | 214 | 12 | 1 | | Prompt® | WalkAway® | 1 | 2 | 265 | 8 | 1 | | Prompt® | autoScan® 4 | 1 | 8 | 223 | 43 | 2 | The data demonstrates that there is very good reproducibility of each method but since the modes of each are used and they may not be the same, this does not demonstrate if there is a difference between methods. The actual data points and the modes did demonstrate that when there was a difference the Prompt® method of inoculation was more resistant if only by one well. This was more apparent in the Staphylococci isolates. These were the same isolates that were used in the colony count inoculum density study which did demonstrate a higher CFU for the Staphylococci which would explain the trend here. b. Linearity/assay reportable range: Not Applicable {4} c. Traceability, Stability, Expected values (controls, calibrators, or methods): Quality Control was performed daily with the turbidity method and with the Prompt® selectively with the following results. The expected ranges are listed in the table also and are also included in the final package insert. | Organism | Conc. In ug/mL | Reference result | Turbidity inoculation with Read methods | Prompt® inoculation with Read methods | | --- | --- | --- | --- | --- | | Manual | Walk-Away® | Auto-Scan® | Manual | Walk-Away® | Auto-Scan® | | S. aureus ATCC 29213 Expected range 0.06-0.25 ug/mL | 0.06 | 16 | 1 | | | | | | | 0.12 | 60 | 87 | 70 | 64 | 40 | 66 | 48 | | 0.25 | 5 | 14 | | 6 | 60 | 3 | 21 | | 0.5 | | | | | | | | | 1 | | | | | | | | | 2 | | | | | | | | | 4 | | 1 | 1 | 1 | | | | | 8 | 1 | | | | | | | | 16 | | | | | | | | | 32 | | 1 | | | | | | | >32 | | | | | | | | | E. faecalis ATCC 29212 Expected range 4-16 ug/mL | 1 | | | | | | | | | 2 | | | | | | | | | 4 | 6 | 22 | 36 | 33 | 2 | | 2 | | 8 | 108 | 82 | 34 | 38 | 33 | 24 | 22 | | 16 | | | | | | | | | 32 | | | | | | | | Quality control results demonstrated the ability of all variables of the procedure (reading and inoculation) to produce acceptable results. The difference in the staphylococci manual turbidity readings and the manual Prompt® reading results demonstrated the same effect observed in the reproducibility where there was a difference of one well more resistance for the Prompt® method of inoculation and even more pronounced with the Prompt® method of inoculation and the manual readings. This would be expected since the Prompt® method of inoculation often produces a higher CFU/ml in the final panel. Inoculum density control: A turbidity meter was used for the turbidity inoculation method with daily checks. The Prompt® method of inoculation had colony counts performed periodically throughout the study to determine the average inoculum density since there is no visual check of the inoculum using this device. The Prompt® method of inoculation had far more variability with average inoculum ranges from 1.1-1.8 $10^{6}$ with a actual data point {5} range of 3.7 x 10⁴ to 5.9 x 10⁶. The inoculum of the Prompt® method of inoculation generally provides a higher number of CFU with more variability than a method using a turbidity meter as demonstrated in this study. The average of the Staphylococci tested was outside the recommended range for the CLSI reference method. The user is referred to the limitation section for the recommendations of when to use an alternate method. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Clinical testing was performed at three sites using fresh isolates supplemented with stock isolates. A comparison of the MicroScan® Dried Gram-Positive test panel results was made to the reference method conducted as recommended in the CLSI standard M7-A6 with the following deviations from that recommendation; Pluronic-F is used as the inoculum in the frozen reference panels. This composed of water which contains a very small amount (0.1) of Pluronic P104 to provide a smoother draw of liquid into the inoculator. Testing of the reference method and the MicroScan panels was performed at the same time. A challenge set was also tested at one site and compared to the reference broth dilution result mode that was determined by previous testing of each isolate multiple times in the recommended reference panel. Only Staphylococcus spp. and beta hemolytic streptococci were considered in the evaluation. All isolates tested grew in the MicroScan panels. | | total | EA | %EA | Total evaluable | EA of evaluable | %EA | CA | %CA | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 307 | 294 | 95.8% | 283 | 271 | 95.8% | 295 | 96.1% | 36 | 8 | 3 | 1 | | Challenge | 75 | 75 | 100% | 75 | 75 | 100% | 75 | 100% | 0 | 0 | 0 | 0 | | Combined | 382 | 369 | 96.6 | 358 | 346 | 96.6% | 370 | 96.9% | 36 | 8 | 3 | 1 | EA-Essential Agreement CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies {6} 7 Evaluable results are those that fall within the test range of the reference method and could also be on-scale with the new device if within the plus/minus one well variability. EA is when there is agreement between the reference method and the MicroScan® within plus or minus one serial two-fold dilution of antibiotic. CA is when the interpretation of the reference method agrees exactly with the interpretation of the MicroScan® result. The challenge set of organisms was also tested using the Prompt® method of inoculation with all reading methods and the turbidity method of inoculation with the WalkAway® and the autoScan®4. This included seventy five challenge isolates that were tested at one site. The inoculum was prepared by the turbidity or Prompt® method and incubated in the WalkAway® instrument. All panels had additional readings performed after the WalkAway® reading was completed using the autoScan®-4 and then manually on the touchSCAN®-SR. The table below demonstrates the numbers that were in exact agreement with the reference method result and those that differed by one or more wells. | Difference in the number of dilutions between the expected reference result and the MicroScan® Result | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Inoculation method | Read method | ≤ minus 2 dilutions | minus 1 dilution | Exact | Plus 1 dilution | ≥ Plus 2 dilutions | % EA | | Turbidity | Manual | | 12 | 42 | 21 | | 100 | | Turbidity | WalkAway ® | | 12 | 50 | 13 | | 100 | | Turbidity | autoScan® 4 | | 11 | 50 | 14 | | 100 | | Prompt® | Manual | | 3 | 45 | 26 | 1 | 98.7 | | Prompt® | WalkAway ® | | 4 | 62 | 8 | 1 | 98.7 | | Prompt® | autoScan® 4 | | 4 | 58 | 13 | | 100 | Although all methods were ≥ 95% essential agreement, there is a slight suggestion that the turbidity method of inoculation has more results in the minus category. This trend to slightly more resistant results for the Prompt® method of inoculation is consistent with the reproducibility data and also the higher CFU/ml in the Prompt® inoculum. Category Agreement was 100% for all but this is meaningless since all results were well below the interpretive criteria. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: {7} Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: ≤ 2 (S), 4 (I), ≥ 8 (R) for Staphylococcus spp. (CLSI comment “For oxacillin resistant staphylococci, report as resistant or do not report”) ≤ 1 for Beta-Hemolytic Streptococcus (CLSI comment: “The current absence of data in resistant strains precludes defining any results other than Susceptible. Strains yielding MIC results suggestive of a “nonsusceptible” category”) The interpretative criteria and Quality Control Ranges are the same as recommended in the FDA approved pharmaceutical package insert and the CLSI. All values are included in the package insert. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...