← Product Code [IYN](/productcode/IYN) · K043559

# SONOSITE HIGH-RESOLUTION DIAGNOSTIC ULTRASOUND SYSTEM (C3 SERIES) (K043559)

_Sonosite,Inc. · IYN · Jan 12, 2005 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K043559

## Device Facts

- **Applicant:** Sonosite,Inc.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jan 12, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The intended uses of the SonoSite High-Resolution Ultrasound System (C3 Series) as defined by FDA guidance documents, are: Fetal - OB/GYN, Abdominal, Intra-operative (Abdominal organs and vascular), Intra-operative (Neuro.), Laparoscopic, Pediatric, Small Organ (breast, thyroid, testicles.), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-urethral, Musculo-skel. (Conventional), Musculo-skel. (Superficial), Cardiac Adult, Cardiac Pediatric, Trans-esophageal (card.), Peripheral vessel.

## Device Story

The SonoSite High-Resolution Ultrasound System (C3 Series) is a software-controlled diagnostic ultrasound system. It utilizes piezoelectric transducers to transmit ultrasound energy into the body and receive reflected signals. The system processes these signals to display real-time images in 2D, M-Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Power Doppler, and Velocity Color Doppler. It supports ECG input for timing M-mode and Doppler measurements. The device includes measurement tools for anatomical structures, fetal biometry, and carotid intima-media thickness (IMT). Used in clinical settings by healthcare professionals, the system provides visual and audio output for diagnostic assessment of pathology, cardiac performance, and hemodynamic status. It also provides ultrasound guidance for interventional procedures like biopsies, drainage, and vascular access. The system features DICOM connectivity for image storage and review, facilitating clinical decision-making and patient management.

## Clinical Evidence

Bench testing only. The device was evaluated for safety and effectiveness through compliance with international standards including EN 60601-1 (electrical safety), EN 60601-1-2 (EMC), and EN 60601-2-37 (ultrasonic safety). Acoustic output measurements were performed in accordance with AIUM/NEMA standards. No clinical data was required for this 510(k) submission.

## Technological Characteristics

System utilizes piezoelectric transducers (2.0-12.0 MHz) for ultrasound generation. Imaging modes: 2D, M-mode, PW/CW Doppler, Color Power Doppler, Velocity Color Doppler, and Tissue Harmonic Imaging. Connectivity: DICOM 3.0. Safety: Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-37. Biocompatible patient-contact materials. Software-controlled architecture.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SonoSite TITAN Ultrasound System ([K030949](/device/K030949.md))
- Philips Ultrasound HD1 5000 Ultrasound System ([K011224](/device/K011224.md))
- GE LOGIQ 9 Diagnostic Ultrasound System ([K011188](/device/K011188.md))
- Siemens OMNIA X/XS Diagnostic Ultrasound System ([K020353](/device/K020353.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K043559

Image /page/0/Picture/2 description: The image shows the logo for SonoSite. The logo consists of a stylized, curved shape above the word "SonoSite." The curved shape is black and appears to be a stylized wave or arc. The word "SonoSite" is in a bold, sans-serif font, with a small trademark symbol to the right of the word.

JAN 1 2 2005

SonoSite, Inc.
21919 30ᵗʰ Drive SE
Bothell, WA 98021-3904 USA

Telephone:1.425.951.1200

Facsimile:1.425.951.1201

www.sonosite.com

# 510(K) Summary of Safety and Effectiveness

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

#### 1) Submitter's name, address, telephone number, contact person:

SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904

| Corresponding Official: | Daina L. Graham                                          |
|-------------------------|----------------------------------------------------------|
|                         | Vice President, Regulatory Affairs and Quality Assurance |
| E-mail:                 | Daina.Graham@sonosite.com                                |
| Telephone:              | (425) 951-1275                                           |
| Facsimile:              | (425) 951-1201                                           |
| Date prepared:          | December 9, 2004                                         |

#### 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:

## Common/ Usual Name

Diagnostic Ultrasound System with Accessories

## Proprietary Name

SonoSite High-Resolution Ultrasound System (C3 Series) (subject to change)

Classification Names

| Name                                        | FR Number | Product Code |
|---------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System    | 892.1550  | 90-IYN       |
| Ultrasonic Pulsed Echo Imaging System       | 892.1560  | 90-IYO       |
| Diagnostic Ultrasound Transducer            | 892.1570  | 90-ITX       |
| Picture Archiving And Communications System | 892.2050  | 90-LLZ       |

#### Identification of the predicate or legally marketed device: 3)

SonoSite, Inc. believes that the System described in this Substantially equivalent to a combination of the SonoSite TITAN™ Ultrasound System (K030949), Philips Ultrasound HD1 5000 Ultrasound System (K011224), the GE LOGIQ 9 Diagnostic Ultrasound System (K011188) and the Siemens OMNIA X/XS Diagnostic Ultrasound System (K020353).

{1}------------------------------------------------

#### 4) Device Description:

The SonoSite High-Resolution Ultrasound System (C3 Series) is a full featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display highresolution, real-time ultrasound data in 2D, M-Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Power Doppler, and Velocity Color Doppler or in a combination of these modes.

The System has an electrocardiography (ECG) display feature and supports a 3-lead ECG cable assembly to collect data for M-mode and Doppler measurements. The System provides measurement capabilities for anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The System has a PW and CW Doppler audio output feature and cine review, image zoom, labeling, biopsy, measurements and calculations, image storage and review, printing, and recording capabilities.

The system includes the ability to measure the intima-media thickness (IMT) of the carotid artery using digital ultrasound images. The IMT measurement of the carotid artery may be used adjunctively with other medical data obtained by a physician to help assess the cardiovascular health of a patient.

The system includes Digital Imaging and Communications (DICOM) capabilities as well as general computer communication capabilities to provide the acceptance, transfer, display, storage, and digital processing of ultrasound images and loops. Security support is also provided to facilitate HIPAA compliance.

The SonoSite High-Resolution Ultrasound System (C3 Series) is designed to accept curved or linear transducers of the types and frequency listed in the table below. All actions affecting the performance of the transducer are activated from the main system control panel.

| Frequency Range:  | 2.0 - 12.0 MHz      |
|-------------------|---------------------|
| Transducer Types: | Linear array        |
|                   | Curved array        |
|                   | Intracavitary array |
|                   | Phased array        |
|                   | Static probes       |

The SonoSite High-Resolution Ultrasound System (C3 Series) is designed to comply with the standards listed below.

| Reference no.                        | Title                                                                                                                                                                             |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN 540                               | Clinical investigation of medical devices for human subjects (1993)                                                                                                               |
| EN 980 A1                            | Graphical symbols for use in the labeling of medical devices (2003)                                                                                                               |
| EN 1041                              | Information supplied by the manufacturer with medical devices (1998)                                                                                                              |
| EN ISO 13485                         | Medical devices - Quality management systems - Requirements for regulatory purposes (2003)                                                                                        |
| EN ISO 14971                         | Medical devices - Application of risk management to medical devices (2000) (ISO 14971:2000) (Superseded standard: EN 1441)                                                        |
| EN ISO 10993-1                       | Biological evaluation of medical devices - Part 1: Evaluation and testing (1997) (ISO 10993-1:1997)                                                                               |
| EN 60601-1                           | Medical electrical equipment. Part 1: General requirements for safety - IEC 601-1:1988                                                                                            |
| EN 60601-1                           | Medical electrical equipment. Part 1: General requirements for safety - IEC 601-1:1988                                                                                            |
| Reference no.                        | Title                                                                                                                                                                             |
| Amendment A1                         | 1:1988/A1:1991                                                                                                                                                                    |
| EN 60601-1                           | Medical electrical equipment. Part 1: General requirements for safety - IEC 601-                                                                                                  |
| Amendment A2                         | 1:1988/A2:1995 + corrigendum June 1995                                                                                                                                            |
| EN 60601-1-1                         | Medical electrical equipment. Part 1: General requirements for safety - 1. Collateral standard: Safety requirements for medical electrical systems (2001) (IEC 601-1-1:2001)      |
| EN 60601-1-2                         | Medical electrical equipment - Part 1: General requirements for safety; 2. collateral standard: electromagnetic compatibility; requirements and tests (2001) (IEC 60601-1-2:2001) |
| EN 60601-2-37                        | Medical electrical equipment - Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment (2001) (IEC 60601-2-37:2001)           |
|                                      | Medical Ultrasound Safety, American Institute of Ultrasound in Medicine (AIUM), 1994                                                                                              |
|                                      | Acoustic Output Measurement and Labeling Standard for Diagnostic Ultrasound Equipment, American Institute of Ultrasound in Medicine, 1993                                         |
|                                      | Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, American Institute of Ultrasound in Medicine, 1998           |
| ANSI/AAMI<br>EC53:1995               | ECG Cables and Electrodes except for sections 4.4 and 4.5.9                                                                                                                       |
| ASTM D5276-98                        | Standard Test Methods for Drop Test of Loaded Containers by Free Fall                                                                                                             |
| ASTM D999-96                         | Standard Methods for Vibration Testing of Shipping Containers                                                                                                                     |
| CEI/IEC 61157:1992                   | International Electrotechnical Commission, Requirements for the Declaration of the Acoustic Output of Medical Diagnostic Ultrasonic Equipment                                     |
| D00017                               | Quality System Manual (SonoSite)                                                                                                                                                  |
| EN 60529                             | Degrees of protection provided by enclosures (IP Code) (1991                                                                                                                      |
| ISO 9001: 1994                       | Quality Systems - Model for quality assurance in design, development, production, installation and servicing                                                                      |
| JIS T 0601-1                         | Medical Electrical Equipment - Part 1: General Requirements for Safety (Japan) (1999)                                                                                             |
| JIS T 1507                           | Electronic Linear Scanning Ultrasonic Diagnostic Equipment (Japan) (1989)                                                                                                         |
| NEMA UD 2-2004                       | Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment                                                                                                          |
| NEMA UD 3-2004                       | Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, American Institute of Ultrasound in Medicine                 |
| RTCA/DO160D:1997                     | Radio Technical Commission for Aeronautics, Environmental Conditions and Test Procedures for Airborne Equipment, Section 21.0 Emission of Radio Frequency Energy, Category B      |
| Title 21 CFR Part 820                | Quality System Regulation - Medical Devices: Current Good Manufacturing Practice (CGMP); Final Rule                                                                               |
| UL 60601-1                           | Underwriters Laboratories, Medical Electrical Equipment-Part 1: General Requirements for Safety (2003)                                                                            |
| ACR-NEMA DICOM<br>version 3.0 - 2003 | Digital Imaging and Communications in Medicine                                                                                                                                    |

510(K) Summary Of Safety And Effectiveness

Page 2 of 6

{2}------------------------------------------------

{3}------------------------------------------------

#### 5) Intended Use:

The intended uses of the SonoSite High-Resolution Ultrasound System (C3 Series) as defined by FDA guidance documents, are:

| Fetal - OB/GYN                               | Trans-rectal                 |
|----------------------------------------------|------------------------------|
| Abdominal                                    | Trans-vaginal                |
| Intra-operative                              |                              |
| (Abdominal organs and vascular)              | Trans-urethral               |
| Intra-operative (Neuro.)                     | Musculo-skel. (Conventional) |
| Laparoscopic                                 | Musculo-skel. (Superficial)  |
| Pediatric                                    | Cardiac Adult                |
| Small Organ<br>(breast, thyroid, testicles.) | Cardiac Pediatric            |
| Neonatal Cephalic                            | Trans-esophageal (card.)     |
| Adult Cephalic                               | Peripheral vessel            |

Typical examinations performed using the SonoSite High-Resolution Ultrasound System (C3 Series) are:

## Abdominal Imaging Applications

This system transmits ultrasound energy into the abdomen of patients using 2D, M Mode, color Doppler (Color), color power Doppler (CPD), Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images. The liver, kidneys, pancreas, spleen, gallbladder, bile ducts, transplanted organs, abdominal vessels, and surrounding anatomical structures can be assessed for the presence or absence of pathology transabdominally.

## Cardiac Imaging Applications

This system transmits ultrasound energy into the thorax of patients using 2D, M Mode, color Doppler (Color), Tissue Harmonic Imaging (THI), pulsed wave (PW) Doppler, and continuous wave (CW) Doppler to obtain ultrasound images. The heart muscle, cardiac valves, great vessels, and surrounding anatomical structures can be assessed for, overall cardiac performance, size, fluid, and the presence or absence of pathology. Quanitification of size, function, and hymodynamic performance can also be assessed. The heart can be imaged transthoracicly or transesophageal imaging can be used to monitor the heart during surgical procedures such as cardiac bypass or valve replacements.

The patient's electrocardiogram (ECG) may be obtained and is used for accurate timing of diastolic and systolic function.

Warning: The ECG is not used to diagnose cardiac arrhythmias and is not designed for long term cardiac rhythm monitoring.

## Gynecology and Infertility Imaging Applications

This system transmits ultrasound energy in the pelvis and lower abdomen using 2D, M Mode, color power Doppler (CPD), color Doppler (Color)Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images. The uterus, ovaries, adnexa, and surrounding anatomical structures can be assessed for the presence or absence of pathology transabdominally or transvaginally.

{4}------------------------------------------------

## Interventional and Intraoperative Imaging Applications

This system transmits ultrasound energy into the various parts of the body using 2D, M-Mode, color Doppler (Color), color power Doppler (CPD), Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images that provide quidance during interventional and intraoperative procedures. This system can be used to provide ultrasound quidance for biopsy and drainage procedures, vascular line placement, peripheral nerve blocks, ova harvesting, amniocentesis and other obstetrical procedures, and provide assistance during abdominal, vascular, and neurological intraoperative procedures.

- Warning: This system is not intended for use in providing guidance for central nerve blocks, i.e., the brain and spinal cord, or for ophthalmic applications.
## Obstetrical Imaging Applications

This system transmits ultrasound energy into the pelvis of pregnant women using 2D, M Mode, color Doppler (Color), color power Doppler (CPD), Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images. The fetal anatomy, viability, estimated fetal weight, gestational age, amniotic fluid, and surrounding anatomical structures can be assessed for the presence or absence of pathology transabdominally or transvaginally. CPD and color Doppler (Color) imaging is intended for high-risk pregnant women. High-risk pregnancy indications include, but are not limited to, multible pregnancy, fetal hydrops, placental abnormalities, as well as maternal hypertension, diabetes, and lupus,

- Warning: To prevent injury or misdiagnosis do not use this system for Percutaneous Umbilical Blood Sampling (PUBS) or in vitro Fertilization (IVF) The system has not been validated to be proven effective for these two uses. CPD, or Color images can be used as an adjunctive method, not as a screening tool, for the detection of structural anomalies of the fetal heart and as an adjunctive method, not as a screening tool for the diagnosis of Intrauterine Growth Retardation (IUGR).
## Pediatric and Neonatal Imaging Applications

This system transmits ultrasound energy into the pediatric or neonatal patients using 2D, M Mode, color Doppler (Color), color power Doppler (CPD), pulsed wave (PW) Doppler, and continuous wave (CW) Doppler to obtain ultrasound images. The pediatric abdominal, pelvic and cardiac anatomy, pediatric hips, neonatal heads, and surrounding anatomical structures can be assessed for the presence or absence of pathology.

## Prostate Imaging Applications

This system transmits ultrasound energy into the prostate of an adult male using 2D, M Mode, color power Doppler (CPD), and pulsed wave (PW) Doppler to obtain ultrasound images. The prostate gland can be assessed for the presence or absence of pathology.

{5}------------------------------------------------

## Superficial Imaging Applications

This system transmits ultrasound energy into various parts of the body using 2D, M Mode, color Doppler (Color), color power Doppler (CPD), and pulsed wave (PW) Doppler to obtain ultrasound images. The breast, thyroid, testicle, lymph nodes, hernias, musculoskeletal structures, soft tissue structures, and surrounding anatomical structures can be assessed for the presence or absence of pathology. This system can be used to provide ultrasound guidance for biopsy and drainage procedures, vascular line placement, and peripheral nerve blocks.

- Warning: This system is not intended for use in providing guidance for central nerve blocks, i.e., the brain and spinal cord, or for ophthalmic applications.
## Vascular Imaging Applications

This system transmits ultrasound energy into the various parts of the body using 2D, M Mode, color Doppler (Color), color power Doppler (CPD), and pulsed wave (PW) Doppler to obtain ultrasound images. The carotid arteries, deep veins in the arms and legs, superficial veins in the arms and legs, great vessels in the abdomen, and various small vessels feeding organs can be assessed for the presence or absence of pathology.

#### 6) Technological Characteristics:

This device operates identically to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as 2D or M-mode images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis. The modes of this device (2D, PW Dopler, CW Doppler, velocity color Doppler, Color Power Doppler, and duplex imaging) are the same as a combination of the predicate devices identified in item 3. Transducer patient contact materials are biocompatible.

This device conforms to the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (AIUM/NEMA UD 3-2004) for an on-screen display feature that provides information on potential thermal and cavitation bioeffect mechanisms. A user education program provides additional information so users may moderate the system's acoustic output in accordance with the ALARA (as low as reasonably achievable) principle.

The device's acoustic output limits are:

All applications:

| ISPTA (d)             | 720 mW/cm² (Maximum)  |
|-----------------------|-----------------------|
| Tls/Tib/Tlc           | 0.1 - 4.0 (Range)     |
| Mechanical Index (MI) | 1.9 (Maximum)         |
| ISPPA (d)             | 0 - 700 W/cm² (Range) |

The limits are the same as predicate Track 3 devices.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized emblem consisting of three abstract shapes resembling human profiles or faces, stacked one above the other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 2 2005

SonoSite, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K043559

Trade Name: SonoSite High-Resolution Diagnostic Ultrasound System (C3 Series) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: December 20, 2004 Received: December 27, 2004

## Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the SonoSite High-Resolution Diagnostic Ultrasound System (C3 Series), as described in your premarket notification:

## Transducer Model Number

2MHZ Doppler Pencil C8/8-5

{7}------------------------------------------------

C11/8-5
C15/4-2
C60/5-2
C60e/5-2
HFL38/12-5
HST/10-5
HST/12-7
HST/12-8
ICT/8-5
LAP/10-5
L25/10-5
L38/10-5
L38e/10-5
L52/10-5
L52-S/10-5
MPT/7-4
P/4-2
P/7-4

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

{8}------------------------------------------------

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouticate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you attire of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronal and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

David A. Dorgenson

Image /page/8/Picture/7 description: The image shows a handwritten symbol or abbreviation that appears to be "fxr". The writing is in black ink and is slightly slanted. The background is plain white.

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{9}------------------------------------------------

| System:                   | SonoSite High-Resolution Ultrasound System (C3 Series)                                |                   |   |     |     |                  |                            |                  |
|---------------------------|---------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|------------------|
| Transducer:               | N/A                                                                                   |                   |   |     |     |                  |                            |                  |
| Intended Use:             | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:    |                   |   |     |     |                  |                            |                  |
| Clinical Application      |                                                                                       | Mode of Operation |   |     |     |                  |                            |                  |
| General<br>(Track I only) | Specific<br>(Tracks I & III)                                                          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.)        | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                                                                            | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Fetal                                                                                 | N                 | N | N   | N   | N                | B+M; B+PWD;<br>B+CWD; B+CD | Note 1           |
|                           | Abdominal<br>Intra-operative<br>(Abdominal organs and<br>vascular)                    | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Intra-operative (Neuro.)                                                              | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
| Fetal Imaging<br>& Other  | Laparoscopic                                                                          | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Pediatric                                                                             | N                 | N | N   | N   | N                | B+M; B+PWD;<br>B+CWD; B+CD | Note 1           |
|                           | Small Organ (breast,<br>thyroid, testicles)                                           | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Neonatal Cephalic                                                                     | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Adult Cephalic                                                                        | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Trans-rectal                                                                          | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)                           | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Musculo-skel. (Convent.)                                                              | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
|                           | Musculo-skel. (Superfic.)<br>Intra-luminal<br>Other (spec.)                           | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
| Cardiac                   | Cardiac Adult                                                                         | N                 | N | N   | N   | N                | B+M; B+PWD;<br>B+CWD; B+CD | Note 1           |
|                           | Cardiac Pediatric                                                                     | N                 | N | N   | N   | N                | B+M; B+PWD;<br>B+CWD; B+CD | Note 1           |
|                           | Trans-esophageal (card.)<br>Other (spec.)                                             | N                 | N | N   | N   | N                | B+M; B+PWD;<br>B+CWD; B+CD | Note 1           |
| Peripheral                | Peripheral vessel<br>Other (spec.)                                                    | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD        | Note 1           |
| System:                   | SonoSite High-Resolution Ultrasound System (C3 Series)                                |                   |   |     |     |                  |                            |                  |
| Transducer:               | Doppler Pencil Transducer 2.0 MHz Dual Element Circular Array                         |                   |   |     |     |                  |                            |                  |
| Intended Use:             | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: |                   |   |     |     |                  |                            |                  |
| Clinical Application      |                                                                                       | Mode of Operation |   |     |     |                  |                            |                  |
| General<br>(Track I only) | Specific<br>(Tracks I & III)                                                          | B                 | M | PWD | CWD | Color Doppler    | Combined<br>(Spec.)        | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                                                                            |                   |   |     |     |                  |                            |                  |
| Fetal Imaging<br>& Other  | Fetal                                                                                 |                   |   |     |     |                  |                            |                  |
|                           | Abdominal                                                                             |                   |   |     |     |                  |                            |                  |
|                           | Intra-operative<br>(Abdominal organs and<br>vascular)                                 |                   |   |     |     |                  |                            |                  |
|                           | Intra-operative (Neuro.)                                                              |                   |   |     |     |                  |                            |                  |
|                           | Laparoscopic                                                                          |                   |   |     |     |                  |                            |                  |
|                           | Pediatric                                                                             |                   |   |     |     |                  |                            |                  |
|                           | Small Organ (breast,<br>thyroid, testicles.)                                          |                   |   |     |     |                  |                            |                  |
|                           | Neonatal Cephalic                                                                     |                   |   |     |     |                  |                            |                  |
|                           | Adult Cephalic                                                                        |                   |   |     |     |                  |                            |                  |
|                           | Trans-rectal                                                                          |                   |   |     |     |                  |                            |                  |
|                           | Trans-vaginal                                                                         |                   |   |     |     |                  |                            |                  |
|                           | Trans-urethral                                                                        |                   |   |     |     |                  |                            |                  |
|                           | Trans-esoph. (non-<br>Card.)                                                          |                   |   |     |     |                  |                            |                  |
|                           | Musculo-skel.<br>(Convent.)                                                           |                   |   |     |     |                  |                            |                  |
|                           | Musculo-skel.<br>(Superfic.)                                                          |                   |   |     |     |                  |                            |                  |
|                           | Intra-luminal                                                                         |                   |   |     |     |                  |                            |                  |
|                           | Other (spec.)                                                                         |                   |   |     |     |                  |                            |                  |
| Cardiac                   | Cardiac Adult                                                                         |                   |   |     | N   |                  |                            |                  |
|                           | Cardiac Pediatric                                                                     |                   |   |     | N   |                  |                            |                  |
|                           | Trans-esophageal<br>(card.)                                                           |                   |   |     |     |                  |                            |                  |
|                           | Other (spec.)                                                                         |                   |   |     |     |                  |                            |                  |
| Peripheral<br>Vessel      | Peripheral vessel                                                                     |                   |   |     |     |                  |                            |                  |
|                           | Other (spec.)                                                                         |                   |   |     |     |                  |                            |                  |

Table 4.3- 1 Diagnostic Ultrasound Indications for Use Form – C3 System

## Additional Comments:

Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, 3-D Imaging, Tissue Hammic maging, Tissue Noter I. Only maddes Ooler Forer Bopper, Solor Doppler includes Velocity Color Doppler. Color Doppler can be Dopper imaging and inisting or guladino of assist in the placement of needles and catheters in vascular on other analonical combined will ally maging mode. This placement of nesser and earlines of previously dearced in K030949. An expanded intended use for imaging guidance for peripheral nerve blook procedures was previously deared in K033367.

Daniel A. Serman

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043559

{10}------------------------------------------------

Table 4.3- 2 Diagnostic Ultrasound Indications for Use Form - 2 MHz Doppler Pencil

Emil A. Bergman

(Division Sign-Off) Division of Reproductive, Abdomi and Radiological Devices 510(k) Number

{11}------------------------------------------------

| SonoSite High-Resolution Ultrasound System (C3 Series) |                                                                                       |                      |   |     |     |                   |                               |                  |
|--------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------|---|-----|-----|-------------------|-------------------------------|------------------|
| System:                                                | SonoSite High-Resolution Ultrasound System (C3 Series)                                |                      |   |     |     |                   |                               |                  |
| Transducer:                                            | C8/8-5 8.0-5.0 MHz Curved Array Transducer                                            |                      |   |     |     |                   |                               |                  |
| Intended Use:                                          | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: |                      |   |     |     |                   |                               |                  |
| Clinical Application                                   |                                                                                       | Mode of Operation    |   |     |     |                   |                               |                  |
| General<br>(Track I only)                              | Specific<br>(Tracks I & III)                                                          | B                    | M | PWD | CWD | Color<br>Doppler  | Combined<br>(Spec.)           | Other<br>(Spec.) |
| Ophthalmic                                             | Ophthalmic                                                                            |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       | N                    | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
| Fetal Imaging<br>& Other                               | Fetal                                                                                 |                      |   |     |     |                   |                               |                  |
|                                                        | Abdominal                                                                             |                      |   |     |     |                   |                               |                  |
|                                                        | Intra-operative<br>(Abdominal organs and<br>vascular)                                 |                      |   |     |     |                   |                               |                  |
|                                                        | Intra-operative (Neuro.)                                                              |                      |   |     |     |                   |                               |                  |
|                                                        | Laparoscopic                                                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Pediatric                                                                             |                      |   |     |     |                   |                               |                  |
|                                                        | Small Organ (breast,<br>thyroid, testicles.)                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Neonatal Cephalic                                                                     |                      |   |     |     |                   |                               |                  |
|                                                        | Adult Cephalic                                                                        |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-rectal                                                                          | N                    | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                        | Trans-vaginal                                                                         | N                    | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                        | Trans-urethral                                                                        |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-esoph. (non-<br>Card.)                                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Musculo-skel.<br>(Convent.)                                                           |                      |   |     |     |                   |                               |                  |
|                                                        | Musculo-skel.<br>(Superfic.)                                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Intra-luminal                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        | Other (spec.)                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        | Cardiac Adult                                                                         |                      |   |     |     |                   |                               |                  |
| Cardiac                                                | Cardiac Pediatric                                                                     |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-esophageal<br>(card.)                                                           |                      |   |     |     |                   |                               |                  |
|                                                        | Other (spec.)                                                                         |                      |   |     |     |                   |                               |                  |
| Peripheral<br>Vessel                                   | Peripheral vessel<br>Other (spec.)                                                    |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       |                      |   |     |     |                   |                               |                  |
| System:                                                | SonoSite High-Resolution Ultrasound System (C3 Series)                                |                      |   |     |     |                   |                               |                  |
| Transducer:                                            | C11/8-5 8.0-5.0 MHz Curved Array                                                      |                      |   |     |     |                   |                               |                  |
| Intended Use:                                          | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       | Clinical Application |   |     |     | Mode of Operation |                               |                  |
| General<br>(Track I only)                              | Specific<br>(Tracks I & III)                                                          | B                    | M | PWD | CWD | Color<br>Doppler  | Combined<br>(Spec.)           | Other<br>(Spec.) |
| Ophthalmic                                             | Ophthalmic                                                                            |                      |   |     |     |                   |                               |                  |
|                                                        | Fetal                                                                                 |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       |                      |   |     |     |                   | B+M; B+PWD;<br>B+CWD;<br>B+CD |                  |
|                                                        | Abdominal                                                                             | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1           |
|                                                        | Intra-operative<br>(Abdominal organs and<br>vascular)                                 | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                        | Intra-operative (Neuro.)                                                              | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
| Fetal Imaging                                          | Laparoscopic                                                                          |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       |                      |   |     |     |                   | B+M; B+PWD;<br>B+CWD;<br>B+CD |                  |
| & Other                                                | Pediatric<br>Small Organ (breast,<br>thyroid, testicles)                              | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1           |
|                                                        | Neonatal Cephalic                                                                     | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                        | Adult Cephalic                                                                        |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-rectal                                                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-vaginal                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-urethral                                                                        |                      |   |     |     |                   |                               |                  |
|                                                        | Trans-esoph. (non-<br>Card.)                                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Musculo-skel.<br>(Convent.)                                                           |                      |   |     |     |                   |                               |                  |
|                                                        | Musculo-skel.<br>(Superfic.)                                                          |                      |   |     |     |                   |                               |                  |
|                                                        | Intra-luminal                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        | Other (spec.)                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        | Cardiac Adult                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       |                      |   |     |     |                   | B+M; B+PWD;<br>B+CWD;<br>B+CD |                  |
| Cardiac                                                | Cardiac Pediatric<br>Trans-esophageal<br>(card.)                                      | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1           |
|                                                        | Other (spec.)                                                                         |                      |   |     |     |                   |                               |                  |
|                                                        |                                                                                       |                      |   |     |     |                   | B+M; B+PWD;<br>B+CD           |                  |
| Peripheral                                             | Peripheral vessel                                                                     | N                    | N | N   | N   | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                        | Other (spec.)                                                                         |                      |   |     |     |                   |                               |                  |

Table 4.3- 3 Diagnostic Ultrasound Indications for Use Form – C8/8-5 Transducer

# Additional Comments:

Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, 3-D Imaging, Tissue Harmonic maging, Tissue Nole 1. Online Includes Color Fore. Bopper, continue a and Gelority Color Doppler. Color Doppler and be Dopper inlaging and inlaging for goldance on blogs). User event of needles and catheters in vascular or other anatomical structures was cleared in K030949.

Daniel A. Snyderman

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043559

{12}------------------------------------------------

Table 4.3- 4 Diagnostic Ultrasound Indications for Use Form - C11/8-5 Transducer

### Additional Comments:

Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging, Tissue Note 1. Only Includes Color Power Dopper, Color Pollor Dolor Pollor Doppler. Color Doppler an be Dopper maging and inloging or gologics or blogs; Uter Popply Marcelles in vasular or there analonical para ha combined will any maging moot. "India in a souse. An expanded intended use for imaging guidance for peripheral nerve blook procedures was previously cleared in K033367

Damil A. Olymm
(Division Sign-Off)

Prescription Use (Per 21 CFR 801.109)

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043559

Indications for Use

**7 of 905 pages**

{13}------------------------------------------------

| SonoSite High-Resolution Ultrasound System (C3 Series)<br>System:                  |                                                                                 |  |                  |   |     |     |                   |                               |                  |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--|------------------|---|-----|-----|-------------------|-------------------------------|------------------|
| C15/4-2 4.0-2.0 MHz Curved Array<br>Transducer:                                    |                                                                                 |  |                  |   |     |     |                   |                               |                  |
| Diagnostic ultrasound imaging or fluid flow analysis of the human<br>Intended Use: |                                                                                 |  |                  |   |     |     |                   |                               |                  |
|                                                                                    |                                                                                 |  | body as follows: |   |     |     |                   |                               |                  |
| Clinical Application                                                               |                                                                                 |  |                  |   |     |     | Mode of Operation |                               |                  |
| General<br>(Track I only)                                                          | Specific<br>(Tracks I & III)                                                    |  | B                | M | PWD | CWD | Color<br>Doppler  | Combined<br>(Spec.)           | Other<br>(Spec.) |
| Ophthalmic                                                                         | Ophthalmic                                                                      |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                                                    | Fetal                                                                           |  | N                | N | N   |     |                   | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1           |
|                                                                                    | Abdominal<br>Intra-operative<br>(Abdominal organs and<br>vascular)              |  | N                | N | N   | N   | N                 | B+CD                          | Note 1           |
|                                                                                    | Intra-operative (Neuro.)<br>Laparoscopic                                        |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
| Fetal Imaging<br>& Other                                                           | Pediatric<br>Small Organ (breast,<br>thyroid, testicles.)                       |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                                                    | Neonatal Cephalic                                                               |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                                                    | Adult Cephalic                                                                  |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                                                    | Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-<br>Card.) |  |                  |   |     |     |                   |                               |                  |
|                                                                                    | Musculo-skel.<br>(Convent.)                                                     |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |
|                                                                                    | Musculo-skel.<br>(Superfic.)<br>Intra-luminal<br>Other (spec.)                  |  |                  |   |     |     |                   |                               |                  |
|                                                                                    | Cardiac Adult                                                                   |  | N                | N | N   | N   | N                 | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1           |
| Cardiac                                                                            | Cardiac Pediatric<br>Trans-esophageal<br>(card.)<br>Other (spec.)               |  | N                | N | N   | N   | N                 | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1           |
| Peripheral<br>Vessel                                                               | Peripheral vessel<br>Other (spec.)                                              |  | N                | N | N   |     | N                 | B+M; B+PWD;<br>B+CD           | Note 1           |

Table 4.3- 5 Diagnostic Ultrasound Indications for Use Form - C15/4-2 Transducer

N= new indication; P= previously cleared by FDA; E= added under Appendix E

# Additional Comments:

A Qurt Includes Color Power Dopler, combined B and Color Power Doppler, 3-D maging, Tissue Hamonic masure, Tissue Note : Other includes Color Power Dopler, combined Band Color Power Chamble hisale hasing.
Dopler maging mor minad B and Color Dopler induses Versity Cher Dopler. Color Dopl combined with any imaging modelin 510(k) K030949.

David A. Lynn

Prescription Use (Per 21 CFR 801.109)

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) gmbe6f-905 pages
K043559 Section 4.3
20

Indications for Use

{14}------------------------------------------------

| SonoSite High-Resolution Ultrasound System (C3 Series)            |                                                       |                   |   |     |     |               |                     |               |
|-------------------------------------------------------------------|-------------------------------------------------------|-------------------|---|-----|-----|---------------|---------------------|---------------|
| System:                                                           |                                                       |                   |   |     |     |               |                     |               |
| C60/5-2 5.0-2.0 MHz Curved Array                                  |                                                       |                   |   |     |     |               |                     |               |
| Transducer:                                                       |                                                       |                   |   |     |     |               |                     |               |
| Diagnostic ultrasound imaging or fluid flow analysis of the human |                                                       |                   |   |     |     |               |                     |               |
| Intended Use:                                                     |                                                       |                   |   |     |     |               |                     |               |
|                                                                   | body as follows:                                      |                   |   |     |     |               |                     |               |
|                                                                   |                                                       | Mode of Operation |   |     |     |               |                     |               |
|                                                                   | Clinical Application                                  | B                 | M | PWD | CWD | Color Doppler | Combined (Spec.)    | Other (Spec.) |
| General<br>(Track I only)                                         | Specific<br>(Tracks I & III)                          |                   |   |     |     |               |                     |               |
| Ophthalmic                                                        | Ophthalmic                                            |                   |   |     |     |               | B+M; B+PWD;<br>B+CD |               |
|                                                                   | Fetal                                                 | N                 | N | N   |     | N             | B+M; B+PWD;<br>B+CD | Note 1        |
|                                                                   | Abdominal                                             | N                 | N | N   |     | N             | B+M; B+PWD;<br>B+CD | Note 1        |
|                                                                   | Intra-operative<br>(Abdominal organs and<br>vascular) | N                 | N | N   |     | N             | B+M; B+PWD;<br>B+CD | Note 1        |
|                                                                   | Intra-operative (Neuro.)                              |                   |   |     |     |               |                     |               |
| Fetal Imaging<br>& Other                                          | Laparoscopic                                          |                   |   |     |     |               | B+M; B+PWD;<br>B+CD | Note 1        |
|                                                                   | Pediatric                                             | N                 | N | N   |     | N             | B+M; B+PWD;<br>B+CD | Note 1        |
|                                                                   | Small Organ (breast,<br>thyroid, testicles.)          |                   |   |     |     |               |                     |               |
|                                                                   | Neonatal Cephalic                                     |                   |   |     |     |               |                     |               |
|                                                                   | Adult Cephalic                                        |                   |   |     |     |               |                     |               |
|                                                                   | Trans-rectal                                          |                   |   |     |     |               |                     |               |
|                                                                   | Trans-vaginal                                         |                   |   |     |     |               |                     |               |
|                                                                   | Trans-urethral                                        |                   |   |     |     |               |                     |               |
|                                                                   | Trans-esoph. (non-<br>Card.)                          |                   |   |     |     |               |                     |               |
|                                                                   | Musculo-skel.<br>(Convent.)                           | N                 | N | N   |     | N             | B+M; B+PWD;<br>B+CD | Note 1        |
|                                                                   | Musculo-skel.<br>(Superfic.)                          |                   |   |     |     |               |                     |               |
|                                                                   | Intra-luminal                                         |                   |   |     |     |               |                     |               |
|                                                                   | Other (spec.)                                         |                   |   |     |     |               |                     |               |
|                                                                   | Cardiac Adult                                         |                   |   |     |     |               |                     |               |
| Cardiac                                                           | Cardiac Pediatric                                     |                   |   |     |     |               |                     |               |
|                                                                   | Trans-esophageal<br>(card.)<br>Other (spec.)          |                   |   |     |     |               |                     |               |
|                                                                   | Peripheral vessel                                     | N                 | N | N   |     | N             | B+M; B+PWD;<br>B+CD | Note 1        |
| Peripheral<br>Vessel                                              | Other (spec.)                                         |                   |   |     |     |               |                     |               |

Table 4.3- 6 Diagnostic Ultrasound Indications for Use Form - C60/5-2 Transducer

## Additional Comments:

A COITI (Offical Color Power Doppler, combined B and Color Power Doppler, 3-D maging, Tissue
Note 1: Other includes Color Power Linery, Color Donner includes Velocity Color D Note 1: Other includes Color Power Doppler, onmined Band Color Color Polor many of Color Dopler Cag he register and e Doppler imaging and imaging for guidanes of biopsy. Color Dopper noder Dobble Ook Bopper Ook Bopper Ook Bopper Ooke Bopper 
 combined with any imaging more in 156045 comotics was previously cleared in 510(K) K030949.

Erich Lyrum

(Division Sign-Off)

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number KD43559

{15}------------------------------------------------

| Table 4.3- 7 Diagnostic Ultrasound Indications for Use Form - C60e/5-2 |
|------------------------------------------------------------------------|
| Transducer                                                             |

| System:                   | SonoSite High-Resolution Ultrasound System (C3 Series)                                |                   |   |     |     |                  |                     |                  |
|---------------------------|---------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer:               | C60e/5-2 5.0-2.0 MHz Curved Array                                                     |                   |   |     |     |                  |                     |                  |
| Intended Use:             | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: |                   |   |     |     |                  |                     |                  |
| Clinical Application      |                                                                                       | Mode of Operation |   |     |     |                  |                     |                  |
| General<br>(Track I only) | Specific<br>(Tracks 1 & III)                                                          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                                                                            |                   |   |     |     |                  |                     |                  |
|                           | Fetal                                                                                 | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD | Note 1           |
|                           | Abdominal                                                                             | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD | Note 1           |
|                           | Intra-operative<br>(Abdominal organs and<br>vascular)                                 | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD | Note 1           |
|                           | Intra-operative (Neuro.)                                                              |                   |   |     |     |                  |                     |                  |
| Fetal Imaging             | Laparoscopic                                                                          |                   |   |     |     |                  |                     |                  |
| & Other                   | Pediatric                                                                             | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD | Note 1           |
|                           | Small Organ (breast,<br>thyroid, testicles.)                                          |                   |   |     |     |                  |                     |                  |
|                           | Neonatal Cephalic                                                                     |                   |   |     |     |                  |                     |                  |
|                           | Adult Cephalic                                                                        |                   |   |     |     |                  |                     |                  |
|                           | Trans-rectal                                                                          |                   |   |     |     |                  |                     |                  |
|                           | Trans-vaginal                                                                         |                   |   |     |     |                  |                     |                  |
|                           | Trans-urethral                                                                        |                   |   |     |     |                  |                     |                  |
|                           | Trans-esoph. (non-<br>Card.)                                                          |                   |   |     |     |                  |                     |                  |
|                           | Musculo-skel.<br>(Convent.)                                                           | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD | Note 1           |
|                           | Musculo-skel.<br>(Superfic.)                                                          |                   |   |     |     |                  |                     |                  |
|                           | Intra-luminal                                                                         |                   |   |     |     |                  |                     |                  |
|                           | Other (spec.)                                                                         |                   |   |     |     |                  |                     |                  |
|                           | Cardiac Adult                                                                         |                   |   |     |     |                  |                     |                  |
| Cardiac                   | Cardiac Pediatric                                                                     |                   |   |     |     |                  |                     |                  |
|                           | Trans-esophageal<br>(card.)                                                           |                   |   |     |     |                  |                     |                  |
|                           | Other (spec.)                                                                         |                   |   |     |     |                  |                     |                  |
| Peripheral<br>Vessel      | Peripheral vessel<br>Other (spec.)                                                    | N                 | N | N   |     | N                | B+M; B+PWD;<br>B+CD | Note 1           |

# Additional Comments:

Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, 3-D maging, Tissue Hammonic maging, Tissue Noter . Other includes Color Fover Dopper, Gard Ookly Color Power Doppler. Color Doppler Color Doppler Color Doppler and be Dopper maging and inlaging to guidative of blogs). Other Doppics from Solor Sopper Ostice Oppion Boller andomical structures was previously cleared in 510(k) K030949.

Dymil A. Lynn

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043559

{16}------------------------------------------------

| SonoSite High-Resolution Ultrasound System (C3 Series)<br>HFL38/12-5 12.0-5.0 MHz Linear Array Transducer |                                                                    |                   |   |     |     |                  |                     |                  |
|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-----------…

---

**Source:** [https://fda-staging.innolitics.com/device/K043559](https://fda-staging.innolitics.com/device/K043559)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com