Xinfeng Heating lamp, including CQ-27, CQ-36 and CQ-55A, is an infrared lamp that emits the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, to temporarily increase local blood circulation, and to temporarily relieve minor muscle and joint pain and stiffness. The lamps may also help to relieve minor pain associated with muscle spasms, minor sprains and strains, and minor muscular back pain.
Device Story
Xinfeng Heating Lamp (models CQ-27, CQ-36, CQ-55A) provides topical infrared heat therapy. Device consists of heating head, swing arm, control box, and pedestal. Operates by emitting infrared spectrum (5-25 microns for CQ-27/36/CQ-55A State I; 2-50 microns for CQ-55A State II) to heat tissue. User positions lamp head 8-12 inches from treatment area; timer controls duration up to 95 minutes. Intended for clinical or home use to relieve minor musculoskeletal pain and stiffness. Healthcare provider or patient monitors treatment distance to maintain skin temperature between 40-45°C (CQ-55A State II requires minimum 10-inch distance). Benefits include temporary pain relief and increased local circulation.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and comparison of technical specifications, materials, and intended use. Safety testing performed according to IEC 60601-1.
Technological Characteristics
Infrared lamp; 110V power; 50/60 Hz; power ≤280W. Components: heating head, swing arm, control box, pedestal. Emission spectrum: 5-25 microns (standard) or 2-50 microns (CQ-55A State II). Timer: up to 95 minutes. Safety: 2 fuses, 5 casters. Tested to IEC 60601-1. Heating plate replacement required after 1,000 hours.
Indications for Use
Indicated for patients requiring topical heating to elevate tissue temperature, increase local blood circulation, and temporarily relieve minor muscle/joint pain, stiffness, muscle spasms, minor sprains, strains, and minor muscular back pain.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Additional Information for Xinfeng Heating Lamp
# 510 (k) Premarket Notification Summary
## Submitter
George Su Crosslinks International 1800 Century Park East, Suite 600 Los Angeles, CA 90067 USA
Tel: 310-229-5748 Fax: 310-388-1067 Email: crosslinks2000@aol.com
### Manufacturer:
Chongqing Xinfeng Medical Instrument Co., Ltd. 12 Da Xi Gou Street 14-6 Hua Xin Tower Chongqing 400013 People's Republic of China
Tel: (86-23) 6361-7000 Fax: (86-23) 6360-3205 E-mail: wfs(@xf-tdp.com.cn
# Sponsor of the 510(k) Submission:
Chongqing Xinfeng Medical Instrument Co., Ltd. 12 Da Xi Gou Street 14-6 Hua Xin Tower Chongqing 400013 People's Republic of China
Tel: (86-23) 6361-7000 Fax: (86-23) 6360-3205 E-mail: wfs@xf-tdp.com.cn
## Device Name:
Proprietary Name: Xinfeng Lamp Common/Usual Name: Infrared Lamp Device Name: Infrared Lamp Product Code: ILY Classification: Class II
# Predicate Device:
TDP CQ-27 Heat Lamp Establishment: LHASA Medical Inc.
{1}------------------------------------------------
Additional Information for Xinfeng Heating Lamp
Regulation Number: 890.5500 Product Code: ILY 510(k) Number: K003538 Registration Number: 1222811 Owner/Operator Number: 9003816
# Description of the Xinfeng Heating Lamps:
Xinfeng Heating Lamp, including 3 different models CQ-27, CQ-36 and CQ-55A, consists of heating head, swing arm, control box and pedestal with extension. It is used to provide topical heating to the body and is specially engineered using a plate. Model CQ-27 and CQ-36 have only one operating mode, while model CQ-55A has two operating modes. Emission spectrum ranges from 5 to 25 microns for models CQ-27, CQ-36 and state I of CQ-55A. Model CQ-55A also includes an additional operating state (State II) having an emission range of 2 to 50 microns. The emission heating plate shall be replaced after 1,000 hours of usage.
#### Indications for use
Xinfeng Heating Lamp, including 3 different models CQ-27, CQ-36 and CQ-55A, is an infrared lamp that emits the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, to temporarily increase local blood circulation, and to temporarily relieve minor muscle and joint pain and stiffness. The lamps may also help to relieve minor pain associated with muscle spasms, minor sprains and strains, and minor muscular back pain.
# Substantial Equivalence Information
Xinfeng CQ-27, CQ-36 and CQ-55A Heating Lamps are generally equivalent to CQ-27 Heat Lamp (K003538) manufactured by Lhasa Medical, Inc. on label and labeling, materials used, specifications and intended use. All the technical data are tested according to GB 9706.1-1995 which is identical to IEC 60601-1
The equivalence information is focused on intended use, materials and specifications. The Xinfeng Heating lamps use 110V voltage, 5 casters and 2 fuses. The use age is 1000 hours and the skin temperature is at 40~45°C in a treatment distance from heating unit to treatment surface 8" to12" for models CQ-27, CQ-36 and state I of CQ-55A, but for state II operating mode of CQ-55A it is 10" to12".
{2}------------------------------------------------
Additional Information for Xinfeng Heating Lamp
Table 1.
Predicate device: TDP CQ-27 Heat Lamp manufactured by Lhasa Medical, Inc.
The new device: Xinfeng Heating Lamp CQ-27
| Item | The predicate device | The new device | | |
|--------------------------|------------------------------------------------------------------|------------------------------------------------------------------|-------------|------------|
| Components | heating head, swing arm, control box and pedestal with extension | heating head, swing arm, control box and pedestal with extension | | |
| Power Frequency | 50/60 Hz | 50/60 Hz | | |
| Power | $\leq 250 W$ | $\leq 250 W$ | | |
| Spectrum Ranges | 2 to 50 microns | 5 to 25 microns | | |
| Warming Up Time | 15 minutes | 15 minutes | | |
| Total Working<br>Hours | 1,200 to 1500 hours | 1,000 hours | | |
| Operating Timer | Up to 60 minutes | Up to 60 minutes | | |
| Number of Fuse | 2 | 2 | | |
| Inner Cover<br>Diameter | 12 cm | 12 cm | | |
| Number of Casters | 4 | 5 | | |
| Skin Temperature<br>(°C) | The distance from the lamp head and the result | | | |
| | 8"<br>45°C | 10"<br>43°C | 12"<br>41°C | 8"<br>45°C |
Table 2.
Predicate device: TDP CQ-27 Heat Lamp manufactured by Lhasa Medical, Inc. The new device: Xinfeng Heating Lamp CQ-36
| Item | The predicate device | The new device | |
|-----------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------|
| Components | heating head<br>swing arm<br>standard control box<br>pedestal with extension | heating head<br>swing arm<br>control box with digital display<br>pedestal with extension | |
| Power Frequency | 50/60 Hz | 50/60 Hz | |
| Power | $\leq 250 W$ | $\leq 250 W$ | |
| Spectrum Ranges | 2 to 50 microns | 5 to 25 microns | |
| Warming Up Time | 15 minutes | 15 minutes | |
| Total Working Hours | 1,200 to 1500 hours | 1,000 hours | |
| Operating Timer | Up to 60 minutes | Up to 95 minutes | |
| Number of Fuse | 2 | 2 | |
| Inner Cover Diameter | 12 cm | 16.6 cm | |
| Number of Casters | 4 | 5 | |
| Skin Temperature (°C) | The distance from the lamp head and the result | | |
| | 8" | 10" | 12" |
| | 45°C | 43°C | 41°C |
| | 8" | 10" | 12" |
| | 44°C | 42°C | 40°C |
{3}------------------------------------------------
4/4
Additional Information for Xinfeng Heating Lamp
Table 3.
Predicate device: TDP CQ-27 Heat Lamp manufactured by Lhasa Medical, Inc. The new device: Xinfeng Heating Lamp CQ-55A
| Item | The predicate device | The new device |
|---------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Components | heating head with one heating unit<br>swing arm<br>standard control box<br>pedestal with extension | heating head with three heating unit<br>swing arm<br>control box with digital display<br>pedestal with extension |
| Power Frequency | 50/60 Hz | 50/60 Hz |
| Power | $\leq 250 W$ | $\leq 280 W$ |
| Warming Up Time | 15 minutes | 15 minutes |
| Total Working<br>Hours | 1,200 to 1500 hours | 1,000 hours |
| Operating Timer | Up to 60 minutes | Up to 95 minutes |
| Number of<br>heating head | 1 | 3 |
| Number of Fuse | 2 | 2 |
| Inner Cover<br>Diameter | 12 cm | 8 cm |
| Number of Casters | 4 | 5 |
| Spectrum Ranges<br>State I | 2 to 50 microns | | | 5 to 25 microns | | |
|-------------------------------------|------------------------------------------------|------|------|-----------------|------|------|
| Skin Temperature<br>State I<br>(°C) | The distance from the lamp head and the result | | | | | |
| | 8" | 10" | 12" | 8" | 10" | 12" |
| | 45°C | 43°C | 41°C | 44°C | 42°C | 41°C |
| Spectrum Ranges<br>State II | 2 to 50 microns | | | 2 to 50 microns | | |
|------------------------------|------------------------------------------------|------|------|-----------------|------|------|
| Skin Temperature<br>State II | The distance from the lamp head and the result | | | | | |
| | 8" | 10" | 12" | 8" | 10" | 12" |
| (°C) | 45°C | 43°C | 41°C | 47°C* | 44°C | 43°C |
*47°C exceeds the safe temperature range and that model CQ-55A should be positioned at least 10 inches from the body surface to be treated.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
JUN 6 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Chongqing Xinfeng Medical Instrument Company, Ltd. C/o Mr. George Su Crosslinks International, Inc. 1800 Century Park East, Suite 600 Century City, California 90067
Re: K043558
- K043338
Trade/Device Name: Xinfeng Heat Lamps, Models CQ-27, CQ-36, CQ-55A Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: May 26, 2005 Received: May 26, 2005
Dear Mr. Su:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 910(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosulty to regally manced provice Amendments, or to commerce prior to May 28, 1776, the enatified in accordance with the provisions of the Federal recognities (1)MAA devices that have been icclassince in acceracy in of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval appli and Cosmelle Act (Act) that do not require approvine of the general controls provisions of the Act. The You may, therefore, market the device, belyer to the one of the only and registration, listing of general Controls provisions of the Procession, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classince (secure) into existing major regulations affecting your device can
may be subject to such additional controls. Existing major research and like o may be subject to such additional bonnons. Title 21, Parts 800 to 898. In addition, FDA may be found in the Code of I casial Hegening your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issualled of a backed on a complies with other requirements of the Act
that FDA has made a determination that your device complies with other re that I DA has made a decemination administered by other Federal agencies. You must of any Pederal Statutes and regulations daministers of 5 mot limited to: registration and issting (21 comply with an the Act 3 requirements, mercesses and actually of the requirements as set CFR Fart 807), adomig (21 OFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 – Mr. George Su
This letter will allow you to begin marketing your device as described in your Section 510(k) i his letter will anow you to oegin manceing your artial equivalence of your device to a legally premarket notheadon: "The FDA miaing of castion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac-1100 cm on one one the regulation entitled,
contact the Office of Compliance at (240) 276-0120. Also, please note the regulation en Colliact the Office of Comphalled at (210) = 16 = 21 807.97). You may obtain Misoranding by reference to premailson ibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-204 or Manufacturers, international and Oolisal http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Az. S. Rhodes
Miriam C. Provost, Ph.
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Ko455556
Additional Information for Xinfeng Heating Lamp
# Indications for Use
510(k) Number (if known): ___K043558__
Infrared Lamp Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
Xinfeng Heating lamp, including CQ-27, CQ-36 and CQ-55A, is an infrared lamp that emits the infrared spectrum to provide topical heating for the purpose of
lamp that emits the infrared spectrum to provide topical heating and to lamp that emiss the mirated spectrain to from the reseal blood circulation, and to elevaling tissue temperature, to temporarily mercule stiffees. The lamps may also help
temporarily relieve minor muscle and joint pain and strains and strains, and temporarily relieve intor muscle and joint pain and strains and strains and strains, and minor muscular back pain.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hytk. Loodu
Division Sion-Division of General, Restorative, and Neurological Devices
KOYJ558 510(k) Number_
5
Page 1 of 1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.