AMPLIFE UPPER ARM BLOOD PRESSURE MONITOR, MODEL M100
Applicant
Amplife Corporation
Product Code
DXN · Cardiovascular
Decision Date
Jan 27, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The AMPLIFE Upper Arm Blood Pressure Monitor, Model M100, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
Device Story
AMPLIFE Model M100 is an automatic non-invasive blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and pulse rate. Input: cuff pressure variations detected by piezoresistive semiconductor sensor; replaces stethoscope/mercury manometer. Operation: automatic cuff inflation to suitable pressure; controlled deflation via linear valve; sensor converts pressure alterations to electrical signals for processing. Output: blood pressure and pulse rate readings displayed to user. Used in home or clinical settings by adult patients or healthcare providers. Includes clock function. Benefits: provides automated, non-invasive blood pressure monitoring without manual inflation pre-setting.
Clinical Evidence
Clinical testing performed in accordance with ANSI/AAMI SP-10 Standard, Section 4.4.2. No specific numerical results (e.g., mean error, SD) provided in the summary document, but testing confirmed safety and effectiveness equivalent to the predicate device.
Technological Characteristics
Piezoresistive semiconductor pressure sensor; oscillometric measurement principle; automatic inflation/deflation system with linear valve; upper arm cuff; electronic display; includes clock function; powered by internal circuitry.
Indications for Use
Indicated for adult individuals for non-invasive measurement of systolic and diastolic blood pressure and pulse rate via an upper arm cuff.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Microlife Automatic Blood Pressure Monitor, Model BP-2BHO (K970211)
Submission Summary (Full Text)
{0}------------------------------------------------
K04-3550
P1/3
# JAN 2 7 2005
EXHIBIT # 1
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
#### 1. Submitter's Identification:
AMPLIFE Corporation 10F, No. 69, Sec. 3, Hue Jung Rd Taichung, Taiwan, R.O.C
Date Summary Prepared: November 29, 2004
Contact: Mr. Laurence Yang Tel: 0422587766 Fax: 0422587558
#### 2. Name of the Device:
AMPLIFE Upper Arm Blood Pressure Monitor, Model M100
### Common Name or Classification:
Non-Invasive Blood Pressure Measurement System
#### 3. Predicate Device Information:
The AMPLIFE Upper Arm Automatic Blood Pressure Monitor, Model M100 is substantially equivalent to the Microlife Automatic Blood Pressure Monitor, Model BP-2BHO, K# 970211.
#### 4. Device Description:
The AMPLIFE Upper Arm Automatic Blood Pressure Monitor, Model M100 is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the Upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating
{1}------------------------------------------------
pulse rate, which is a well - known technique in the market called the "oscillometric method".
#### Intended Use: 5.
The AMPLIFE Upper Arm Blood Pressure Monitor, Model M100 is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
#### Comparison to Predicate Devices: e.
Both the subject and predicate devices use the well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. For the subject device, the cuff is inflated to a suitable pressure automatically without the need to pre-set the inflation pressure value by hand, as is the case with the predicate device; deflate rate is controlled, but a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. The subject device, Model M100, uses a semiconductor pressure sensor instead of a capacitive pressure sensor (predicate device) to translate the pressure variations to electrical signals that can be interpreted by an integrating circuit. Once the reading is determined, each unit operates the linear valve to release the pressure to zero. In addition, the subject device has the added clock function which is not included in the predicate device.
Unlike the predicate device, the subject device, Model M100, uses a piezoresistive pressure sensor to detect alternation of cuff pressure.
### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the AMPLIFE Automatic Blood Pressure Monitor, Model M100 in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.
The following testing was conducted:
- General Functions Test a.
- Reliability Test Operation Conditions b.
- Reliability Test Drop Testing C.
- Reliability Test Storage ರು.
- Reliability Test Vibrating Testing e.
{2}------------------------------------------------
- f. EMC Test
- IEC 60601-1 Safety Test ப் ப
None of the testing demonstrated any design characteristics that violated the Norte of the Reviewer Guidance or resulted in any safety hazards. It was requirements of the Ronomatic Blood Pressure Monitor, Monitor, Model our ocholder. that relevant requirements of the aforementioned tests.
#### Discussion of Clinical Tests Performed: 8.
We have performed the clinical test on AMPLIFE Upper Arm Blood Pressure We have portormou the "Clinical Data and Analysis ANSI/AAMI -- SP-10 Standard, Section 4.4.2" which outlines and summarizes clinical testing performed.
#### Conclusions: ு.
We have demonstrated that the AMPLIFE Automatic Blood Pressure Monitor, Model M100, is as safe and effective as the predicate, the Microlife Automatic Model Mr. Co, re Monitor, Model BP-2BHO, based on electrical, mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and, the ANSI/AAMI Voluntary Standard, SP10-1992 testing results.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 7 2005
AMPLIFE Corporation c/o Mr. Ned Devine Entela, Inc. 3033 Madison Ave. SE Grand Rapids, MI 43548
Re: K043550
Trade Name: AMPLIFE Upper Arm Blood Pressure Monitor, Model M100 Regulation Number: 21 CFR 870.1130 Regulation Name: Blood Pressure Monitor Regulatory Class: Class II Product Code: DXN Dated: January 14, 2005 Received: January 18, 2005
Dear Mr. Devine:
ﻏﻴ
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Ned Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding bv reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Blinima for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Exhibit B
Page _ 1
K043550 510(k) Number (if known):
Device Name: AMPLIFE Upper Arm Blood Pressure Monitor, Model M100
Indications For Use:
The AMPLIFE Upper Arm Blood Pressure Monitor, Model M100, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use X (21 CFT 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Z.H. HERRMANN, JR.
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number U Cardiovascular Devices
510(k) Number U 510(k) Number K 0 4 న
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.