← Product Code [KWA](/productcode/KWA) · K043547

# COMPRESS SEGMENTAL FEMORAL REPLACEMENT SYSTEM (K043547)

_Biomet, Inc. · KWA · Aug 5, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K043547

## Device Facts

- **Applicant:** Biomet, Inc.
- **Product Code:** [KWA](/productcode/KWA.md)
- **Decision Date:** Aug 5, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3330
- **Device Class:** Class 3
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Tumor resections. Revision of previously failed total joint arthroplasty. Trauma. The ComPreSs® Segmental Femoral Replacement System components are intended for uncemented use only.

## Device Story

The ComPreSs® Segmental Femoral Replacement System is a modular orthopedic implant used for femoral reconstruction in the proximal or distal femur. The system comprises three primary components: an anchor plug, a spindle assembly, and an intercalary segment. The anchor plug is embedded into the medullary canal; the spindle assembly attaches to the anchor plug and is compressed against the bone/implant interface at the osteotomy site; the intercalary segment attaches to the spindle assembly to complete the femoral stem. The device is used by orthopedic surgeons in clinical settings to restore skeletal integrity following tumor resection, trauma, or revision of failed arthroplasty. By providing a stable, compressed interface at the bone-implant junction, the system facilitates structural replacement of bone segments, aiming to improve patient mobility and function.

## Clinical Evidence

No new clinical testing was performed for this submission. The manufacturer relied on clinical data previously submitted in support of the predicate ComPreSs® Distal Femoral Replacement System to establish safety and efficacy.

## Technological Characteristics

Modular segmental femoral replacement system consisting of an anchor plug, spindle assembly, and intercalary segment. Designed for uncemented fixation. Materials, design, and processing are identical to predicate components. Mechanical fixation is achieved via compression at the bone/implant interface.

## Regulatory Identification

A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).

## Predicate Devices

- ComPreSs® Distal Femoral Replacement System - Biomet Inc. ([K031804](/device/K031804.md), [K041352](/device/K041352.md))
- Orthopedic Salvage System (OSS) - Biomet Inc. ([K002757](/device/K002757.md))
- Modular Replacement System (MRS) - Howmedica ([K952970](/device/K952970.md), [K972401](/device/K972401.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 5 - 2005

AUG 5 - 2005 K043547

Image /page/0/Picture/3 description: The image shows the word "BIOMET" in a stylized, bold font. The letters are white with a thick black outline. The "O" in the word is a perfect circle, and the "E" and "T" have a unique, blocky design.

# 510(k) Summary

| Applicant/Sponsor:    | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:       | Gary Baker<br>Biomet Manufacturing Corp.<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587<br>Phone: (574) 267-6639<br>FAX: (574) 372-1683                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Proprietary Name:     | ComPreSs® Segmental Femoral Replacement System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common or Usual Name: | Segmental Femoral Stem Component.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Name:  | Hip joint metal/metal semi-constrained, with a cemented acetabular component,<br>prosthesis (21 CFR §888.3320)<br>Hip joint metal/metal semi-constrained, with an uncemented acetabular<br>component, prosthesis (21 CFR §888.3330)<br>Knee Joint, Femorotibial, Metal/Polymer Constrained, Cemented Prosthesis (21<br>CFR §8883510).<br>The Compress® Proximal Femoral components, compatible shells, heads, and<br>liners included in this submission have the following classifications:<br>1. Hip joint metal/polymer constrained cemented or uncemented prosthesis (21 CFR §888.3310)<br>2. Hip joint metal/polymer semi-constrained cemented prosthesis (21 CFR §888.3350)<br>3. Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21 CFR §888.3353)<br>4. Hip joint metal/polymer/metal semi-constrained, porous-coated, uncemented prosthesis (21 CFR §888.3358)<br>5. Hip joint (hemi-hip) acetabular metal cemented prosthesis (21 CFR §888.3370)<br>6. Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis (21 CFR §888.3390) |

Device Product Codes: JDL, KWA, KRO, KWZ, JDI, LZO, MEH, LPH, LZY and KWY.
126

 MAILING ADDRESS  

 P.O. Box 587  

 Warsaw, IN 46581-0587

 SHIPPING ADDRESS  

 56 E. Bell Drive  

 Warsaw, IN 46582
���רכ�� ээг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэвэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг хэрэг х

-

()[ff[[C][ 574.267.6639

l·AX 574.267.8137

■

Warsaw, IN 46582 E-MAIL .

biomet@biomet.com

{1}------------------------------------------------

K043547

2/2

## Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

ComPreSs® Distal Femoral Replacement System - Biomet Inc. (K031804, K041352) Orthopedic Salvage System (OSS) - Biomet Inc. (K002757) Modular Replacement System (MRS) - Howmedica (K952970, K972401)

### Device Description:

The ComPreSs® Segmental Femoral Replacement System consists of three main components, the anchor plug, the spindle assembly, and the intercalary segment. The Anchor Plug is embedded within the medullary canal, the Spindle Assembly attaches to the Anchor Plug and is compressed against the bone / implant interface at the osteotomy site. The Intercalary Segment attaches to the Spindle Assembly, and completes the Femoral Stem Component.

This system is intended for use in both the proximal and distal femur.

## Intended Use:

Indications:

1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.

- 2. Tumor resections.
- 3. Revision of previously failed total joint arthroplasty
- 4. Trauma

The ComPreSs* Segmental Femoral Replacement System components are intended for uncemented use only.

Summary of Technologies: All components of this system are identical in material, processing, function, and design to their respective predicate components.

Non-clinical Testing: The engineering rationale with the attached literature and mechanical test reports support the claim for substantial equivalence to the predicate devices.

Clinical Testing: Clinical testing in support of the predicate Compress® Distal Femoral Replacement System is sufficient to ensure safety and efficacy. No further clinical testing is necessary to support the claim for substantial equivalence to the predicate devices.

All trademarks are property of Biomet, Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, conveying the organization's focus on health and human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 23 2009

Biomet Manufacturing Corp. % Mr. Gary Baker Regulatory Specialist 56 East Bell Drive Warsaw, Indiana 46581-0587

Re: K043547

Trade/Device Name: Compress Segmental Femoral Replacement System Regulation Number: 21 CFR 888.3320 Regulation Name: Hip joint metal/metal semi-constrained, with a cemented acetabular component, prosthesis Regulatory Class: III Product Code: KWA, JDL, KRO, KWZ, JDI, LZO, MEH, LPH, KWY Dated: July 25, 2005 Received: July 26, 2005

Dear Mr. Baker:

This letter corrects our substantially equivalent letter of August 5, 2005.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not

{3}------------------------------------------------

Page 2 - Mr. Gary Baker

limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
Chardan Ruenup

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Page I of I

#### Statement of Indications For Use

510(k) Number (IF KNOWN): _KO43547

Device Name: Compress Segmental Femoral Replacement System

#### Indications for Use:

Indications:

1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.

2. Tumor resections.

3. Revision of previously failed total joint arthroplasty

4. Trauma

・・

The Compress Segmental Femoral Replacement System components are intended for uncemented use only.

Prescription Use X (Per 21 CFR 801 Subpart D) OR

Over-the-Counter Use (Per 21 CFR 807 Subpart C)

LEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Pavlovich

Division Sign-Restorative, Division of Gene and Neurological

**510(k) Number** K043547

---

**Source:** [https://fda-staging.innolitics.com/device/K043547](https://fda-staging.innolitics.com/device/K043547)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).