# SOLEUS POWERED WHEELCHAIR, MODEL FAD-2 (K043511)

_Soleus Champ International Co., Ltd. · ITI · Mar 14, 2005 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K043511

## Device Facts

- **Applicant:** Soleus Champ International Co., Ltd.
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Mar 14, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

## Device Story

The Soleus FAD-2 is a battery-operated indoor/outdoor powered wheelchair. It features a four-wheeled base and a seat. The rider operates the device using hand controls located on a steering column. The wheelchair is designed to be disassembled for transport and includes an onboard battery charger. It is intended for use by individuals with mobility restrictions to provide independent movement. The device is operated by the patient.

## Clinical Evidence

Bench testing only. The device underwent electromagnetic compatibility (EMC) and safety testing according to ANSI/RESNA WC/Vol.2-1998, CISPR 11:1990, EN61000-3-2:1995, and IEC61000-3-3:1995. Additionally, frame and upholstery materials passed resistance ignition tests.

## Technological Characteristics

Battery-operated powered wheelchair; four-wheeled base; hand-controlled steering column; foldable frame; onboard battery charger. Complies with ANSI/RESNA WC/Vol.2-1998, CISPR 11, EN61000-3-2, IEC61000-3-3, and EN 12184 standards. Electronic systems are UL certified.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- TEH LIN POWERED WHEELCHAIR MDG-201 ([K022696](/device/K022696.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K043511

## SOLEUS CHAMP INTERNATIONAL CO., LTD.

No.1-6, Kung Fu Rd., Liao Bao Village, Ta Ya Hsiang, Taichung Hsien, Taiwan, 428, R.O.C. Tel: 886-4-2560-1853 Fax: 886-4-2560-1142 E-mail: kspeed97@ms27.hinet.net

MAR 1 4 2005

#### દ દ 510(k) SUMMARY "

### Submitter's Name: SOLEUS CHAMP INTERNATIONAL CO., LTD.

No.1-6, Kung Fu Rd., Liao Bao Village,Ta Ya Hsiang, Taichung Hsien, Taiwan, 428, R.O.C.

Date summary prepared:

December 9, 2004

Device Name:

Proprietary Name: Common or Usual Name: Classification Name:

SOLEUS POWERED WHEELCHAIR, FAD-2 Powered Wheelchair Powered Wheelchair, Class II, 21 CFR 890.3860

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

Description of the device:

The SOLEUS POWERED WHEELCHAIR, FAD-2 is an indoor / outdoor Powered Wheelchair that is battery operated. It has a base with four-wheeled with a seat. The movement of the Wheelchair is controlled by the rider who uses hand controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.

Performance Testing:

EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically Powered Wheelchairs, controller, and their chargers - requirements and test methods)

Legally marketed device for substantial equivalence comparison: TEH LIN POWERED WHEELCHAIR MDG-201 (K022696)

{1}------------------------------------------------

# SOLEUS CHAMP INTERNATIONAL CO., LTD.

No.1-6, Kung Fu Rd., Liao Bao Village, Ta Ya Hsiang, Taichung Hsien, Taiwan, 428, R.O.C. Tel: 886-4-2560-1853 Fax: 886-4-2560-1142 E-mail: kspeed97@ms27.hinet.net

Summary for substantial equivalence comparison:

The electronic systems between two devices are the same suppliers and all passed by the UL certificated, for instance the batteries and recharge. Thus the same safety level for the two devices is assured. Besides, the foldable frame and back upholstery are the same material that also be passed the resistance ignition test by The major differences existing of the two Powered Wheelchairs are the SGS. different overall dimension, and the sizes of tires are differences between the two The overall appearance differences are not safety aspect. Besides the two devices. devices use the different electric controllers but all passed by EN 12184, WC / Vol.2:1998, Radiated Immunity: 30 V/m. So the new device is substantially equivalent to the predicate devices in this aspect.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three swooping lines representing its wings. The text "DEPARTMENT OF HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

MAR 1 4 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Dr. Ke Min Jen Official Correspondence Soleus Champ International Co., LTD. No. 1-6, Kung Fu Rd., Liao Bao Village Ta Ya Hsiang, Taichung Hsien, Taiwan, 428, ROC

Re: K043511

Trade/Device Name: Soleus Power Wheelchair, FAD-2 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: January 27, 2005 Received: February 22, 2005

Dear Dr. Ke Min Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device is died in we nave leviewed your Section 910(x) prematitilly equivalent (for the indications for use stated in the above and have determined the devices marketed in interstate commerce prior to May 28, 1976, enclosure) to regarly marketod produce Amendments, or to devices that have been reclassified in the chacinent date of the Frederal Food, Drug, and Cosmetic Act (Act) that do not require accondited with the provisions of the PMA). You may, therefore, market the device, subject to approval of a premaince approvial approvince to the general controls provisions of the Act include the general controls pro risins of an i seing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be If your device is elassified (soc above) into major regulations affecting your device can be found in the Subject to Suen additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act of Statules and legalatons administered of other in the mand listing (21 CFR Part 807); labeling (21 CFR requirements, meruding, but not mines es virements as set forth in the quality systems (QS) regulation I all 601), good minutectaring product radiation control product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2 -- Dr. Ke Min Jen

This letter will allow you to begin marketing your device as described in your Section 5 I0(k) premarket This letter will allow you to begin marketing your device of your device to a legally market. notification. The FDA Inding of substantal equivalience or your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact If you desire specific advice for your device on the regulation entitled, "Mishranding by the Office of Compliance at (240) 270-0120 . Yuso, production other general information on
reference to premarket notification" (21CFR Part 807.97). You may obtain other an reference to premiarket nonitation (21ct I F in Correst Small Manufacturers, International and your responsibilities under the Act from the Division of Shan Manada 44-47-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark Melbern

Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

043511 510 (K) Number ( If Known ):_

Device Name: SQLEUS Power Wheelchair, FAD-2_

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted 1 a sitting position.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use √

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark. A. Millham

Page_ 1__of __1

Division of Genera , Restorative, and Neurological Devices

**510(k) Number** K043511

---

**Source:** [https://fda-staging.innolitics.com/device/K043511](https://fda-staging.innolitics.com/device/K043511)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).