K043435 · Sunrise Medical · ITI · Feb 11, 2005 · Physical Medicine
Device Facts
Record ID
K043435
Device Name
QUICKIE INTERCHANGE FWD W/LIFT POWER WHEELCHAIR
Applicant
Sunrise Medical
Product Code
ITI · Physical Medicine
Decision Date
Feb 11, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The Quickie Interchange FWD w/ Lift Power Wheelchair's intended use is to provide mobility to persons limited to a seating position, that have the capability of opprating a powered wheelchair. The Quickie Interchange FWD w/ Lift Power Wheelchair provides an optional means of mobility for physically challenged people.
Device Story
Quickie Interchange FWD w/ Lift is a modular, high-end/mid-range rehabilitation power wheelchair base. Design allows configuration as either front-wheel drive or rear-wheel drive using common frame parts. System includes rigid seat frame, backrest, armrests, cushion, footrests, casters, controller, joystick, motor, brakes, and 24V DC rechargeable batteries. Accessories include positioning belts, backpacks, seat pouches, oxygen tank holders, and IV poles. Operated by end-user via joystick or equivalent command mode. Used for indoor and moderate outdoor mobility. Provides independent mobility for physically challenged individuals. No therapeutic claims made.
Clinical Evidence
Bench testing only. Device tested in accordance with ISO 7176-14:1997 (Power and Control Systems for Electric Wheelchairs) and other applicable ISO/ANSI/RESNA standards. Results demonstrated the device met required performance criteria and functioned as intended.
Technological Characteristics
Modular power wheelchair base; steel and aluminum frame; standard foam/fabric cushions. Energy source: 24V DC rechargeable batteries. Control: Programmable controller with joystick interface. Connectivity: None. Sterilization: N/A. Software: Embedded firmware for motor/controller management.
Indications for Use
Indicated for physically challenged individuals limited to a seating position who have the capability to operate a powered wheelchair independently.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains a handwritten string of characters. The string appears to be "K osl3435". The characters are written in a cursive style, and the image is in black and white.
#### Section 1.0 510(k) Summary
## Administrative Information and Device Identification
| Name and address of the manufacturer and<br>sponsor of the 510(k) submission: | Sunrise Mobility Products Division<br>2842 Business Park Avenue<br>Fresno, California 93727-1328 |
|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA registration number of the manufacturer<br>of the new device: | 2082643 |
| Official contact person for all correspondence: | Joseph E. Olsavsky<br>Director - Regulatory Affairs<br>Sunrise Medical<br>100 DeVilbiss Drive<br>Somerset, PA 15501<br>Phone: 814-443-7690<br>Fax: 814-443-7597<br>Cell: 814-521-9152<br>Email: joe.olsavsky@sunmed.com |
| Date Prepared: | November 18, 2004 |
| Device Name: | Quickie Model Powered<br>Wheelchair |
| Proprietary name of new device: | Quickie Interchange FWD w/ Lift |
| Generic name of the device: | Quickie Powered Wheelchair |
| Classification of the predicate device: | Class II |
| Classification of new device: | Class II |
| Classification Panel: | Physical Medicine |
| Panel Code: | ITI |
| CFR Regulation Number: | 21 CFR 890.3860 |
| Predicate Device Name(s) and 510(k)<br>number(s): | Arrow FWD Power Wheelchair<br>(K991168) |
{1}------------------------------------------------
#### Description of Device:
The Quickic Interchange FWD w/ Lift Power Wheelchair is a High End/Mid-Range Rehabilitation type Power Chair Base design that offers the functionality to be used as a Front Wheel drive or a Rear Wheel drive base without changing major frame parts. The system allows for reassembling parts that will for either front or rear wheel drive function. The design will offer a modular approach within the drive base as intve adopt various seating & wiring elements. The device utilizes components typically found on most wheelchairs, including but not limited to rigid seat frame, backrest, product armrests, cushion, footrests and casters. Accessories that may be added after market include items such as positioning belts, backpacks, seat pouches, oxygen taak holders, IV poles, etc. As a motorized wheelchair, it also contains a controller, joystick, motor, brakes, drive wheel and batteries. This product is appropriate for use by any individual who has the ability to drive a powered wheelchair without having to utilize the services of an attendant. Sunrise Medical makes no claim as to the therapeutic effectiveness of the product. Sunrise Medical's only claim relates to the ability of the product to provide an optional means of mobility for physically challenged people.
## Comparison of Device Technological Characteristics to Predicate Devices:
This device has similar technological characteristics as the predicate devices. They all use steel and aluminum in their frames and components, and standard foams and covers for the slings, backs and cushions. Microprocessors are typically used with a programmable controller, and the end-user controls the chair by using a joystick or other equivalent command mode. Motors use 24 volt DC rechargeable batteries for an energy source. The operating speeds, maneuverability, power modules, hand controls, seat types, drive wheels, and climbing ability are substantially equivalent and are recommended for indoor and moderate outdoor use. The standard accessories and components are common to all power wheel chair devices.
See Section 8.0.
#### Statement of Intended Use:
The Quickie Interchange FWD w/ Lift Power Wheelchair's intended use is to provide mobility to persons limited to a seating position, that have the capability of opprating a powered wheelchair. The Quickie Interchange FWD w/ Lift Power Wheelchair provides an optional means of mobility for physically challenged people.
#### Non-Clinical Testing:
This device has been tested to appropriate ISO & ANSI/RESNA standards and other applicable requirements passing all test protocols.
As required by FDA's July 26, 1995, draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanice and
{2}------------------------------------------------
Powered Wheelchairs, and Motorized Three-Wheeled Vehicles", the Quickic Interchange FWD w/ Lift Power Wheelchair was tested in accordance with the ISO EMC Standard "ISO 7176-14:1997. Wheelchairs - Part 14. Power and Control Systems for Electric Wheelchairs - Requirements and Test Methods". The Quickie Interchange FWD v/ Lift Power Wheelchair met the required performance criteria and functioned is intended.
See Section 10 Test Reports and Attachment A.
### Statement of Safety and Effectiveness:
Analysis of comparison of design, function and features of the Sunrise Medical Quickie Interchange FWD w/ Lift Power Wheelchair to the Invacare Arrow FWD Power Wheelchair, together with the results of testing demonstrates the device to be substantially equivalent to the predicate device in terms of meeting performance criteria and functioning as intended.
Conclusion: The Sunrise Medical Quickie Interchange FWD w/ Lift Power Wheelchair is substantially equivalent to the predicate device listed in this Summary and the device, as changed, does not raise any new issues of safety and effectivenss.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's emblem in the center. The emblem is a stylized image of an eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion.
FEB 1 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Joseph E. Olsavsky Director, Regulatory Affairs Sunrise Medical 100 DeVilbiss Drive Somerset, Pennslyvania 15501-2125
Re: K043435
Device Name: Quickie Interchange FWD w/Lift Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: II Product Code: ITI Dated: January 28, 2005 Received: January 31, 2005
Dear Mr. Olsavsky
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Mr. Joseph E. Olsavsky
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and your your finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spocific acrioliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionnal and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Milliman
Celia M. Witten, Ph.D., M.D. Director Division Of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K043435
Device Name:
Quickie Interchange FWD w/ Lift Power Wheelchair
Indications for Use:
The Quickie Interchange FWD w/ Lift Power Wheelchair's intended use is to empower physically challenged individuals by providing a means of mobility.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
ﺮ ﺍﻟﻤﺮﺍ
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark H. Mellaver
Division of General, Restorative, and Neurological Devices
**510(k) Number** K043435
Page 1 of of -----------------------------------------------------------------------------------------------------------------------------------------------------------------
(
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.