AUTOSTAT II ANTI-TISSUE TRANSGLUTAMINASE IGG ELISA

K043433 · Hycor Biomedical , Ltd. · MVM · Mar 7, 2005 · Immunology

Device Facts

Record IDK043433
Device NameAUTOSTAT II ANTI-TISSUE TRANSGLUTAMINASE IGG ELISA
ApplicantHycor Biomedical , Ltd.
Product CodeMVM · Immunology
Decision DateMar 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

Enzyme linked immunosorbent assay method for the semi-quantitative determination of specific IgG autoantibodies to tissue transglutaminase (tTg) in human serum. Uses: The results of the anti-fTg assay can be used as an aid in the diagnosis of Coeliac Disease. Levels of these autoantibodies are one indicator in a multi-factorial diagnostic regime. In addition to the manual assay protocol, this device has been validated for use with the HYCOR HY.TEC automated EIA instrument. For in vitro diagnostic use only.

Device Story

Device is semi-quantitative ELISA for detection of anti-tTg IgG antibodies in human serum; utilizes microplate wells coated with recombinant human tissue transglutaminase. Input: human serum samples. Process: patient serum added to antigen-coated wells; anti-tTg IgG binds to antigen; horseradish peroxidase-conjugated anti-human IgG added; TMB substrate added to produce colorimetric reaction proportional to antibody concentration. Output: optical density measured at 450nm via spectrophotometer; results calculated as arbitrary units/mL. Used in clinical laboratories; manual protocol or automated via HYCOR HY.TEC EIA instrument. Output aids clinicians in Celiac Disease diagnosis as part of multi-factorial diagnostic regime.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Solid-phase ELISA; 96-well microtiter plates coated with recombinant tissue transglutaminase (E. coli derived). Detection via horseradish peroxidase-conjugated anti-human IgG and TMB substrate. Energy source: spectrophotometer (450nm). Connectivity: standalone or integrated with HYCOR HY.TEC automated EIA instrument. Software: automated dilution and calculation functions.

Indications for Use

Indicated for use as an aid in the diagnosis of Celiac Disease in human patients. The device detects IgG autoantibodies to tissue transglutaminase (tTg) in serum.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k043433 B. Purpose for Submission: New Device C. Measurand: Anti- Tissue Transglutaminase D. Type of Test: Semi- quantitative ELISA E. Applicant: Hycor Biomedical Ltd. F. Proprietary and Established Names: AUTOSTAT™ II Anti-Tissue Transglutaminase IgG ELISA G. Regulatory Information: 1. Regulation section: 21 CFR 866.5660, Multiple autoantibodies immunological test system 2. Classification: II 3. Product code: MVM, Autoantibodies, Endomysial (Tissue Transglutaminase) 4. Panel: Immunology 82 H. Intended Use: 1. Intended use(s): Enzyme linked immunosorbent assay method for the semi-quantitative determination of specific IgG autoantibodies to tissue transglutaminase (tTg) in human serum. The results of the anti-tTg assay can be used as an aid in the diagnosis of Celiac Disease. Levels of these autoantibodies are one indicator in a multi-factorial diagnostic regime. In addition to the manual assay protocol, this device has been validated for use with the HYCOR HY.TEC automated EIA instrument. For in vitro diagnostic use only. 2. Indication(s) for use: Same as intended use. 3. Special conditions for use statement(s): The device is for prescription use only. 4. Special instrument requirements: {1} This device has been validated for use with the HYCOR HY.TEC automated EIA instrument (k941278). # I. Device Description: The Hycor AUTOSTAT™ II Anti-Tissue Transglutaminase IgG ELISA test is an enzyme-linked immunosorbent assay for detection of anti-tTg IgG antibodies in human serum. It consists of microplate wells coated with purified antigen, horseradish peroxidase conjugated mouse anti-human IgG, four standards, positive and negative controls, TMB substrate solution, sample diluent, wash buffer and stop solution. The conjugate, substrate and stop solution are supplied in ready to use format. # J. Substantial Equivalence Information: 1. Predicate device name(s): AUTOSTAT™ II Anti-Tissue Transglutaminase IgA ELISA 2. Predicate 510(k) number(s): K033744 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | New Device | Predicate Device | | Indications for use | The results of the anti-tTg assay can be used as an aid in the diagnosis of Celiac Disease. Levels of these autoantibodies are one indicator in a multi-factorial diagnostic regime. In addition to the manual assay protocol, this device has been validated for use with the HYCOR HY.TEC automated EIA instrument. | Same | | Technology | ELISA | Same | | Assay Format | Semi-quantitative | Same | | Sample type and dilution | Serum at 1/100 | Same | | Enzyme-Conjugate | Horseradish Peroxidase | Same | | Sample diluent, wash buffer, substrate and stop solution | Same | Same | | Incubation times | 30, 15 and 15 minutes | Same | | Antigen | Recombinant tissue transglutaminase, expressed in E.coli | Same | | Platform | 96 well microtitre plates | Same | # Differences {2} | Item | Device | Predicate | | --- | --- | --- | | Intended use/indications for use | IgG anti-tTg autoantibodies | IgA anti-tTg autoantibodies | | Standards | 10, 20, 50, 100 arbitrary u/mL | 6, 12.5, 50, 150 arbitrary u/mL | | Cut-off | 17 u/mL | 7 u/mL | # K. Standard/Guidance Document Referenced (if applicable): None referenced. # L. Test Principle: The Autostat™II assay for detection of autoantibodies is a solid phase immunosorbent assay (ELISA) in which the analyte is indicated by a color reaction of an enzyme and substrate. The Autostat™II wells are coated with purified antigen. Standards, controls and diluted patient sera are added to separate wells and during incubation anti-tTg IgG antibodies in the sample will bind to the antigen. After incubating at room temperature and washing away unbound material, horseradish peroxidase conjugated anti-IgG antibody is added, which binds to the immobilised antibodies. Following further incubation and washing, tetra-methyl benzidine substrate (TMB) is added to each well. The presence of the antigen-antibody-conjugate complex turns the substrate to a dark blue color. Addition of the stop solution turns the color to yellow. The color intensity is measured at $450\mathrm{nm}$ by a spectrophotometer and is directly proportional to the amount of autoantibodies present in the original serum sample. In the automated instrument, the dilutions of controls and samples are performed automatically and the assay results are calculated automatically. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: a. Precision/Reproducibility: i. Study design: Three samples (high, medium and near cut-off) were assayed in duplicates over 20 assays by the manual and by the automated method. The $\% \mathrm{CV}$ for each duplicate in each assay was calculated, and the mean over 20 assays was taken. This is the intra-assay precision (variation). Inter-assay precision (variation) is calculated by calculating the mean concentration value between the twenty assays for each sample. ii. Results/Acceptance criteria: Hycor's acceptance criteria for intra-assay variation and inter-assay variation are $&lt; 10\%$ and $&lt; 15\%$ respectively. | Manual Method | Intra-assay Variation | | | | --- | --- | --- | --- | | | Sample 1 | Sample 2 | Sample 3 | | CV range | 0.4 – 9.6 | 0.4 -22.7 | 0.3 – 20.9 | | Mean CV over 20 assays | 4.4 | 5.8 | 7.4 | | Automated Method | | | | | CV range | 0.5 – 17.2 | 0.4 – 12.7 | 0.3 – 15.3 | | Mean CV over 20 assays | 5.4 | 4.1 | 5.0 | | Manual Method | Inter-assay Variation | | --- | --- | {3} | | Sample 1 | Sample 2 | Sample 3 | | --- | --- | --- | --- | | Concentration range | 15.0 - 20.3 | 39.3 -52.1 | 62.2 – 81.2 | | Mean CV over 20 assays | 7.4 | 6.5 | 8.2 | | Automated Method | | | | | Concentration range | 50.7 – 71.1 | 35.2 – 44.1 | 17.6 – 20.7 | | Mean CV over 20 assays | 8.0 | 6.6 | 5.2 | Total precision was calculated on the basis of 20 runs over 6 days. By the manual method, the total precision for sample 1 was 11.44%, for sample 2 was 14.21% and for sample 3 was 16.34%. By the automated method, the total precision for sample 1 was 13.69%, for sample 2 was 10.43% and for sample 3 was 10.83%. b. Linearity/assay reportable range: i. Study design: Three samples known to contain high levels of anti-tTg IgG antibody were chosen and serially diluted to determine the linearity of the assay. From an initial dilution of 1/100, further dilutions of ½, ¼, 1/6, 1/8, 1/10, 1/12 and 1/14 were made. Each dilution was run in duplicate and concentration values were plotted against dilution ratios. ii. Results/Acceptance criteria: The results showed that the assay was linear up to and beyond the standard 4 at 100 u/mL. Based on this study, the dynamic range of the assay is 1 to 100 u/mL. c. Traceability, Stability, Expected values (controls, calibrators, or methods): The standards are prepared in-house. International Reference Preparation (IRP) is not available for tTg, hence, arbitrary units are assigned during the development process. These values are verified and quality control tested against previously approved batches. In addition, a panel of positive and negative samples is run to verify sensitivity of the standards. The positive and negative controls are prepared in-house from commercially purchased sera. The controls are quality control tested by running them at the assay dilution (1/100) on three different batches of plates using an approved batch of standards. The controls are also tested in the final assay of the complete kit prior to release to verify that the controls are within their respective acceptable ranges. d. Detection limit: The analytical sensitivity of the assay was established by calculation of the mean plus three standard deviations of a minimum of 20 replicates of the zero standard which gave a value of 0.534 u/mL for manual method and 0.598 u/ml for automated method. e. Analytical specificity: Interference by endogenous substances: No interference study data was given. The package insert states that grossly hemolysed, lipemic or microbiologically contaminated samples should not be used for this assay. Also, it states that samples with abnormally elevated levels of hemoglobin, bilirubin and especially EDTA may interfere with assay performance and accuracy. f. Assay cut-off: The cut-off value for this device was selected on the basis of statistical calculation of the mean value of 68 normal samples. The mean plus three standard deviations of this normal population was 13.5 u/ml. {4} The assay cut-off shown below was based on testing of 68 normal samples. | | tTG IgG (u/mL) | | --- | --- | | Negative | <17 | | Positive | ≥17 | # 2. Comparison studies: a. Method comparison with predicate device: i. Study design: The manual assay was compared with predicate device which was an anti-tTg IgA kit. A total of 152 samples were assayed. 68 of the samples were from normal subjects, 47 samples were from patients with confirmed Celiac Disease and 37 samples from patients with other autoimmune conditions. Normal samples were from males and females aged between 18 and 56 years. The Celiac disease positive test panel was from a commercial supplier and included samples from males $(n = 13)$ and females $(n = 31)$ aged between 14 and 50 years. ii. Results: All Celiac diseased samples tested positive by the new device and by the predicate device. One normal sample and an SLE sample also tested positive by both new and predicate devices. | | Autostat II tTg IgG | | | | | --- | --- | --- | --- | --- | | | | + | - | Total | | Predicate Device | + | 49 | 0 | 49 | | | - | 0 | 103 | 103 | % positive agreement $= 100\%$ % negative agreement $= 100\%$ % total agreement $= 100\%$ # Manual and Automated correlation: A linear regression analysis of the correlation between manual and automated tTg IgG assay was performed by running 72 samples on both systems. The correlation coefficient is 0.986. The slope is 0.9761 (95% CI: 0.95651 to 0.9957) and the intercept is 0.132 (95% CI:-06644 to 0.9284). {5} ![img-0.jpeg](img-0.jpeg) b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: A panel of 47 confirmed Celiac Disease samples was run in this study. All samples tested positive by this assay. b. Clinical specificity: Of a panel of 68 normal samples, 67 samples gave negative results. c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable. 4. Clinical cut-off: Same as Assay cut-off. 5. Expected values/Reference range: Same as Assay cut-off. # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...