K043412 · Impac Medical Systems, Inc. · LLZ · Dec 20, 2004 · Radiology
Device Facts
Record ID
K043412
Device Name
MODIFICATION TO: VIEWSTATION
Applicant
Impac Medical Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 20, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ViewStation supports image and information flow among health care facility personnel. ViewStation can be used whenever digital images and associated data are the means for communicating information. ViewStation is not intended for use in diagnosis. The intended use of ViewStation is to provide health care facility personnel with an efficient and effective means to utilize patient images during the course of therapy or treatment. ViewStation allows users to import, view, annotate, manipulate, enhance, manage, and archive patient images. The images and associated information are stored in a database, providing users access to the information necessary to perform their functions.
Device Story
ViewStation is a software-only image management system used by healthcare personnel in clinical environments to facilitate image and data flow. It imports digital images generated by third-party products, storing them in a central database. Users interact with the system to view, annotate, manipulate, and enhance these images to support therapy, treatment, and follow-up workflows. By centralizing access to digital images, the device eliminates the need for physical film transfer, increasing departmental productivity. The system includes tools for image registration to calculate relative reference position shifts and provides security features for HIPAA compliance. It does not perform automated diagnosis; it serves as a tool for clinicians to access and utilize existing patient information during care.
Clinical Evidence
No clinical data provided. Bench testing only. Verification and validation activities, including system hazard analysis and software testing, confirmed that all functional requirements and hazard mitigations were successfully implemented.
Technological Characteristics
Software-only image management system. Functions include image import, viewing, annotation, manipulation, enhancement, and archiving. Features image registration for relative position shifts. Compliant with ISO 9001:2000, ISO 13485:2003, ISO 14971:2000, EN 60601-1-4:1996, and ISO/IEC 9003:2004.
Indications for Use
Indicated for use by healthcare facility personnel to manage, view, annotate, and archive digital patient images and associated data during therapy, treatment, or follow-up. Not intended for diagnostic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
KO434/2
DEC 20 2004
### Thursday, December 09, 2004
# ViewStation, Image Management System Premarket Notification (510(k)) 510(K) Summary
#### Introduction
This document provides a summary of the safety and effectiveness information contained in the Viewstation Premarket Notification (510(k)). The Summary of Safety and Effectiveness contains no confidential or trade secret information and is intended for full public disclosure and distribution. For additional information, feel free to contact the submitter's Management Representative listed below.
#### Premarket Notification Information
- 1. Previous Notification Information:
- Previous Premarket Notification # K011694, K 942346 a.
- Previous Submission Date 05/31/01, 05/16/94 b.
- 08/29/01, 05/15/95 Previous FDA Clearance Date c.
- ViewStation/Image RT d. Product Name
#### 2. Product Information
| a. | Product Name | ViewStation |
|----|------------------------|-------------|
| b. | Release Version Number | Version 8.2 |
| c. | Product Size | 4.9 MB |
#### Classification Information 3.
b.
C.
- Picture Archiving and Communications System a. Classification Name
Class II
Radiology
21 CFR 892.2050
- Common/Usual Name Image Management System
- Product Classification ದಿ.
- 90 LLZ Product Code b.
- CFR Reference
- Review Panel d.
#### 4. Establishment Information
- Submitter IMPAC Medical Systems, Inc. e. f. Manufacturer (no sterilization) Submitter Type 2950347 Establishment Number g. Contact Thomas H. Faris, VP RA/QA h. 650-623-8807 i. Contact Phone Contact Email TFaris@impac.com j
{1}------------------------------------------------
### Predicate Device
ViewStation is substantially equivalent to the original ViewStation product, which is currently marketed by IMPAC Medical Systems, Inc., and was the subject of Premarket Notifications K011694 and K942346. The new ViewStation and previous ViewStation products are equivalent in intended use and safety and effectiveness.
#### ViewStation Indications for Use
ViewStation supports image and information flow among health care facility personnel. ViewStation can be used whenever digital images and associated data are the means for communicating information. ViewStation is not intended for use in diagnosis.
The intended use of ViewStation is to provide health care facility personnel with an effective means to utilize patient images during the course of therapy or treatment. ViewStation allows users to import, view, annotate, manipulate, enhance, manage, and archive patient images and associated information are stored in a database, providing users access to the information necessary to perform their functions.
#### Description of the Product
The primary function of ViewStation is to provide a means to more effectively manage image information in a therapy or treatment environment. ViewStation provides the ability to import, view, annotate, manipulate, enhance, and archive patient images during the course of therapy, treatment, and follow-up.
ViewStation imports existing digital images acquired or generated by other products. ViewStation retains the original image, which was acquired or generated by a third party product. With these facts in mind, the goal of ViewStation is to make electronic patient image information more accessible throughout the department. IMPAC is providing a tool to increase department productivity since digital images, unlike films, do not have to be physically transferred from one station to another.
#### Summary of the Product Change
The only modifications made to the product are the following evolutionary software changes:
- . Image viewing and image management feature improved annotation layer overlay, and image registration for calculation of relative reference position shifts.
- . Privacy and security functionality to support customer HIPAA compliance.
- User interface and product registration improvements. .
- . Other minor workflow enhancements that do not affect product safety or efficacy.
### Clinical Demonstration of Efficacy
No additional or changed diagnostic or therapeutic claims arise as the result of the ViewStation product. Therefore, demonstration of clinical efficacy is not a required element of this Premarket Notification. Further, clinical performance data is not required for determination of substantial equivalence for this type and class of device.
#### Substantial Equivalence Comparison
ViewStation is substantially equivalent to the original ViewStation product, which is currently being marketed by IMPAC Medical Systems, Inc. This was the subject of Premarket Notifications K011694 and K942346.
{2}------------------------------------------------
This Premarket Notification is a resubmission for ViewStation, triggered by enhancement of existing functionality. A I his realler incremental changes that are typical with routine software maintenance also accompany this change, but are too insignificant to merit discussion. The total sum of all feature enhancements does not affect the intended use of ViewStation. The technological characteristics remain principally the same. The evolutionary product changes does not raise any new questions of safety and effectiveness, nor do the changes require novel methods of verification or validation. The product change does not diminish the safety or effectiveness of ViewStation.
The modified ViewStation Image Management System has the following similarities to the previous product that received FDA 510(k) clearance:
- Identical indications for use; .
- . The same operating principles;
- Incorporation of the same basic product design; .
- Same materials a software only product; and .
- Sterility, biocompatibility, materials, mechanical safety, energy use, environmental . compatibility, and electrical safety all remain non-applicable issues.
The sum of the changes does not affect the basic functionality of ViewStation remains dedicated to providing healthcare personnel with a means to import, view, annotate, manage, and archive patient More simply, the changes do not affect the premise of the basic intended use of ViewStation: to provide images. healthcare facility personnel with an effective means to utilize patient images during the course of therapy or treatment. Once cleared by the FDA, IMPAC will continue to market and distribute ViewStation to be used in this manner.
In summary, the ViewStation Image Management System changes described in this submission are substantially equivalent to the predicate ViewStation device.
#### Device Safety
ViewStation is a medical device that is to be used in a treatment or therapy setting under the use of appropriately trained health care professionals who are responsible for ensuring the correct and accurate use of medical images.
The ViewStation System Hazard Analysis was performed to determine and evaluate all areas that represent potential safety or health hazards during ViewStation system operation. For all system hazards, effects, and mitigations have all been documented (SHA2102), reviewed, and implemented. This System Hazard Analysis is reviewed with every change and release of the product, including the above-mentioned changes. Validation and verification activities trace the hazard identification through evaluation, design, specification, implementation, and testing. The Design Review Team has reviewed the product change and System Hazard Analysis and has determined that the product change does not increase health or safety risk to patients, users, or other third parties.
#### Quality System
The fundamental goal of IMPAC's quality program is to provide value to customers and internal operations by producing better and safer products that are less expensive to build and maintain, simpler to use, and easier to support. IMPAC has implemented the IMPAC Ouality System to operate in a manner that has proven to be the most effective. Organizational experience and expertise is built into the management system to ensure that each process consistently meets defined specifications and continuously seeks improvement. All employees receive Quality System training and take pride in the value that they contribute to IMPAC products and to the final customer and their patients.
{3}------------------------------------------------
ViewStation was developed according to the IMPAC Software Design Control Procedure (SDCP). This procedure governs the process by which system and software development are to be planned, dested, and released.
| DOC. ID | TITLE |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 9001:2000 | Quality Management Systems - Requirements |
| ISO 13485:2003 | Medical Devices - Quality Management Systems - System Requirements for<br>Regulatory Purposes |
| ISO 14971:2000 | Medical Devices - Application of Risk Management to Medical Devices |
| EN 60601-1-4 :1996 | Medical Electrical Equipment - Part 1: General Requirements for Safety - 4.<br>Collateral Standard: Programmable Electrical Medical Systems |
| ISO/IEC 9003 :2004 | Software Engineering - Guidelines for the Application of ISO 9001:2000 to<br>Computer Software |
| 93/42/EEC | Medical Device Directive |
The IMPAC Quality System was developed in compliance with all of the following standards and regulations:
#### Verification and Validation Testing
A Traceability Matrix has been created, based upon the project plan, to ensure the completion of the specification, implementation, and testing of all requirements of the feature enhancement, including performance of full system hazard mitigation and basic operational testing. The System Test Plan for completing full application, integration, and system testing of ViewStation, while the Test Procedures capture the detailed testing parameters, results, and certification. A test certification statement certifies that the planned testing requirements were completed successfully. Design Reviews have been performed at the conclusion of each software design and development phase to review and validate the fulfillment of all of the phase requirements and deliverables. All of the above have been completed for ViewStation and representative documents are available for review.
#### Summary of Test Conclusions
IMPAC's Quality Engineering department has completed all product operation and hazard mitigation testing and has certified passing test results. Engineering was also performed to ensure that the algorithms and all other technical changes function exactly as intended. The testing demonstrated that the algorithms and all other functionality of ViewStation were successfully implemented.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## DEC 2 0 2004
Mr. Thomas H. Faris Vice President RA/QA, Management Representative IMPAC Medical Systems, Inc. 100 West Evelyn Avenue MOUNTAIN VIEW CA 94041 Re: K043412 Trade/Device Name: ViewStation, Image Processing System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: December 9, 2004 Received: December 10, 2004
Dear Mr. Faris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approvisions of the Act. The general controls provisions of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter witi anow you to oegh manceing of substantial equivalence of your device to a legally premarket nothreation. The I DA maing of batism for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice of your correst on our of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation on your responsibilities under the Act from the 807.97). You may obtain other general missistance at its toll-free number (800) Division of 301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsmaldsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# ViewStation Indications for Use Statement
510(k) Number (if known): _ic o4 3412
Device Name: ViewStation, Image Processing System
Indications for Use:
ViewStation supports image and information flow among health care facility personnel. ViewStation can be used whenever digital images and associated data are the means for communicating information. ViewStation is not intended for use in diagnosis.
The intended use of ViewStation is to provide health care facility personnel with an efficient and effective means to utilize patient images during the with an of therapy or treatment. ViewStation allows users to import, view, annotate, manipulate, enhance, manage, and archive patient images. The images and associated information are stored in a database, providing users access to the information necessary to perform their functions.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
(Posted July 1, 1998)
*Prescription Use*
(Division Sign-Off Division of Reproductive, A and Radiological Dev 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.