K043403 · Exactech, Inc. · LOD · Dec 28, 2004 · Orthopedic
Device Facts
Record ID
K043403
Device Name
CEMEX SYSTEM GENTA FAST BONE CEMENT
Applicant
Exactech, Inc.
Product Code
LOD · Orthopedic
Decision Date
Dec 28, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cemex System Genta Fast Bone Cement is indicated for the fixation of prostheses to living bone in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.
Device Story
Cemex System Genta Fast is a PMMA bone cement system for orthopedic revision surgery. Device consists of a powder component (PMMA, barium sulphate, benzoyl peroxide, gentamicin sulphate) and a liquid monomer component (methylmethacrylate, N-N dimethyl p-toluidine, hydroquinone). Operation involves breaking an internal glass ampoule to mix components within a closed manual device; mixing achieved by striking device against palm. Closed system minimizes volatile release and contamination. Once mixed, device attaches to an application gun to dispense semi-fluid cement as a syringe. Transparency allows visual inspection of mixing and application. Used by surgeons in clinical settings. Gentamicin release provides local antibiotic delivery. Benefits include secure prosthesis fixation and infection management in revision arthroplasty.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 5833 and ASTM 451-99 standards. Testing confirmed that the reduced powder-to-liquid ratio achieves a faster setting time while maintaining mechanical properties and gentamicin release rates comparable to the predicate device.
Technological Characteristics
PMMA bone cement. Powder: PMMA, barium sulphate, benzoyl peroxide, gentamicin sulphate. Liquid: methylmethacrylate, N-N dimethyl p-toluidine, hydroquinone. Powder-to-liquid ratio 2.4:1. Closed manual mixing/delivery system. Sterilization: Ethylene oxide (powder), membrane filtration (liquid). Complies with ISO 5833 and ASTM 451-99.
Indications for Use
Indicated for fixation of prostheses to living bone during second stage of two-stage revision total joint arthroplasty following clearance of initial infection. Contraindicated in primary orthopaedic procedures, patients with hypersensitivity to components (including Gentamicin Sulphate or aminoglycosides), active/incompletely treated infection at application site, or limb conditions where surgery is unjustifiable. Use caution in patients with myasthenia gravis or those receiving concurrent ototoxic drugs/loop diuretics.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
K043403.
2320 NW 66TH COURI GAINESVILIE, FL 32653
352-377-1140 FAX 352-378-2617
## Tecres Cemex® System Genta Fast bone cement Special 510(k)
DEC: 2 8 2004
# Summary of Safety and Effectiveness
| Trade Names: | Cemex System Genta Fast bone cement |
|----------------------|----------------------------------------------|
| Common Name: | Bone Cement |
| Classification Name: | Polymethylmethacrylate (PMMA)<br>Bone Cement |
:
### Legally Marketed Device for Substantial Equivalence Comparison:
| Model | Manufacturer | 510(k) Number |
|-----------------------------------|----------------|---------------|
| Cemex Genta<br>Cemex System Genta | Tecres, S.p.A. | #K033596 |
#### INDICATIONS FOR USE
Cemex System Genta Fast Bone Cement is indicated for the fixation of prostheses to living bone in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.
#### CONTRAINDICATIONS
Cemex System Genta Fast Bone Cement is contraindicated in primary orthopaedic musculoskeletal surgical procedures.
Cemex System Genta Fast Bone Cement is contraindicated in patients who are allergic or sensitive to any of its components, including Gentamicin Sulphate.
If the patient has a history of hypersensitivity or serious toxic reactions to aminoglycosides, the use of any other aminoglycosides may also be contraindicated due to the known cross-sensitivity of patients to drugs in this class.
Cemex System Genta Fast Bone Cement is contraindicated in the presence of active or incompletely treated infection, at the site where the bone cement is to be applied.
rev. 12/09/04
Section 4 Page 2 of 4
{1}------------------------------------------------
## Tecres Cemex® System Genta Fast bone cement Special 510(k)
352 377-1140 י FAX 352-378-26 ד
## Summary of Safety and Effectiveness
Cemex System Genta Fast Bone Cement is contraindicated where the loss of musculature or neuromuscular compromise in the affected limb would render the surgical procedure unjustifiable.
Cemex System Genta Fast Bone Cement must be considered carefully in the presence of mvastenia gravis.
There may be increased risk of oto:oxicity from gentamicin, if other ototoxic drugs such as cisplatin (an antineoplatic agent) and vancomycin (another antibiotic) are given at the same time. There also appears to be a synergistic effect of loop diuretics, such as furosemide or ethacrynic acid, and also loud noise, when combined with gentamicin.
### GENERAL DESCRIPTION - Substantial Equivalency Information
The individual chemical constituents in Cemex System Genta Fast bone cement are identical to those in the predicate Centa / Cemex System Genta bone cement (#K033596). The liquid component contains methylmethacrylate, N-N dimethyl p-toluidine, and hydroquinone. The dry powder component contains polymethylmethacrylate, barium sulphate, benzoyl peroxide and gentamicin sulphate.
The only difference is that the predicate Cemex Genta / Cemex Genta System has a 2.7:1 powder-to-liquid ratio compared to the 2.4:1 ratio the new Cemex System Genta Fast product. The powder-to-liquid ratio reduction results in a quicker setting time to accommodate various application techniques,
Performance testing shows that the proposed device is equivalent to the predicate and meet the requirements of ISO 5833 and ASTM 451-99. Performance testing shows that the rate of gentamicin release is within the safety range and comparable to the predicate Cemex Genta.
#### PACKAGING
Cemex System Genta Fast bone cement is packaged similarly to the predicate Cemex System Genta bone cement. The device is contained into the same double blister pack used for the predicate. The blisters are sealed with Tyvek® lids. The double blister pack is positioned inside an aluminum bag. The outer packaging is a heavy weight cardboard box.
### MIXING & APPLICATION
Use of Cemex System Genta Fast bone cement takes place in two consecutive stages. During the first stage, the glass ampoule containing the liquid monomer is broken to allow the component to flow into the chamber containing the powdered constituents. The device is then
rev. 12/09/04
Section 4 Page 3 of 4
{2}------------------------------------------------
## Tecres Cemex® System Genta Fast bone cement Special 510(k)
# Summary of Safety and Effectiveness
used as a closed manual mixing device. Mixing is accomplished by firmly striking the device against the palm of the hand and mating at each strike. Because no direct contact is made between the components and the use , volitile release into the local environment and possibility of contamination is minimized. In re second stage, the container is attached to an application gun device and used as a syringe while the cement is still in a semi-fluid state. The transparency of the mixing device provides for preliminary inspection of the suitability of the cement components and the visualization of he mixing and application stages as required by ISO 5833. Detailed instructions for use and pr >aution/warning information is outlined in the instruction leaflet provided with the product.
### STERILITY ASSURANCE
The powdered component is steriliz. d by ethylene oxide (EO) to a Sterility Assurance Level (SAL) of 10 °. The liquid component is sterilized by a membrane filtration technique to a SAL of 10-3.
rev. 12/09/04
Section 4 Page 4 of 4
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 8 2004
Ms. Lisa Simpson Sr. Regulatory Representative Exactech, Inc. 2320 NW 66th Court Gainesville, Florida 32653
Re: K043403
Trade Name: Cemex System Genta Fast Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Code: LOD Dated: December 16, 2004 Received: December 20, 2004
Dear Ms. Simpson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
b. Mark A. Melkman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Tecres Cemex® System Genta Fast Bone Cement Indications for Use
510(k) Number:
K043403
Device Names:
Cemex System Genta Fast Bone Cement
#### INDICATIONS FOR USE
Cemex System Genta Fast Bone Cement is indicated for the fixation of prostheses to living bone in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.
#### CONTRAINDICATIONS
Cemex System Genta Fast Bone Cement is contraindicated in primary orthopaedic musculoskeletal surgical procedures.
Cemex System Genta Fast Bone Cement is contraindicated in patients who are allergic or sensitive to any of its components, including Gentamicin Sulphate.
If the patient has a history of hypersensitivity or serious toxic reactions to aminoglycosides, the use of any other aminoglycosides may also be contraindicated due to the known cross-sensitivity of patients to drugs in this class.
Cemex System Genta Fast Bone Cement is contraindicated in the presence of active or incompletely treated infection, at the site where the bone cement is to be applied.
Cemex System Genta Fast Bone Cement is contraindicated where the loss of musculature or neuromuscular compromise in the affected limb would render the surgical procedure unjustifiable.
Cemex System Genta Fast Bone Cement must be considered carefully in the presence of myastenia gravis.
There may be increased risk of ototoxicity from gentamicin, if other ototoxic drugs such as cisplatin (an antineoplatic agent) and vancomycin (another antibiotic) are given at the same time. There also appears to be a synergistic effect of loop diuretics, such as furosemide or ethactynic acid, and also loud noise, when combined with gentamicin.
X Over the Counter Use Prescription Use Or
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
rev. 12/09/04
Page 1 of 1
510(k) Number K043403
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.