The PLANET 50 multiparameter Patient Monitoring system is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intrahospital transport along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3 lead), Plethysmograph Respiration and Capnography (CO2). It can also display the digital values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EtCO2 and FiCO2 readings. The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device, which can also be used as a portable device, permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.
Device Story
PLANET 50 is a multiparameter patient monitor; inputs include ECG (3-lead), respiration, capnography (CO2), plethysmography, NIBP, and temperature sensors. Device processes signals to display real-time waveforms and digital values for HR/PR, SpO2, RR, NIBP (systolic, diastolic, mean), EtCO2, and FiCO2. Features include graded color-coded alarms, 24-hour tabular/graphical trends (except NIBP), and 16-event alarm recall. Optional two-channel thermal array recorder provides hard-copy trends and waveforms. Used in clinical settings (bedside or intra-hospital transport) by professional healthcare providers. Output allows clinicians to assess patient status and adjust care based on monitored vital signs. Benefits include continuous patient surveillance and timely notification of physiological changes via adjustable alarm limits.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through comparison of technological characteristics, parameter range/accuracy, and adherence to international safety and EMC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-27).
Technological Characteristics
TFT color display; 3-lead ECG, respiration, capnography, NIBP, temperature, and pulse oximetry sensing. Powered by two sealed lead-acid batteries. Weight: 5kg. Includes optional two-channel thermal array recorder. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-27 (ECG).
Indications for Use
Indicated for monitoring vital signs in adult, pediatric, or neonatal patients at the bedside or during intra-hospital transport. Parameters include ECG (3-lead), respiration, capnography (CO2), plethysmography, HR/PR, SpO2, NIBP, and temperature. Not intended for home use.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Eagle 3000 patient Monitoring System (Marquette Electronic) (K952474)
{0}------------------------------------------------
FEB 1 5 2005
Image /page/0/Picture/1 description: The image shows a logo consisting of the letters 'L' and 'T' intertwined within a circle. The letters are stylized with sharp, angular lines, giving them a modern and geometric appearance. The logo is presented in black against a white background, creating a high contrast and making it easily recognizable. The overall design is simple yet distinctive, suggesting a corporate or brand identity.
## LARSEN & TOUBRO LIMITED
ELECTRICAL BUSINESS GROUP - ELECTRONIC PRODUCTS
:
:
.
Mysore Works, KIADB Industrial Area, Hebbal-Hootagalii, Mysore - 571 186 · Tel : (91) - 821 - 402561 · Fax : (91) - 821 - 402468
Ref :
E-Mall :
.
25th April, 2004 Page : 01 of 04
### 510(K) SUMMARY
(Per section 807.92 O)
| CONTACT DATA | | | |
|-------------------------------|--------------------------------------------------------------------------------|-------|--------------------------|
| Submitter's Name | Larsen & Toubro Limited | | |
| Address | KIADB Industrial Area, Hebbal Hootagalli,<br>Mysore – 570018, Karnataka, INDIA | | |
| Telephone | 91-821-2402561 | Fax | 91-821-2402468 |
| Contact Person | A.B.Deshpande | Title | Head – Quality Assurance |
| E-Mail address | DeshpandeAB@myw.ltindia.com | | |
| Date the summary was prepared | April 25th,2004 | | |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a black and white logo of Larsen & Toubro (L&T). The logo consists of the letters 'L' and 'T' interlocked within a circle. The 'L' is on the left side, and the 'T' is on the right side, with both letters sharing a common horizontal line. The logo is simple and modern, representing the company's brand identity.
KC43370/S1
P 2/4
# LARSEN & TOUBRO LIMITED
ELECTRICAL BUSINESS GROUP - ELECTRONIC PRODUCTS
Mysore Works, KIADB Industrial Area, Hebbal-Hootagalli, Mysore - 571 186 • Tel : (91) - 821 - 402561 • Fax : (91) - 821 - 402468
Ref :
.
.
:
E-Mail :
25th April, 2004 Page : 02 of 04
| DEVICE | |
|---------------------|---------------------------|
| Trade name | PLANET 50 |
| Common name | Patient Monitoring System |
| Classification name | Vital Signs Monitor |
| PREDICATE DEVICE IDENTIFICATION | | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------|-----|
| CFR21 Section | Product code (optional) | MWI |
| 870.2300 | | |
| Classification panel | Cardiovascular | |
| Device Class | Class II | |
| Legally marketed Comparison Device / K# | Eagle 3000 patient Monitoring System (Marquette Electronic) / K952474 Vital signs monitor Model 8100 (CSI) / K001020 | |
{2}------------------------------------------------
K04337c/51
\$\mathbb{P}^{3/4}\$
Image /page/2/Picture/2 description: The image shows a logo consisting of the letters "L" and "T" intertwined within a circle. The "L" is positioned on the upper left, while the "T" is on the lower right. The letters are bold and connected, creating a unified design within the circular border.
- **ELECTRICAL BUSINESS GROUP - ELECTRONIC PRODUCTS** -
Mysore Works, KIADB Industrial Aree, Hebbal-Hootagalli, Mysore - 571 186 • Tel : (91) - 821 - 402561 • Fax : (91) - 821 - 402462
Ref :
E-Mail :
2500 April, 2004 Page : 03 of 04
### DEVICE DESCRIPTION
This PLANET 50 unit is a multiparameter Patient monitor System (TFT color monitor) with ECG(3 lead), Respiration, Capnography, Temperature, NIBP and Pulse oximetry. with an optional inbuilt two channel thermal array recorder for printing of Tabular trends & waveforms.
PLANET 50 is a three channel monitor with waveform display capability for ECG (3 lead), Respiration, Capnography (CO2) and Plethysmograph. It also displays the digital values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EtCO2 and FiCO2 readings. It has graded and color coded alarms It has 24 hours tabular and graphical trends for all parameters except NIBP. It has special tabular trend for NIBP to store the last 240 readings. Alarm recall feature offers last 16 alarm conditions.
### INTENDED USE OF THE DEVICE
The PLANET 50 multiparameter Patient Monitoring system is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intrahospital transport along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3 lead), Plethysmograph Respiration and Capnography (CO2). It can also display the digital values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EtCO2 and FiCO2 readings.
The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device, which can also be used as a portable device, permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.
Regd. Off : L & T House, Ballard Estate, P. O. Box 278, Mumbai 400 001 ● Phone : 261 8181 / 82 · Fax : 91 - 22 · 262 0222
{3}------------------------------------------------
E-Mage : 04 Of 04
Image /page/3/Picture/1 description: The image shows a logo consisting of the letters 'L' and 'T' inside of a circle. The letters are stylized and appear to be connected. The logo is black and white and has a simple, clean design.
### BRO LI
- **ELECTRICAL BUSINESS GROUP - ELECTRONIC PRODUCTS** -
Mysore Works, KIADB Industrial Area, Hebbal-Hootagalli, Mysore - 571 186 • Tel : (91) - 821 - 402561 • Fax : (91) - 821 - 402468 2500 April, 2004
Ref :
### TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
Device : Larsen & Toubro limited make PLANET 50 Patient Monitoring System.
Predicate device :
Eagle 3000 patient Monitoring System (Marquette Electronic), K# K952474 Vital signs monitor Model 8100 (CSI), K# K001020
The parameters available with the Larsen & Toubro Limited make PLANET 50 Patient monitoring system (ECG-3 lead, Respiration, Temperature, NIBP, Capnography and Pulse oximetry) are also available with these predicate devices. The range and accuracy of the parameters & method of sensing are similar to the predicate devices. In PLANET 50 monitor audible & visual alarms are provided similar to those in the Predicate devices.
PLANET 50 has got TFT color display like CSI Model 8100. PLANET 50 has got thermal array recorder similar to that available in Marquette Eagle 3000. Weight of the PLANET 50 (5Kg) is less than that of the predicate devices. Battery (2 sealed lead acid) is provided in PLANET 50 monitor like that of the predicate device CSI Model 8100.
Comparison of all the parameters of PLANET 50 to that of the predicate devices is given in the "Substantial Equivalence Equipment comparison" document.
### Compliance to standards :
The following international standards are referred. IEC 60601-1 Medical Electrical safety IEC 60601-1-2 EMC compliance IEC 60601-2-27 ECG safety
### Conclusion :
Based on the Technological characteristics of PLANET50 and its comparison with those of a predicate device CSI Model 8100 and Marquettee Eagle 3000 monitors, Larsen & Toubro Limited believes that their device is substantially equivalent to these Monitors and doesn't pose any additional risk on safety & effectiveness of the device.
(Mohan G.R.)
Head – Design & Development
Regd. Off : L & T House, Ballard Estate, P. O. Box 278, Mumbai 400 001 • Phone : 261 8181 / 82 • Fax : 91 - 22 - 262 0223
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 5 2005
Larsen & Toubro LTD c/o Mr. Ned Devine Entela, Inc. 3033 Madison Ave. SE Grand Rapids, MI 49548
Re: K043370
Trade Name: PLANET 50 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: II (two) Product Code: MWI Dated: December 07, 2004 Received: December 08, 2004
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bection b re(s) perceived is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated in the encreativent of the enactment date of the Medical Device Amendments, or to commerce prof to May 20, 1978, as secordance with the provisions of the Federal Food, DNIg. devices mat have been recuire approval of a premarket approval application (PMA). and Cosmetic Ac. (11ct) that to not requently of the general controls provisions of the Act. The I ou may, therefore, manel the Act include requirements for annual registration, listing of general controls provisions and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (soc a c roy ols. Existing major regulations affecting your device can may oc subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Ned Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Blummenfor
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### K043370/51
### Indications for Use
510(k) Number (if known): ' Ko43370
Device Name: PLANET 50
Indications for Use:
The PLANET 50 multi-parameter Patient Monitoring system is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport along with the appropriate accessories mentioned/supplied with the unit. Vital signs parameters include ECG (3 lead), Plethysmograph Respiration and Capnography (CO2). It can also display the digital values of HR/PR, SpO2, non-invasive blood pressure ( systolic, diastolic and mean) Temperature, EtCO2 and FiCO2 readings.
The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device, which can also be used as a portable device, permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.
Blummen
(Division Sign-Cit) Division of Cardiovascular Devices 510(k) Number_1043
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.