OMEGA EXCEL/XP LASER SYSTEM

K043353 · Advanced Medical Technologies, Inc. · NHN · Jul 1, 2005 · Physical Medicine

Device Facts

Record IDK043353
Device NameOMEGA EXCEL/XP LASER SYSTEM
ApplicantAdvanced Medical Technologies, Inc.
Product CodeNHN · Physical Medicine
Decision DateJul 1, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Omega Xp Laser System is intended for use as an adjunctive treatment modality to provide temporary relief of pain associated with rotator cuff tendonitis.

Device Story

The Omega Xp Laser System is a non-invasive, portable therapeutic laser device. It consists of a base unit housing electronics and controls, connected via umbilical cord to laser probes containing laser diodes. The system operates on AC or battery power. It delivers infrared light energy to target tissues to provide temporary pain relief for rotator cuff tendonitis. The device is intended for use as an adjunctive treatment modality. Healthcare providers apply the laser probes to designated treatment areas. The system is designed to comply with 21 CFR 1010 and 1040 performance standards.

Clinical Evidence

Evidence includes functional performance testing, electrical safety testing, and a randomized, blinded clinical trial. The device complies with ISO 9000:2000, EN46001, and Directive 89/336 (electromagnetic compatibility).

Technological Characteristics

Infrared lamp (21 CFR 890.5500); utilizes infrared diodes for light energy delivery. Portable base unit with umbilical-connected laser probes. Powered by AC or battery. Complies with ISO 9000:2000, EN46001, and electromagnetic compatibility standards. Designed to meet 21 CFR 1010 and 1040 performance standards.

Indications for Use

Indicated for patients with rotator cuff tendonitis requiring adjunctive treatment for temporary pain relief.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 1 - 2005 # APPENDIX D ### SUMMARY OF SAFETY AND EFFICACY # Advanced Medical Technologies, Inc. Omega Xp Therapeutic Laser System (as per 21 CFR Part 807.92) November 11, 2004 #### GENERAL INFORMATION I. Device Generic Name: Low Level Laser Therapy (LLLT) Therapeutic Laser Trade Name: Omega XP Laser System Device Classification: Class II, Performance Standards 21CFR Part 890.5500 - Infrared Lamp, Non-heating NHN Product Code: Advanced Medical Technologies Applicant Name and Address: 101 Waterside Drive Centerville, MA 02632 508 / 790-9300 Nell Camera, President lasertherapeutics@comcast.net 510(k) Number: K043353 #### DEVICE DESCRIPTION II. The Omega Xp Laser System is a non-invasive, portable therapeutic medical laser designed to deliver light energy to the target tissue. The System operates by either AC or battery power and can be used with a variety of laser probes. The Omega Xp Laser System is intended to emit energy in the infrared spectrum to provide temporary relief of pain associated with rotator cuff tendonitis. The Omega Xp Laser System is comprised of a Base Unit and select Laser Probes. The Base Unit houses the electronics, circuits and controls to power the Laser Probes. The Laser Probes are connected to the Base Unit by an umbilical cord. The Laser Probes house the laser diode and circuitry to deliver the light energy to the designated treatment areas. {1}------------------------------------------------ K04333)3 #### III. INDICATIONS FOR USE The Omega Xp Laser System is indicated for use as an adjunctive treatment modality to provide temporary relief of pain associated with rotator cuff tendonitis. #### IV. Predicate Devices Predicate devices to the Omega Xp Laser System are the Photo Thera, Inc. Acculaser Pro 4 (K023060) and the Stargate International, Inc. Excalibur Light Therapy System (K041530). ### V. Summary of the Technical Characteristics of the Omega XP Laser System as Related to the Referenced Predicate Devices. The Omega System and the aforementioned predicate devices are infrared lamps as defined in 21 CFR 890.5500. These devices utilize infrared diodes to emit infrared light energy to human tissues. The Omega System and the named predicate devices have the same intended uses and similar technical and performance characteristics. #### VI. Testing Testing of the System includes functional performance testing and electrical safety testing. The Omega Xo Laser System is manufactured to comply with the following international standards: ISO 9000:2000 EN46001 Directive 89/336 regarding electromagnetic compatibility #### VII. Conclusions Pursuant to the testing and comparison to the predicate devices, the Omega XP Laser has the same intended uses, with similar functional and performance characteristics. The System is designed to comply with applicable performance standards promulgated by Federal Food and Drug Administration, such as 21 CFR 1010 and 1040. Further evidence of the safety and efficacy of the Omega System was assessed by the randomized, blinded clinical trial conducted. The Omega Xp Laser System performs as intended and do not raise any new safety or efficacy issues. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an eagle or bird with three curved lines forming its body and wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 - 2005 Advanced Medical Technologies Incorporated C/o Ms. Joyce Heinrich Regulatory Consultant Texas Applied Biomedical Services Incorporated 12101 Cullen Boulevard # A Houston, Texas 77047 Re: K043353 Trade/Device Name: Omega Xp laser System Regulation Number: 21 CFR 890.5500 Regulation Name: Lamp, Non-heating for adjunction use in pain therapy Regulatory Class: II Product Code: NHN Dated: March 31, 2005 Received: April 4, 2005 Dear Ms. Heinrich: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 -- Ms. Joyce Heinrich This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. # # Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # APPENDIX A ## Indications for Use Statement 510(k) Number (if known): K043353 # Device Name: Advanced Medical Technologies, Inc. Omega Xp Laser System ### Indications for Use: The Omega Xp Laser System is intended for use as an adjunctive treatment modality to provide temporary relief of pain associated with rotator cuff tendonitis. AND/OR Over the Counter Use: Prescription Use: X X (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODDE) Zele ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------... | Division of Careera, Restorative ond Neurological Devices 3553 - (1-)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...