K043350 · Apollo Medical Imaging Technology Pty. , Ltd. · LLZ · Dec 15, 2004 · Radiology
Device Facts
Record ID
K043350
Device Name
MISTAR, MODEL V3.2
Applicant
Apollo Medical Imaging Technology Pty. , Ltd.
Product Code
LLZ · Radiology
Decision Date
Dec 15, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
MIStar is a software package that provides manipulation, visualization and processing of medical images in a diagnostic imaging setting. It allows transfer of DICOM 3.0 images over a medical imaging network, and can receive digital images and data from various sources (including but not limited to CT, MR, NM, and PT). MIStar supports the analysis of dynamic CT and MR images acquired during and after the injection of a compact bolus of contrast media. It allows visual inspection of time intensity curves and calculation of parametric parameters (i.e., area under curve, time to peak, maximum slope of enhancement, etc). It also calculates various blood perfusion related parameters such as blood flow (BF), blood volume (BV), mean transit time (MTT) and capillary permeability from dynamic CT images. This software provides supplemental information to those images extracted from CT and MR temporal datasets and will aid in the assessment of the extent and type of perfusion, blood volume and capillary permeability changes related but not limited to stroke or tumor angiogenesis and be helpful in therapy monitoring. MIStar is intended to be used by trained medical professionals, including but not limited to licensed radiologists, technologists and clinicians, and for the rendering clinical diagnosis.
Device Story
MIStar is a standalone software package running on Intel-based PCs; processes DICOM 3.0 medical images from CT, MR, NM, and PT scanners. Input: dynamic CT/MR image datasets acquired post-contrast bolus injection. Processing: calculates time-intensity curves and perfusion parameters (blood flow, blood volume, mean transit time, capillary permeability, area under curve, time to peak, maximum slope of enhancement). Output: parametric images and quantitative data displayed in graphic format. Used in diagnostic imaging settings by radiologists, technologists, and clinicians. Provides supplemental information to assist in clinical diagnosis, assessment of stroke/tumor angiogenesis, and therapy monitoring. Requires physician intervention for parameter setting and image assessment. Does not contact patients or control image acquisition hardware.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, software verification, and validation procedures.
Technological Characteristics
Standalone software package; operates on Intel-based PC platform. Supports DICOM 3.0 standard. Performs image post-processing and perfusion parameter calculation. No patient contact; no hardware control. Software-based analysis.
Indications for Use
Indicated for trained medical professionals (radiologists, technologists, clinicians) to manipulate, visualize, and process medical images (CT, MR, NM, PT) in diagnostic settings. Supports analysis of dynamic CT/MR images post-contrast bolus injection to assess perfusion, blood volume, and capillary permeability changes related to conditions such as stroke or tumor angiogenesis and therapy monitoring.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K043350
| Submitter: | Apollo Medical Imaging Technology Pty. Ltd<br>26 Hood Street, North Balwyn<br>Melbourne, Vic 3104, Australia<br>Phone: 61-3-9859 9316<br>Fax: 61-3-9859 5701 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | September 30, 2004 |
| Contact Person: | Qing Yang |
| Device Trade Name: | MIStar |
| Device Common Name: | Picture Archiving Communications System (PACS) |
| Regulation Number: | 892.2050 |
| Device Classification: | Class II |
| Classification Name: | Image Processing System |
#### Predicate Device:
| Manufacturer | Product | 510(k) Number | Clearance Date |
|---------------------|----------------------------------------|---------------|----------------|
| Efilm Medical, Inc. | Efilm Workstation with Modules | K020995 | 04/12/2002 |
| General Electric | CT Perfusion 2 | K010042 | 01/30/2001 |
| GE Medical Systems | Advantage Windows with Functool Option | K960265 | 07/03/1996 |
### Device Description
MIStar is a software package, which runs on an Intel-based PC platform. It allows manipulation, visualization and processing of medical images acquired with various clinical scanners and stored in DICOM and for other proprietary formats. MIStar allows transfer of DICOM 3.0 images over a medical imaging network.
MIStar also provides post-processing of dynamic CT and MR images acquired during and after the injection of a compact bolus of contrast media, where the contrast media acts as a pure intravascular tracer. It allows visual inspection of time intensity curves and calculation of parameters (i.e., area under curve, time to peak, maximum slope of enhancement, etc). It also allows calculation of various perfusion related parameters (i.e. regional blood flow, regional blood volume, mean transit time and capillary permeability) from dynamic CT data. The results are displayed in a user-friendly graphic format as parametric images that provide supplementary information for diagnosis purposes.
{1}------------------------------------------------
#### Indications for Use
MIStar is a software package that provides manipulation, visualization and processing of medical images in a diagnostic imaging setting. It allows transfer of DICOM 3.0 images over a medical imaging network, and can receive digital images and data from various sources (including but not limited to CT, MR, NM, and PT).
MIStar supports the analysis of dynamic CT and MR images acquired during and after the injection of a compact bolus of contrast media. It allows visual inspection of time intensity curves and calculation of parametric parameters (i.e., area under curve, time to peak, maximum slope of enhancement, etc). It also calculates various blood perfusion related parameters such as blood flow (BF), blood volume (BV), mean transit time (MTT) and capillary permeability from dynamic CT images. This software provides supplemental information to those images extracted from CT and MR temporal datasets and will aid in the assessment of the extent and type of perfusion, blood volume and capillary permeability changes related but not limited to stroke or tumor angiogenesis and be helpful in therapy monitoring.
MIStar is intended to be used by trained medical professionals, including but not limited to licensed radiologists, technologists and clinicians, and for the rendering clinical diagnosis.
#### Conformance to Standard
MIStar provides DICOM functionality in conformance to the ACR-NEMA DICOM 3.0 standard as documented in the MIStar DICOM 3.0 Conformance Statement.
#### Software Development and Testing
The MIStar software package was designed, developed, verified and validated according to the procedures described in the Software Information section: Software Development Process.
#### Comparison with Predicate
MIStar is a stand-alone software package that operates on an Intel-based PC platform that satisfies the minimum system requirements. MIStar provide digital image processing and measurement capability. It can also communicate DICOM images with remote viewing stations or server systems over a medical imaging network.
MIStar also provides image post-processing as such does not have contact with the patient, nor does it control any devices that acquire medical images or sustain life. The algorithms used to calculate parametric parameters are similar to CT perfusion2 and Functool. It further requires the intervention of a physician in setting necessary parameters and in assessing resultant images.
The functional features of this package are substantially equivalent to that of eFilm Workstation with Modules (K020995), CT Perfusion 2 (K010042) and Advantage Windows with Functool option (K960265). The use of M1Star does not result in any additional potential hazards when compared to predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes extending from its wing, symbolizing health, hope, and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 5 2004
Apollo Medical Imaging Technology Pty Ltd. % Ms. Laura Danielson Responsible Third Party Official 510(k) Program Manager TÜV Product Service 1775 Old Highway 8 NW, Suite 104 NEW BRIGHTON MN 55112-1891 Re: K043350 Trade/Device Name: MIStar Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 12, 2004 Received: December 6, 2004
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number: _Koy3350
Device Name: MIStar
Indications For Use:
MIStar is a software package that provides manipulation, visualization and processing of medical images in a diagnostic imaging setting. It allows transfer of DICOM 3.0 images over a medical images in a diagnostic maging couligital images and data from various sources (including but not limited to CT, MR, NM, and PT).
MIStar supports the analysis of dynamic CT and MR images acquired during and after the injection of a compact bolus of contrast media. It allows visual inspection of time intensity curves and calculation a Computer of contines freana under curve, time to peak, maximum slope of enhancement, etc). It also calculates various blood perfusion related parameters such as blood flow (BF), blood volume (BV), mean transit time (MTT) and capillary permeability from dynamic CT images. This software provides supplemental information to those images extracted from CT and MR temporal datasets and will aid in the assessment of the extent and type of perfusion, blood volume and capillary permeability whates related but not limited to stroke or tumor angiogenesis and be helpful in therapy monitoring.
MIStar is intended to be used by trained medical professionals, including but not limited to licensed radiologists, technologists and clinicians, and for the rendering clinical diagnosis.
Prescription Use (Part 21 CFR 801 Subpart D) ATTB/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Lyonn
(Division Sian-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.