The ONLINE DAT Benzodiazepines Plus is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.
Device Story
In vitro diagnostic assay for benzodiazepine detection in human urine; utilizes Kinetic Interaction of Microparticles in Solution (KIMS) technology. Input: human urine sample; Principle: competitive immunoassay where drug in sample competes with drug-conjugated microparticles for sheep polyclonal antibody binding sites; Output: change in light transmission/absorbance measured by automated clinical chemistry analyzers. Absence of drug leads to particle aggregation and increased absorbance; presence of drug inhibits aggregation, decreasing absorbance relative to cutoff. Used in clinical laboratories by technicians/pathologists to assess benzodiazepine presence. Results assist healthcare providers in clinical decision-making regarding drug screening. Benefits include rapid, automated, semi-quantitative assessment of drug presence.
Clinical Evidence
No clinical data provided; substantial equivalence based on analytical performance characteristics and technological comparison to predicate.
Technological Characteristics
KIMS (Kinetic Interaction of Microparticles in a Solution) immunoassay. Reagents: monoclonal/polyclonal antibodies, substrate, enzyme-labeled drug conjugate. Automated clinical chemistry analyzer platform (Hitachi 917, 911, 912, Modular P/DAT). Liquid reagents stored at 2-8°C. Qualitative/semi-quantitative output.
Indications for Use
Indicated for qualitative and semi-quantitative detection of benzodiazepines in human urine for individuals requiring drug testing. Used on automated clinical chemistry analyzers at 100, 200, and 300 ng/ml cutoffs.
Regulatory Classification
Identification
A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K043327
B. Purpose for Submission:
New submission
C. Measurand:
Benzodiazepine
D. Type of Test:
Qualitative/Semi-Quantitative/ Immunoassay
E. Applicant:
Roche Diagnostics
F. Proprietary and Established Names:
ONLINE DAT Benzodiazepines Plus
G. Regulatory Information:
1. Regulation section:
21 CFR section 862.3170 - Enzyme Immunoassay, Benzodiazepine
2. Classification: Class II
3. Product code: JXM
4. Panel: 91 (Toxicology)
{1}
H. Intended Use:
1. Intended use(s):
The ONLINE DAT Benzodiazepines Plus is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.
Benzodiazepines: 100 Cutoff: 2.1 ng/mL nordiazepam
200 Cutoff: 3.8 ng/mL nordiazepam
300 Cutoff 7.6 ng/mL nordiazepam
2. Indication(s) for use:
The ONLINE DAT Benzodiazepines Plus is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.
3. Special conditions for use statement(s):
The ONLINE DAT Benzodiazepines Plus Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/Mass spectrometry is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
The assay is for Prescription Use.
Semi-quantitative results may be helpful in estimating the concentrations of drug(s) in samples. This can aid users when they are preparing dilutions of the samples for further analysis.
4. Special instrument requirements:
Performance was demonstrated in this submission on the Roche Hitachi 917 analyzer. In addition to the Hitachi 917 reagent application included in this
{2}
filing, Roche Diagnostics plans to apply the ONLINE DAT Benzodiazepines Plus reagent to the following Hitachi analyzers:
* Hitachi 911 - A II level application
* Hitachi 912 - A I level application
* Modular P - A II level application
* Modular DAT - A I level application
Reagent performance validations have been performed with the ONLINE DAT Benzodiazepines Plus reagent in accordance with Roche Diagnostics minimum requirements application validation protocol. Master application level work as described in this submission was completed on the Hitachi 917 analyzer for the Benzodiazepines Plus assay, following FDA's Replacement Reagent and Instrument Family Policy guidance and Roches internal application protocol. The Benzodiazepines Plus assay was validated on other members of the Hitachi family with successful completion of A I or A II level application works.
# I. Device Description:
The device consists of two wet reagents which contain the key components of the immunoassay; monoclonal/ polyclonal antibody against the drug, substrate, and enzyme-labeled drug (conjugate).
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Abuscreen OnLine Benzodiazepines Assay
2. Predicate $510(\mathbf{k})$ number(s):
k983702
3. Comparison with predicate:
The following table illustrates the similarities and differences between the Roche ONLINE DAT Benzodiazepines Plus assay and the predicate device.
| Feature | Roche Abuscreen OnLine Benzodiazepines (Predicate) | Roche ONLINE DAT Benzodiazepines Plus |
| --- | --- | --- |
| Methodology | KIMS, Kinetic interaction of microparticles in a solution | KIMS, Kinetic interaction of microparticles in a solution |
| Reagents | Antibody Working Solution: Benzodiazepines polyclonal antibody (sheep) in buffer with bovine serum albumin, and | Antibody Working Solution: Benzodiazepines polyclonal antibody (sheep) in buffer with bovine serum albumin, and 0.09% sodium azide |
{3}
| Feature | Roche Abuscreen OnLine Benzodiazepines (Predicate) | Roche ONLINE DAT Benzodiazepines Plus |
| --- | --- | --- |
| | preservative. | Microparticle Working Solution: Microparticles coated with conjugated benzodiazepine derivatives in a buffer and preservative. |
| Sample Type | Urine | Urine |
| Controls | Abuscreen OnLine Positive Control and Negative Control, and Abuscreen OnLine Benzodiazepine Special Controls | Control Sets DAT I, DAT II and DAT III |
| Calibrators | Abuscreen OnLine Preciset DAT I and DAT II and Abuscreen OnLine Cfas DAT Qualitative calibrator sets. | Preciset DAT Plus I and Plus II, Cfas DAT Qualitative Plus and Cfas DAT Qualitative Clinical calibrator set. |
| Calibrator Levels ng//mL | 100 ng/mL Qualitative (1 point calibration): Abuscreen OnLine Preciset DAT I, Cal 3 or Cfas DAT Qualitative: 100 ng/mL
100 ng/mL Semiquantitative: A.O. Preciset DAT I, Cal 1,0 ng/mL
A.O. Preciset DAT I, Cal 2, 50 ng/mL
A.O. Preciset DAT I, Cal 3, 100 ng/mL
A.O. Preciset DAT I, Cal 4, 200 ng/mL
200 ng/mL Qualitative (1 point calibration): Abuscreen OnLine Preciset DAT I, Cal 4: 200 ng/mL
200 ng/mL Semiquantitative: A.O. Preciset DAT I, Cal 1,0 ng/mL
A.O. Preciset DAT I, Cal 3, 100 ng/mL
A.O. Preciset DAT I, Cal 4, 200 ng/mL
A.O. Preciset DAT I, Cal 7, 400 ng/mL
300 ng/mL Qualitative: (1 point calibration): Abuscreen OnLine Preciset DAT II, Cal 5: 300 ng/mL
300 ng/mL Semiquantitative: A.O. Preciset DAT I, Cal 1,0 ng/mL
A.O. Preciset DAT I, Cal 6, 150 ng/mL
A.O. Preciset DAT I, Cal 5, 300 ng/mL
A.O. Preciset DAT I, Cal 8, 600 ng/mL | 100 ng/mL Qualitative (1 point calibration): Preciset DAT II, Cal 3, or Cfas DAT Qualitative Clinical, Cal 2: 100 ng/mL
100 ng/mL Semiquantitative: Preciset DAT Plus II, Cal 1, 0 ng/mL
Preciset DAT Plus II, Cal 2, 50 ng/mL
Preciset DAT Plus II, Cal 3, 100 ng/mL
Preciset DAT Plus II, Cal 4, 200 ng/mL
200 ng/mL Qualitative: (1 point calibration): Preciset DAT Plus II, Cal 4: 200 ng/mL
200 ng/mL Semiquantitative: Preciset DAT Plus II, Cal 1, 0 ng/mL
Preciset DAT Plus II, Cal 3, 100 ng/mL
Preciset DAT Plus II, Cal 4, 200 ng/mL
Preciset DAT Plus II, Cal 5, 400 ng/mL
300 ng/mL Qualitative: (1 point calibration): Preciset DAT Plus I, Cal 3, or Cfas DAT Qualitative Plus, or Cfas DAT Qualitative Clinical, Cal 1: 300 ng/mL
300ng/mL Semiquantitative: Preciset DAT Plus II, Cal 1, 0 ng/mL
Preciset DAT Plus II, Cal 2, 150 ng/mL
Preciset DAT Plus II, Cal 3, 300 ng/mL
Preciset DAT Plus II, Cal 4, 600 ng/mL |
| Sensitivity | 100 Cutoff: 17 ng/mL nordiazepam | 100 Cutoff: 2.1 ng/mL nordiazepam |
{4}
| Feature | Roche Abuscreen OnLine Benzodiazepines (Predicate) | | | | | Roche ONLINE DAT Benzodiazepines Plus | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| (LOD) Limit of Detection | 200 Cutoff: 40 ng/mL nordiazepam
300 Cutoff: 36 ng/mL nordiazepam | | | | | 200 Cutoff: 3.8 ng/mL nordiazepam
300 Cutoff: 7.6 ng/mL nordiazepam | | | | |
| Method comparison, 100 ng/mL Cutoff | OnLine Benzodiazepines
+ 50
- 0 | | | | | Benzodiazepines
Plus
+ 82
- 10
GC/MS
+ 82
- 10 | | | | |
| Method Comparison, 200 ng/mL Cutoff | OnLine Benzodiazepines
+ 50
- 0 | | | | | Benzodiazepines
Plus
+ 76
- 12
GC/MS
+ 76
- 12 | | | | |
| Method Comparison, 300 ng/mL Cutoff | OnLine Benzodiazepines
+ 50
- 0 | | | | | Benzodiazepines
Plus
+ 72
- 10
GC/MS
+ 72
- 10 | | | | |
| SQ Precision | 100 ng/mL | | | | | 100 ng/mL | | | | |
| Within Run | 50 | 75 | 100 | 125 | 150 | 200 | 50 | 75 | 100 | 125 |
| Target: Mean (ng/mL) CV% | 51 | 75 | 102 | 130 | 148 | 202 | 47.5 | 75.2 | 99.9 | 130.7 |
| Day to day | 50 | 75 | 100 | 125 | 150 | 200 | 50 | 75 | 100 | 125 |
| Target: Mean (ng/mL) CV% | 52 | 76 | 101 | 128 | 152 | 202 | 47.3 | 76.9 | 100.9 | 131.6 |
| SQ Precision | 200 ng/mL | | | | | 200 ng/mL | | | | |
| Within Run | 100 | 150 | 200 | 250 | 300 | 100 | 150 | 200 | 250 | 300 |
| Target: Mean (ng/mL) CV% | 96 | 141 | 200 | 251 | 315 | 97.7 | 154.4 | 197.2 | 258.2 | 314.4 |
| Day to day | 100 | 150 | 200 | 250 | 300 | 100 | 150 | 200 | 250 | 300 |
| Target: Mean (ng/mL) CV% | 100 | 145 | 205 | 258 | 320 | 98.8 | 155.3 | 197.3 | 259.1 | 313.8 |
| SQ Precision | 300 ng/mL | | | | | 300 ng/mL | | | | |
| Within Run | 150 | 225 | 300 | 375 | 450 | 150 | 225 | 300 | 375 | 450 |
| Target: Mean (ng/mL) CV% | 152 | 224 | 301 | 388 | 465 | 154.5 | 227.7 | 305.4 | 384 | 434.8 |
| Day to day | 150 | 225 | 300 | 375 | 450 | 150 | 225 | 300 | 375 | 450 |
| Target: Mean (ng/mL) CV% | 154 | 228 | 306 | 394 | 470 | 153.1 | 223.8 | 301.2 | 377.3 | 438.3 |
| Indications for use | The Roche Abuscreen OnLine assay for Benzodiazepines is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL, 200 ng/mL and 300 ng/mL. Semi-quantitative test results may be obtained that permit | | | | | Benzodiazepines Plus is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL, 200 ng/mL and 300 ng/mL. Semi-quantitative test results may be obtained that permit | | | | |
{5}
| Feature | Roche Abuscreen OnLine Benzodiazepines (Predicate) | Roche ONLINE DAT Benzodiazepines Plus |
| --- | --- | --- |
| Indications for use (cont.) | ng/mL. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. | laboratories to assess assay performance as part of a quality control program. |
| Qualitative and Semi-Quantitative cutoffs | 100, 200, and 300 ng/mL | 100, 200, and 300 ng/mL |
K. Standard/Guidance Document Referenced (if applicable):
FDA’s Replacement Reagent and Instrument Family Policy Guidance.
L. Test Principle:
The test is an enzyme immunoassay for use on automated clinical chemistry analyzers. Calibrators, ranging in concentration from 0 to 600 ng/mL, are run with the assay. Enzyme-labeled drug and drug present in the sample compete for limited antibody binding sites. Binding of the enzyme-labeled drug inhibits its reaction with the substrate, thereby influencing the rate of absorbance change measured by the instrument. The rate of absorbance change is proportional to the concentration of drug in the sample. Concentrations of controls and unknowns are calculated from the standard curve.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
(All performance was established on the Roche Hitachi 917 analyzer unless otherwise noted.)
The imprecision of the ONLINE DAT Benzodiazepines Plus assay at the 100, 200 and 300 ng/mL cutoffs was determined by running a series of calibrators and controls in replicates of 20 per day for five days on a Hitachi 917 analyzer. The following results were obtained:
Acceptance criteria: 100 ng/mL Cutoff
Within-run imprecision ≤ 8% CV
(≤ 10% CV for 50 & 150 ng/mL)
Run-to-run Imprecision ≤ 10% CV
(≤ 12% CV for 50 & 150 ng/mL)
Near cutoff Imprecision ≤ 5% Crossovers
{6}
Within-run Imprecision (N = 20), 100 ng/mL cutoff
| | Semi-quantitative | | Qualitative | |
| --- | --- | --- | --- | --- |
| Cals)/ Controls (ng/mL) | Mean (ng/mL) | CV% | Mean (mAbs) | CV% |
| .50 X (50) | 47.5 | 2.1% | 575.4 | 0.6% |
| .75 X (75) | 75.2 | 1.3% | 473.8 | 0.6% |
| Cutoff (100) | 99.9 | 1.5% | 400.2 | 1.0% |
| 1.25 X (125) | 130.7 | 0.7% | 312.8 | 1.0% |
| 1.50 X (150) | 157.8 | 0.7% | 249.1 | 1.1% |
Day-to-day Imprecision (N = 100), 100 ng/mL cutoff
| | Semi-quantitative | | Qualitative | |
| --- | --- | --- | --- | --- |
| Cals)/ Controls (ng/mL) | Mean (ng/mL) | CV% | Mean (mAbs) | CV% |
| .50 X (50) | 47.3 | 2.7% | 580.8 | 0.8% |
| .75 X (75) | 76.9 | 1.7% | 473.9 | 0.7% |
| Cutoff (100) | 100.9 | 1.7% | 400.8 | 1.0% |
| 1.25 X (125) | 131.6 | 1.1% | 312.7 | 1.0% |
| 1.50 X (150) | 158.6 | 1.1% | 246.8 | 1.2% |
Near cutoff Imprecision, 100 ng/mL
| Controls (ng/mL) | Number Tested | Correct Results | Confidence Level |
| --- | --- | --- | --- |
| .75 X (75) | 100 | 100 | >95% negative reading |
| 1.25 X (125) | 100 | 100 | >95% positive reading |
Acceptance criteria: 200 ng/mL Cutoff
Within-run imprecision $\leq 8\%$ CV
(≤10% CV for 100 & 300 ng/mL)
Run-to-run Imprecision $\leq 10\%$ CV
(≤12% CV for 50 & 150 ng/mL)
Near cutoff Imprecision $\leq 5\%$ Crossovers
Within-run Imprecision (N = 20), 200 ng/mL cutoff
| | Semi-quantitative | | Qualitative | |
| --- | --- | --- | --- | --- |
| Cals)/ Controls (ng/mL) | Mean (ng/mL) | CV% | Mean (mAbs) | CV% |
| .50 X (100) | 97.7 | 2.1% | 624.0 | 0.7% |
| .75 X (150) | 154.4 | 1.0% | 501.2 | 0.8% |
| Cutoff (200) | 197.2 | 1.2% | 431.8 | 1.1% |
| 1.25 X (250) | 258.2 | 0.7% | 323.2 | 0.9% |
| 1.50 X (300) | 314.4 | 0.8% | 248.1 | 1.2% |
{7}
Day-to-day Imprecision (N = 100), 200 ng/mL cutoff
| | Semi-quantitative | | Qualitative | |
| --- | --- | --- | --- | --- |
| Cals)/ Controls (ng/mL) | Mean (ng/mL) | CV% | Mean (mAbs) | CV% |
| .50 X (100) | 98.8 | 2.7% | 623.4 | 0.7% |
| .75 X (150) | 155.3 | 1.2% | 501.4 | 0.7% |
| Cutoff (200) | 197.3 | 1.6% | 423.8 | 2.0% |
| 1.25 X (250) | 259.1 | 0.8% | 320.7 | 1.4% |
| 1.50 X (300) | 313.8 | 0.9% | 245.0 | 1.6% |
Near cutoff Imprecision, 200 ng/mL
| Controls (ng/mL) | Number Tested | Correct Results | Confidence Level |
| --- | --- | --- | --- |
| .75 X (150) | 100 | 100 | >95% negative reading |
| 1.25 X (250) | 100 | 100 | >95% positive reading |
Acceptance criteria: 300 ng/mL Cutoff
Within-run imprecision $\leq 8\%$ CV
(≤10% CV for 150 & 450 ng/mL)
Run-to-run Imprecision $\leq 10\%$ CV
(≤12% CV for 150 & 450 ng/mL)
Near cutoff Imprecision $\leq 5\%$ Crossovers
Within-run Imprecision $(\mathbf{N} = 20)$ , $300~\mathrm{ng / mL}$ cutoff
| | Semi-quantitative | | Qualitative | |
| --- | --- | --- | --- | --- |
| Cals)/ Controls (ng/mL) | Mean (ng/mL) | CV% | Mean (mAbs) | CV% |
| .50 X (150) | 154.5 | 1.3% | 599.0 | 0.8% |
| .75 X (225) | 227.7 | 1.0% | 475.6 | 0.7% |
| Cutoff (300) | 305.4 | 1.0% | 365.0 | 0.9% |
| 1.25 X (375) | 384.0 | 0.8% | 276.3 | 0.9% |
| 1.50 X (450) | 434.8 | 0.8% | 219.9 | 1.8% |
Day-to-day Imprecision $(\mathbf{N} = 100)$ , $300~\mathrm{ng / mL}$ cutoff
| | Semi-quantitative | | Qualitative | |
| --- | --- | --- | --- | --- |
| Cals)/ Controls (ng/mL) | Mean (ng/mL) | CV% | Mean (mAbs) | CV% |
| .50 X (150) | 153.1 | 1.8% | 598.1 | 1.0% |
| .75 X (225) | 223.8 | 1.5% | 474.7 | 1.0% |
| Cutoff (300) | 301.2 | 1.4% | 365.2 | 1.2% |
| 1.25 X (375) | 377.3 | 1.2% | 274.5 | 1.4% |
| 1.50 X (450) | 438.3 | 1.9% | 216.3 | 3.9% |
{8}
Near cutoff Imprecision, 300 ng/mL
| Controls (ng/mL) | Number Tested | Correct Results | Confidence Level |
| --- | --- | --- | --- |
| .75 X (225) | 100 | 100 | > 95% negative reading |
| 1.25 X (375) | 100 | 100 | > 95% positive reading |
b. Linearity/assay reportable range:
Not applicable. The assay is intended for qualitative use.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Unassayed multi-analyte controls used for these qualitative and semi-quantitative applications are Control Set DAT I, Control Set DAT II, and Control Set DAT III, targeted at 0.75x for the negative control and 1.25x for the positive control relative to each of the 100, 200, and 300, ng/mL assay cutoffs.
The multi-analyte calibrator sets used for the Benzodiazepines Plus application are Preciset DAT Plus I (K031775) and Preciset DAT Plus II (K033306). In addition, C.f.a.s. DAT Qualitative Plus (K033306), a single level, multi-analyte qualitative calibrator can also be used for the calibration of the qualitative assay at a 300 ng/mL cutoff. Finally, C.f.a.s. DAT Qualitative Clinical, a 5 level, multi-analyte calibrator set, can also be used for the benzodiazepines qualitative assay at 100 or 300 ng/mL cutoffs. This calibrator set is comprised completely of qualified bottles into a new kit configuration from 5 existing Roche DAT calibrator sets listed as follows: Preciset DAT Plus I, level 4 (K031775), Preciset DAT Plus II, level 3 and C.f.a.s. DAT Qualitative Plus (K033306), Preciset DAT THC, level 3 (K033246) and Preciset DAT Amphetamine, level 4 (K033265)
Calibrators and controls are required with this assay and are specifically identified in the labeling. They are both provided separately, and have been previously cleared.
STABILITY
Product shelf-life claims for the R1 reagent (bottle 1 diluent and bottle 1a antibody) and the R2 conjugated drug derivative/ microparticle reagent will be established based on real-time stability studies. At least one lot of the reagent will be tested up to or exceeding the product shelf life claim (expiration date) before commercializing the product.
Real time stability testing is conducted on product stored at 2° - 8° C over the claimed shelf life. Product is tested at time 0, and at approximately 50%, 75% and
{9}
100% of the shelf life claim. Testing includes ≤ 5% crossovers at the cutoff when tested at 0.75x and 1.25x relative to the 300 ng/mL cutoff. All protocols, acceptance criteria, and data will be kept on file at Roche Diagnostics.
d. Detection limit:
Sensitivity of the assay at cutoffs of 100, 200 and 300 ng/mL.
The limit of detection (or analytical sensitivity) of the ONLINE DAT Benzodiazepines Plus at a 100 ng/mL cutoff was performed by testing 21 replicates of the 0 ng/mL standard. Two standard deviations above the mean yields a limit of detection of 2.1 ng/mL benzodiazepines.
| Calculation | | |
| --- | --- | --- |
| | Std 1 | Std 2 |
| Conc | 0 | 50 |
| Mean | 776.9 | 559.0 |
| SD | 4.8 | |
| 2 SD | 9.2 | |
| 3 SD | 13.8 | |
| Slope | -0.2294 | |
| Analytical Sensitivity | | |
| @ 2 SD | 2.1 | |
| @ 3 SD | 3.2 | |
The limit of detection (or analytical sensitivity) of the ONLINE DAT Benzodiazepines Plus at a 200 ng/mL cutoff was performed by testing 21 replicates of the 0 ng/mL standard. Two standard deviations above the mean yields a limit of detection of 3.8 ng/mL benzodiazepines.
| Calculation | | |
| --- | --- | --- |
| | Std 1 | Std 2 |
| Conc | 0 | 100 |
| Mean | 860.4 | 610.1 |
| SD | 4.7 | |
| 2 SD | 9.5 | |
| 3 SD | 14.2 | |
| Slope | -0.3995 | |
| Analytical Sensitivity | | |
| @ 2 SD | 3.8 | |
| @ 3 SD | 5.7 | |
{10}
The limit of detection (or analytical sensitivity) of the ONLINE DAT Benzodiazepines Plus at a 300 ng/mL cutoff was performed by testing 21 replicates of the 0 ng/mL standard. Two standard deviations above the mean yields a limit of detection of 7.6 ng/mL benzodiazepines.
| Calculation | | |
| --- | --- | --- |
| | Std 1 | Std 2 |
| Conc | 0 | 150 |
| Mean | 895.8 | 600.1 |
| SD | 7.5 | |
| 2 SD | 15.0 | |
| 3 SD | 22.5 | |
| Slope | -0.5074 | |
| Analytical Sensitivity | | |
| @ 2 SD | 7.6 | |
| @ 3 SD | 11.4 | |
e. Analytical specificity:
The specificity of Benzodiazepines Plus for various benzodiazepines and benzodiazepine metabolites was determined by generating inhibition curves for each of the compounds listed and determining the approximate quantity of each compound that is equivalent in assay reactivity to a 100, 200, and 300 ng/mL nordiazepam assay cutoff.
Benzodiazepines
| Compound | Response equivalent to 100 ng/mL Nordiazepam | Approximate Percent Cross-reactivity |
| --- | --- | --- |
| Alprazolam | 108 | 93 |
| Chlordiazepoxide | 148 | 68 |
| Diazepam | 106 | 94 |
| Oxazepam | 122 | 82 |
| Clonazepam | 148 | 68 |
| Flunitrazepam | 142 | 71 |
| Nitrazepam | 114 | 88 |
Benzodiazepines
| Compound | Response equivalent to 200 ng/mL Nordiazepam | Approximate Percent Cross-reactivity |
| --- | --- | --- |
| Alprazolam | 219 | 91 |
| Chlordiazepoxide | 318 | 63 |
| Diazepam | 215 | 93 |
| Oxazepam | 259 | 77 |
| Clonazepam | 307 | 65 |
| Flunitrazepam | 283 | 71 |
| Nitrazepam | 246 | 81 |
{11}
Benzodiazepines
| Compound | Response equivalent to 300 ng/mL Nordiazepam | Approximate Percent Cross-reactivity |
| --- | --- | --- |
| Alprazolam | 338 | 89 |
| Chlordiazepoxide | 486 | 62 |
| Diazepam | 340 | 88 |
| Oxazepam | 398 | 75 |
| Clonazepam | 445 | 67 |
| Flunitrazepam | 424 | 71 |
| Nitrazepam | 359 | 84 |
f. Assay cut-off:
Assay cutoff 100, 200, and 300 ng/mL. (See precision/reproducibility section above.
2. Comparison studies:
a. Method comparison with predicate device:
Because the candidate device was compared to a reference method, GC/MS, it was not compared to a predicate device.
Accuracy – Clinical Samples – 100 ng/mL Cutoff
Accuracy of the ONLINE DAT Benzodiazepines Plus assay at a 100 ng/mL cutoff was evaluated in method comparison studies between the Benzodiazepines Plus assay and GC/MS results.
Negative Urine Study – One hundred (100) urine samples, obtained from a clinical laboratory where they screened negative in a drug test panel, were evaluated with Benzodiazepines Plus. Ten of these samples were also confirmed negative by GC/MS. Of the 100 samples evaluated, 100 were negative relative to the 100 ng/mL cutoff.
| Qualitative Negatives | 100 | True Negatives | 100 |
| --- | --- | --- | --- |
| Qualitative Positives | 0 | False Positives | 0 |
| Qualitative Total | 100 | | |
| GC/MS Confirmed Negatives | 10 | | |
| SemiQuant Negatives | 100 | True Negatives | 100 |
| Semi Quant Positives | 0 | False Positives | 0 |
| SemiQuant Total | 100 | | |
| SQ Average of True Negs | 10.1 | | |
| SQ Std Dev of True Negs | 13.0 | | |
{12}
Positive Urine Study – Eighty-two urine samples were obtained from a clinical laboratory where they were screened positive for benzodiazepines by a commercially available immunoassay. They were subsequently confirmed positive by GC/MS for Oxazepam and/or other benzodiazepines or metabolites (i.e., Lorazepam, Lordiazepam, Nordiazepam, Alprazolam, and Norfludiazepam). Of the 82 samples evaluated, 82 were positive relative to the 100 ng/mL cutoff.
| Roche/ Hitachi 917
SemiQuant vs GC/MS | | + Test +
- Test - | Roche/ Hitachi 917
Qualitative vs GC/MS | |
| --- | --- | --- | --- | --- |
| GC/MS + | GC/MS - | | GC/MS + | GC/MS - |
| 82 | 0 | | 82 | 0 |
| 0 | 10 | | 0 | 10 |
## Acceptance criteria:
Correlation with GC/MS Positives/ Negatives >95%
## Accuracy – Clinical Samples – 200 ng/mL Cutoff
Accuracy of the ONLINE DAT Benzodiazepines Plus assay at a 200 ng/mL cutoff was evaluated in method comparison studies between the Benzodiazepines Plus assay and GC/MS results.
Negative Urine Study – One hundred (100) urine samples, obtained from a clinical laboratory where they screened negative in a drug test panel, were evaluated with Benzodiazepines Plus. Ten of these samples were also confirmed negative by GC/MS. Of the 100 samples evaluated, 100 were negative relative to the 200 ng/mL cutoff.
| Qualitative Negatives | 100 | True Negatives | 100 |
| --- | --- | --- | --- |
| Qualitative Positives | 0 | False Positives | 0 |
| Qualitative Total | 100 | | |
| GC/MS Confirmed Negatives | 10 | | |
| SemiQuant Negatives | 100 | True Negatives | 100 |
| Semi Quant Positives | 0 | False Positives | 0 |
| SemiQuant Total | 100 | | |
| SQ Average of True Negs | 15.2 | | |
| SQ Std Dev of True Negs | 16.8 | | |
{13}
Positive Urine Study – Seventy-eight urine samples were obtained from a clinical laboratory where they were screened positive for benzodiazepines by a commercially available immunoassay. They were subsequently confirmed positive by GC/MS for Oxazepam and/or other benzodiazepines or metabolites (i.e., Lorazepam, Lordiazepam, Nordiazepam, Alprazolam, and Norfludiazepam). Of the 78 samples evaluated, 76 were positive relative to the 200 ng/mL cutoff. Two results were false negatives by both the semiquantitative and qualitative 200 ng/mL cutoff assays. These results are still within the specifications of 95% agreement with GC/MS reference results.
| Roche/ Hitachi 917
SemiQuant vs GC/MS | | + Test +
- Test - | Roche/ Hitachi 917
Qualitative vs GC/MS | |
| --- | --- | --- | --- | --- |
| GC/MS + | GC/MS - | | GC/MS + | GC/MS - |
| 76 | 0 | | 76 | 0 |
| 2 | 10 | | 2 | 10 |
## Acceptance criteria:
Correlation with GC/MS Positives/ Negatives >95%
## Accuracy – Clinical Samples – 300 ng/mL Cutoff
Accuracy of the ONLINE DAT Benzodiazepines Plus assay at a 300 ng/mL cutoff was evaluated in method comparison studies between the Benzodiazepines Plus assay and GC/MS results.
Negative Urine Study – One hundred (100) urine samples, obtained from a clinical laboratory where they screened negative in a drug test panel, were evaluated with Benzodiazepines Plus. Ten of these samples were also confirmed negative by GC/MS. Of the 100 samples evaluated, 100 were negative relative to the 300 ng/mL cutoff.
| Qualitative Negatives | 100 | True Negatives | 100 |
| --- | --- | --- | --- |
| Qualitative Positives | 0 | False Positives | 0 |
| Qualitative Total | 100 | | |
| GC/MS Confirmed Negatives | 10 | | |
| SemiQuant Negatives | 100 | True Negatives | 100 |
| Semi Quant Positives | 0 | False Positives | 0 |
| SemiQuant Total | 100 | | |
{14}
SQ Average of True Negs 18.1
SQ Std Dev of True Negs 17.6
Positive Urine Study – Seventy-two urine samples were obtained from a clinical laboratory where they were screened positive for benzodiazepines by a commercially available immunoassay. They were subsequently confirmed positive by GC/MS for Oxazepam and/or other benzodiazepines or metabolites (i.e., Lorazepam, Lordiazepam, Nordiazepam, Alprazolam, and Norfludiazepam). Of the 72 samples evaluated, 72 were positive relative to the 300 ng/mL cutoff.
| Roche/ Hitachi 917
SemiQuant vs GC/MS | | + Test +
- Test - | Roche/ Hitachi 917
Qualitative vs GC/MS | |
| --- | --- | --- | --- | --- |
| GC/MS + | GC/MS - | | GC/MS + | GC/MS - |
| 72 | 0 | | 72 | 0 |
| 0 | 10 | | 0 | 10 |
## Acceptance criteria:
Correlation with GC/MS Positives/ Negatives >95%
b. Matrix comparison:
Not applicable. The assay is indicated only for urine specimens.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable. Clinical studies are not typically submitted for this device type.
b. Clinical specificity:
Not applicable. Clinical studies are not typically submitted for this device type.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable.
{15}
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.