D 905 EOS HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY RESERVOIR
Applicant
Dideco S.R.L.
Product Code
DTZ · Cardiovascular
Decision Date
Feb 1, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Dideco D 905 EOS Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation to assure the proper oxygenation capability of the device. The device is intended to be used for 6 hours or less.
Device Story
D 905 EOS is a hollow fiber membrane oxygenator with integrated heat exchanger and hardshell venous/cardiotomy reservoir. Device receives venous or suctioned blood as input; performs gas exchange (oxygenation/CO2 removal) and temperature regulation; outputs oxygenated blood to patient. Used in OR during cardiopulmonary bypass surgery; operated by perfusionists/surgeons. Smaller dimensions than predicate D 903 Avant 2, optimized for pediatric and small adult patients. Reservoir collects aspirated blood and venous return (gravity/vacuum assisted) for autotransfusion. Benefits include efficient extracorporeal circulation and blood management during surgery.
Clinical Evidence
Bench testing only. No clinical data. In vitro testing performed per ISO 7199 (1996) and FDA guidance documents. Evaluated gas transfer, pressure drop, plasma leakage, heat exchanger performance, hemolysis/cell depletion, mechanical integrity, and reservoir characterization (filtration efficiency, defoaming capacity). Testing conducted on devices aged up to 3 years. Results met established specifications and were comparable to predicate devices.
Technological Characteristics
Hollow fiber membrane oxygenator with integrated heat exchanger and hardshell venous/cardiotomy reservoir. Materials include PmM2/phosphorycholine coating. Ethylene oxide sterilized; single-use. Dimensions reduced compared to predicate. Operates via passive gas exchange and heat transfer. Complies with ISO 10993-1:1995 and ISO 7199:1996.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring extracorporeal circulation with max blood flow rate of 5 L/min. Used for oxygenation, CO2 removal, blood temperature control, and collection/autotransfusion of shed blood/chest drainage. Duration limited to 6 hours or less.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
D 903 Avant 2 Ph.I.S.I.O. Adult Hollow Fiber Oxygenator with Integral Hardshell Cardiotomy / Venous Reservoir with Biocompatible Treatment Surface (Ph.I.S.I.O.) (K033323)
{0}------------------------------------------------
Abbreviated 510(k) December 1, 2004
ber 1, 2004
# 510(k) SUMMARY
| SUBMITTER: | Dideco S.r.l.<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) Italy |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Luigi Vecchi<br>Phone: 011 39 0535 29811<br>Fax: 011 39 0535 25229 |
| DATE PREPARED: | December 1, 2004 |
| DEVICE TRADE NAME: | D 905 EOS: Hollow Fiber Oxygenator With Integrated<br>Hardshell Venous/Cardiotomy Reservoir |
| COMMON NAMES: | Hollow Fiber Oxygenator with Hardshell<br>Venous/Cardiotomy Reservoir<br>Hollow Fiber Oxygenator<br>Hardshell Venous/Cardiotomy Reservoir |
| CLASSIFICATION NAMES: | Cardiopulmonary Bypass Oxygenator<br>Cardiopulmonary Bypass Heat Exchanger<br>Cardiopulmonary Bypass Blood Reservoir<br>Cardiopulmonary Bypass Defoamer |
| PREDICATE DEVICES: | D 903 Avant 2 Ph.I.S.I.O. Adult Hollow Fiber<br>Oxygenator with Integral Hardshell Cardiotomy /<br>Venous Reservoir with Biocompatible Treatment<br>Surface (Ph.I.S.I.O.) (K033323)<br><br>Cobe Optimin Hollow Fiber Membrane Oxygenator<br>(K991452) |
#### DEVICE DESCRIPTION:
The D 905 EOS Hollow Fiber Oxygenator With Integrated Hardshell Venous/Cardiotomy Reservoir (hereafter referred to as the D 905 EOS) is a high efficiency microporous hollow fiber membrane oxygenator integrated with an heat exchanger and connected to an hardshell venous/cardiotomy reservoir. The device is a modified version of the currently marketed D 903 Avant 2 Ph.I.S.I.O (K033323) predicate device (hereafter referred to as the Avant Ph.1.S.I.O.). The modification is limited to an overall reduction in the size of the device. The reduction in size enables the device to be better suited for the pediatric and small adult patient population.
{1}------------------------------------------------
## INDICATION FOR USE:
The D 905 EOS Hollow Fiber Oxygenator With Integrated Hardshell Venous/Cardiotomy Reservoir hereinafter called the D 905 EOS, is a sterile, nonpyrogenic device intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters/minute. The device provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia, or aids in the maintenance of normothermia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from the patient's veins (gravity or vacuum assisted) during normal operation to assure the proper oxygenation capability of the device. Following intraoperative use, the reservoir is used for the collection and autotransfusion of shed blood and/or chest drainage. The device is intended to be used for 6 hours or less. Contact with blood for longer periods is inadvisable.
# TECHNOLOGICAL CHARACTERISTICS:
The D 905 EOS is essentially identical to the Avant Ph.I.S.I.O. predicate device with respect to design, materials, biocompatibility of the PmM2 coating, operating principles, technological characteristics and manufacturing process. The hardshell cardiotomy/venous reservoir present in both D 905 EOS and Avant Ph.I.S.I.O. share the same technological characteristics, operating principles, materials and basic design of the Avant Ph.I.S.I.O. except for a reduction of its dimensions. The D 905 EOS oxygenating module shares the same basic design, operating principles and control mechanisms of the Avant Ph.I.S.I.O. module. The only modifications consist of a reduction of its dimensions and consequently in less hollow fiber membrane material with respect to the Avant Ph.I.S.I.O. predicate device. These differences make the D 905 EOS more suited for smaller adult and pediatric patients like the Cobe Optimin predicate device. The D 905 EOS is substantially equivalent to the Cobe Optimin predicate device in intended use, patient population and performance specifications.
The coating of the oxygenating module is identical to the phosphorycholine coating used on the Avant Ph.I.S.I.O. predicate device.
The oxygenator is ethylene oxide sterilized and has a nonpyrogenic fluid path. It is for single use only.
## BIOCOMPATIBILITY TEST RESULTS:
A complete battery of tests were carried out in accordance with the requirements of ISO 10993-1:1995 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. Testing were performed on the D 905 EOS. (accelerated aging). The device aged up to three years was tested for Hemolysis, Cytotoxicity, Irritation, Acute Systemic Toxicity and Mutagenicity. Sterlity, Pyrogenicity, ETO residuals and package integrity testing were also conducted. The results of this testing met established specifications.
## IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the requirements of "Guidance for Cardiopylmonary Bypass Oxygenalors 510(k) submissions - Final Guidance for Industry and FDA Staff" issued on November 13, 2000 - "Guidance for Blood Extracorporeal Blood Circuit Defoamer 510(k) Submission" Final Guidance for industry and FDA issued on November 29, 2000 - "Guidance for Cardiopylmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for industry and FDA issued on November 29, 2000 and when applicable, following the ISO 7199 (1996) standard for "Cardiovascular Implants and Artificial Organs - Extra Corporeal Blood-Gas Exchangers (Oxygenator)" for providing the data necessary to demonstrate both the substantial equivalence with the predicate devices and also complying with safety and effectiveness requirements. The device aged up to 3 years was tested for gas transfer characteristics, pressure drop, plasma leakage data, operating blood volumes, heat exchanger performance evaluation, hemolysis/cell depletion, mechanical integrity, venous cardiotomy reservoir characterization (including breakthrough times and volumes, reservoir graduated scale accuracy, residual blood volume, defoaming capacity, filtration efficiency and reservoir housing integrity) and leaching studies characterization. The results of these tests met established
{2}------------------------------------------------
specifications. For comparative purposes, the same testing, when applicable, has been conducted also on the D 903 Avant 2 Ph.1.S.I.O. and on the Cobe Optimin predicate device. The result of the study showed that the device is comparable to the predicate devices concerning with all characteristics.
#### MARKETING HISTORY:
Up to now the D 905 EOS is in commercial distribution in Europe since March 2003. No reports of adverse events involving patient safety due to malfunctioning have been received.
#### CONCLUSIONS:
The results of in vitro studies demonstrate that the D 905 EOS performs in a manner substantially equivalent to the Cobe Optimin hollow fiber oxygenator with respect to the relevant functional parameters. Furthermore, the D 905 EOS performs in a manner substantially equivalent to the Avant Ph.I.S.I.O. predicate device with respect to the venous reservoir characterization. Biocompatibility studies demonstrate that the phosphorycholine coating is biocompatible and functional tests demonstrate that its performance is equivalent to the Cobe Optimin predicate device, according to its intended use. Additional testing has also demonstrated the effectiveness of production techniques to assure that the oxygenator is sterile and non-pyrogenic.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 1 2005
Dideco S.R.L. c/o Mr. Barry Sall Parexel International Corporation 195 West Street Waltham, MA 02451
Re: K043323
D 905 EOS Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir Regulation Number: 21 CFR 868.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II (two) Product Code: DTZ Dated: December 1, 2004 Received: December 2, 2004
Dear Mr. Sall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Mr. Barry Sall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dwana R. Lochner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): 上の4ろ323
Device Name: Dideco D 905 EOS Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir
Indications For Use:
The Dideco D 905 EOS Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation to assure the proper oxygenation capability of the device. The device is intended to be used for 6 hours or less.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mina P. Lacher
Division Sian-C Division of Cardiovascular Devices
510(k) Number K043323
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.