The TERATECH Model 8IOC4, 8IOL4, and 10LAP4 Probes are intended for diagnostic ultrasound imaging or fluid flow analysis of the human body; specific indications for use a tabulated in Section 4.3 of this submission.
Device Story
Diagnostic ultrasound transducers (8IOC4, 8IOL4, 10LAP4) designed for use with Teratech Model 2000 portable imaging system. Probes convert electrical signals to ultrasonic pulses and receive echoes to generate images; system processes signals for B-mode, M-mode, PWD, and Color Doppler imaging. Operated by clinicians in hospitals, clinics, or military field settings. Output displayed on system monitor for real-time visualization of anatomy and fluid flow; assists in diagnostic assessment and image-guided procedures (e.g., biopsies, needle placement). Benefits include portable, high-resolution imaging for diverse clinical environments.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence established through comparison of technological characteristics and intended use to legally marketed predicate devices.
Technological Characteristics
Diagnostic ultrasound transducers with 128 elements. 8IOC4: 6.0 MHz, curved array. 8IOL4: 7.5 MHz, linear array. 10LAP4: 7.0 MHz, linear array. Pitch 0.30-0.32 mm, elevation width 5.0 mm. Connectivity via Teratech Model 2000 portable system. Standard ultrasound imaging modes (B, M, PWD, Color Doppler).
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in clinical applications including intra-operative (specialized and neuro), abdominal, fetal, pediatric, small organ (thyroid, breast, testes), neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal (conventional and superficial), cardiac (adult/pediatric), and peripheral vessel. Includes ultrasound guidance for needle/catheter placement, cryosurgery, brachytherapy, amniocentesis, and infertility monitoring. Used in hospital, clinic, and military field settings.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
# 6043278
# DEC 1 3 2004
### 510(k) Summary for TERATECH Model 810C4, 810L4, and 10LAP4 Probes
#### SPONSOR 1.
Teratech Corporation 77-79 Terrace Hall Rd. Burlington, MA 01803
Contact Person: Charles F. Hottinger, Ph.D., RAC, Regulatory Affairs Consultant
Telephone: 206-780-7945
Date Prepared: November 5, 2004
#### 2. DEVICE NAME
Proprietary Name: TERATECH Model 8IOC4, 8IOL4, and 10LAP4 Probes Common/ Usual Name: Diagnostic Ultrasound Transducer
Classification Name:
Diagnostic Ultrasound Transducer (21 CFR 892.1570, 90-ITX)
#### 3. PREDICATE DEVICES
| Subject Device | Predicate 1 | Predicate 2 |
|----------------|------------------|---------------|
| 8IOLA | Philips LI9-5 | TERATECH 10V5 |
| 8IOC4 | Philips CT8-4 | TERATECH 10V5 |
| 10LAP4 | Philips LAP L9-5 | |
The Philips probes are marketed for use with the Philips HDI 5000; that system has been cleared in following 510(k) submissions, among possibly others: K961459, K991671, K994373, K002003, and K011224.
The TERATCH predicate probes, as well as the subject devices, are used with the TERATECH Model 2000 portable imaging system. This system has been cleared under the following 510(k) submissions: K992595,
Page 4.2-4
{1}------------------------------------------------
### K010883, K012191, K030191, and K040840.
#### INTENDED USE ರ್.
The TERATECH Model 8IOC4, 8IOL4, and 10LAP4 Probes are intended for diagnostic ultrasound imaging or fluid flow analysis of the human body; specific indications for use a tabulated in Section 4.3 of this submission.
#### DEVICE DESCRIPTION ર-
The TERATECH Model 810C4, 81OL4, and 10LAP4 Probes are intended for use with the Model TERATECH2000, a portable ultrasoundimaging system. Technical specificationsfor the Model 810C4, and 10LAP4 Probes with the Model 2000 are as follows:
| Model | 8IOC4 | 8IOLA | 10LAP4 |
|----------------------|---------|---------|---------|
| Frequency/ | 6.0 MHz | 7.5 MHz | 7.0 MHz |
| # Elements | 128 | 128 | 128 |
| Array type | Curved | Linear | Linear |
| Pitch (mm) | 0.32 | 0.30 | 0.30 |
| Elevation width (mm | 5.0 | 5.0 | 5.0 |
| Geometric focus (mm) | | 25 | 25 |
| Azimuth radius (mm | 40 | N/A | N/A |
| Azimuth length (mm | 50.0 | 38.4 | 38.4 |
#### BASIS FOR SUBSTANTIAL EQUIVALENCE 6.
The TERATECH Model 810C4, 81OL4, and 10LAP4 Probes are substantially equivalent to the above cited Philips transducers, which are currently in commercial distribution in the United States. The TERATECH Model 810C4, 81OL4, and 10LAP4 Probes are believed to be identical in mechanical design and materials to the respective Philips, and are intended for the same clinical applications.
{2}------------------------------------------------
#### 4.3 INDICATIONS FOR USE
.
The TERATECH Model 8IOC4, 8IOL4, and 10LAP4 Probes are intended for the uses described in
the for the filiance of adjustical From is provided below I he TEKATECH MOGELBIOCS, STOCH, and 10EM IST is I v is provided below.
the Diagnostic Ultrasound Indications For Use Form is provided below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized bird-like design with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 3 2004
TERATECH Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services 1394 25th Street NW BUFFALO MN 55313
Re: K043278
Trade Name: Terason (Teratech) Model 2000 Portable Ultrasound System and Teratech Model 8IOC4, 8IOL4, 10LAP4 Probes Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed Doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: November 22, 2004 Received: November 26, 2004
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been rear 20, 1770, and vith the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general may also rovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with 1 the Terason (Teratech) Model 2000 Portable Ultrasound System, as described in your premarket notification:
Transducer Model Number
Teratech Model 8IOC4 Teratech Model 8IOL4 Teratech Model 10LAP4
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be If your device to such additional controls. Existing major regulations affecting your device can be found in the Sublest to suen additions. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I least be advisou that 1221 b 1804at your device complies with other requirements of the Act or any I DA mas made a adversions administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) er N Fart 801); good manally; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act);-21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first I mis occernmination of ourselearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing (Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by the Ornee of Gomphanes that in the Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
{5}------------------------------------------------
Page 3 – Mr. Job
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
| System: | Terason Model 2000 Portable Ultrasound | | | | | | | |
|-----------------------------|------------------------------------------------------------------------------------|-------------------|------|------|-----|---------------|----------------|-------|
| Transducer: | (see comments) | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Dopp | Comb.<br>Modes | Other |
| Ophthalmic | Ophthalmic | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Fetal4 | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Abdominal4: | P1 | P2,4 | P2,4 | | P2,4 | P3 | P2,4 |
| | Intra-operative (Spec.)3,4 | P1 | P2 | P2 | | P2 | P3 | P2 |
| | Intra-operative (Neuro) | P3 | P3 | P3 | | P3 | P3 | P3 |
| | Laparoscopic | N | N | N | | N | N | N |
| | Pediatric4: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| Fetal<br>Imaging<br>& Other | Small Organ (Thyroid,<br>Breast, Testes, etc.)3,4: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Neonatal Cephalic4: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Adult Cephalic4: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Trans-rectal: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Trans-vaginal: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)4: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Musculo-skel. (Superfic)4: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | P1 | P1 |
| | Cardiac Adult | P1 | P2 | P2 | | P2 | P2 | P2 |
| Cardiac | Cardiac Pediatric | P1 | P2 | P1 | | P2 | P1 | P1 |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Peripheral vessel4: | P1,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P1,4 |
| Peripheral<br>Vessel | Other (Specify) | | | | | | | |
Terason Model 2000 Portable Ultrasono System_
N= new indication: P= previously cleared by FDA; E= added under Appendix E
V= new indication; P= previously clearce by FDA; E= added thour Applerior C
Includes Color Doppler (CD), Directional Power Doppler (OPD), and (non-directional) Power Doppler
B+M: B+PWD; B+CD; B+DPD; B+PD.
Harmonic Imaging (HI)
Harmonic Imaging (HI)
Includes ultrasound guidance for placement of needles, catheters.
Abdominal organs and peripheral vessel.
" Abdominal organs and periorient of needles, catheles, catheres, cayosurgery, and brechytherepy
' lockes und guidance of transvaglial biopsy. Inforlik, morely, and breany deny deny delay delay dela
" Includes uitane of transvaglial biopsy. Inferlily monitor development.
"Includes guidance of amniocentesis, infertility monitoring of follicle development."
System uses previously cleared under K992505 with 3 MHz Model L3 (Linear).
" System uses previously cleared under K012191.
· System uses praviously cleared under K010883.
· System uses previously cleared under K030191.
System uses previously cleared under K040840.
े System uses previously cleaned under K040840.
Includes uses in milltary field settings in addition to hospitalicinic settings.
{PLEASE DO NOT WRITE BELOW THIS LINE-CONTINU
(FEESE DO NOT WRITE BECOW THE BROOM PHO DATE Of Device Evaluation
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescipion Use (Per 2
Daniel b. Legron
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number KD43278
1 2/09/04 TERATECH Corp. 510(k) Teratech Model 8IOC4, 810L4, and 10LAP4 Probes
{7}------------------------------------------------
Tarason Model 2000 Portable Ultrascund System System:
BIOC4 Transducer: Intended Use: Diagnostic ultraging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|--------------------------------------------------|-------------------|---|-----|-----|---------------|----------------|-------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Dopp | Comb.<br>Modes | Other |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal* | | | | | | | |
| | Abdominal*: | | | | | N* | N* | N* |
| | Intra-operative (Spec.) | N | N | N | | N* | N* | N* |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric*: | | | | | | | |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)*: | | | | | | | |
| | Neonatal Cephalic*: | | | | | | | |
| | Adult Cephalic*: | | | | | | | |
| | Trans-rectal*: | | | | | | | |
| | Trans-vaginal*: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)* | | | | | | | |
| | Musculo-skel. (Superfic)*: | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel*: | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E= added under Appendix E
Includes Color Doppier (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler
*B+M; B+PWD; B+CD; B+DPD; B+PD.
Harmonic Imaging (HI)
Includes ultrasound guidance for placement of needles, cathelers
Abdominal organs and peripheral vessel.
Indudes ultrasound guidance for placement of needles, cather of felicle development
Includes ultrasound guidance of transvaginal bloosy, infectility montantonograph
- Includes guidance of amniocentesis, infertility monitoring of follicle development.
micudes unistophe of variates, infrilly moltong of folloc development.
Addes guismer ef anince in the monthing of folloce development.
Addional Comments P: uses previousl
under K030191
Urban, non-metro
for regular uses in military field settings, in addition to hospital/clinic settings.
Includes uses in military lield settings in addition to hospitalicing settings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Daniel A. Bergman
51
TERATECH Corp. 510(k) 1 2/09/04 Teratech Model 810C4, 810L4, and 10LAP4 Probes
{8}------------------------------------------------
| System: | Terason Model 2000 Portable Ultrasound System | | | | | | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------|----|-------------------|-----|-----|----------------|-----------------|--------|--|
| Transducer: | 8I0L4 | | | | | | | | |
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | |
| General | Specific | B | M | PWD | CWD | Color<br>Dapp' | Comb.<br>Modes' | Other' | |
| (Track I Only) | (Tracks I & III) | | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal' | | | | | | | | |
| | Abdominal': | | | | | | | | |
| | Intra-operative (Spec.) <span style="font-size: smaller;">a</span> | N' | N' | N' | | N' | N' | N' | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric': | | | | | | | | |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.) <span style="font-size: smaller;">4</span> : | | | | | | | | |
| | Neonatal Cephalic": | | | | | | | | |
| | Adult Cephalic": | | | | | | | | |
| | Trans-rectal': | | | | | | | | |
| | Trans-vaginal': | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.)': | | | | | | | | |
| | Musculo-skel. (Superfic)": | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel': | | | | | | | | |
| | Other (Specify) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
V= new Indication; P= previously cleared by FDA; E= saded under Appendix E
lincludes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler
B+M; B+PWD; B+CD; B+DPD; B+PD.
Harmonic Imaging (HI)
Includes ultrasound guidance for placement of needles, catheters.
Abdominal organs and peripheral vessal.
Indudes ultrasound guldance for placement of needies, cathelers, cryosurgery, and brachytherapy
"Includes ultrasound guidance of transvaginal bloosy, informations of follicie development.
System uses previously cleared under K992505 with 3 Milze development
System uses previously cleared under K992505 with 3 Mitz Model L3 (Linear).
System uses previously cleared under K012191.
System uses previously cleared under K010883.
· System uses previously cleared under K030191.
Includes uses in military field settings in addition to hospitaliclinic settings.
(PLEASE DO NOT WRITE BELOW THIS LINEA to HOSpitation a Catalys.
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation Prescription Use (Per 21 CFR 801.109)
David R. Seymore
(Division Sign-Off) Division of Re and Radiological Det 510(k) Number
TERATECH Corp. 510(k) 1 2/09/04 Teratech Model 810C4, 810L4, and 10LAP4 Probes
{9}------------------------------------------------
| System: | Terason Model 2000 Portable Ultrasound System | | | | | | | |
|--------------------------------------------------------------------------------------------------|-------------------------------------------------|---|---|-----|-----|---------------|----------------|-------|
| Transducer: | 10LAP4 | | | | | | | |
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | B | M | PWD | CWD | Color<br>Dopp | Comb.<br>Modes | Other |
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal: | | | | | | | |
| | Intra-operative (Spec.): | N | N | N | | N | N | N |
| | Intra-operative (Neuro) | N | N | N | | N | N | N |
| Fetal<br>Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric: | | | | | | | |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.): | | | | | | | |
| | Neonatal Cephalic: | | | | | | | |
| | Adult Cephalic: | | | | | | | |
| | Trans-rectal: | | | | | | | |
| | Trans-vaginal: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic): | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Peripheral vessel: | | | | | | | |
| Peripheral | Other (Specify) | | | | | | | |
on Model 2000 Portable Utraspund System -
N= new indication; P= previously cleand by FDA; E= added under Appendired
l= new indication; Pe previously desired by FDF E= added under Apperionely Power Doppler.
Includes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directlonal)
· B+M; B+PWD; B+CD; B+DPD; B+PD.
Harmonic Imaging (HI)
" Includes ultrasound guidance for placement of needles, cathelers.
Abdominal organs and peripheral vessel.
° Abdominal organizat of needles cathelers, cathelers, axpsurgery, and brachyherapy
' lnekdes titance dransmit of necility moritorial
2. Includes ultrasound evidence of transvaginal biopsy, infertility monitoring or IUI/IVF de
- Includes evidence of amniocentesis, infertility monitoring or follicle development.
¹System uses previously cleared under K982505 with 5 Mhz Model L3-101
2 System uses previously cleared under K012191.
System uses proviously deared under K0106B3. 2
· System uses previously cleared under K030191.
Includes uses in military field settings in addition to hospitalicinic settings. Includes uses in mlitary field actings in addition to nosphareines sellings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANDTHER PAGE IF NEEDED)
Concurrence of Canter for Devices and Radiological Health, Office of Device Evaluation
Concerner Starter (B = 01 SED =01 100) Prescription Use (Per 21 CFR 801.109)
David A. Lippson
(Division Sign-Off) eproductive, Abdomin Division of and Radi 510(k) Numb
12/09/04 TERATECH Corp. 510(k) Teratech Model 810C4, 810C4, and 10LAP4 Probes
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.